Ultraschall Med 2009; 30(4): 383-389
DOI: 10.1055/s-0028-1109673
Original Article

© Georg Thieme Verlag KG Stuttgart · New York

Contrast-Enhanced Ultrasound (CEUS) for the Characterization of Focal Liver Lesions – Prospective Comparison in Clinical Practice: CEUS vs. CT (DEGUM Multicenter Trial) Parts of this Manuscript were presented at the Ultrasound Dreiländertreffen 2008, Davos

Kontrastverstärkte Sonografie (CEUS) zur Charakterisierung fokaler Leberläsion im klinischen Alltag: CEUS vs. CTK. Seitz1 , D. Strobel2 , T. Bernatik2 , W. Blank3 , M. Friedrich-Rust4 , A. von Herbay5 , C. F. Dietrich6 , H. Strunk7 , W. Kratzer8 , A. Schuler9
  • 1Medizinische Klinik, Kreiskrankenhaus, Sigmaringen
  • 2Innere Medizin I, Universitätskrankenhaus Erlangen
  • 3Medizinische Klinik, Klinikum am Steinenberg, Reutlingen
  • 4Innere Medizin I, J. W. Goethe-Universität, Frankfurt a. M.
  • 5Innere Medizin, Universität Tübingen
  • 6Innere Medizin 2, Caritas-Krankenhaus, Bad Mergentheim
  • 7Radiologische Klinik, Universitätsklinik, Bonn
  • 8Klinik für Innere Medizin I, Universitätsklinikum Ulm
  • 9Medizinische Klinik, Helfensteinklinik, Geislingen
Weitere Informationen

Publikationsverlauf

received: 24.6.2009

accepted: 22.7.2009

Publikationsdatum:
17. August 2009 (online)

Zusammenfassung

Ziel: Ziel der Studie war es, den diagnostischen Stellenwert der Kontrastmittelsonografie (CEUS) in der Differenzialdiagnose von fokalen Leberläsionen prospektiv und multizentrisch im klinischen Alltag zu evaluieren. Hierzu ist der Vergleich mit der Spiral-CT (SCT), dem radiologischen Standardverfahren geeignet und sinnvoll. Material und Methoden: 1349 Patienten mit einem im fundamentalen Ultraschall einschließlich Farbdoppleruntersuchung ungeklärtem Lebertumor wurden von Mai 1004 bis Dezember 2006 in 14 Krankenhäusern prospektiv mit einer standardisierten CEUS (Pulsinversionstechnik, mechanischer Index < 0,4) untersucht. Die Kontrastmittelanreicherung der Lebertumoren wurde anhand bekannter tumortypischer Vaskularisationsmuster dokumentiert und analysiert. Ziel der Analyse war neben der Differenzierung der Dignität, möglichst die definitive Zuordnung zu einer Tumorentität. Ein Subkollektiv von 267 Patienten wurde zusätzlich standardisiert mit SCT untersucht. Die definitive Diagnose stützte sich auf die Histologie, bei typischem Hämangiom und FNH auf CT oder MRT, gesicherte klinische Befunde und zusätzliches Follow-up. Ergebnisse: Das Subkollektiv von 267 Patienten wurde in zwei Subgruppen unterteilt. Bei 110 dieser Patienten (Subkollektiv A) wurde aufgrund eindeutiger SCT-Diagnosen in 79 Fällen von Hämangiom oder FNH sowie bei weiteren 20 Fällen mit sicherer klinischer Diagnose auf die histologische Sicherung verzichtet. In 6 (5,5 %) Fällen wurde keine definitive klinische Enddiagnose gestellt. In diesem Subkollektiv zeigte sich bei der Dignitätsbeurteilung 90-mal ein konkordanter Befund mit der CEUS, 19-mal war der Befund diskordant, die Entität der Raumforderung wurde 82-mal-übereinstimmend und 27-mal-unterschiedlich beurteilt. Bei weiteren 158 Fällen (Subkollektiv B) lag neben der CEUS und SCT eine Histologie vor, 4 Fälle blieben ohne definitive Diagnose, nur in 23 Fällen lag eine FNH oder ein Hämangiom vor. Die Dignität wurde mit CEUS und SCT in dieser Gruppe in 124 Fällen übereinstimmend und in 30 Fällen diskordant beurteilt (CEUS/CT: Sensitivität 94,0 / 90,7 %, Spezifität 83,0 / 81,5 %, PPV 91,6 / 91,5 %, NPV 87,5 / 80,0 %, Accuracy 90,3 / 87,8 %). Die Tumorentititäten wurde in 103 Fällen übereinstimmend beurteilt, in 51 Fällen war die Beurteilung diskrepant (CEUS/SCT: Sensitivität 95,3 / 90,6 %, Spezifität 83,7 / 81,6 %, PPV 92,7 / 91,4 %, NPV 89,1 / 80,0 %, Accuracy 91,6 / 87,7 %). Ein statistisch signifikanter Unterschied ließ sich nicht nachweisen. Auch die Analyse für die einzelnen Tumorentitäten ergab für Hämangiome, FNH, HCC und Metastasen lediglich eine tendenzielle Überlegenheit der CEUS in der Tumordifferenzierung. Schlussfolgerung: Im multizentrischen Einsatz unter klinischen Routinebedingungen ist in der vorliegenden prospektiven Studie die CEUS der CT hinsichtlich der Beurteilung der Tumordignität und Tumordifferenzierung ebenbürtig. Statistisch signifikante Unterschiede waren nicht nachweisbar. Die CEUS sollte zur Tumordifferenzierung vor der SCT eingesetzt werden, da sich bei sorgfältiger klinischer Bewertung strahlenbelastende Computertomografien und auch invasive Biopsien in erheblicher Zahl vermeiden lassen.

