Key words mammography - image processing - noise equivalent quanta - modulation transfer function
- system power - CDMAM
Introduction
Up to now, automated methods may not be used to evaluate mammographic test phantom
images according to the regulations for acceptance and constancy tests [1 ]
[2 ]
[3 ]
[4 ]. The MammoControl DIANA software developed by the Reference Center for Mammography
in Muenster for mammographic screening already offers a platform for automated, centralized
quality control [5 ]
[6 ]. This software enables, in principle, both the transmission of measurement values
and test phantom images as well as automated evaluation. Various analysis algorithms
can be implemented due to the modular structure of the software. The software gives
immediate feedback to the operator. Preparations for automated evaluation of test
phantom images in mammography screening are far advanced [7 ].
The constancy test for screening including a visual evaluation of CDMAM test phantom
images must be conducted on an annual basis (in the field of curative mammography
every two years). This image quality test that is decisive in obtaining/extending
approval to conduct patient examinations has the shortcoming that the foreseen analysis
by three viewers contradicts ICRU recommendations [8 ] due to the lack of randomization and learning effects as well as the switching back
and forth between recognizing circular test objects and noise analysis. The test objects
themselves can cause significant variations in the detection rate. There is reasonable
doubt whether this method offers adequate accuracy and whether it is suitable for
this application. A significant change in the detection rate can only be observed
if the dose is increased by a factor of 2 [9 ]. The CDCOM program [10 ] offered by EUREF [11 ] does not meet the requirements for a diagnostic method as it does not provide values
for specificity and detection accuracy nor has the detection process been clearly
described. It is a noise analysis-based method that enables further processing of
its primary detection rates in various ways. An alternative method, the freely available
CDIC [12 ], is a diagnostic method, and its detection method has been clearly described [13 ]. No limiting values that could be used for image quality testing are available for
either of these methods. Nevertheless, the phantom was used in other studies to compare
image quality [14 ].
Therefore, comparative examinations must be made with equipment systems that comply
with the existing PAS 1054 and statutory requirements (CDMAM) to determine and compare
values in order to extrapolate limiting values that were obtained with different methods.
Physical methods provide values based on the IEC Standard [15 ] for the modulation transfer function (MTF), noise power spectrum (NPS), noise equivalent
quanta (NEQ) and detective quantum efficiency (DQE). Diagnostic methods provide parameters
such as sensitivity, specificity and accuracy [16 ]
[17 ].
Materials and Methods
Additional test phantom images were made during the course of the annual constancy
tests of digital mammographic X-ray equipment according to PAS 1054 (with 8 CR and
12 DR detector systems; the tests with the Slanted Edge insert started later. This
resulted in a weak data basis for the physical methods in respect to the CDMAM analysis).
The tests according to PAS 1054 and the visual evaluation of the CDMAM test phantom
images were conducted by employees of the Prüfstelle für Strahlenschutz (pfs) (a private
German Test Center for Radiation Protection). A test plate for the PAS 1054 test phantom
was developed for these additional exposures. It contains two lead edges that are
imaged with a slight tilt toward the detector matrix (ROI – Region of Interest, of
10 × 30 mm). The test phantom used is described in detail in [9 ]. When the test phantom is correctly positioned, the angle to the edge of the Bucky
table on the chest wall side equals about 3°. To determine the characteristic curve,
a series of exposures was made with different tube loads [mAs] and the target-filter
combination normally used for patient exposures together with a “manual exposure technique”
with each of the tested digital mammographic systems. The field that does not attenuate
the primary radiation was analyzed. The estimated detector dose was used as the corresponding
dose; the entrance surface air kerma (ESAK; without backscattering) on the test phantom
was corrected by the extended distance to the detector as well as with a Bucky factor
of 2. Depending on the detector type, a curve was fitted through the data points with
a linear, logarithmic or potential function. Prior to further evaluation, all images
were transformed into a “dose image,” that is, linearized.
