The perimodiolar CI532-electrode (Cochlear Ltd.) enables the preservation of high levels of residual hearing due to a low diameter and flexible electrode design. This study’s aim was to assess the levels of frequency-dependant residual hearing after implantation.

A number of 57 Cochlear Implants with CI532-electrodes were inserted at our clinic from 2015-2018. In six cases a revision surgery was necessary due to misplacement of the electrode. The implant was replaced by a different implant. Accordingly, 51 implantations were retrospectively analysed (age at surgery 58±16 years). Residual hearing was tested at 3 appointments: audiograms were done preoperatively (T0), within 30 days after surgery (T1) and at a second aftercare appointment (15,5 ±10,4 months, T2).

Within the time interval T0-T1 there was no significant deterioration of residual hearing at the frequency levels 125 and 250 Hz. However, at frequency levels >250 dB a significant decline of on average 10 dB (p <0,05) was shown.

Within the time interval until the second aftercare appointment (T1-T2) residual hearing at 250Hz declined significantly by 17 dB. At the remaining frequency levels, no differences were shown.

The CI532-electrode allows the preservation of residual hearing at low frequencies. However, in this study’s observation period a significant deterioration at 250 Hz was shown. Further research is necessary to analyse residual hearing in long-term follow-up.

Poster-PDF A-1637.PDF



Publication History

Article published online:
10 June 2020

© 2020. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/).