Abstract

Aim: The aim of our study was to evaluate the diagnostic value of Contrast-Enhanced Ultrasound (CEUS) for the characterization of focal liver lesions in a prospective multi-center study in clinical practice. For this purpose CEUS was compared with the spiral-CT (SCT), the standard radiological method. Material and Methods: 1349 patients with unclear liver lesions after fundamental ultrasound diagnostics including color doppler analysis were examined with standardized CEUS (pulse inversion method, mechanical index < 0.4) from May 2004 to December 2006 in 14 hospitals in a prospective study. The enhancement of contrast medium in the liver tumors was analyzed according to known tumor-specific vascular patterns, using standardized documentation and analysis methods for the differentiation of tumor differentiation (malign or benign) and tumor specification (entity). A subcollective of 267 patients was additionally examined by standardized SCT method. Final diagnosis was based on histology, SCT or MRI in typical findings of liver hemangioma and FNH and on proved clinical data and additional follow up. Results: The subcollective of 267 patients was divided in two subgroups. In 109 of these patients (subgroup A) there was no histological verification, diagnoses based on clear SCT-findings in 79 cases of hemangioma or FNH, as well as in 20 cases with a clear clinical diagnosis. 6 cases (5.5 %) remained unclear. In this subgroup the assessment of tumor differentiation was concordant with CEUS in 90 cases, discordant in 19 cases and the assessment of tumor specification was concordant in 82 and and discordant in 27 cases. In 158 patients (subgroup B) a histological finding was also present, only in 4 cases no definitive tumor diagnosis was achieved. In this subgroup assessment of tumor differentiation with CEUS and SCT was concordant in 124 cases and discordant in 30 cases (CEUS/SCT: sensitivity 94.0 / 90.7 %, specificity 83.0 / 81.5 %, PPV 91.6 / 91.5 %, NPV 87.5 / 80.0 %, accuracy 90.3 / 87.8 %). Tumor specification matched in 103cases and were different in 51 cases (CEUS/SCT: sensitivity 95.3 / 90.6 %, specificity 83.7 / 81.6 %, PPV 92.7 / 91.4 %, NPV 89.1 / 80.0 %, accuracy 91.6 / 87.7 %). A statistically significant difference could not be established. The analysis of particular tumor specification showed a statistically non significant slight advantage in tumor differentiation for CEUS in the case of hemangioma, FNH, HCC and metastases. Conclusion: In a multi-center approach under routine clinical conditions, this prospective study demonstrates CEUS to be of equal rank to the CT-scan in regard to the assessment of tumor differentiation and specification. No statistically significant differences could be established. CEUS should be employed before computed tomography is performed for the differentiation of liver tumors, because radiation exposure and invasive biopsies can be avoided in veritable numbers of cases, when precise clinical evaluation of the findings is implemented.