In addition to [18 ], a window function (Hanning Window [19 ]) was applied to the central ROI (see below). The MTF-INDEX is determined as the
average value of the frequency range from 0 lp/mm to the limiting frequency of the
detector. The center of the test plate, 6 cm away from the chest wall with the test
phantom positioned in the middle of the Bucky table, is exposed by the central X-ray
beam. This area of 30 × 30 [mm] is used to determine the NPSQC . A flat field correction (polynomial fit) is applied to the NPSQC ROI while maintaining the average value. The 2-dimensional noise spectrum is calculated
according to [20 ]. To create a 1-dimensional display, the NPSQC values of the same frequency (with the same radius in the 2-D spectrum, center = Frequency
0) are averaged. The median of all NPSQC values was used for the NPS-INDEX to dampen the influence of fixed pattern noise
(FPN) that primarily occurs in the lower frequency ranges.
The terms and definitions of IEC, like DQE, NPS, MTF and NEQ are indexed with QC as
“Quality Control” to stress on the one hand the methodical closeness and on the other
hand the existence of some minor, experimentally based differences to establish these
values.
Based on the DQE measurement, a modified formula according to [20 ] was used to determine the detective quanta efficiency [DQEQC ]: [ ]
with the MTF-Index of the respective better MTF curve (usually horizontal to the chest
wall side), ESAKPMMA is the dose calculated using the characteristic curve from the average pixel value
in the ROI to determine the NPS, and Φin is the interpolated quanta fluence ([Table 1 ]) and the NPS-Index is the median of the NPSQC values across the frequency range under review.
Table 1
The values for the quanta flux, which are presented in the table, are calculated with
an open accessible program of Siemens [25 ] based on the procedure of J. M. Boone [23 ]. The used coefficients of interaction are sourced by NIST [24 ]. The calculations consider the tube filtering, 550 mm air in the course of beam,
1 mm PMMA for the paddle, 46 mm PMMA for the test device, 1 mm for the carbon fiber
laminate of the bucky table and 1 mm Al for the influence of the grid on the beam
quality.
Tab. 1 Die in der Tabelle angegebenen Werte für die Quantenfluenz am Detektoreingang wurden
mit einem öffentlich zugänglichen Computerprogramm von SIEMENS [25 ] berechnet, das auf dem Verfahren von J. M. Boone [23 ] beruht. Die verwendeten Wechselwirkungskoeffizienten stammen von NIST [24 ]. Die Berechnungen berücksichtigen die Zusatzfilterung, 550 mm Luft im Strahlengang,
1 mm PMMA für das Paddel, 46 mm PMMA für den Prüfkörper, 1 mm CFK für den Buckytisch
(Auflageplatte) und 1 mm Al für den Einfluss auf die Strahlenqualität durch das Raster.
quanta fluence [1/(mm2* mGy)]
tube voltage [kV]
target
filter
filter [µm]
26
28
30
32
34
36
Mo
Mo
32
10,624,633
11,654,676
13,072,705
14,639,916
16,372,435
18,093,313
Mo
Rh
25
11,700,324
12,240,600
13,053,155
14,083,138
15,365,921
16,766,246
Rh
Rh
25
12,405,710
12,945,572
13,557,945
14,313,211
15,157,267
16,056,749
W
Rh
50
12,440,765
12,846,260
13,468,413
14,491,935
15,909,125
17,689,924
W
Ag
50
*
*
*
*
*
*
W
Al
250
13,043,945
14,431,527
15,825,607
17,303,299
18,769,454
20,217,760
W
Al
500
13,210,110
14,632,300
16,068,760
17,595,597
19,106,050
20,594,917
* incalculable by the used program, but required in the future
* mit dem verwendeten Programm nicht berechenbar, aber zukünftig benötigt.
In preparing the reference test image, the dose was measured with a calibrated dosimeter
at the position in the phantom provided for this purpose.
