Geburtshilfe Frauenheilkd 2013; 73(5): 422-427
DOI: 10.1055/s-0032-1328600
GebFra Science
Original Article
Georg Thieme Verlag KG Stuttgart · New York

Full Field Digital Mammography (FFDM) versus CMOS Technology, Specimen Radiography System (SRS) and Tomosynthesis (DBT) – Which System Can Optimise Surgical Therapy?

Digitale Vollfeldmammografie (FFDM), CMOS-Technologie, Specimen Radiography System (SRS) versus Tomosynthese (DBT) – welches System optimiert die operative Therapie?
R. Schulz-Wendtland
1   Gynäkologische Radiologie, Radiologisches Institut, Universitätsklinikum Erlangen, Erlangen
,
G. Dilbat
2   Radiologische Praxis Roth und Weißenburg, Roth
,
M. Bani
3   Frauenklinik, Universitätsklinikum Erlangen, Erlangen
,
P. A. Fasching
3   Frauenklinik, Universitätsklinikum Erlangen, Erlangen
,
K. Heusinger
3   Frauenklinik, Universitätsklinikum Erlangen, Erlangen
,
M. P. Lux
3   Frauenklinik, Universitätsklinikum Erlangen, Erlangen
,
C. R. Loehberg
3   Frauenklinik, Universitätsklinikum Erlangen, Erlangen
,
B. Brehm
1   Gynäkologische Radiologie, Radiologisches Institut, Universitätsklinikum Erlangen, Erlangen
,
M. Hammon
1   Gynäkologische Radiologie, Radiologisches Institut, Universitätsklinikum Erlangen, Erlangen
,
M. Saake
1   Gynäkologische Radiologie, Radiologisches Institut, Universitätsklinikum Erlangen, Erlangen
,
P. Dankerl
1   Gynäkologische Radiologie, Radiologisches Institut, Universitätsklinikum Erlangen, Erlangen
,
S. M. Jud
3   Frauenklinik, Universitätsklinikum Erlangen, Erlangen
,
C. Rauh
3   Frauenklinik, Universitätsklinikum Erlangen, Erlangen
,
C. M. Bayer
3   Frauenklinik, Universitätsklinikum Erlangen, Erlangen
,
M. W. Beckmann
3   Frauenklinik, Universitätsklinikum Erlangen, Erlangen
,
M. Uder
4   Radiologisches Institut, Universitätsklinikum Erlangen, Erlangen
,
M. Meier-Meitinger
1   Gynäkologische Radiologie, Radiologisches Institut, Universitätsklinikum Erlangen, Erlangen
› Author Affiliations
Further Information

Publication History

received 12 March 2013
revised 16 April 2013

accepted 16 April 2013

Publication Date:
06 June 2013 (online)

Abstract

Aim: This prospective clinical study aimed to evaluate whether it would be possible to reduce the rate of re-excisions using CMOS technology, a specimen radiography system (SRS) or digital breast tomosynthesis (DBT) compared to a conventional full field digital mammography (FFDM) system. Material and Method: Between 12/2012 and 2/2013 50 patients were diagnosed with invasive breast cancer (BI-RADS™ 5). After histological verification, all patients underwent breast-conserving therapy with intraoperative imaging using 4 different systems and differing magnifications: 1. Inspiration™ (Siemens, Erlangen, Germany), amorphous selenium, tungsten source, focus 0.1 mm, resolution 85 µm pixel pitch, 8 lp/mm; 2. BioVision™ (Bioptics, Tucson, AZ, USA), CMOS technology, photodiode array, flat panel, tungsten source, focus 0.05, resolution 50 µm pixel pitch, 12 lp/mm; 3. the Trident™ specimen radiography system (SRS) (Hologic, Bedford, MA, USA), amorphous selenium, tungsten source, focus 0.05, resolution 70 µm pixel pitch, 7.1 lp/mm; 4. tomosynthesis (Siemens, Erlangen, Germany), amorphous selenium, tungsten source, focus 0.1 mm, resolution 85 µm pixel pitch, 8 lp/mm, angular range 50 degrees, 25 projections, scan time > 20 s, geometry: uniform scanning, reconstruction: filtered back projection. The 600 radiographs were prospectively shown to 3 radiologists. Results: Of the 50 patients with histologically proven breast cancer (BI-RADS™ 6), 39 patients required no further surgical therapy (re-excision) after breast-conserving surgery. A retrospective analysis (n = 11) showed a significant (p < 0.05) increase of sensitivity with the BioVision™, the Trident™ and tomosynthesis compared to the Inspiration™ at a magnification of 1.0 : 2.0 or 1.0 : 1.0 (tomosynthesis) (2.6, 3.3 or 3.6 %), i.e. re-excision would not have been necessary in 2, 3 or 4 patients, respectively, compared to findings obtained with a standard magnification of 1.0 : 1.0. Conclusion: The sensitivity of the BioVision™, the Trident™ and tomosynthesis was significantly (p < 0.05) higher and the rate of re-excisions was reduced compared to FFDM using a conventional detector at a magnification of 2.0 but without zooming.