References

  • 1 Meire H, Cosgrove D, Dewbury K. et al .Clinical Ultrasound a comprehensive text. Abdominal and General Ultrasound London; Churchill Livingstone 2001 2nd ed
  • 2 Seitz K, Schuler A, Rettenmaier G. Klinische Sonographie und Differenzialdiagnose. Stuttgart; Thieme Verlag 2007 2nd ed
  • 3 Oldenburg A, Albrecht T. Sonografische Leberdiagnostik bei Tumorpatienten ohne und mit Kontrastmittel.  Ultraschall in Med. 2008;  29 488-498
  • 4 Seitz K. CEUS for Liver Tumors: Facts, Studies, Relevance and Reality in the Clinical Routine.  Ultraschall in Med. 2008;  29 485-487
  • 5 Lanka B, Jang H J, Kim T K. et al . Impact of Contrast-Enhanced Ultrasonography in a Tertiary Clinical Practice.  J Ultrasound Med. 2007;  26 1703-1714
  • 6 Tranquart F, Correas J M, Ladam M V. et al . Real-time contrast-enhanced ultrasound in the evaluation of focal liver lesions: diagnostic efficacy and economical issues from a French multicentric study.  J Radiol. 2009;  90 109-122
  • 7 Strobel D, Seitz K, Blank W. et al . Contrast-enhanced Ultrasound for the Characterization of Focal Liver Lesions – Diagnostic Accuracy in Cinical practice (DEGUM multicenter trial).  Ultraschall in Med. 2008;  29 499-505
  • 8 Greis C. Technical Overview: SonoVue (Bracco, Milan).  Eur Radiol. 2004;  14 P11-P15
  • 9 Albrecht T, Blomley M, Bolondi L. et al . Guidelines for the use of ultrasound. January 2004.  Ultraschall in Med. 2004;  25 249-256
  • 10 Prokop M, Galanski M. Spiral and Multislice Computed Tomography. Stuttgart; Thieme Verlag 2003
  • 11 Quaia E, Calliada F, Bertolotto M. et al . Characterization of focal liver lesions with contrast-specific US modes and a sulphur hexafluoride-filled microbubble contrast agent: diagnostic performance and confidence.  Radiology. 2004;  232 420-430
  • 12 Dietrich C F. Update on Contrast Enhanced Ultrasound in Abdominal and Vascular Imaging.  Ultraschall in Med. 2008;  29 (S4) S 188-202
  • 13 Bleuzen A, Huang C, Olar M. et al . Diagnostic Accuracy of Contrast-Enhanced Ultrasound in Focal Lesions of the Liver Using Cadence Contrast pulse Sequencing.  Ultraschall in Med. 2006;  27 40-48
  • 14 Wilson S R, Jang H J, Kim T K. et al . Diagnosis of focal liver masses on ultrasonography: comparison of unenhanced and contrast-enhanced scans.  Am J Roentgenol. 2007;  189 W7-W12
  • 15 Quaia E, D’onofrio M, Palumba E. et al . Comparison of contrast-enhanced ultrasound and contrast-enhanced computer tomography in metastatic disease of the liver: diagnostic performance and confidence.  Eur Radiol. 2006;  16 1599-1609
  • 16 Quaia E, Stacul F, Gaiani S. et al . Comparison of diagnostic performance of unenhanced vs. SonoVue-enhanced ultrasonograpgy in focal liver lesions characterization. The experience of three Italian Centers.  Radiol Med. 2004;  108 71-81
  • 17 Konopke R, Bunk A, Kersting S. Contrast-Enhanced Ultrasonography in Patients with Colorectal Liver Metastases after Chemothrapy.  Ultraschall in Med. 2008;  29 S4 S /203-209
  • 18 Catala V, Nicolau C, Vilana R. et al . Characterization of focal liver lesions: comparative study of contrast-enhanced ultrasound versus spiral computer tomography.  Eur Radiol. 2007;  17 1066-1073
  • 19 Celli N, Gaiani S, Piscaglia F. et al . Characterization of liver lesions by real-time contrast-enhanced ultrasonography.  Eur J Gastroenterol Hepatol. 2007;  19 3-14
  • 20 Dietrich C F, Mertens J C, Braden B. et al . Contrast-Enhanced Ultrasound of Histologically Proven Liver Hemangiomas.  Hepatology. 2007;  45 1139-1145
  • 21 Strobel D, Kleinecke C, Hänsler J. et al . Contrast-Enhanced Sonography of Hepatocellular Carcinomas – Correlation with Histological Differentiation.  Ultraschall in Med. 2005;  26 270-276
  • 22 Burns P, Wilson S. Focal Liver Masses: Enhancement patterns of contrast-enhanced images – concordance of US scans with CT scans and MR images.  Radiology. 2007;  242 162-174
  • 23 Dietrich C F, Kratzer W, Strobel D. et al . Assessment of metastatic liver disease in patients with primary extrahepatic tumors by contrast-enhanced sonography versus CT and MRI.  World J Gastroenterol. 2006;  12 1699-1705
  • 24 Xu H X, Liu G J, Lu M D. et al . Characterization of small focal liver lesions using real-time contrast-enhanced. Diagnostic performance analysis in 200 patients.  J Ultrasound Med. 2006;  25 349-361
