Bildqualität aufaddierter Schichten bei MSCT-Untersuchungen des Thorax: Bildnachverarbeitung versus Direktrekonstruktion

Image Quality of Thickened Slabs in Multislice CT Chest Examinations: Postprocessing vs. Direct ReconstructionU. Wedegärtner1 , J. Yamamura1 , H.-D Nagel2 , D. Aldefeld1 , C. Brinkmann1 , S. Popovych1 , R. Buchert3 , C. Weber1 , G. Adam1
  • 1Klinik und Poliklinik für Diagnostische und Interventionelle Radiologie, Universitätsklinikum Hamburg-Eppendorf
  • 2Philips Medizin Systeme GmbH, Philips
  • 3Klinik und Poliklinik für Nuklearmedizin, Universitätsklinikum Hamburg-Eppendorf
Weitere Informationen

Publikationsverlauf

eingereicht: 5.5.2006

angenommen: 6.10.2006

Publikationsdatum:
07. März 2008 (online)

Zusammenfassung

Ziel: Die Bildnachverarbeitung bietet den Vorteil der interaktiven Erstellung aufaddierter Schichten, sodass ein unklarer Befund in Echtzeit problemlos in einer dünneren Schichtdicke abgeklärt werden kann. Um dieses Verfahren in der klinischen Routine einzusetzen, muss eine ausreichende Bildqualität gewährleistet sein. In dieser Arbeit erfolgte ein Vergleich der Bildqualität von mittels Bildnachverarbeitung aufaddierten Bildern mit direktrekonstruierten Bildern von Mehrschicht-Spiral-CT(MSCT)-Untersuchungen des Thorax. Material und Methoden: Es wurden MSCT-Untersuchungen (Mx8000IDT, Philips, Best, Niederlande) des Thorax von 20 Patienten (Durchschnitt: 56 Jahre) unter Verwendung folgender Parameter durchgeführt: 120 kV, 94 effektive mAs, 16 × 1,5 mm Kollimation, Pitch 0,575, 512 × 512 Matrix, Field of View 371 × 371 mm, CTDIvol = 6,3 mGy, DLP = 210 mGyxcm. Von sämtlichen Untersuchungen wurden axiale Schichten mit 3 und 5 mm Schichtdicke (SD) über Direktrekonstruktion und mittels Bildnachverarbeitung erstellt. Die korrespondierenden Bilder aus Direktrekonstruktion und Bildnachverarbeitung (Lungen-/Weichteilfenster) wurden paarweise von 2 Radiologen geblindet hinsichtlich der folgenden 5 Kriterien beurteilt: Organstrukturen, Randkontur, Kontrast, Bildrauschen, Artefakte. Unterschiede zwischen Direktrekonstruktion und Bildnachverarbeitung hinsichtlich dieser Kriterien wurden mittels Wilcoxon-Test analysiert. Das Bildrauschen wurde quantitativ mittels einer Region of Interest in der Aorta untersucht. Ergebnisse: Für Lungen-/Weichteilfenster, die beiden Untersucher sowie alle Qualitätskriterien zusammenfassend wurden Direktrekonstruktion und Bildnachverarbeitung in 96 % (3 mm SD) bzw. in 95 % (5 mm SD) der Fälle gleich bewertet. In den Fällen mit diskrepanter Bewertung lag der Unterschied bei maximal einem Punkt auf der 5er-Skala. Die statistische Auswertung mittels Wilcoxon-Test und die quantitative Analyse des Bildrauschens ergaben keinen Unterschied zwischen den beiden Verfahren. Schlussfolgerung: Die interaktive Erstellung aufaddierter Schichten in Echtzeit bietet eine Verbesserung des Arbeitsablaufes in der Befundung. Für die in dieser Studie verwendeten Untersuchungsparameter unterschied sich das Bildnachverarbeitungsverfahren hinsichtlich der Bildqualität nicht von der herkömmlichen Direktrekonstruktion.

Abstract

Purpose: Postprocessing offers the possibility of real-time creation of thickened slabs from a set of thin slices. This allows the interactive change from thick to thin slices for better evaluation of unclear lesions. As a result the clinical workflow of MSCT evaluation can be improved. However, to be able to apply this postprocessing software in the clinical routine, degradations in the image quality (compared to standard original reconstructed images) have to be avoided. The purpose of this study was to compare the image quality of thickened slabs from MSCT chest examinations that have either been directly reconstructed from the raw data or have been retrospectively generated via postprocessing. Materials and Methods: Chest MSCT examinations of 20 patients (mean age: 56 years) were performed on a 16-slice MSCT scanner (Mx8000IDT16, Philips, Best, Netherlands) using the following scan parameters: 120 kV, 94 effective mAs, 16 × 1.5 mm collimation, 512 × 512 matrix, field of view 371 × 371 mm, CTDIvol = 6.3 mGy, DLP = 210 mGyxcm). Slices with a thickness of 3 and 5 mm were generated for each examination both directly from the raw data and via postprocessing. Corresponding images from postprocessing and direct reconstruction (lung/soft tissue window) were evaluated by two radiologists with respect to 5 criteria on the basis of a five-point scale: organ structure, contour of small objects, contrast, image noise and artifacts. Differences between both data sets regarding image quality were assessed for each of the 5 criteria using a Wilcoxon test with Bonferroni correction. In addition, image noise was analyzed quantitatively in a region of interest in the aorta. Results: For the lung and soft tissue window, both reviewers and all criteria, no differences in image quality were detected between the thickened slices obtained via direct reconstruction and the postprocessing method. In 96 % and 95 % of the cases images of the two reconstruction methods were graded identically for 3 mm and 5 mm slices. In the remaining 4 % and 5 %, the evaluations differed only by one point on the five-point scale. The median grade of the first reviewer was 1 and that of the second reviewer was 2. There were no differences in the quantitative analysis of image noise between both methods. Conclusion: The interactive creation of thickened slices is an effective tool for the evaluation of MSCT examinations. For the defined scan parameters in this study there were no differences in image quality between postprocessing methods (e. g. slab viewer) and direct image reconstruction.

