CC BY-NC-ND 4.0 · Eur J Dent 2019; 13(04): 556-562
DOI: 10.1055/s-0039-1698369
Original Article
Dental Investigation Society

In Vivo Evaluation of Operative Torque Generated by Two Nickel-Titanium Rotary Instruments during Root Canal Preparation

Gianluca Gambarini
1   Department of Oral and Maxillo-Facial Sciences, Sapienza University of Rome, Italy
,
Massimo Galli
1   Department of Oral and Maxillo-Facial Sciences, Sapienza University of Rome, Italy
,
Marco Seracchiani
1   Department of Oral and Maxillo-Facial Sciences, Sapienza University of Rome, Italy
,
Dario Di Nardo
1   Department of Oral and Maxillo-Facial Sciences, Sapienza University of Rome, Italy
,
Marco A. Versiani
2   Department of Restorative Dentistry, Dental School of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil
,
Lucila Piasecki
3   Department of Periodontics and Endodontics, University at Buffalo, Buffalo
,
Luca Testarelli
1   Department of Oral and Maxillo-Facial Sciences, Sapienza University of Rome, Italy
› Institutsangaben
Weitere Informationen

Publikationsverlauf

Publikationsdatum:
31. Dezember 2019 (online)

Abstract

Objectives This in vivo study evaluated the operative torque and preparation time of ProTaper NEXT (Dentsply Maillefer; Ballaigues, Switzerland) and EdgeFile X7 (EdgeEndo; Albuquerque, New Mexico, United States) rotary systems during root canal preparation of maxillary premolars.

Materials and Methods Ten double-rooted maxillary premolars with independent canals were selected. Each canal in each tooth was prepared with one of the rotary systems (n = 10), ProTaper NEXT or EdgeFile X7. The instruments were rotated at 300 rpm with maximum torque set at 2 N.cm using an electric motor (KaVo; Biberach, Germany) that automatically recorded torque values at every 1/10th of a second (ds).

Statistical Analysis Operative torque (N.cm) and preparation time (s) of the first shaping instrument (size 17/.04) of both rotary systems were recorded and statistically compared using the Mann–Whiney U test with a significance level set at 5%.

Results No instrument exhibited flute deformation or underwent intracanal failure. No differences were found between the instruments regarding the maximum (peak) torque values (p > 0.05). EdgeFile X7 17/.04 required significantly less preparation time (3.75 seconds interquartile range [IQR]: 3.2–9.0) than ProTaper NEXT X1 (15.45 seconds IQR: 8.35–21.1) (p < 0.05). The median operative torque values of ProTaper NEXT X1 (0.26 N.cm; IQR: 0.18–0.49) were significantly higher compared with EdgeFile X7 17/.04 (0.09 N.cm; IQR: 0.05–0.17) (p < 0.05).

Conclusions Although no difference was found between the median peak torque values of ProTaper NEXT X1 and EdgeFile X7 17/.04 instruments, the operative torque and instrumentation time results were impacted by their different designs and alloys during clinical preparation of root canals.