Zusammenfassung

Ziel: Prospektive Untersuchung, ob mit Hilfe der CMOS-Technologie, dem Specimen Radiography System bzw. der digitalen Brusttomosynthese (DBT) im Vergleich mit einem herkömmlichen FFDM-System eine Senkung der Rate an Reexzisionen möglich ist. Material und Methode: In der Zeit von 12/2012 bis 2/2013 wurde bei 50 Patientinnen ein invasives Mammakarzinom (BI-RADS™ 5) diagnostiziert. Nach histologischer Sicherung erfolgte die brusterhaltende Therapie mit intraoperativem Präparateradiogramm mit 4 unterschiedlichen Systemen und Vergrößerungen: 1. Inspiration™ (Siemens, Erlangen, Deutschland), amorphes Selen, W-Anode, Fokus 0,1 mm, Ortsauflösung 85 µm Pixelpitch, 8 Lp/mm. 2. BioVision™ (Bioptics, Tucson, AZ, USA), CMOS-Technologie, Photodioden Array, Flat Panel, W-Anode, Fokus 0,05, Ortsauflösung 50 µm Pixelpitch, 12 Lp/mm. 3. Trident™ (Hologic, Bedford, MA, USA), Specimen Radiography System (SRS System), amorphes Selen, W-Anode, Fokus 0,05, Ortsauflösung 70 µm Pixelpitch, 7,1 Lp/mm. 4. Tomosynthese (Siemens, Erlangen, Deutschland), amorphes Selen, W-Anode, Fokus 0,1 mm, Ortsauflösung 85 µm Pixelpitch, 8 Lp/mm, Winkelbereich 50 Grad, 25 Projektionen, Scandauer > 20 s, Geometrie: gleichförmige Abtastung, Rekonstruktion: gefilterte Rückprojektion. Die jeweils 600 Präparateradiogramme wurden prospektiv 3 Radiologen auf einem Befundungsmonitor präsentiert. Ergebnisse: Bei den 50 Patientinnen mit histologisch gesichertem Mammakarzinom (BI-RADS™ 6) war bei 39 Patientinnen nach erfolgter brusterhaltender Therapie keine weitere operative Therapie (Reexzision) indiziert. Die retrospektive Analyse (n = 11) ergab eine signifikante (p < 0,05) Steigerung der Sensitivität des BioVision™-, des Trident™-Systems und der Tomosynthese gegenüber dem Inspiration™ bei einer Vergrößerung von 1,0 : 2,0 bzw. 1,0 : 1,0 (Tomosynthese) (2,6, 3,3 bzw. 3,6 %) d. h. eine Reexzision wäre bei 2, 3 bzw. 4 Patientinnen nicht notwendig gewesen im Vergleich zur Standardvergrößerung 1,0 : 1,0. Schlussfolgerung: Sowohl der Einsatz der Präparateradiografie-Systeme BioVision™ und Trident™, aber auch die Tomosynthese zur Präparateradiografie erreichen eine signifikant (p < 0,05) bessere Sensitivität und Reduzierung der Reexzisionsrate im Vergleich zu einem FFDM-System mit herkömmlichem Detektor unter der Voraussetzung einer reellen Vergrößerung mit dem Faktor 2,0, jedoch keinem Zooming.

 
  • References

  • 1 Ruckhaberle E, Solbach C, Kaufmann M. Conference report on the 31th annual San Antonio breast cancer symposium, December 10–14, 2008 “Continuing change in diagnostics and therapy of breast cancers – from molecular basics to tailored therapy”. Geburtsh Frauenheilk 2009; 69: 202-211
  • 2 Kaufmann M, Rody A. Breast cancer: reduced mortality by early detection and adjuvant therapy. Geburtsh Frauenheilk 2009; 69: 218-232
  • 3 Katalinic A. Breast cancer: declining mortality despite its increasing incidence. Geburtsh Frauenheilk 2009; 69: 237-239
  • 4 Beckmann MW, Blohmer JU, Costa SD et al. Zurich consensus: German expert opinion on the St. Gallen votes on 15 March 2009 (11th international conference at St. Gallen: primary therapy of early breast cancer). Geburtsh Frauenheilk 2009; 69: 377-383
  • 5 Ruckhaberle E, Rody A, Kaufmann M. Report of the 32nd annual St. Antonio breast cancer symposium, December 10–13, 2009 “An international scientific symposium for interaction and exchange among basic scientists and clinicians in breast cancer”. Geburtsh Frauenheilk 2010; 70: 177-183
  • 6 Fasching PA, Fehm T, Janni W et al. Breast cancer therapy – a state of the art review. Geburtsh Frauenheilk 2010; 70: 875-886
  • 7 Fasching PA, Ekici AB, Adamietz BR et al. Breast cancer risk – genes, environment and clinics. Geburtsh Frauenheilk 2011; 71: 1056-1066
  • 8 Perry NM. Quality assurance in the diagnosis of breast disease. Eur J Cancer 2001; 37: 159-172
  • 9 Wallis M, Tardivon A, Helbich T et al. Guidelines from the European Society of Breast Imaging for diagnostic interventional breast procedures. Eur Radiol 2007; 17: 581-588
  • 10 Albert US Hrsg. S3-Leitlinie Brustkrebs-Früherkennung in Deutschland. Aktualisierung 2012. München: W. Zuckschwerdt-Verlag; 2012. http://www.senologie.org last access: Juli 2012
  • 11 Kreienberg R, Kopp I, Albert US et al. S3-Leitlinie Diagnostik, Therapie und Nachsorge des Mammakarzinoms. 1. Aktualisierung. München: W. Zuckschwerdt-Verlag; 2008. http://www.senologie.org last access: Juli 2008
  • 12 Bani MR, Lux MP, Heusinger K et al. Factors correlating with reexcision after breast-conserving therapy. EJSO 2009; 35: 32-37
  • 13 DʼOrsi CJ. Management of the breast specimen. Radiology 1995; 194: 297-302
  • 14 Grunert J-H, Wiechmann U, Gmelin E. Kompressionsfreie Mammapräparateradiographie in 2 Ebenen mit Hilfe eines Radiographiecontainers. Fortschr Röntgenstr 2003; 175: 413-417
  • 15 Kürzl R, Baltzer J, Lohe KJ. Intraoperative specimen radiography in mammographically suspected, non palpable breast lesions. Experiences with the Faxitron unit. Fortschr Med 1979; 97: 1688-1690
  • 16 Kürzl R, Lohe KJ, Baltzer J et al. Radiography used during operations for biopsies of non-palpable lesions of the breast which have been suspected because of mammography. J Gynecol Obstet Biol Reprod 1981; 10: 431-434
  • 17 Moritz JD, Westerhof JP, Funke M et al. Ein neues Präparatradiographiegerät mit maximal 20facher Vergrößerung für die Mammadiagnostik. Fortschr Röntgenstr 1998; 168: 133-138
  • 18 Grunert JH, Borchert B, Farber A et al. Mikrokalknachweis in Mamma-Exzisionspräparaten mit Hilfe der DIMA-Vergrößerungspräparateradiographie vierfach im Vergleich zur konventionellen Vergrößerungsradiographie eineinhalbfach. Fortschr Röntgenstr 1999; 171: 302-306
  • 19 Grunert J-H, Flemming P, Borchert B et al. Vergrößerungspräparateradiographie als Qualitätskontrolle der Mikrokalkerfassung im Rahmen der histopathologischen Aufarbeitung von Mammabiopsiegewebe. Fortschr Röntgenstr 2000; 172: 68-72
