Endoscopy 2022; 54(05): E199-E200
DOI: 10.1055/a-1480-6846
E-Videos

Stretch occlusion of stone extraction balloon catheter

Shou-jiang Tang
Division of Digestive Diseases, Department of Medicine, University of Mississippi Medical Center, Jackson, MS
› Institutsangaben
 

    A 65-year-old woman underwent endoscopic retrograde cholangiopancreatography (ERCP) for choledocholithiasis, and a standard stone extraction balloon (Multi-3V; Olympus, Tokyo, Japan) was used initially to remove the stone after biliary sphincterotomy. After several attempts of downward pulling on the balloon catheter, aiming for stone expulsion, we were unable to deflate the balloon ([Fig. 1]). No elevator manipulation was used during stone removal. We suspected that the balloon catheter had developed stretch occlusion: total collapse of the air channel within the balloon catheter caused by the stretched surrounding plastic housing catheter. Through this mechanism the air inside the balloon becomes totally trapped above the stretched occlusion site. The balloon could not be deflated even after we cut the catheter close to the handle ([Fig. 2]). So, the inflated balloon was pulled to the biliary opening. Next to the cut catheter, an endoscopic needle-knife without cauterization was used to puncture the inflated balloon ([Fig. 3]) under endoscopic and fluoroscopic guidance. The balloon deflated immediately and was removed ([Video 1]). We removed all stone fragments during the same ERCP session. All ERCP endoscopists need to be aware of this rare potential complication when using stone extraction balloons and know how to manage it accordingly.

    Zoom Image
    Fig. 1 Fluoroscopic image: after several attempts of downward pulling on the balloon catheter, aiming for stone expulsion, we were unable to deflate the balloon.
    Zoom Image
    Fig. 2 Endoscopic image: the balloon could not be deflated even after we cut the catheter close to the handle.
    Zoom Image
    Fig. 3 Endoscopic image: we used an endoscopic needle-knife without cauterization to puncture the inflated balloon.

    Video 1 Stretch occlusion of stone extraction balloon catheter.


    Qualität:

    Endoscopy_UCTN_Code_CPL_1AK_2AF

    Endoscopy E-Videos
    https://eref.thieme.de/e-videos

    Endoscopy E-Videos is an open access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online. Processing charges apply (currently EUR 375), discounts and wavers acc. to HINARI are available.

    This section has its own submission
    website at
    https://mc.manuscriptcentral.com/e-videos


    #

    Competing interests

    The authors declare that they have no conflict of interest.


    Corresponding author

    Shou-jiang Tang, MD
    Division of Digestive Diseases
    Department of Internal Medicine
    University of Mississippi Medical Center
    2500 North State Street
    Jackson, MS 39216
    USA   

    Publikationsverlauf

    Artikel online veröffentlicht:
    12. Mai 2021

    © 2021. Thieme. All rights reserved.

    Georg Thieme Verlag KG
    Rüdigerstraße 14, 70469 Stuttgart, Germany


    Zoom Image
    Fig. 1 Fluoroscopic image: after several attempts of downward pulling on the balloon catheter, aiming for stone expulsion, we were unable to deflate the balloon.
    Zoom Image
    Fig. 2 Endoscopic image: the balloon could not be deflated even after we cut the catheter close to the handle.
    Zoom Image
    Fig. 3 Endoscopic image: we used an endoscopic needle-knife without cauterization to puncture the inflated balloon.