Endurance and Performance of Two Different Concepts for Left Ventricular Stimulation with Bipolar Epicardial Leads in Long-Term Follow-Up[*]

H. Burger
1   Department of Cardiac Surgery, Kerckhoff-Klinik, Bad Nauheim, Germany
,
J. Kempfert
1   Department of Cardiac Surgery, Kerckhoff-Klinik, Bad Nauheim, Germany
,
A. van Linden
1   Department of Cardiac Surgery, Kerckhoff-Klinik, Bad Nauheim, Germany
,
Z. Szalay
1   Department of Cardiac Surgery, Kerckhoff-Klinik, Bad Nauheim, Germany
,
M. Schoenburg
1   Department of Cardiac Surgery, Kerckhoff-Klinik, Bad Nauheim, Germany
,
T. Walther
1   Department of Cardiac Surgery, Kerckhoff-Klinik, Bad Nauheim, Germany
,
T. Ziegelhoeffer
1   Department of Cardiac Surgery, Kerckhoff-Klinik, Bad Nauheim, Germany
› Institutsangaben
Weitere Informationen

Publikationsverlauf

17. Januar 2011

19. Mai 2011

Publikationsdatum:
25. Juli 2011 (online)

Abstract

Background Epicardial left ventricular (LV) leads represent an alternative for CRT therapy if transvenous lead implantation fails. Data on endurance, performance, the impact of the surgical approach (lateral minithoracotomy vs. median sternotomy simultaneously with other cardiac surgery), and the optimal technical concept (screw-in vs. suture-on) is limited.

Methods Over a period of 48 months we evaluated 130 consecutive patients with comparable characteristics. A total of 54 screw-in (MyoDex™ 1084T, SJM) and 76 suture-on (Capture Epi 4968, Medtronic) bipolar epicardial steroid-eluting LV leads were implanted either via a left lateral or a median thoracotomy. Sensing, pacing threshold, impedance and NYHA class were recorded at defined time points.

Results No surgery-related death or major complication was observed. At the time of implantation, the pacing threshold, sensing and NYHA class did not differ significantly between the two groups. The impedances of screw-in leads were significantly lower compared to those of suture-on leads. Suture-on leads showed a moderate initial drop in their pacing threshold but afterwards remained stable. Screw-in leads were characterized by a moderate but significant increase in the pacing threshold in the first year followed by a continuous decrease thereafter. Twenty-four months post-implantation no differences between both lead types could be detected. Sensing and NYHA class improved in both groups. The surgical approach had no significant impact on lead functionality.

Conclusion Our study showed that the implantation of epicardial leads was safe with very low complication rates. There was no superior technical epicardial lead concept (screw-in vs. suture-on leads) and all epicardial leads demonstrated an excellent long-term performance and durability. Therefore, it seems that epicardial leads represent a good alternative to transvenous leads and surgeons should be encouraged to implant epicardial leads during concomitant cardiac surgery when the indications for CRT are present.

* This paper was presented at the annual meeting of the German Society for Thoracic and Cardiovascular Surgery, Stuttgart, February 2010.


 
  • References

  • 1 St John Sutton MG, Plappert T, Abraham WT , et al. Effect of cardiac resynchronization therapy on left ventricular size and function in chronic heart failure. Circulation 2003; 107: 1985-1990
  • 2 Auricchio A, Stellbrink C, Sack S , et al. Long-term clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delay. J Am Coll Cardiol 2002; 39: 2026-2033
  • 3 Cazeau S, Leclercq C, Lavergne T , et al. Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. N Engl J Med 2001; 344: 873-880
  • 4 Abraham WT, Fisher WG, Smith AL , et al. Cardiac resynchronization in chronic heart failure. N Engl J Med 2002; 346: 1845-1853
  • 5 Young JB, Abraham WT, Smith AL , et al. Combined cardiac resynchronization and implantable cardioversion defibrillation in advanced chronic heart failure: the MIRACLE ICD Trial. JAMA 2003; 289: 2685-2694
  • 6 Bristow MR, Saxon LA, Boehmer J , et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 2004; 350: 2140-2150
  • 7 Cleland JG, Daubert JC, Erdmann E , et al. Longer-term effects of cardiac resynchronization therapy on mortality in heart failure [the CArdiac REsynchronization-Heart Failure (CARE-HF) trial extension phase]. Eur Heart J 2006; 27: 1928-1932
  • 8 Auricchio A, Stellbrink C, Butter C , et al. Clinical efficacy of cardiac resynchronization therapy using left ventricular pacing in heart failure patients stratified by severity of ventricular conduction delay. J Am Coll Cardiol 2003; 42: 2109-2116
  • 9 Lau EW. Achieving permanent left ventricular pacing-options and choice. PACE 2009; 32: 1466-1477
  • 10 Hansky B, Vogt J, Gueldner H , et al. Linksventrikuläre Stimulation und CRT: Welche Elektrode passt zu welcher Vene? [Left ventricular pacing and CRT. What CV lead fits into which vein?]. Herzschrittmacherther Elektrophysiol 2006; 17 (Suppl. 01) I14-I19
  • 11 Mizuno T, Tanaka H, Makita S, Tabuchi N, Arai H, Sunamori M. Biventricular pacing with coronary bypass and Dor's ventriculoplasty. Ann Thorac Surg 2003; 75 (3) 998-999
  • 12 Golzio PG, Anselmino M, Comoglio C, Vinci M, Rinaldi M, Trevi GP. Biventricular pacing concomitant to on-pump heart surgery: a case series. J Cardiovasc Surg (Torino) 2009; 50 (6) 801-806
  • 13 Mair H, Sachweh J, Meuris B , et al. Surgical epicardial left ventricular lead versus coronary sinus lead placement in biventricular pacing. Eur J Cardiothorac Surg 2005; 27: 235-242
  • 14 Doll N, Opfermann UT, Rastan AJ , et al. Facilitated minimally invasive left ventricular epicardial lead placement. Ann Thor Surg 2005; 79 (3) 1023-1025
  • 15 Saggau W, Sack FU, Lange R , et al. Superiority of endocardial versus epicardial implantation of the implantable cardioverter defibrillator (ICD). Eur J Cardiothorac Surg 1992; 6 (4) 195-200
  • 16 Doll N, Piorkowski C, Czesla M , et al. Epicardial versus transvenous left ventricular lead placement in patients receiving cardiac resynchronization therapy: results from a randomized prospective study. Thorac Cardiovasc Surg 2008; 56 (5) 256-261
  • 17 Tomaske M, Gerritse B, Kretsers L , et al. A 12-year experience of bipolar steroid-eluting epicardial pacing leads in children. Ann Thorac Surg 2008; 85: 1704-1711
  • 18 Romeyer-Bouchard C, Da Costa A, Dauphinot V , et al. Prevalence and risk factors related to infections of cardiac resynchronization therapy devices. Eur Heart J 2010; 31 (2) 203-210