 
  • References

  • 1 Plotino G, Grande NM, Mercadé Bellido M, Testarelli L, Gambarini G. Influence of temperature on cyclic fatigue resistance of ProTaper Gold and ProTaper Universal Rotary files. J Endod 2017; 43 (02) 200-202
  • 2 Testarelli L, Plotino G, Al-Sudani D. et al. Bending properties of a new nickel-titanium alloy with a lower percent by weight of nickel. J Endod 2011; 37 (09) 1293-1295
  • 3 Pedullà E, Grande NM, Plotino G, Gambarini G, Rapisarda E. Influence of continuous or reciprocating motion on cyclic fatigue resistance of 4 different nickel-titanium rotary instruments. J Endod 2013; 39 (02) 258-261
  • 4 Gambarini G, Plotino G, Piasecki L, Al-Sudani D, Testarelli L, Sannino G. Deformations and cyclic fatigue resistance of nickel-titanium instruments inside a sequence. Ann Stomatol (Roma) 2015; 6 (01) 6-9
  • 5 Gao Y, Shotton V, Wilkinson K, Phillips G, Johnson WB. Effects of raw material and rotational speed on the cyclic fatigue of ProFile Vortex rotary instruments. J Endod 2010; 36 (07) 1205-1209
  • 6 Dentsply Sirona. ProTaper NEXT directions for use. Available at: http://www.dentsplymaillefer.com/product-category/glide-path-shaping/protaper-next Accessed July 20th 2018
  • 7 EdgeEndo. EdgeFile X7 directions for use. Available at: https://edgeendo.com/wp-content/uploads/2015/08/DFU-EdgeFile-x7.pdf Accessed July 20th 2018
  • 8 Dosanjh A, Paurazas S, Askar M. The effect of temperature on cyclic fatigue of nickel-titanium rotary endodontic instruments. J Endod 2017; 43 (05) 823-826
  • 9 Elnaghy AM. Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary files. Int Endod J 2014; 47 (11) 1034-1039
  • 10 Nguyen HH, Fong H, Paranjpe A, Flake NM, Johnson JD, Peters OA. Evaluation of the resistance to cyclic fatigue among ProTaper Next, ProTaper Universal, and Vortex Blue rotary instruments. J Endod 2014; 40 (08) 1190-1193
  • 11 Yared GM, Bou Dagher FE, Machtou P. Cyclic fatigue of ProFile rotary instruments after clinical use. Int Endod J 2000; 33 (03) 204-207
  • 12 Lee W, Song M, Kim E, Lee H, Kim HC. A survey of experience-based preference of Nickel-Titanium rotary files and incidence of fracture among general dentists. Restor Dent Endod 2012; 37 (04) 201-206
  • 13 Iqbal MK, Kohli MR, Kim JS. A retrospective clinical study of incidence of root canal instrument separation in an endodontics graduate program: a PennEndo database study. J Endod 2006; 32 (11) 1048-1052
  • 14 Gambarini G, Piasecki L, Di Nardo D. et al. Incidence of deformation and fracture of Twisted File Adaptive instruments after repeated clinical use. J Oral Maxillofac Res 2016; 7 (04) e5
  • 15 Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod 2000; 26 (03) 161-165
  • 16 Wu J, Lei G, Yan M, Yu Y, Yu J, Zhang G. Instrument separation analysis of multi-used ProTaper Universal rotary system during root canal therapy. J Endod 2011; 37 (06) 758-763
  • 17 Gambarini G, Seracchiani M, Piasecki L. et al. Measurement of torque generated during intra-canal instrumentation in vivo . Int Endod J 2019; 52 (05) 737-745
  • 18 Gambarini G, Seracchiani M, Piasecki L. et al. The effect of a brushing motion inside a sequence; an in vivo study. Ann Stomatol (Roma) 2018; 9: 72-76
  • 19 Gambarini G, Tucci E, Bedini R. et al. The effect of brushing motion on the cyclic fatigue of rotary nickel titanium instruments. Ann Ist Super Sanita 2010; 46 (04) 400-404
  • 20 Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol 1971; 32 (02) 271-275
  • 21 Yared GM, Bou Dagher FE, Machtou P. Influence of rotational speed, torque and operator’s proficiency on ProFile failures. Int Endod J 2001; 34 (01) 47-53
  • 22 Schrader C, Peters OA. Analysis of torque and force with differently tapered rotary endodontic instruments in vitro. J Endod 2005; 31 (02) 120-123
  • 23 Blum JY, Machtou P, Micallef JP. Location of contact areas on rotary Profile instruments in relationship to the forces developed during mechanical preparation on extracted teeth. Int Endod J 1999; 32 (02) 108-114
  • 24 Jamleh A, Komabayashi T, Ebihara A. et al. Root surface strain during canal shaping and its influence on apical microcrack development: a preliminary investigation. Int Endod J 2015; 48 (12) 1103-1111
  • 25 Sattapan B, Palamara JE, Messer HH. Torque during canal instrumentation using rotary nickel-titanium files. J Endod 2000; 26 (03) 156-160
  • 26 Baek SH, Lee CJ, Versluis A, Kim BM, Lee W, Kim HC. Comparison of torsional stiffness of nickel-titanium rotary files with different geometric characteristics. J Endod 2011; 37 (09) 1283-1286
  • 27 Gambarini G, Testarelli L, Milana V. et al. Angular deflection of rotary nickel titanium files: a comparative study. Ann Ist Super Sanita 2009; 45 (04) 423-426
  • 28 Boessler C, Paque F, Peters OA. The effect of electropolishing on torque and force during simulated root canal preparation with ProTaper shaping files. J Endod 2009; 35 (01) 102-106
  • 29 Gambarini G, Giansiracusa Rubini A, Sannino G. et al. Cutting efficiency of nickel-titanium rotary and reciprocating instruments after prolonged use. Odontology 2016; 104 (01) 77-81
  • 30 Gambarini G, Pompa G, Di Carlo S, De Luca M, Testarelli L. An initial investigation on torsional properties of nickel-titanium instruments produced with a new manufacturing method. Aust Endod J 2009; 35 (02) 70-72
  • 31 Dane A, Capar ID, Arslan H, Akçay M, Uysal B. Effect of different torque settings on crack formation in root dentin. J Endod 2016; 42 (02) 304-306
  • 32 Kim HC, Cheung GS, Lee CJ, Kim BM, Park JK, Kang SI. Comparison of forces generated during root canal shaping and residual stresses of three nickel-titanium rotary files by using a three-dimensional finite-element analysis. J Endod 2008; 34 (06) 743-747
  • 33 Gambarini G. Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors. J Endod 2001; 27 (12) 772-774
  • 34 Kwak SW, Ha JH, Cheung GS, Kim HC, Kim SK. Effect of the glide path establishment on the torque generation to the files during instrumentation: an in vitro measurement. J Endod 2018; 44 (03) 496-500
  • 35 Gambarini G, Testarelli L, Galli M, Tucci E, De Luca M. The effect of a new finishing process on the torsional resistance of twisted nickel-titanium rotary instruments. Minerva Stomatol 2010; 59 (07) (08) 401-406
  • 36 Yared G, Bou Dagher F, Kulkarni K. Influence of torque control motors and the operator’s proficiency on ProTaper failures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2003; 96 (02) 229-233