  • 25 Wermke W. Sonographische Differenzialdiagnose: Leberkrankheiten. Köln; Deutscher Ärzte Verlag 2007
  • 26 Bauditz J, Schade T, Wermke W. Sonografische Diagnostik des hilären cholangiozellulären Karzinoms mittels Echosignalverstärker.  Ultraschall in Med. 2007;  28 161-167
  • 27 Dörffel Y, Wermke W. Neuroendocrine Tumors: Characterization with Contrast-Enhanced Ultrasonography.  Ultraschall in Med. 2008;  29 506-524
  • 28 Nicolau C, Vilana R, Catala V. et al . Importance of evaluting all vascular phases on contrast-enhanced sonography in the differentition of benign from malignant focal liver lesions.  Am J Roentgenol. 2006;  186 158-167
  • 29 Konopke R, Kersting S, Saeger H D. et al . Detection of liver lesions by contrast-enhanced ultrasound – comparison to intraoperative findings.  Ultraschall in Med. 2005;  26 107-113
  • 30 Albrecht T, Hoffmann C, Schmitz S. et al . Detection of liver Metastases: comparison of contrast-enhanced phase inversion ultrasound and dual phase spiral CT with intraoperative sonographic correlation.  Radiology. 2000;  207 459
  • 31 Wilson S R, Jang H J, Kim T K. et al . Enhancement patterns of focal liver masses: discordance between contrast-enhanced sonography and contrast-enhanced CT and MRI.  J Ultrasound Med. 2007;  26 145-1556
  • 32 Heide R, Strobel D. Kontrastmittelsonographie der Leber – Mittelklassegerät versus High-End-Gerät.  Ultraschall in Med. 2007;  28 S1-S34
  • 33 Murpy-Lavalle J, Jang H J, Kim T K. et al . Are metastases really hypovascular in the arterial phase? The perspective based on contrast-enhanced Ultrasonography.  J Ultrasound Med. 2007;  26 1545-1556
  • 34 Albrecht T. Contrast-Enhanced Ultrasound has come of Age.  Ultraschall in Med. 2008;  29 S4-S187
  • 35 Claudon M, Cosgrove D, Albrecht T. et al . Guidelines and good clinical practice recommendations for contrast-enhanced ultrasound (CEUS) – update 2008.  Ultraschall in Med. 2008;  29 28-44
  • 36 Romanini L, Passamonti M, Aiani L. Economic assessment of contrast-enhanced ultrasonography for evaluation of focal liver lesions: a multicentre Italian experience.  Eur Radiol. 2007;  17 S6 F 99-106
  • 37 Faccioli N, D’Onofrio M, Comai A. et al . Contrast-enhanced ultrasonography in the characterization of benign liver lesions: activity-based cost analysis.  Radiol Med. 2007;  112 810-820
  • 38 Nürnberg D, Jung A, Schmieder C. et al . Was kostet die Sonographie – Ergebnisse einer Kosten- und Prozessanalyse an einem Versorgungskrankenhaus.  Ultraschall in Med. 2008;  29 405-417
  • 39 Giesel F L, Delorme S, Ribbel R. et al . Contrast-Enhanced Ultrasound for the Characterization of Incidental Liver Lesions – An Economical Evaluation in Comparison with Multi-Phase Computed Tomography.  Ultraschall in Med. 2009;  30 259-268
  • 40 Brenner D J, Hall E J. Computed tomography – an increasing source of radiation exposure.  N Engl J Med. 2007;  357 2277-2284
  • 41 Thomsen H S, Morcos S K. Radiographic contrast media.  BJU Int. 2000;  86 1-10
  • 42 US Food and Drug Administration . Gadolinium-containing contrast agents for magnetic resonance imaging (MRI): Omniscan, OmniMARK, Magnevist, ProHance, and MultiHance.  FDA Alert,. 2006; 
  • 43 Dillmann J R, Ellis J H, Cohan R H. et al . Frequency and severity of acute allergic-like reactions to gadolinium-containing i. v. contrast media in children and adults.  Am J Roentgenol. 2007;  189 1533-1538
  • 44 Li A, Wong C S, Wong M K. et al . Acute adverse reactions to magnetic resonance contrast media – gadolinium chelates.  Br J Radiol. 2006;  79 368-371
  • 45 Jakobsen J A, Oyen R, Thomsen H S. et al . Safety of ultrasound contrast agents.  Eur Radiol. 2005;  15 941-945
  • 46 Piscaglia F, Bolondi L. Italian Society for Ultrasound in Medicine and Biology (SIUMB) Study Group on Ultrasound Contrast Agents . The Safety of Sonovue in abdominal applications: retrospective analysis of 23 188 investigations.  Ultrasound Med Biol. 2006;  32 1369-1375
  • 47 ter Haar G. Bubble Trouble?.  Ultraschall in Med. 2008;  29 550-551

PD Dr. Karlheinz Seitz

Medizinische Klinik Kreiskrankenhaus

Hohenzollernstr. 40

72488 Sigmaringen

Telefon: ++ 49/7 57 11 00 22 91

Fax: ++ 49/7 57 11 00 22 83

eMail: k.seitz@klksig.de

    >