Abstract
Background: Anterior cervical discectomy and fusion (ACDF) using bone graft or a cage with plate
fixation is an accepted technique for the treatment of symptomatic degenerative disc
disease. It is, however, debatable whether a plate is really necessary to increase
the progress of fusion. Thus, the aim of this randomized and controlled prospective
study was to evaluate whether ACDF with a cage and anterior plate fixation results
in a greater progress of fusion compared with ACDF using a stand-alone cage.
Methods: 37 candidates for ACDF were treated either with a stand-alone cage (study group)
or with a cage+plate fixation (control group). 19 patients were randomized to be stabilized
with a stand-alone cage and 18 patients were treated with a cage and additional anterior
plate fixation. The progress of cervical fusion over time was compared by radiostereometric
analysis (RSA). Follow-up examinations pre- and postoperatively were done using the
Visual Analogue Scale (VAS) for neck and arm pain. Radiographic assessment of fusion
using an RSA-control was done after one, six and twelve weeks, as well as after six
months, and one and two years postoperatively. Mann-Whitney test for unpaired values
was used to determine the statistical differences in residual intervertebral motion.
Results: Three-dimensional analysis of segmental motion (left-right, cranio-caudal, and posterior-anterior)
did not reveal any statistical differences between both groups at any examination
time postoperatively (p >0.05). The VAS score did not differ between the groups (p >0.05).
Conclusion: Anterior plate fixation did not demonstrate an improvement in the progress of fusion
in one-level ACDF.
Zusammenfassung
Hintergrund: Die anteriore zervikale Diskektomie und Fusion ist eine etablierte Behandlungstechnik
bei symptomatischen degenerativen Halswirbelsäulenerkrankungen mit entsprechender
OP-Indikation. Die Fusion kann mit einem Cage als Bandscheibenersatz und zusätzlicher
ventraler Plattenosteosynthese bzw. auch ohne zusätzliche Verplattung durchgeführt
werden. Trotzdem ist es umstritten, ob eine zusätzliche Plattenosteosynthese die Fusionsprogression
des behandelten Segments beschleunigt. Das Ziel dieser prospektiven, randomisierten
und kontrollierten Studie war es zu untersuchen, ob die zusätzliche Plattenosteosynthese
zu einer beschleunigten Fusion bei mono-segmentalen Fusion führt.
Methoden: 37 Patienten wurden entweder mit „alleinstehendem” Cage (19 Patienten) oder mit zusätzlicher
Plattenosteosynthese (18 Patienten) behandelt. Zur Beurteilung der funktionellen Stabilität
wurde die Röntgenstereometrische Analyse (RSA) verwendet. Dadurch konnten postoperative
segmentale Bewegungen des fusionierten Segments dreimensional mit einer Genauigkeit
von <=0,4 mm bestimmt werden. Verlaufsuntersuchungen erfolgten 1 Woche, 6 und 12 Wochen
sowie 6, 12 und 24 Monate postoperativ. Für die klinische Beurteilung des Therapieerfolgs
(Nuchalgie, Brachialgie) wurde die Visuelle Schmerzanalogskala (VAS) gewählt. Die
statistische Signifikanz konnte mit dem Mann-Whitney-u-Test ermittelt werden.
Ergebnisse: Seitens der radiologischen Befunde konnte zu keinem Untersuchungszeitpunkt ein Unterschied
hinsichtlich der segmentalen Restbeweglichkeit des fusionierten Segments festgestellt
werden. Klinisch bestand ebenfalls zu keinem Zeitpunkt ein Unterschied zwischen den
untersuchten Behandlungsgruppen.
Zusammenfassung: Bei der mono-segmentalen zervikalen Fusion wird die Fusionsgeschwindigkeit durch
die zusätzliche Plattenosteosynthese nicht beschleunigt.
Key words
spine - bone - fusion - implants - clinical study - radiostereometric analysis
Schlüsselwörter
Wirbelsäule - Knochen - Fusion - Implantate - klinische Studie - Röntgenstereometrische
Analyse
References
1
Barlocher CB, Barth A, Krauss JK, Binggeli R, Seiler RW.
