Successful Retrieval of Retained Rotablator Burr in the Mid Left Anterior Descending Coronary Artery using a Retrograde Approach: A Case Report

Abstract
We present a rare complication of Rotablator (Boston Scientific) atherectomy during percutaneous coronary intervention in a 67-year-old patient with a history of coronary artery disease and prior coronary artery bypass graft (CABG) surgery. The fracture of the Rotablator drive shaft and retention of the Rotablator burr in the mid left anterior descending coronary artery posed significant challenges in patient management. This case demonstrates the successful extraction of the retained Rotablator burr using retrograde total occlusion recanalization and highlights the importance of understanding the potential complications and management strategies in complex interventional cardiology procedures.
Publication History
Article published online:
08 February 2024
© 2024. International College of Angiology. This article is published by Thieme.
Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA
-
References
- 1 Lin C-P, Wang J-H, Lee W-L. et al. Mechanism and management of burr entrapment: a nightmare of interventional cardiologists. J Geriatr Cardiol 2013; 10 (03) 230-234
- 2 Chiang C-H, Liu S-C. Successful retrieval of an entrapped rotablator burr by using a guideliner guiding catheter and a snare. Acta Cardiol Sin 2017; 33 (01) 96-98
- 3 Mechery A, Jordan PJ, Doshi SN, Khan SQ. Retrieval of a stuck Rotablator burr (“Kokeshi phenomenon”) and successful percutaneous coronary intervention. J Cardiol Cases 2015; 13 (03) 90-92
- 4 Lata K, Grines CL. An Innovative Retrieval Technique for an Entrapped Rotational Atherectomy Device: A Case Discussion from the 2015 SCAI Fellows Course. 2015 Accessed on August 30, 2023 at: https://www.hmpgloballearningnetwork.com/site/cathlab/article/innovative-retrieval-technique-entrapped-rotational-atherectomy-device-case-discussion-2015