J Reconstr Microsurg 2007; 23(4): 189-197
DOI: 10.1055/s-2007-974655
Copyright © 2007 by Thieme Medical Publishers, Inc., 333 Seventh Avenue, New York, NY 10001, USA.

Noninvasive Tissue Oximetry for Flap Monitoring: An Initial Study

Alex Keller1 , 2
  • 1Department of Surgery, Division of Plastic Surgery, North Shore Long Island Jewish Health System, New Hyde Park, New York
  • 2Department of Surgery, Division of Plastic Surgery, New York University Medical Center, New York, New York
Weitere Informationen

Publikationsverlauf

Publikationsdatum:
25. Mai 2007 (online)

ABSTRACT

The ideal monitoring tool to evaluate free flap success should be noninvasive, continuous, and reliable. A new device, the ViOptix Tissue Oximeter (ODISsey™) based on near-infrared spectroscopy was evaluated in 30 patients undergoing autologous tissue perforator free flap breast reconstruction with continuous monitoring of the flap during elevation, transfer, and the postoperative period. The device accurately reflected the ischemic drop in oxygen saturation during flap transfer, and the hyperemic response after flow was reestablished. There were no flap failures, but in two patients, the device indicated a venous thrombosis before it was clinically obvious and allowed for expeditious flap salvage. The noninvasive nature of the device, the ability to replace and move the probe in the postoperative period, and the measurement of end organ oxygenation are all advantages over other currently available techniques.

REFERENCES

  • 1 Hirigoyen M B, Urken M L, Weinberg H et al.. Free flap monitoring: a review of current practice.  Microsurgery. 1995;  16 723-727
  • 2 Yuen J C, Feng Z. Reduced cost of extremity free flap monitoring.  Ann Plast Surg. 1998;  41 36-40
  • 3 Khouri R K. Avoiding free flap failure.  Clin Plast Surg. 1992;  19 773
  • 4 Jones N F. Intraoperative and postoperative monitoring of microsurgical free tissue transfers.  Clin Plast Surg. 1992;  19 783
  • 5 Swartz W M, Izquierdo R, Miller M J. Implantable venous Doppler microvascular monitoring: laboratory investigation and clinical results.  Plast Reconstr Surg. 1994;  93 152-163
  • 6 Swartz W M, Jones N F, Cherup L et al.. Direct monitoring of microvascular anastomoses with the 20 mHz ultrasonic Doppler probe: an experimental and clinical study.  Plast Reconstr Surg. 1988;  81 149-158
  • 7 Jones N F, Rocke A M, Swartz W M et al.. Experimental and clinical monitoring of free jejunal transfers using an implantable ultrasonic Doppler probe.  Br J Plast Surg. 1989;  42 274-280
  • 8 Hirigoyen M B, Blackwell K E, Zhang W X et al.. Continuous tissue oxygen tension measurement as a monitor of free-flap viability.  Plast Reconstr Surg. 1997;  99 763-773
  • 9 Bruce Chwatt A J. Free flap monitoring using a microcomputer linked to a laser Doppler flowmeter.  Br J Plast Surg. 1986;  39 229-238
  • 10 Yoshino K, Nara S, Endo M et al.. Intraoral free flap monitoring with a laser Doppler flowmeter.  Microsurgery. 1996;  17 337-340
  • 11 Hunt T K, Rabkin J, Jensen J A et al.. Tissue oximetry: an interim report.  World J Surg. 1987;  11 126-132
  • 12 Mao J M, Wright L, Elmandjra M. Application of near-infrared spectroscopy in monitoring blood perfusion in digital replantation.  Proc SPIE. 2005;  5692 382-389
  • 13 Thorniley M S, Sinclair J S, Barnett N J et al.. The use of near-infrared spectroscopy for assessing the viability during reconstructive surgery.  Br J Plast Surg. 1998;  51 218-226
  • 14 Blondeel P N, Beyens G, Verhaeghe R et al.. Doppler flowmetry in the planning of perforator flaps.  Br J Plast Surg. 1998;  51 202-209
  • 15 Hellekes D, Hettich R. Continuous laser-Doppler free flap monitoring via the telephone line: preliminary experiences.  Int J Clin Monit Comput. 1995;  12 241-244
  • 16 Kamolz L P, Giovanoli P, Haslik W et al.. Continuous free-flap monitoring with tissue-oxygen measurements: three-year experience.  J Reconstr Microsurg. 2002;  18 487-491
  • 17 Graham B H, Gordon L, Alpert B S et al.. Serial quantitative skin surface fluorescence: a new method for postoperative monitoring of vascular perfusion in revascularized digits.  J Hand Surg [Am]. 1985;  10 226-230
  • 18 Hjortdal V E, Hauge E, Hansen E S. Differential effects of venous stasis and arterial insufficiency on tissue oxygenation in myocutaneous island flaps: an experimental study in pigs.  Plast Reconstr Surg. 1992;  89 521-529
  • 19 Kind G M, Buntic R F, Buncke G M et al.. The effect of an implantable Doppler probe on the salvage of microvascular tissue transplants.  Plast Reconstr Surg. 1998;  101 1268-1273
  • 20 Hovius S E, van Adrichem L N, Mulder H D et al.. Comparison of laser Doppler flowmetry and thermometry in the postoperative monitoring of replantations.  J Hand Surg Am. 1995;  20 88-93

Alex KellerM.D. 

900 Northern Boulevard

Great Neck, NY 11021

    >