  • 20 Diekmann F, Grebe S, Bick U et al. Digitale intraoperative Präparateradiographie bei diagnostischen Exzisionen nicht palpabler Läsionen der Brust. Fortschr Röntgenstr 2000; 172: 969-971
  • 21 Muttalib M, Tisdall M, Scawn R et al. Intra-operative specimen analysis using faxitron microradiography for excision of mammographically suspicious, non-palpable breast lesions. The Breast 2004; 13: 307-315
  • 22 Schulz-Wendtland R, Wenkel E, Wacker T et al. Quo vadis? Trends in digital mammography. Geburtsh Frauenheilk 2009; 69: 108-117
  • 23 Schulz-Wendtland R, Bani MR, Lux MP et al. CMOS-Technologie für intraoperative digitale Präparateradiografien. Geburtsh Frauenheilk 2011; 71: 525-529
  • 24 Schulz-Wendtland R, Dilbat G, Bani MR et al. Use of tomosynthesis in intraoperative digital specimen radiography – is a reduction of breast re-excision rates possible?. Geburtsh Frauenheilk 2011; 71: 1080-1084
  • 25 Niklason LT, Christian BT, Niklason LE et al. Digital tomosynthesis in breast imaging. Radiology 1997; 205: 399-406
  • 26 Schulz-Wendtland R, Wenkel E. Experimental phantom lesion detectability study using a digital breast tomosynthesis prototype system. Röfo 2006; 178: 1219-1223
  • 27 Poplack SP, Tosteson TD, Kogel CA et al. Digital breast tomosynthesis: initial experience in 98 women with abnormal digital screening mammography. AJR 2007; 189: 616-623
  • 28 Good WF, Abrams GS, Catullo VJ et al. Digital breast tomosynthesis: a pilot observer study. AJR 2008; 190: 865-869
  • 29 Gur D, Abrams GS, Chough DM et al. Digital breast tomosynthesis: observer performance study. AJR 2009; 193: 586-591
  • 30 Kontos D, Bakic PR, Carton AK et al. Parenchymal texture analysis in digital breast tomosynthesis for breast cancer risk estimation: a preliminary study. Acad Radiol 2009; 16: 283-298
  • 31 Schulz-Wendtland R, Fuchsjäger M, Wacker T et al. Digital mammography: an update. Eur J Radiol 2009; 72: 258-265
  • 32 Teertstra HJ, Loo CE, van den Bosch MA et al. Breast tomosynthesis in clinical practice: initial results. Eur Radiol 2010; 20: 16-24
  • 33 Schulz-Wendtland R, Hermann K-P, Wenkel E et al. Vergleich der klinischen Wertigkeit von Zielaufnahmen mit reduzierter Pixelgröße des Detektors und digitalem Monitorzooming bei der Beurteilung von Mikrokalzifikationen in der digitalen Mammographie. Fortschr Röntgenstr 2007; 179: 492-496
  • 34 Haygood TM, Arribas E, Brennan PC et al. Conspicuity of microcalcifications on digital screening mammograms using varying degrees of monitor zooming. Acad Radiol 2009; 16: 1509-1517
  • 35 Kim MJ, Youk JH, Kang DR et al. Zooming method (x 2.0) of digital mammography vs. digital magnification view (x 1.8) in full-field digital mammography for the diagnosis of microcalcifications. Br J Radiol 2010; 83: 486-492
  • 36 Koutalonis M, Delis H, Pascoal A et al. Can electronic zoom replace magnification in mammography?. BJR 2010; 83: 569-577
  • 37 Schulz-Wendtland R, Dilbat G, Bani M et al. Full field mammography (FFDM) versus CMOS technology versus tomosynthesis (DBT) – which system increases the quality of intraoperative imaging?. Geburtsh Frauenheilk 2012; 72: 532-538
  • 38 Schulz-Wendtland R, Bani MR, Lux MP et al. Pilot study on the detection of simulated lesions using a 2D and 3D digital full-field mammography system with a newly developed high resolution detector based on two shifts of a-Se. Geburtsh Frauenheilk 2012; 72: 408-411