The evaluation included only systems that provided access to original data (according
to IEC Definition [15 ], or to DICOM images “for processing” [21 ], or to raw data). In the case of some CR systems, DICOM images “for presentation”
were saved in the constancy testing mode (linear or related designations) without
recognizable image processing. Some of the tested CR units deliver original data,
but nevertheless they were processed in such a way that establishing the characteristic
curve was not possible (Cropping) but the evaluation of the CDMAM images was. The
requirements for visual recognition of gold-plated mammo detail test objects in the
CDMAM test images had to be fulfilled, and the responsible radiologist had to agree
with the preparation of the test images.
Automated Evaluation Methods
All test images were evaluated with the help of an internally developed batch processing
program in which the three methods were integrated: CDCOM (Version 1.5.2), CDIC (Version
3.10) and SE (Slanted Edge). Certain tags from the DICOM File Meta Information (header)
had to be used during processing. The headers of the individual test images were supplemented
as needed whenever tags had not been automatically filled in. The characteristic curve,
pixel value versus detector dose, was automatically determined from the “mAs series.”
The parameters of the linear smoothing functions (linear or logarithmic equations
and/or potential function (y = a*xb )) were recorded in a database. These values are needed for the dose linearization
of the original data.
The evaluation method described in [9 ] for the SE method was modified. Instead of using the average glandular dose (AGD)
to replace the quanta fluence (Φin ) in the detector surface, the quanta fluence at the detector was estimated. Normally,
Φin is determined using a dose and an aluminum half value layer (Al HVL) measurement
at the detector. However, such measurements cannot be made on site due to a lack of
access, at least with DR systems. As a result, Φin was estimated as follows: The entrance surface air kerma (ESAK) was corrected by
the extended distance to the detector (inverse square law, specification of the focal
spot-detector distance in the manufacturer’s documentation, or, in the case of a lack
of access, a fixed value of 65 cm) and an attenuation factor of two was assumed for
the Bucky table (see [22 ]) and grid attenuation. This factor represents an upper limit and most mammographic
X-ray equipment exhibits an effective value that is very close to this value. The
error or measurement uncertainty that occurs due to a possibly overestimated attenuation
or variation between different systems is probably only in the range of a few percent
and is small compared with the uncertainty of the dose measurement. A pixel value
is determined in the open area of the PAS 1054 test phantom that is correlated with
a detector dose ESAKDet. . According to Boone [23 ], the quanta fluence can be calculated from the detector dose and the manufacturer’s
specifications regarding target-filter combination and tube voltage. That means that
by determining the MTFQC , the NPSQC and the Φin , a value can be calculated that roughly approximates the DQE. The value determined
in this way is designated as system power (DQEQC ) to differentiate it from the DQE according to IEC [15 ]. All physical parameters are marked with the index QC, for Quality Control, to differentiate
them from the IEC parameters: DQEQC , NEQQC etc. A correction of the fixed pattern noise to compensate for low frequency interference
as required by IEC cannot be made due to the limited number of images made and the
corresponding area.
The system power at discrete frequencies (1, 2, 3, 4, 5, … lp/mm) according to IEC
[15 ] was also calculated.
Results
Diagnostic Methods
All important data are summarized in [Table 2 ]. As shown in [Fig. 1a ], the values for sensitivity, specificity and accuracy that were determined with
the CDIC program are all at about the same level regardless of the equipment system
used. DR systems exhibit the trend that a higher dose leads to an increase in the
3 parameters. The average values for the three parameters cannot be statistically
separated between the DR and CR systems as the uncertainty ranges overlap. It is striking
that the standard deviation for the sensitivity, the suitability of the method to
detect circular-shaped structures, is significantly greater than for the other two
parameters: 13 to about 2 %. With about 10 analyzed CDMAM test images, the uncertainty
of the average value of the specificity and accuracy parameters is so small that the
plateau value is almost reached. However, the uncertainty of the average value of
the sensitivity parameter remains at a high level regardless of the number of averaged
test images. The average values reduced by the standard deviation (with a confidence
interval of 1σ) may be used as limiting values for acceptance tests. Based on the
above, the following values apply when using the CDIC program: sensitivity 24, specificity
93, and accuracy 66. These values can be achieved if the dose used with CR systems
is approximately double the dose used with DR systems: ca. 6:12 and if the entrance
surface air kerma (ESAK) of DR systems amounts to at least 3 mGy.
