Endoscopy 2016; 48(S 01): E334-E335
DOI: 10.1055/s-0042-117712
Cases and Techniques Library (CTL)
© Georg Thieme Verlag KG Stuttgart · New York

A forward-pulling technique for complicated Trapezoid basket impaction during retrieval of common bile duct stones

Tadahisa Inoue
Department of Gastroenterology, Aichi Medical University School of Medicine, Nagakute, Japan
,
Kiyoaki Ito
Department of Gastroenterology, Aichi Medical University School of Medicine, Nagakute, Japan
,
Norimitsu Ishii
Department of Gastroenterology, Aichi Medical University School of Medicine, Nagakute, Japan
,
Yuji Kobayashi
Department of Gastroenterology, Aichi Medical University School of Medicine, Nagakute, Japan
,
Masashi Yoneda
Department of Gastroenterology, Aichi Medical University School of Medicine, Nagakute, Japan
› Author Affiliations
Further Information

Publication History

Publication Date:
14 October 2016 (online)

Although basket impaction rarely occurs during retrieval of common bile duct stones, it is a potentially serious event [1]. The Trapezoid RX wire-guided retrieval basket (Boston Scientific Corp., Marlborough, Massachusetts, USA) has an emergency release feature, whereby the tip is designed to disengage, thereby facilitating basket removal and reducing the risk of impaction [2]. However, we encountered a rare case of trapezoid basket impaction that was complicated by difficulties in detaching the outer sheath from the basket wires.

A man in his 50s was admitted with multiple common bile duct stones ([Fig. 1]). After endoscopic sphincterotomy and endoscopic papillary large-balloon dilation, mechanical lithotripsy was attempted using a Trapezoid basket. However, basket impaction occurred because the stone was too hard and the tip of the Trapezoid failed to disengage.

Zoom Image
Fig. 1 Endoscopic retrograde cholangiography revealing multiple common bile duct stones.

We attempted to remove the duodenoscope and outer sheath simultaneously, but were unable to remove the outer sheath at the stomach because it was entwined with the inner basket wires. As a result, only the duodenoscope was removed, following which a forward-viewing endoscope was inserted along the basket catheter. Using rat-tooth forceps, we grasped, twisted, and pulled the basket wires into the working channel of the scope. Next, we simultaneously pushed the scope to the anal side and pulled the outer sheath to the oral side, dissociating it from the basket wires, so that it could be successfully removed ([Video 1]). The impaction was subsequently successfully reversed using an emergency mechanical lithotriptor handle (Olympus Medical Systems, Tokyo, Japan).


Quality:
The outer sheath could not be removed as it was entwined with the basket wires. After a forward-viewing endoscope was inserted along the basket catheter, rat-tooth forceps were used to grasp and push the basket wires to the anal side, while the outer sheath was pulled to the oral side; however, adequate traction could not be obtained because the rat-tooth forceps slipped. The basket wires were therefore grasped, twisted, and pulled into the working channel of the scope, which was then pushed to the anal side, while the outer sheath was simultaneously pulled to the oral side (arrows). This successfully dissociated the outer sheath from the basket wires (arrow), so that it could be removed.

Therefore, even when using a Trapezoid retrieval basket, basket impaction can occur and an emergency mechanical lithotriptor handle should be available on standby. Additionally, entanglement between the outer sheath and the inner basket wires can occur because the outer sheath of the Trapezoid is thinner than some other mechanical lithotriptors. The technique described here may be useful in such cases.