CC BY-NC-ND 4.0 · Laryngorhinootologie 2021; 100(S 02): S220-S221
DOI: 10.1055/s-0041-1728439
Abstracts
Otology / Neurotology / Audiology

First experience with an active bone conduction implant system fixed by selfdrilling screws

A Koitschev
1   Klinikum Stuttgart, HNO-Klinik, Abt. Kinder-HNO und Otologie, Stuttgart
,
N Berger
1   Klinikum Stuttgart, HNO-Klinik, Abt. Kinder-HNO und Otologie, Stuttgart
,
A Ulmer
1   Klinikum Stuttgart, HNO-Klinik, Abt. Kinder-HNO und Otologie, Stuttgart
,
P Amrhein
1   Klinikum Stuttgart, HNO-Klinik, Abt. Kinder-HNO und Otologie, Stuttgart
,
C Koitschev
1   Klinikum Stuttgart, HNO-Klinik, Abt. Kinder-HNO und Otologie, Stuttgart
› Author Affiliations
 

Objective The active bone conduction hearing implant (BoneBridge) is since few years in use. The floating mass transducer has to be fixed with two screws, which require predrilled openings. Selfdrilling screws could accelerate the implantation procedure. We present our first experience with the implant fixation by selfdrilling screws.

Study design Prospective study, single-subject repeated-measures design, one single tertiary referral center (Olgahospital, Stuttgart). All implants fixed by selfdrilling screws were evaluated.

Patients and methods 18 patients aged 5 to 58 years received one implant each. Intraoperative observations were evaluated according to a standard data set.

Air and bone conduction threshold and improvement in word recognition scores in aided and unaided condition were measured.

Results The variety of individual surgical situations could be summarized in 3 indication groups: a. malformations, b. revision surgery and c. single sided deafness.

Intraoperative complications were not observed. Implant fixation was sufficient in all cases. Significant speech discrimination improvement was found in all patients tested after 3 months.

In one case revision surgery was required due to skin dehiscence. In this case the skin was affected by prior surgery.

Conclusion Bone conduction implant fixation by selfdrilling screws is safe and reliable. Additional advantages are time saving and simplification of the surgery. Audiological improvement was achieved in all cases.

Poster-PDF A-1376.pdf

Ein Teil der Patienten wurde im Rahmen von Studien der Fa. MedEl, Innsbruck, Österreich ausgewertet.



Publication History

Article published online:
13 May 2021

© 2021. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/).

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany