ABSTRACT
Continuous monitoring by pulse oximetry is a common practice for preterm and critically
ill newborns. A new generation of motion-tolerant pulse oximeters have been designed
for improved clinical performance with a substantial reduction in alarm frequency.
However, little is known about the differences among these new-generation pulse oximeters
in the neonatal intensive care unit (NICU). The purpose of this study is to assess
the clinical performance of two new-generation pulse oximeters in the NICU. Two new-generation
pulse oximeters were used simultaneously to monitor 36 patients in the NICU. The two
devices studied were the Philips FAST and the Masimo SET. Patients were randomly assigned
for their digit selection and data were collected only when waveforms were of good
quality and/or the pulse oximeter's pulse rate (PR) correlated with the electrocardiogram
heart rate (HR). The data for oxygen saturation measurements, number of true and false
alarms, and number of dropouts as well as the duration of dropouts for each pulse
oximeter were recorded by the primary investigator at 5-minute intervals for a period
of 2 hours on each patient. Dropouts are instances when the pulse oximeter alarm sounds
due to its inability to identify the arterial pulse and provide an oxygen saturation
reading. The mean gestational age for the study group was 32 weeks (rang, 24 to 42
weeks). Repeated-measures analysis of variance indicated no difference between the
two devices across all time measurements (p = 0.357). In addition, paired t -tests for true alarms and false alarms were not significant, with p -values of 0.151 and 0.869, respectively. There was a difference in the number of
data dropouts (p < 0.001): the Philips device had a nearly six-fold increase in the number of dropouts
(Philips 247 versus Masimo 38). The duration of dropouts was also significant; the
Philips device had three times longer duration of dropouts. Physiologic monitoring
in the critical care setting requires accurate data measurements. The two new-generation
pulse oximeters, the Philips FAST and Masimo SET, are equally sensitive in their ability
to identify true and false alarms. Masimo pulse oximeter with its signal extraction
technology, however, provides more consistent and accurate reporting of SpO
2 values as demonstrated by its markedly decreased incidence of data dropouts. The
Masimo unit appears to be more resistant to the effects of motion artifact.
KEYWORDS
Pulse oximetry - neonate - patient monitoring - medical technology
REFERENCES
1
Elander G, Hellstrom G.
Reduction of noise levels in intensive care units for infants: evaluation of an intervention.
Heart Lung.
1995;
24
376-379
2
Lawless S T.
Crying wolf: false alarms in the pediatric intensive care unit.
Crit Care Med.
1994;
22
981-985
3
Sabar R, Zmora E.
Nurses response to alarm from monitoring systems in NICU.
Pediatr Res.
1997;
41
174A
4
Bonhorst B, Peter C, Poets C F.
Pulse oximeters' reliability in detecting hypoxemia and bradycardia: comparison between
a conventional and two new generation pulse oximeters.
Crit Care Med.
2000;
28
1565-1568
5
Hay Jr W W, Rodden D J, Collins S M, Melara D L, Hale K A, Fashaw L M.
Reliability of conventional and new pulse oximetry in neonatal patients.
J Perinatol.
2002;
22
360-366
6
Poets C F, Urschitz M S, Bonhorst B.
Pulse oximetry in the neonatal intensive care unit (NICU): detection of hyperoxemia
and false alarm rates.
Anesth Analg.
2002;
94
S41-S43
7
Goldstein M R, Barnum P T, Vogt J, Gangitano E S, Stephenson C G, Liberman R L.
Conventional pulse oximetry can give spurious data in a neonatal population at risk
for retinopathy of prematurity (ROP).
Pediatr Res.
1998;
43
216A
8
Gehring H, Hornberger C, Matz H.
The effects of motion artifact and low perfusion on the performance of new generation
pulse oximeters in volunteers undergoing hypoxemia.
Respir Care.
2002;
47
48-60
9
Lutter N O, Urankar S, Kroeber S.
False alarm rates of three third generation pulse oximeters in PACU, ICU and IABP
patients.
Anesth Analg.
2002;
94
S69-S75
10
Gerstmann D, Berg R, Haskell R et al..
Operational evaluation of pulse oximetry in NICU patients with arterial access.
J Perinatol.
2003;
23
378-383
11
Brockway J, Hay W W.
Prediction of arterial partial pressure of oxygen with pulse oxygen saturation measurements.
J Pediatr.
1998;
133
63-66
12
Hay W W, Brockway J M, Eyzaguirre M.
Neonatal pulse oximetry: accuracy and reliability.
Pediatrics.
1989;
83
717-722
13
STOP-ROP: The STOP-ROP Multicenter Study Group .
Supplemental therapeutic oxygen for prethreshold retinopathy of prematurity (STOP-ROP),
a randomized, controlled trial.
Pediatrics.
2000;
105
295-310
Fegegta A WorkieM.D.
Department of Neonathology, 111 Michigan Avenue, N.W.
Washington, DC, 20010