Comparative evaluation of microdiscectomy only, autograft fusion, polymethylmethacrylate
interposition, and threaded titanium cage fusion for treatment of single-level cervical
disc disease: a prospective randomized study in 125 patients.
Neurosurg Focus.
2002;
15
E4
2
Blumenthal SL, Gill K.
Can lumbar spine radiographs accurately determine fusion in postoperative patients?
Correlation of routine radiographs with a second surgical look at lumbar fusions.
Spine.
1993;
18
1186-1189
3
Bohlman HH, Emery SE, Goodfellow DB, Jones PK.
Robinson anterior cervical discectomy and arthrodesis cervical for cervical radiculopathy.
J Bone Joint Surg [Am].
1993;
75
1298-1307
4
Caspar W, Geisler FH, Pitzen T, Johnson TA.
Anterior cervical plate stabilisation in one- and two level degenerative disease:
overtreatment or benefit?.
J Spinal Disord.
1998;
11
1-11
5
Chen JF, Wu CT, Lee Sc, Lee ST.
Use of a polymethylmethacrylate cervical cage in the treatment of single-level cervical
disc disease.
J Neurosurg Spine.
2005;
3
24-28
6
Cho DY, Lee WY, Sheu PC.
Treatment of multilevel cervical fusion with cage.
Surg Neurol.
2004;
62
378-385
7
Cloward RD.
Treatment of acute fractures and fracture dislocation of cervical spine by vertebral
body fusion: A report of 11 cases.
J Neurosurg.
1961;
18
205-209
8
Emery SE, Fischer JR, Bohlman HH.
Three level anterior cervical discectomy and fusion: radiographic and clinical results.
Spine.
1998;
22
2622-2624
9
Hwang SL, Lin CL, Lieu AS, Lee KS, Kuo TH, Hwang YF, Su YF, Howng SL.
Three-level and four-level anterior cervical discectomies and titanium cage-augmented
fusion with and without plate fixation.
J Neurosurg Spine.
2004;
1
160-167
10
Johnsson R, Stromqvist B, Axelsson P, Selvik G.
Influence of spinal immobilization on consolidation of posterolateral lumbosacral
fusion. A roentgen stereophotogrammetric and radiographic analysis.
Spine.
1992;
17
16-21
13
Karrholm J.
Roentgen stereophotogrammetry: review of orthopedic applications.
Acta Orthop Scand.
1989;
60
491-503
11
Karrholm J, Hansson LI, Selvik G.
Mobility of the lateral malleolus: A roentgen stereophotogrammetric analysis.
Acta Orthop Scand.
1985;
56
479-483
12
Karrholm J, Jonsson H, Nilsson KG, Soderqvist I.
Kinematics of successful knee prostheses during weight-bearing: three-dimensional
movements and positions of screw axes in the Tricon-M and Miller-Galante designs.
Knee Surg Sports Traumatol Arthrosc.
1994;
2
50-59
14
Lang P, Chafetz N, Genant HK, Morris JM.
Lumbar spinal fusion. Assessment of functional stability with magnetic resonance imaging.
Spine.
1990;
15
581-588
15
Langley GB, Sheppeard H.
The Visual Analogue Scale: its use in pain measurement.
Rheumatol Int.
1985;
5
145-148
16
Lowey GL, MacDonough RF.
The significance of hardware failure in anterior cervical plate fixation. Patients
with 2- to 7-year follow-up.
Spine.
1998;
15
181-186
17
Majd ME, Vadhva M, Holt RT.
Anterior cervical recostruction using titanium cage with anterior plating.
Spine.
1999;
24
1604-1610
18
Panjabi MM, Isomi T, Wagner JL.
Loosening at the screw-vertebra junction in multilevel anterior cervical plate constructs.
Spine.
1999;
15
2383-2388
19
Pape D, Adam F, Fritsch E, Muller K, Kohn D.
Primary lumbosacral stability after open posterior and endoscopic anterior fusion
with interbody implants: A roentgen stereophotogrammetric analysis.
Spine.
2000;
25
2514-2518
20
Pointillart V, Cernier A, Vital JM. et al .
Anterior discectomy without interbody fusion for cervical disc herniation.
Eur Spine J.
1995;
4
45-51
21
Pompili A, Canitano S, Caroli F, Caterino M, Crecco M, Raus L. et al .
