Keywords endoluminal therapy - laser therapy - laser probe complication - varicose vein surgery
Introduction
Endoluminal thermal ablation of truncal varicosis of the great saphenous vein and
the
small saphenous vein using laser or radiofrequency techniques is widely used worldwide
and
has been well documented in the literature since 1999 [1 ]. An advantage of the procedure is the low complication rates and the low postoperative
discomfort of patients [1 ]. The errors reported in the USA for endoluminal therapy/stripping procedures (various
laser systems, radiofrequency catheters, vein stripping) are listed in the freely
accessible
Maud database (Maude Adverse Event Report: Symmetry Surgical, Inc, Symmetry Vein Stripper
(fda.gov). Here, problems are found numerically more frequently with radiofrequency
techniques
than with laser systems. Recently, Scholl and Leitz [2 ] described such a case in the journal Phlebologie.
Case report
We report below a rare complication of endoluminal thermal laser ablation that has
not
been previously described for this type of laser and fiber.
A 52-year-old man with a BMI of 28.4 underwent thermal ablative treatment of an
insufficient left saphenous vein using a 1470 nm laser with single-ring radial fiber
(1470 nm
Laser Leonardo and ELVeS Radial fiber, Biolitec, Germany). The laser probe was inserted
from
the inside of the knee after careful inspection for integrity and advanced up to the
saphenofemoral junction zone in a duplex-controlled fashion. The laser probe tip was
placed at
a distance of 1.1 mm ([Fig. 1 ]) from the femoral vein.
Fig. 1 Placement of the radial laser probe in the saphenofemoral junction zone; distance
to
the roof of the femoral vein is 1.1 mm.
The application of the TLA solution cooled down to 4 degrees Celsius was performed
in a
duplex-controlled manner using a 0.9 × 70 mm cannula (20 G). Ninety-five ml of tumescent
local
anesthetic (TLA) solution was injected at the saphenofemoral junction, and 382 ml
of TLA was
injected at the saphenous trunk of the thigh. In addition, another 280 ml of TLA was
applied
in preparation for a miniphlebectomy in the area of side branch varices (total amount
775
ml).
The set laser power was 10 watts. The proximal 3 cm of the great saphenous vein were
treated with a total of 551 joules, i.e., 183 J/cm, following the recommendations
of Spreafico
et al [3 ] and Hartmann et al [4 ]. Hereafter, the probe was continuously withdrawn with an average energy of 84.4
J/cm.
The total treatment distance was 33 cm, and the total energy delivered was 2919.9
joules. The
treatment time was 55.4 sec at the saphenofemoral junction and 292.5 sec in the saphenous
trunk.
Sticking of the laser probe tip in the segment of the great saphenous vein near the
orifice, which we frequently observe and which we counteract from the second centimeter
distal
to the crosse by gently moving the laser probe tip back and forth while delivering
energy, was
not present in this case. After the proximal 3 cm of the great saphenous vein was
treated with
3 complete cycles totaling 551 joules, continuous retraction of the radial laser probe
was
performed with the patient in the head-down position to the puncture site at the medial
knee
without intraoperative features. Postoperative inspection of the laser probe tip revealed
that
the proximal segment of the optical fiber was missing, over the length of 1 cm. Clinically,
the patient offered no abnormalities. A subsequent repeat Doppler examination was
technically
difficult to assess due to the infiltrated amount of TLA.
However, a 4-mm-long echo-rich district was seen approximately 5 mm distal to the
saphanofemoral junction and then an 8–10-mm-long echo-rich district was seen mid-thigh.
To
rule out embolization, although unlikely, we ordered a CT scan of the thorax, abdomen,
left
groin, and left thigh. The examination revealed no pathologic abnormalities. Since
the
endoluminal surgery was performed late in the afternoon, the CT diagnostics were not
completed
until about 10:00 p.m., and the patient offered no clinical abnormalities, we did
nothing more
that evening and waited for the TLA solution to resorb until the following morning
and then
repeated the duplex diagnostics. The groin showed the same findings as the previous
evening.
In the middle of the thigh, the double contour of the glass fiber could now be clearly
detected ([Fig. 2 ]).
Fig. 2 Detached probe tip of the radial laser probe mid-thigh 1 day post-op.
Thereupon, in consultation with the patient, we excised the probe tip in the thigh,
which
was possible without any problems ([Fig. 3 ]).
Fig. 3 Detached laser probe tip mid-thigh intraoperatively.
After it then became apparent intraoperatively that parts of the glass sheath were
missing
from the detached probe tip ([Fig. 4 ]), which corresponded in size to the 4-mm, echo-dense finding in the crosse region
described by duplex sonography, we decided to additionally explore the crosseregion.