Table 2
Results of the analysis with physical and diagnostic methods of the systems under
investigation.
Tab. 2 Ergebnisse der Analyse mit physikalischen und diagnostischen Verfahren bei den untersuchten
Systemen.
system
MTFvertical
MTFhoriz.
DQEQC
NEQQC
diagnostic methods
reader
generator
type
U [kV]
target
filter
MTF-Index
σ %
MTF-Index
σ %
SL
σ (%)
intercept
slope
ESAKDet. [mGy]
for NEQQC
370 000
NEQQC for
CDMAM
expos.
ESAKDet. [mGy]
for CDMAM expos.
sensitivity
specificity
accuracy
S CDCOM
PMS: PCR Eleva
PMS System
CR
26
Mo
Rh
0.22
0.0
0.26
0.0
0.09
0.0
4,288
4,4450
8.23
251,874
5.57
Agfa CR 85
Planmed Sophie Nuance
CR
28
Mo
Rh
0.22
0.7
0.26
0.6
0.12
9.3
69,564
76,757
3.91
391,650
4.20
28.6
94.9
68.4
68.1
Agfa CR 85
No data
CR
28
Mo
Mo
0.23
1.8
0.28
2.3
0.14
1.4
959
106,156
3.48
Agfa CR 35
GE Senographe DR
CR
28
Mo
Rh
0.24
0.5
0.28
0.4
0.12
8.5
61,485
63,664
4.85
367,560
4.81
31.7
95.0
69.6
73.0
Agfa CR 85
SAG Mammomat 3000
CR
27
Mo
Rh
0.25
1.0
0.26
0.8
0.11
13.9
91,905
49,610
5.61
344,918
5.10
31.0
95.4
69.6
73.6
Agfa CR 85
No data
CR
28
Mo
Rh
0.25
1.2
0.27
0.8
0.10
8.0
25,017
48,474
7.12
Agfa DX M
SAG Mammomat 3000
CR
27
Mo
Rh
0.27
0.4
0.29
0.3
0.29
3.2
36,134
132,880
2.51
616,716
4.37
42.8
94.8
74.0
71.4
Agfa DX M
GE DMR
CR
26
Mo
Rh
0.27
0.6
0.34
0.4
0.33
2.1
49,723
158,743
2.02
average CR Systems
0.24
0.9
0.27
0.8
0.19
5.8
4.30
394,543
4.92
30.42
95.09
69.22
71.56
standard deviation
0.03
0.03
0.11
standard Dev. (%)
39
34
12
5
0
1
4
modality
Planmed SophieNuance
DR
28
Mo
Mo
0.66
0.2
0.70
0.4
0.17
10.7
–57,317
76,330
5.60
SAG Mammomat Novation DR
DR
28
Mo
Mo
0.70
0.3
0.70
1.1
0.10
41.9
–92,006
50,093
9.22
Planmed SophieNuance
DR
28
Mo
Mo
0.66
0.3
0.70
0.4
0.17
7.3
–57,317
76,330
5.60
SAG Mammomat Inspiration
DR
26
Mo
Mo
0.68
0.1
0.74
0.2
0.11
5.4
–202,015
98,991
5.78
SAG Mammomat Inspiration
DR
28
Mo
Mo
0.69
0.3
0.74
0.2
0.11
8.1
–190,001
116,416
4.81
SAG Mammomat Inspiration
DR
30
Mo
Mo
0.69
0.6
0.74
0.2
0.09
3.7
–221,958
237,248
2.50
GE Senographe 2000
DR
28
Rh
Rh
0.57
4.2
0.60
2.5
0.13
43.0
–198,254
128,994
4.41
117,782
2.45
40.1
95.3
73.3
76.9
GE Senographe 2000 D
DR
28
Rh
Rh
0.57
4.2
0.57
0.2
0.16
18.9
–232,713
179,099
3.37
GE Senographe Essential
DR
26
Rh
Rh
0.60
0.60
0.12
*
GE Senographe Essential
DR
28
Rh
Rh
0.65
0.0
0.65
7.2
0.17
29.1
–235,693
138,048
4.39
GE Senographe Essential
DR
30
Rh
Rh
0.64
0.63
n. b.