Literatur

  • 1 Bonmassari R, Muraglia S, Centonze M. et al . Noninvasive detection of coronary artery stenosis with 16-slice spiral computed tomography in a population at low to moderate risk for coronary artery disease.  J Cardiovasc Med (Hagerstown). 2006;  7 817-25
  • 2 Van Mieghem C A, Cademartiri F, Mollet N R. et al . Multislice spiral computed tomography for the evaluation of stent patency after left main coronary artery stenting: a comparison with conventional coronary angiography and intravascular ultrasound.  Circulation. 2006;  114 645-653
  • 3 Korner M, Krotz M, Kanz K G. et al . Development of an accelerated MSCT protocol (Triage MSCT) for mass casualty incidents: comparison to MSCT for single-trauma patients.  Emerg Radiol. 2006;  12 203-209
  • 4 Lim C M, Shridhar I, Tan L. et al . Contrast CT in localization of acute lower gastrointestinal bleeding.  Asian J Surg. 2006;  29 92-94
  • 5 Mahnken A H, Günther R W, Krombach G A. Grundlagen der linksventrikulären Funktionsanalyse mittels MRT und MSCT.  Fortschr Röntgenstr. 2004;  176 1365-1379
  • 6 Blobel J, Baartmann H, Rogalla P. et al . Räumliche und zeitliche Auflösung für die Herzdiagnostik mit 16-Schicht-Computertomographie.  Fortschr Röntgenstr. 2003;  175 1264-1274
  • 7 Fischbach R, Knollmann F, Griesshaber V. et al . Detection of pulmonary nodules by multislice computed tomography: improved detection rate with reduced slice thickness.  Eur Radiol. 2003;  13 2378-2383
  • 8 Das M, Ley-Zaporozhan J, Gietema H A. et al . Accuracy of automated volumetry of pulmonary nodules across different multislice CT scanners.  Eur Radiol. 2007 Jan 6; [Epub ahead of print] PMID: 17206420 [PubMed; 
  • 9 Marten K, Engelke C. Computer-aided detection and automated CT volumetry of pulmonary nodules.  Eur Radiol. 2007;  17 888-901
  • 10 Tsushima K, Sone S, Hanaoka T. et al . Comparison of bronchoscopic diagnosis for peripheral pulmonary nodule under fluoroscopic guidance with CT guidance.  Respir Med. 2006;  100 737-745
  • 11 Ko J P. Lung nodule detection and characterization with multi-slice CT.  J Thorac Imaging. 2005;  20 196-209. Review
  • 12 Nishino M, Hatabu H. Volumetric expiratory HRCT imaging with MSCT.  J Thorac Imaging. 2005;  20 176-185. Review
  • 13 Chooi W K, Morcos S K. High resolution volume imaging of airways and lung parenchyma with multislice CT.  Br J Radiol. 2004;  77 S98-S105. Review
  • 14 Aziz Z A, Padley S P, Hansell D M. CT techniques for imaging the lung: recommendations for multislice and single slice computed tomography.  Eur J Radiol. 2004;  52 119-136. Review
  • 15 Marten K, Seyfarth T, Auer F. et al . Computer-assisted detection of pulmonary nodules: performance evaluation of an expert knowledge-based detection system in consensus reading with experienced and inexperienced chest radiologists.  Eur Radiol. 2004;  14 1930-1938
  • 16 Zaporozhan J, Ley S, Gast K K. et al . Visuelle Abschätzung der funktionellen Lungenanteile in der HRCT und 3He-MRT bei Patienten nach Einzel-Lungentransplantation: Vergleich mit der absoluten Volumetrie.  Fortschr Röntgenstr. 2005;  177 516-523
  • 17 Wedegärtner U, Lorenzen M, Nagel H D. et al . Dünn- und dickschichtige MSCT-Untersuchungen bei Niedrigkontrastobjekten (Leberläsionen): Vergleich der Bildqualität bei gleicher Dosis.  Fortschr Röntgenstr. 2004;  176 1676-1682
  • 18 Kalender W. Computertomographie. München; Publicis MCD Verlag 2000
  • 19 Nagel H D. Strahlenexposition in der Computertomographie. Grundlagen, Einflußfaktoren, Dosisermittlung, Optimierung, Zahlenwerte, Begriffe. Info: ctb-publications@gmx.de CTB Publications 2002

PD Dr. Ulrike Wedegärtner

Klinik und Poliklinik für Diagnostische und Interventionelle Radiologie, Universitätsklinikum Hamburg-Eppendorf

Martinistraße 52

20246 Hamburg

Telefon: ++49/40/4 28 03 40 10

Fax: ++49/40/4 28 03 67 99

eMail: wedegaer@uke.uni-hamburg.de

    >