Fig. 4 Tip of the laser probe with glass defect in the roof area.
The then possible intraoperative evidence of discontinuous carbonization reactions
of the
treated great saphenous vein with intervening rough vein segments we consider as additional
evidence that the probe tip had broken in the crosseous area ([Fig. 5 ]).
Fig. 5 Intraoperative crosse findings with central ligation. Normal postoperative laser
aspect in the central 1 cm saphenous segment. Then distally complete carbonization
of the
great saphenous vein.
The immediate crosseregion shows no signs of carbonization. These are seen only about
1 cm
distal to the junction of the great saphenous vein with the femoral vein. Consequently,
the
probe tip has experienced partial damage at this level. The probe tip did not tear
off until
mid-thigh. We performed a correct, level crossectomy [5 ]. The surgical procedure corresponded to that of a recurrent crossectomy, i.e., the
femoral vein was dissected proximally and then distally to the saphenous vein orifice.
The
saphenous vein was then double ligated with Ethibond at the same level as the femoral
vein
without touching the central saphenous vein segment, and the saphenous vein was resected
distally over a length of 5–7 cm and preserved for histopathological examination.
Palpatory
findings were unremarkable in the proximal 8-mm-long saphenous vein segment that was
double
ligated. In particular, we could not palpate any glass particles in the great saphenous
vein
crosse. Histopathologically, no glass remnants were detectable either.
Histopathologically, the crosseregion showed necrobiosis of the vein wall as well
as
transection of the vein wall ([Fig. 6 ]).
Fig. 6 Rupture of the vein wall, as known from studies on barefiber-treated veins.
Discussion
To our knowledge, this is the first publication of an intraoperative detachment of
the
probe tip of a radial fiber for endovenous thermal laser ablation. According to Biolitec,
there is no other reported case of such a complication with the ELVeS Radial 1ring
fiber
used.
In contrast, complications due to lodged sections of other types of lasers and fibers
have
already been described [6 ]
[7 ].
What can be considered as the cause of such damage to a radial fiber? There are two
scenarios to be discussed here:
According to the manufacturer, a “lack of mechanical integrity of the fiber before
usage
(perhaps due to transport damage or mishandling during unpacking and preparation of
the fiber
before use”) is to be considered. In our case, we can largely exclude this due to
our routine,
careful preoperative inspection of the laser fiber.
On the other hand, intraoperative damage to the fiber is possible.
The destruction of the probe tip observed here due to mechanical effects (pulling
on the
probe or damage to the probe by the injection needle) is extremely unlikely.
Intraoperatively, during the application of the tumescent local anesthetic solution
by the
injection needle described above in the crosse area, contact of the needle tip with
the glass
envelope of the laser probe occurs more frequently. Under the premise that the laser
probe tip
is almost completely in contact with the vein wall in the trendlenburg position, and
the vein
as such is embedded in a surrounding fluid bed of a large amount of TLA solution,
intraoperative contact of the injection needle with the laser probe tip, which occurs
in a
dosed manner, is seen to yield to the needle contact. Destruction of the glass tip
by such a
maneuver is inconceivable. Also, the manufacturing company, to which we handed over
the
defective probe for the analysis of the damage process, comes after subtle diagnostics
to the
result, "that the blasting off of the dome roof mechanically by the tumenescence needle
is to
be doubted very much". Such a mechanism would have to lead to such complications much
more
frequently, given the presumed frequency of conscious or unintentional contact of
the needle
with the laser probe tip. However, this is defacto not the case. This is the first
description
of such a complication. The situation is completely different when the injection needle
comes
into contact with the vulnerable laser fiber. Contact of the injection needle with
the laser
fiber should be avoided at all costs. In this case, detachments of the optical fiber
have been
described intraoperatively [8 ]
[9 ].
Also, in vitro experiments we performed after the laser probe detachment described
above
show that damage to the glass envelope of the laser tip cannot be achieved using the
20 G
puncture needle. We repeatedly tried to damage the laser probe tip on a hard surface
(metal
instrument table, desk) using the 20 G needle. This did not succeed. The needle slips
very
easily on the glass envelope, so damage per se is not possible. Even if one presses
the tip of
the needle into the glass or tries to drill into the glass by turning the needle,
a procedure
that is never performed intraoperatively, it is not possible to damage the tip of
the glass
even with massive pressure against a hard surface, and here too there is ultimately
a
difference from soft tissue intraoperatively – here the laser probe tip floats in
the vein or
in the surrounding tumescent solution. In this respect, the purely mechanical idea
that the
laser tip is shattered by the injection needle is certainly very unlikely. This statement
is
also supported by the manufacturing company.
