Endoscopy 2023; 55(03): v7
DOI: 10.1055/a-1944-8392
Author commentary

Author commentary on Stefano Francesco Crinò et al.

Stefano Francesco Crinò et al. Wet-suction versus slow-pull technique for endoscopic ultrasound-guided fine-needle biopsy: a multicenter, randomized, crossover trial

This randomized crossover study of 210 individuals with pancreatic and nonpancreatic lesions compared wet-suction versus stylet slow-pull techniques for endoscopic ultrasound-guided fine-needle biopsy with a Franseen or fork-tip needle. Tissue cores, defined as intact tissue of ≥ 550 μm in length, were obtained more frequently with the wet-suction technique (71.4 % vs. 61.4 %; P = 0.03). Diagnostic accuracy was similarly high, regardless of sampling technique.

Qualität:


Publikationsverlauf

Artikel online veröffentlicht:
28. Februar 2023

© 2023. Thieme. All rights reserved.

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany