Thorac Cardiovasc Surg 2019; 67(S 01): S1-S100
DOI: 10.1055/s-0039-1678875
Oral Presentations
Monday, February 18, 2019
DGTHG: Kathetergestützte Herzklappentherapie (TAVI)
Georg Thieme Verlag KG Stuttgart · New York

How to Prepare an Easy Device Landing Zone for Transcatheter Aortic Valve Implantation in Pure Aortic Regurgitation—Proof of Concept

A. Unbehaun
1   Department of Cardiothoracic and Vascular Surgery, German Heart Center Berlin, Berlin, Germany
,
S. Buz
1   Department of Cardiothoracic and Vascular Surgery, German Heart Center Berlin, Berlin, Germany
,
C. Klein
2   Department of Internal Medicine - Cardiology, German Heart Center Berlin, Berlin, Germany
,
M. Kofler
1   Department of Cardiothoracic and Vascular Surgery, German Heart Center Berlin, Berlin, Germany
,
A. Meyer
1   Department of Cardiothoracic and Vascular Surgery, German Heart Center Berlin, Berlin, Germany
,
E. Potapov
1   Department of Cardiothoracic and Vascular Surgery, German Heart Center Berlin, Berlin, Germany
,
M. Kukucka
3   Institute for Anesthesiology, German Heart Center Berlin, Berlin, Germany
,
V. Falk
1   Department of Cardiothoracic and Vascular Surgery, German Heart Center Berlin, Berlin, Germany
,
J. Kempfert
1   Department of Cardiothoracic and Vascular Surgery, German Heart Center Berlin, Berlin, Germany
› Author Affiliations
Further Information

Publication History

Publication Date:
28 January 2019 (online)

Objectives: Transcatheter aortic valve implantation (TAVI) in pure aortic regurgitation (AR) is challenging because of the subsequent difficulty in anchoring the transcatheter valve in a noncalcified device landing zone (DLZ). Prestenting can help prepare a stable DLZ for TAVI in pure AR, prevent valve migration, and improve device success. In an initial case series, we report the novel technique of uncovered stent implantation into a noncalcified aortic valve as a prestenting strategy to prepare a neo DLZ for TAVI in pure AR.

Methods: Between July 2017 and May 2018, five consecutive patients with pure AR of their native aortic valve and high risk for surgical aortic valve replacement underwent a TAVI procedure (Table 1). To prevent valve migration, an uncovered stent (sinus XL, 40 mm length, OptiMed-Medizinische-Instrumente-GmbH, Ettlingen, Germany) was implanted into the DLZ just before a SAPIEN-3 valve (Edwards Lifesciences, Irvine, California, United States) was deployed in a standard manner. In patients without permanent mechanical circulatory support (n = 2), TAVI was performed under a short duration of cardiopulmonary bypass for safety reasons using percutaneous peripheral cannulation.

Results: According to Valve Academic Research Consortium-II (VARC-II) criteria, device success was achieved in all patients (100%). Postprocedural aortic valve regurgitation was absent in three patients (60%) and trace in two patients (40%). There was no in-hospital mortality. Postprocedural computed tomography scans confirmed a stable position of the SAPIEN-3 valve and the stent inside the DLZ. Beside acute renal failure in one patient (20%), there were no further procedure-related VARC-II events.

Conclusion: TAVI has been shown to be effective in eliminating pure AR but was associated with a high necessity of implantation of a second valve. In our initial case series of five patients, prestenting prepared an easy DLZ for TAVI in pure AR and prevented valve migration. We consider this prestenting technique a useful novel strategy that may be applicable also in patients without mechanical circulatory support, at least as long as specific TAVR devices for pure AR are lacking.

[Table 1]

Pat.

Gender

Age (y)

TAVI indication

EuroSCORE II (%)

Mean annulus diameter (mm)

Sinus XL stent diameter (mm)

TAVI strategy

Device success

Status after TAVI

1

Male

71.8

s/p LVAD

62.4

28.0

32

TF SAPIEN-3 29 mm

Yes

Death after 105 d (right heart failure)

2

Male

77.6

s/p LVAD

10.4

25.8

30

TF SAPIEN-3 29 mm

Yes

Alive after 211 d

3

Male

69.9

s/p LVAD

39.6

25.1

30

TF SAPIEN-3 29 mm

Yes

Death after 13 d (septic MOF)

4

Male

75.6

s/p aortic dissection

2.4

27.3

32

TF SAPIEN-3 29 mm

Yes

Alive after 120 d

5

Female

67.7

Frailty, immobility

6.7

24.5

30

TA SAPIEN-3 29 mm

Yes

Alive after 106 d