Asymptomatic esophageal perforation caused by late screw migration after anterior
cervical plating: report of a case and review of relevant literature.
Spine.
2002;
27
499-502
22
Price DD, Bush FM, Long S, Harkins SW.
A comparison of pain measurement characteristics of mechanical visual and simple numerical
rating scales.
Pain.
1994;
56
217-226
23
Ragnarsson JI, Eliasson P, Karrholm J, Lundstrom B.
The accuracy of measurements of femoral neck fractures. Conventional radiography versus
roentgen stereophotogrammetric analysis.
Acta Orthop Scand.
1992;
63
152-156
24
Ryu SI, Michelle M, Daniel K.
A prospective randomized study comparing a cervical carbon fiber cage to the Smith-Robinson
technique with allograft and plating: up to 24 months follow-up.
Eur Spin J.
2005;
25
25
Samartzis D, Shen FH, Lyon C, Philips M, Goldberg EJ, An HS.
Does rigid instrumentation increase fusion rate in one-level anterior cervical discectomy
and fusion?.
Spine J.
2004;
4
636-643
26
Selvik G, Alberius P, Aronson AS.
A roentgen stereophotogrammetric system: construction, calibration and technical accuracy.
Acta Radiol Diagn (Stockh).
1983;
24
343-352
27
Selvik G.
Roentgen stereophotogrammetric analysis.
Acta Radiol.
1990;
31
113-126
28
Shad A, Leach JC, Teddy PJ, Cadoux-Hudson TA.
Use of the Solis cage and local autologous bone graft for anterior cervical discectomy
and fusion: early technical experience.
J Neurosurg Spine.
2005;
2
116-122
29
Simmons E, Bhalla S.
Anterior cervical discectomy and fusion.
J Bone Joint Surg [Br].
1996;
51
225-237
30
Song KJ, Lee KB.
A preliminary study of the use of cage and plating for single-segment fusion in degenerative
cervical spine disease.
J Clin Neurosci.
2006;
13
181-187
31
Tancredi A, Agrillo A, Delfini R, Fiume D, Frati A, Rinaldi A.
Use of carbon fiber cages for treatment of cervical myeloradiculopathy.
Surg Neurol.
2004;
61
221-226
32
Thalgott JS, Xiongsheng C, Giuffre Jm.
Single-stage anterior cervical reconstruction with titanium mesh cages, local bone
graft, and anterior plating.
Spine J.
2003;
3
294-300
33
Thome C, Leheta O, Krauss JK, Zevgaridis D.
A prospective randomized comparison of the rectangular titanium cage fusion and iliac
crest autograft fusion in patients undergoing anterior cervical discectomy.
J Neurosurg Spine.
2006;
4
1-9
34
Van Jonbergen HP, Spruit M, Anderson PG, Pavlov PW.
Anterior cervical interbody fusion with a titanium box cage: early radiological assessment
of fusion and subsidence.
Spine J.
2005;
5
645-649
35 Whitecloud T. Management of radiculopathy and myelopathy by the anterior approach. In
Cervical Spine Research Society (eds)
The Cervical Spine . Philadelphia: Lippincott 1983 pp. 411-424
36
Whitecloud TS, LaRoocca SH.
Fibular strut graft in reconstructive surgery of the cervical spine.
Spine.
1976;
1
33-43
37
Yen CP, Hwang TY, Wagner CJ, Howng SL.
Fracture of anterior cervical plate -report of two cases.
Acta Neurochir (Wien).
2005;
147
665-667
38
Zevgaridis D, Thome C, Krauss JK.
Prospective controlled study of rectangular titanium cage fusion compared with iliac
crest autograft fusion in anterior cervical discectomy.
Neurosurg Focus.
2002;
15
E2
39
Zoega B, Kärrholm J, Lind B.
Plate fixation adds stability to two-level anterior fusion in the cervical spine:
a randomised study using radiostereometry.
Eur Spine J.
1998;
7
302-307
Correspondence
Dr. A. Nabhan
Department of Neurosurgery
University of Saarland
66421 Homburg
Germany
Phone: +49/6841/162 4429
Fax: +49/6841/162 4439
Email: dr_a_nabhan@hotmail.com