0.19
*
GE Senographe Essential
DR
32
Rh
Rh
0.67
0.65
n. b.
0.18
*
Sectra Imtec AB
DR
35
W
Al
0.39
2.1
0.49
0.3
0.13
0.005
161,845
579,948
0.36
1,381,074
2.10
38.9
97.0
73.8
80.2
SAG Mammomat Novation DR
DR
28
W
Rh
0.70
0.2
0.73
0.5
0.22
16.4
30,878
82,154
4.13
252,063
2.69
35.2
96.7
72.1
72.7
Hologic Selenia Dimensions
DR
28
W
Rh
0.61
0.4
0.60
0.8
0.10
31.2
–108,991
113,216
4.23
SAG Mammomat Inspiration
DR
29
W
Rh
0.66
1.3
0.72
6.6
0.19
31.9
–159,409
226,354
2.34
317,675
2.11
29.9
96.8
70.1
77.6
PMS Mammodiagnost DR
DR
26
W
Rh
0.68
0.4
0.74
0.2
0.31
3.4
–78,637
287,033
1.56
PMS Mammodiagnost DR
DR
28
W
Rh
0.68
0.0
0.74
0.0
0.30
4.8
–53,542
303,226
1.40
774,498
2.73
35.5
95.5
71.5
78.2
PMS Mammodiagnost DR
DR
30
W
Rh
0.68
0.0
0.74
0.0
0.29
2.9
–91,140
398,704
1.16
average DR Systems*
0.66
0.9
0.67
1.3
0.18
16.4
3.64
365,504*
2.50
35.9
96.3
72.1
77.1
standard deviation
0.04
0.07
0.07
standard Dev. (%)
59
78
11.5
3.98
0.81
1.47
2.75
*without Sectra
*without Sectra
* Due to preprocessing not determinable
* Aufgrund von Vorverarbeitung nicht bestimmbar
Fig. 1 a Comparison of the results of an automatic evaluation of CDMAM test images of different
DR and CR systems, which are yielded with the program CDIC (CUAS), b Comparison of the results of an automatic evaluation of CDMAM test images of different
DR and CR systems, which are yielded with the program CDCOM (EUREF).
Abb. 1 a Vergleich der Ergebnisse einer automatischen Auswertung von CDMAM Prüfkörperaufnahmen
mit dem Programm CDIC (FHK) von unterschiedlichen DR- und CR-Systemen, b Vergleich der Ergebnisse einer automatischen Auswertung von CDMAM Prüfkörperaufnahmen
mit dem Programm CDCOM (EUREF) von unterschiedlichen DR- und CR-Systemen.
In evaluations performed with the CDCOM program ([Table 2, ]
[Fig. 1b ]), DR systems do not exhibit a recognizable dose dependence. The distribution around
the average value equals ca. 6 %. A value of 70 should be used as the lower limiting
value for sensitivity SCDCOM . There is no need to mathematically adjust the results to visual evaluation methods
as such adjustments significantly depend upon the visual evaluation method selected.
Physical Methods
Values for NEQQC and DQEQC can be determined with the values measured for MTF-INDEX, NPS-INDEX and dose as well
as by using the values for quanta fluence ([Table 1 ]). The uncertainty in determining the MTF is very small (standard deviation of 6
test images of about 1 %), and the values determined are independent of dose, target-filter
combination (TFC) and tube voltage ([Table 3 ]). Nearly identical values are achieved for the respective detector types of the
different equipment systems ([Table 3 ]). The idea of horizontal and vertical edges that was originally developed for testing
CR systems also proved effective in testing the Sectra System in which the MTF characteristic
curves differ significantly from one another ([Fig. 2 ]). The uncertainty in determining the NPSQC was estimated using the standard deviation of the values of a potential fit function
over the upper half of the NPSQC spectrum. Depending on the system, relative fluctuations of 2 – 5 % of the polynomial
fit can be observed. The uncertainty of the value for NPS-INDEX can be reduced to
under 1 % by using the quantity of up to one hundred interpolation points. The greatest
uncertainty in determining detective quanta efficiency DQEQC results from the dose measurement. The uncertainty from the calibration and the determination
of corrective factors for different radiation qualities (TFC, Al HVL, X-ray tube voltage
U) dominate the overall uncertainty of the method. The estimation of input fluence
is subject to slighter uncertainties but is primarily dominated by PMMA filtration.
Once NEQ and dose are linked linearly with each other, the sensitivity to changes
in dose or detector sensitivity lies in a range of just a few percentage points.
Table 3
Results of the determination of the MTF-Index with respect to different tube loadings
and voltages on the system: MammoDiagnostDR (manufacturer: Philips Medical Systems).
Tab. 3 Darstellung der Unabhängigkeit der Werte für den MÜFvert. -Index von unterschiedlichen Spannungen und Ladungen am Beispiel des MammoDiagnostDR
(Hersteller: Phillips Medical Systems).
Tube loading Q
Tube voltage U
ESAK
MTFvert.-Index
[mAs]
[kVp]
[mGy]
71
28
2.16
0.744
80
28
2.43
0.740
90
28
2.74
0.737
100
28
3.04
0.739
110
28
3.34
0.737
125
28
3.8
0.737
125
26
3.07
0.738
140
26
3.44
0.742
160
26
3.93
0.744
80
30
2.61
0.738
90
30
2.93
0.734
100
30
3.26
0.737
average
0.739
standard dev.
0.003
standard dev.%
0.41
uncertainty U
0.001
U%
0.12
Fig. 2 Vertical and horizontal MTF of the Sectra System as an example for an asymmetric
MTF of a DR system.
Abb. 2 Vertikale und horizontale MTF des Sectra Systems als ein Beispiel für eine asymetrische
MTF eines DR Systems.
The values of the physical parameters of the CR systems exhibit very similar values,
especially the NEQQC value. The values of the Agfa DM X System with “Needle Phosphor” technology are also
very close to the other CR results when one considers that the system has a significantly
better DQE and is operated with entrance doses that are used with “Powder Phosphor”
systems. The DR systems exhibit wider ranges; the Sectra System should be retested
due to the suspicion of image processing (see also asymmetric MTF results); the GE
Senographe 2000 seems to be an example of an underexposed DR System.
The minimum limiting value for NEQQC that must be achieved is approximately 370,000. This value is not an averaged value
from measured data, but rather it’s a preliminary fixed one under special respect
of the values of CR systems. Some DR systems showed some irregularities which could
not be explained without new test series and a broader data base. If, in addition,
the detector-specific values for MTF-INDEX are achieved, one can be sure that the
equipment systems will produce diagnostic-level image quality regardless of the respective
detector technology.
Discussion
Automated analysis of test phantom images must fulfill three criteria: objectivity,
reliability and validity. Automated analysis provides objectivity for all physical
methods. Reliability is also ensured as uniform threshold values, algorithms, etc.
are used. Validity is achieved by determining the dose (in line with IEC written as
d), the MTFQC and the NPSQC as they are clearly linked with the DQEQC and the NEQQC :[ ]
with d2 = Signal (mGy)
The first two criteria are fulfilled by diagnostic methods. The third criteria can
only be determined by comparison with physical methods.
In performing automated evaluation, the tags in the DICOM header should be used to
accelerate the method. Missing or incorrect tags significantly disrupt automated evaluation.
We determined during this study that one cannot assume that tags are complete and
correct and that possible tag entries in certain fields according to DICOM do not
always lead to a logically clear classification of the systems: e. g. tag 0008;0060
– Modality. In this case, one cannot clearly determine whether a DR or a CR system
is involved.