Hypotheses on the damage process of the laser tip:
In the case of optical waveguides, and especially in those probes that deflect and
reshape the beam, great care must be taken to ensure that the optical structures that
deflect and shape the beam absorb the beam only slightly. Because of the very high
intensities, even small spots of absorption (impurities that cannot be seen by the
human
eye) in the glass can heat the material to a significant degree. This leads to melting
and
subsequent destruction of the glass material and detachment of the glass dome. Once
the
dome is detached, the concentric beam exits the probe, since it is no longer converted
by
the Axicon structure into a “harmless” beam with low intensity. After the roof of
the
laser probe, which is about 4 mm long and about 1 mm wide, flaked off, the laser has
completely changed its laser properties. In connection with the then also occurring
“contamination” of the exit surface with body fluids, the absorption of the laser
radiation increases further and can lead to a complete destruction of the probe, which
obviously is what happened. The laser has thus developed properties that are comparable
to
those of a barefiber in the broadest sense. This is evidenced by the carbonization
phenomena of the saphenous segment near the cross ([Fig. 5 ]), as well as the histology with typical rupture of the vein wall ([Fig. 6 ]). The completely altered laser properties then caused the tip of the laser probe
to melt off at the weld approx. 20 cm below the groin, i.e., after delivery of approx.
2000 joules of energy.
Alternatively, the following damage mechanism is also conceivable: If the probe tip
was manufactured using a glass fusion process, it is also possible that stresses are
introduced into the glass during this process (in very rare cases) and then these
glass
structures become very sensitive to heating, but also to slight mechanical stresses.
Only
the manufacturer can answer whether such a damage mechanism is possible or has occurred.
He knows his manufacturing process very well.
A simple form of quality control of the probe could be to briefly apply a significantly
higher power to the probe before use and then visually check whether the probe has
been
damaged. However, this is clearly the responsibility of the manufacturer and not the
user. In
addition, the user who acts in this way exposes himself to the accusation that any
destruction
of the probe that may occur was only caused by the power being too high and not in
accordance
with the specifications.
The above-mentioned hypotheses are based on an intensive discussion of the facts with
a
renowned laser physicist of a German university, who explicitly asked not to be named.
His
damage hypothesis is based on the Fig. 1–6 sent to him and the operational procedure
described by us). Like the manufacturer, the
laser physicist concludes that the observed destruction of the probe was probably
not caused
by mechanical influences (pulling on the probe or injury to the probe by an impinging
injection needle).
After intensive testing, the manufacturing company to which the used probe was provided
concludes that an unfortunate chain of effects, which can no longer be traced, led
to the
partial detachment of the distal quartz glass cap. The essential aspects of this multi-page
test report are summarized below:
“we use pure quartz glass as the material for the fiber core, this glass is also used
for
high power laser with power of 4–8 KW, the commonly used power of 10W is very low
for the
material. The breaking point of the fiber cap has sharp edges, thus the melting of
the glass
can be excluded with certainty. The fiber was subjected to a tensile test of 100 kpsi
during
manufacture, so the contamination of the surface of the fiber can be excluded, otherwise
the
fiber would have broken during the test. The fiber itself has no welded seam. The
cap and the
fiber are made of the same high purity fused silica. Furthermore, the fracture site
does not
show any melted off areas. Softening/melting of fused quartz occurs at approximately
1600° at
the earliest. The clear smooth edges indicate spalling. In our opinion, the break-off
you
assume occurred due to a very unfortunate break-off of the glass dome roof. In the
course of
the laser action, a light leakage occurred at the transition from the glass fiber
to the rigid
glass cap. We assume that initially there was a break in the cap and that this crack
led to a
detachment of a part of the roof and the formation of a micro-gap. Liquid penetrated
through
this gap. This was followed by laser power leakage in the transition area between
the
remaining fused silica cap and the fused silica fiber. In the course of the distal
thigh, the
cap came off or the cap was pulled off the fiber. The now naked glass fiber no longer
has any
light conducting property so that more and more energy was pulled out in the transition
region
of liquid. We dare to doubt very much whether the blowing off of the dome roof is
mechanically
possible by the tumescent needle and we agree with the authors here. The case was
closely
examined, and no anomalies were found in the manufacture and testing of the probe
in question.
