Eur J Pediatr Surg 1981; 34(10): 147-149
DOI: 10.1055/s-2008-1063320
© Georg Thieme Verlag KG Stuttgart · New York

The Evaluation of Non-invasive Fibre-optical Monitoring of Intracranial Pressure (ICP) over the Fontanelle

K. von Wild
  • From the Neurochirurg. Klinik, Städt. Krankenanstalten, Hannover, Federal Republic of Germany
Further Information

Publication History

Publication Date:
19 May 2008 (online)

Abstract

Non-invasive ICP monitoring over the fontanelle with the pressure-activated fibre-optical Ladd transducer has proven to be a reliable, accurate, and simple method. It was used in 140 continuous measurements in pre-term infants (800 gm) and infants up to 16 months with small fontanelles (6 mm in diameter). After determination of the aplanation pressure, the sensor is kept in place by a simple plaster dressing ( 6 × 6 cm) up to 72 hours.

Zusammenfassung

Die nichtinvasive Hirndruckmessung über der Fontanelle mit dem fiberoptischen Ladd Hirndruckmeßsystcm hat sich bei 140 Messungen als zuverlässig, genau und einfach durchführbar erwiesen bei Frühgeborenen (800 G) und auch bei Säuglingen mit sehr kleiner Fontanelle (6 mm im Durchmesser). Nach Bestimmung des Aplanationsdrucks wird der Sensor mit einem einfachen Gipsverband ( 6 × 6 cm) sicher bis zu 72 Stunden befestigt.

    >