Thromb Haemost 1998; 79(01): 122-128
DOI: 10.1055/s-0037-1614246
Review Article
Schattauer GmbH

The European Concerted Action on Anticoagulation (ECAA) Evaluation of a Set of Lyophilized Normal Plasmas to Establish the Normal Prothrombin Time for Coagulometer Systems

L. Poller
1   From the Central Facility, ECAA, Department of Pathological Sciences, The University of Manchester, UK
,
T. W. Barrowcliffe
2   From the Division of Haematology, National Institute for Biological Standards and Control, UK
,
A. M. H. P. van den Besselaar
3   From the Haemostasis and Thrombosis Research Centre, Leiden University Hospital, The Netherlands
,
J. Jespersen
4   From the Department of Clinical Biochemistry, Centralsygehuset, Esbjerg, Denmark
,
A. Tripodi
5   From the A. Bianchi Bonomi, Hemophilia and Thrombosis Centre, University & IRCCS Maggiore Hospital, Milan, Italy
,
D. Houghton
1   From the Central Facility, ECAA, Department of Pathological Sciences, The University of Manchester, UK
,
ECAA by the above steering committee › Author Affiliations
Further Information

Publication History

Received 26 February 1997

Accepted after resubmission 11 August 1997

Publication Date:
08 December 2017 (online)

Summary

Establishing the mean normal prothrombin time (MNPT) from fresh samples for prothrombin ratios and INR often presents difficulty in selection and collection of donors. A set of seven lyophilized normal plasmas has therefore been prepared at the ECAA Central Facility and studied at 143 laboratories in sixteen European states using coagulometers in serial field exercises. All centres tested either the high ISI ECAA rabbit or low ISI ECAA human reference thromboplastin. The MNPT of fresh plasmas and means of the lyophilized samples were closely comparable with most routine rabbit thromboplastins. Using human thromboplastins means with the lyophilized normals were marginally but significantly longer and with the bovine Thrombotest significantly shorter than the MNPT of fresh plasmas causing alterations in INR. There was no appreciable effect on INR of 2.5 and 3.5 when lyophilized normals were substituted for fresh normals with the rabbit reagents.

 
  • References

  • 1 WHO Expert Committee on Biological Standardisation.. 33rd Report, WHO Technical Series. 1983 687. 81-105.
  • 2 van den Besselaar AMHP. The use of lyophilized calibration plasmas and control blood for INR calculation on external quality control of the prothrombin time.. Am J Clin Pathol 1993; 102: 123-7.
  • 3 European Concerted Action on Anticoagulation (ECAA).. Poller L, Barrowcliffe TW, van den Besselaar AHMP, Jespersen J, Tripodi A, Houghton D. Minimum Lyophilized Plasma Requirement for ISI Calibration.. Am J Clin Path 1998; in press.
  • 4 Poller L, Barrowcliffe TW, van den Besselaar AHMP, Jespersen J, Tripodi A, Houghton D. European Concerted Action on Anticoagulation (ECAA). The Multicentre Calibration of Rabbit and Human ECAA Reference Thromboplastins.. Thromb Haemost 1996; 76: 977-82.
  • 5 Poller L, Houghton D, Carroll J. The reliability of the mean normal prothrombin time of fresh plasmas and of the normal value from a lyophilized “normal” plasma in prothrombin ratio determination.. Brit J Haemat 1994; 88: 866-73.
  • 6 Peters RHM, van den Besselaar AMHP, Olthuis FMFG. Determination of the Mean Normal Prothrombin Time for Assessment of International Normalized Ratios. Usefulness of Lyophilized Plasma.. Thromb Haemost 1991; 66: 442-5.