The manufacturers’ claim of “DICOM conformity” for all of the equipment systems does
not ensure that automated evaluation is possible. We also determined at the same time
that the term “original images” definitely has different meanings. For instance, operations
such as histogram cropping are performed making the plotting of a characteristic curve
impossible and preventing further physical analyses.
To enable testing of the physical quality of the detector, all manufacturers must
comply with the IEC requirements for original images and the DICOM standard for header
tags. The EUREF Group will include the requirement in the new supplement (to be published
probably by year 2011) that the original images must be made available.
Diagnostic Methods
Automated diagnostic methods exclude the influence of the viewer and have high reproducibility.
However, the values for standard deviation in evaluating different images produced
with the same exposure parameters and slight shifts in phantom position are finite:
for sensitivity with CDIC about 10 % and for accuracy about 1 – 2 % and about 5 %
with CDCOM. This requires that a sufficiently large number of CDMAM test images must
be made to ensure that the uncertainty of the result remains at a reasonable level:
ca. 10 test images also for automated evaluation. All three parameters, the dose measurement,
the phantom and the method influence the accuracy, that is, influence the cross-comparability
of the results. It should be noted that there are no systematic studies about how
diagnostic methods using different detection methods react to interference (FPN and/or
TN, MTF changes, etc.), i. e., about the validity of these methods. See [Fig. 3 ] for two sample results of CDIC in which the results of an analysis are presented:
gray shaded circles represent the true positions of gold discs (type 3.14). Colored
dots – blue in the center and red in corners – represent findings (TP and FP).
Fig. 3 Results of CDIC, a Agfa CR 85 Planmed Sophie Nuance, b Siemens Mammomat Novation DR.
Abb. 3 Ergebnisse von CDIC, a Agfa CR 85 Planmed Sophie Nuance, b Siemens Mammomat Novation DR.
The EPQC method uses two different limiting values. The ratings “acceptable” and “achievable”
should be used with diagnostic methods. “Achievable” should correspond with the average
value and “acceptable” with the average value reduced by the standard deviation.
Physical Methods
Characteristic curve: We determined that when using standard exposure parameters for
the CDMAM test image, the open area in the aluminum step wedge region was frequently
overexposed and could not be used for the evaluation. Therefore, the exposure parameters
for the CDMAM test image should be used as maximum values, and subsequent test images
be made with a lower dose.
Determination of the MTFQC or the NPSQC does not dominate the uncertainty of NEQ QC , but rather the uncertainty of the dose measurement is the limiting factor. The value
of NEQQC is a required parameter to describe detector-related image quality, but has to be
completed by an additional value which gives information regarding the imaged dynamic
range of glandular tissue respective to the contrast in the image. A specific test
procedure is under development.
Test equipment: The edge can be produced with an accuracy that fulfills the IEC requirements
(5 – 8 µm according to manufacturer's data from PEHA med. Geräte GmbH, Sulzbach).
That means that the edge testing device does not notably influence the test result.
This means in practice that as a rule, only a few test images that can be used for
the evaluation must be made and that they are very dose-sensitive. As modern generators
and dosimeters have very high reproducibility, a deviation of just a few percentage
points in NEQQC can indicate that the ROI was not correctly positioned when evaluating the image.
In such cases, additional test images must be made to exclude incorrect measurements.
Cross-Comparability
Theoretically, one would expect that systems that fulfill the requirements for image
quality and dose would exhibit a close correlation between the dose required to produce
CDMAM test images and the NEQQC calculated using the same dose. However, as the CDMAM test is not particularly dose-sensitive,
one can only expect a rough correlation between the dose required for a certain NEQQC and the dose with which the CDMAM test images were made. This expectation was confirmed
([Fig. 4 ]). One may assume that a minimum NEQQC value of 370,000 must be achieved.
Fig. 4 Comparison of NEQQC yielded with the dose of CDMAM exposures with the results of diagnostic methods (Sectra
results included).
Abb. 4 Vergleich der NEQQC die mit der Dosis erzielt wird, mit der die CDMAM Aufnahmen angefertigt wurden mit
den Ergebnissen der diagnostischen Verfahren (Sectra Ergebnisse eingeschlossen).