The darkening of the adhesive at the junction between the cap and the optical fiber
indicates
that the adhesive was very hot during the process, but it is not possible to determine
whether
the overheating caused the probe to fail or whether the overheating occurred after
the
failure. Due to effects, we can no longer trace, partial loss of the fused silica
cap occurred
due to the dome roof blowing off. Micro-cracks then led to the formation of light
absorption
in the quartz glass composite, resulting in the failure of the joint.”
Histologically, there was rupture of the vein wall after crossectomy, as has been
shown
with the use of barefiber [10 ]. Whether the missing roof of the laser tip was melted away or blown off by the altered
properties of the laser remains unanswered. We found no glass remnants either palpatorily
or
histologically. The fiber tip did not detach until midthigh. This indicates that the
weld seam
between the laser tip and the glass fiber only became detached due to the altered
radiation of
the laser.
Holdstock et al. [9 ] already demonstrated damage to the laser fiber in vitro experiments with an 810
nm
laser system (Varilase) using a 21 G needle in 2008. Damage to the glass tip of a
radial laser
probe has not yet been described.
In summary, the following conclusions can be drawn from this case (see also [11 ]):
Pre- and especially postoperatively, the integrity of the laser fiber should be
checked.
If postoperative defects of the laser fiber appear, sonographic detection of blown-off
glass particles in the treated vein may be difficult due to the applied TLA. Therefore,
repeat sonography is indicated after the TLA has been drained off.
Duplex diagnostics were superior to CT diagnostics in the case described here.
The manufacturing company recommends the following procedure:
What to do after localization of the position of the cap?
If the cap is in
junction area and the junction is open proximal to the cap, consider whether to perform
a
ligation cranial to the cap and possibly the surgical removal of the cap
itself. Balance the difficulty of ligature and the risk of displacement of the cap in
the venous system. If the cap is in the trunk, included between occluded vein
segments, up and down, or in a perforator, there is no risk of embolization into the
venous system. If the cap is in a favorable position, for example: superficial in a
thin, skinny patient palpable through the skin, the cap can be easily removed, through
a
small incision, under local anesthesia and ultrasound guidance. Conversely, in a
disadvantageous position, for example: deep, near to the junctions or nerves in an
obese
patient, the cap can be left, and the patient should be warned that he has a small
inert
foreign body, which will probably be asymptomatic, has a modest chance to move because
it
is embedded in a scar and removal can create more problems than benefits. If the cap
is
not removed, it is indicated to explain to the patient what the signs and symptoms
of deep
vein thrombosis, pulmonary embolism, and cap dislocation may be, so that in case of
a
suspect he can be quickly reassessed.
What to do if the cap could
not be identified?
Patients should be warned that it was not possible to
locate the cap fragment with the X-ray and ultrasound (also indicating the possibility
that it is not within the limb as was assumed) and it is necessary to wait a few months
in
order to be able to detect the body's reaction (foreign body granuloma), which should
help
locate the fragment and eventually remove it. The fragment is small in size, of an
inert
material (quartz) and, in the formation of granuloma, is blocked within the inflammatory
tissue. Therefore, there are no particular dangers in this waiting time. In addition,
the
inflammatory reaction around the foreign body could cause a nuisance that could help
to
better focus on the search for the foreign body in the limb. Our therapeutic procedure
can certainly be discussed diversely. It is by no means mandatory to follow the
manufacturer's recommendations listed above in English. It would certainly have been
wrong to do nothing and not inform the patient about the complication that occurred
during
the procedure. Whether the torn probe tip could not have been left in situ must be
clarified with the patient in each individual case. We had informed the patient that
the
probe tip was superficial and could be easily removed surgically with little risk.
Our
advice to the patient was to remove the foreign body and to accept an approximately
1.5 cm
long scar. Waiting and leaving the foreign body in situ, according to the
manufacturer's recommendations, combined with the need for repeated sonographic controls,
may be possible in principle. We have decided against this. In our view, the decision
to re-explore the inguinal region is completely uncontroversial. The intraoperative
aspect
of broken glass particles from the roof of the probe tip in the immediate vicinity
of the
saphenofemoral junctional zone and the risk of possible carryover of foreign body
material
to the central region made open surgical intervention with performance of a level
crossectomy imperative. The authors suggest that this rare complication be included
in the appropriate surgical education protocols.
Conclusion
The authors describe for the first time the intraoperative detachment of a laser probe
tip
during endoluminal therapy of truncal varicosis. This complication was not caused
by faulty
medical action. A causal relationship between the clinically not so rare contact of
the
tumenescent needle with the laser probe tip and the very rare complication described
here, is
as described above, extremely unlikely. The different treatment options resulting
from this
are discussed. The authors have decided on an open surgical approach.