ABSTRACT

We evaluated interference screw fixation in a plug-tunnel construct using bioabsorbable screws as void fillers with different percentages of the screw removed. Nine-millimeter tunnels in a closed-cell foam block were filled with a 10-mm bioabsorbable screw, and 10-mm revision tunnels were placed in parallel with tunnel overlap resulting in removal of 10%, 25%, or 50% of the screw diameter. Synthetic bone plugs were fashioned to fit 10-mm tunnels. In all groups, the plugs were secured in standard interference fixation with a 9-mm metal screw between the void-filling bioabsorbable screw and plug. Failure loads for the control group (no revision tunnel) averaged 926 ± 44 N, 10% (1024 ± 129 N) and 25% (932 ± 129 N) groups were not significantly different; failure load in the 50% diameter group (780 ± 72 N) was significantly lower than all other groups (p < 0.001). Using a bioabsorbable screw as void filler provided mean load to failure not different from that of standard reconstruction when 10 and 25% of the diameter of the void-filling screw was removed. Load to failure was significantly lower when 50% of the void-filling screw diameter group was removed. This may be applicable in anterior cruciate ligament reconstruction where a previous tunnel void has to be addressed.

REFERENCES

  • 1 Brown Jr C H, Carson E W. Revision anterior cruciate ligament surgery.  Clin Sports Med. 1999;  18 (1) 109-171
  • 2 Johnson D L, Coen M J. Revision ACL surgery. Etiology, indications, techniques, and results.  Am J Knee Surg. 1995;  8 (4) 155-167
  • 3 Battaglia T C, Miller M D. Management of bony deficiency in revision anterior cruciate ligament reconstruction using allograft bone dowels: surgical technique.  Arthroscopy. 2005;  21 (6) 767
  • 4 Greis P E, Johnson D L, Fu F H. Revision anterior cruciate ligament surgery: causes of graft failure and technical considerations of revision surgery.  Clin Sports Med. 1993;  12 (4) 839-852
  • 5 O'Neill D B. Arthroscopically assisted reconstruction of the anterior cruciate ligament. A follow-up report.  J Bone Joint Surg Am. 2001;  83-A (9) 1329-1332
  • 6 Azar F M. Revision anterior cruciate ligament reconstruction.  Instr Course Lect. 2002;  51 335-342
  • 7 Wolf R S, Lemak L J. Revision anterior cruciate ligament reconstruction surgery.  J South Orthop Assoc. 2002;  11 (1) 25-32
  • 8 Noyes F R, Barber-Westin S D. Revision anterior cruciate ligament reconstruction: report of 11-year experience and results in 114 consecutive patients.  Instr Course Lect. 2001;  50 451-461
  • 9 Herbenick M, Gambardella R. Revision anterior cruciate ligament reconstruction using a unique bioabsorbable interference screw for malpositioned tunnels.  Am J Orthop. 2008;  37 (8) 425-428
  • 10 Allen C R, Giffin J R, Harner C D. Revision anterior cruciate ligament reconstruction.  Orthop Clin North Am. 2003;  34 (1) 79-98
  • 11 Getelman M H, Friedman M J. Revision anterior cruciate ligament reconstruction surgery.  J Am Acad Orthop Surg. 1999;  7 (3) 189-198
  • 12 Jaureguito J W, Paulos L E. Why grafts fail.  Clin Orthop Relat Res. 1996;  325 (325) 25-41
  • 13 Kaplan K S, Hayes W C, Keaveny T M et al.. Form and function of bone. In: Simon S R, ed. Orthopaedic Basic Science. Rosemont, IL: American Academy of Orthopaedic Surgeons; 1994: 127-184
  • 14 Firoozbakhsh K K, Moneim M S, DeCoster T A. Pullout strength of power- and hand-driven staples in synthetic bone: effect of design parameters.  J Orthop Trauma. 1992;  6 (1) 43-49
  • 15 McNamara B P, Cristofolini L, Toni A, Taylor D. Evaluation of experimental and finite element models of synthetic and cadaveric femora for pre-clinical design-analysis.  Clin Mater. 1994;  17 (3) 131-140
  • 16 Lemos M J, Jackson D W, Lee T Q, Simon T M. Assessment of initial fixation of endoscopic interference femoral screws with divergent and parallel placement.  Arthroscopy. 1995;  11 (1) 37-41
  • 17 Kousa P, Järvinen T L, Kannus P, Järvinen M. Initial fixation strength of bioabsorbable and titanium interference screws in anterior cruciate ligament reconstruction. Biomechanical evaluation by single cycle and cyclic loading.  Am J Sports Med. 2001;  29 (4) 420-425
  • 18 Harvey A, Thomas N P, Amis A A. Fixation of the graft in reconstruction of the anterior cruciate ligament.  J Bone Joint Surg Br. 2005;  87 (5) 593-603
  • 19 Rowden N J, Sher D, Rogers G J, Schindhelm K. Anterior cruciate ligament graft fixation. Initial comparison of patellar tendon and semitendinosus autografts in young fresh cadavers.  Am J Sports Med. 1997;  25 (4) 472-478
  • 20 Kurosaka M, Yoshiya S, Andrish J T. A biomechanical comparison of different surgical techniques of graft fixation in anterior cruciate ligament reconstruction.  Am J Sports Med. 1987;  15 (3) 225-229
  • 21 Matthews L S, Lawrence S J, Yahiro M A, Sinclair M R. Fixation strengths of patellar tendon-bone grafts.  Arthroscopy. 1993;  9 (1) 76-81
  • 22 Seil R, Rupp S, Krauss P W, Benz A, Kohn D M. Comparison of initial fixation strength between biodegradable and metallic interference screws and a press-fit fixation technique in a porcine model.  Am J Sports Med. 1998;  26 (6) 815-819
  • 23 Woo S L, Gomez M A, Woo Y K, Akeson W H. Mechanical properties of tendons and ligaments. II. The relationships of immobilization and exercise on tissue remodeling.  Biorheology. 1982;  19 (3) 397-408
  • 24 Pavlik A, Hidas P, Czigány T, Berkes I. Biomechanical evaluation of press-fit femoral fixation technique in ACL reconstruction.  Knee Surg Sports Traumatol Arthrosc. 2004;  12 (6) 528-533

Richard HintonM.D. 

Department of Orthopaedics, Union Memorial Hospital, 3333 N. Calvert Street

Suite 400, Baltimore, MD 21218

Email: lyn.camire@medstar.net

    >