Proximal biliary stent migration is a rare and unwanted complication. Various endoscopic
methods have been used to remove migrated biliary stents, including the use of direct
cholangioscopy for plastic biliary stent migration [1]
[2]
[3]
[4]. Here, we describe a case of cholangioscopy-assisted removal of a proximally migrated
fully covered metal biliary stent by inversion technique.
A 78-year-old woman with metastatic pancreatic head adenocarcinoma initially presented
to an outside facility with jaundice due to biliary compression. Endoscopic retrograde
cholangiopancreatography (ERCP) with placement of two fully covered self-expandable
metal stents (FCSEMS) was performed at the outside facility. Several days later, the
patient presented to our facility with persistent jaundice. Serum bilirubin was 20 g/dL.
A repeat ERCP was then performed at our facility.
Initial fluoroscopy showed two biliary stents in the common duct. Endoscopic view,
however, revealed only one stent emanating from the ampulla. Cholangiogram showed
that that the second stent had migrated upward into the common hepatic duct (CHD)
adjacent to the nonmigrated stent. The nonmigrated stent was removed, and cholangioscopy
using SpyGlass DS Direct Visualization System (Boston Scientific Corp., Marlborough,
Massachusetts, USA) was performed to visualize the migrated stent ([Fig. 1]). The migrated stent was in the oblique position, partially obstructing the CHD.
In order to remove the migrated stent, SpyBite Biopsy Forceps (Boston Scientific Corp.)
via cholangioscopy was used to grasp the proximal portion of the migrated stent and
invert it inward while pulling the stent downward ([Fig. 2], [Fig. 3], [Video 1]). A Radial Jaw 4 Pediatric Biopsy Forceps (Boston Scientific Corp.,) was then used
to fully remove the migrated stent ([Fig. 4]).
Fig. 1 Fluoroscopic image of proximally migrated stent and cholangioscope.
Fig. 2 Proximal portion of the stent being grasped by SpyBite Biopsy Forceps (Boston Scientific
Corp., Marlborough, Massachusetts, USA).
Fig. 3 Inverted biliary stent seen on fluoroscopy.
Video 1 Cholangioscope-assisted removal of proximally migrated fully covered metal biliary
stent. The stent was grasped at its proximal end by SpyBite Biopsy Forceps (Boston
Scientific Corp., Marlborough, Massachusetts, USA) and inverted while being pulled
downward. Pediatric forceps then removed the inverted stent.
Fig. 4 Removed migrated biliary stent in inverted state.
Endoscopy_UCTN_Code_CPL_1AK_2AC
Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques
in gastroenterological endoscopy. All papers include a high quality video and all
contributions are freely accessible online.
This section has its own submission website at https://mc.manuscriptcentral.com/e-videos