Key words
intramammary clip marker system - ultrasound-guided core needle biopsy - digital breast
tomosynthesis (DBT)
Schlüsselwörter
intramammäre Clipmarkierung - sonografisch gesteuerte Stanzbiopsie - digitale Brusttomosynthese
(DBT)
Introduction
Around 55 000 women develop breast cancer and 20 000 women die of it in Germany every
year [1]. The individual prognosis of patients with breast cancer depends in the first instance
on the extent of disease at the time of diagnosis and on the implemented treatment
plan [2], [3], [4].
According to both European treatment recommendations and the German interdisciplinary
S3 Guideline for the Diagnosis, Treatment and Follow-up of Breast Cancer, at least
70 % of all breast lesions suspicious for malignancy (BI-RADS™ 4/5) should be verified
histologically prior to surgery; the ideal would be to verify 90 % [5], [6]. The goal of preoperative investigation of suspected malignant processes is to ensure
that only one surgical procedure will be required subsequently. In addition, all non-palpable
breast lesions should be marked prior to the actual surgical procedure (e.g. with
ultrasound-guided wire marking) [5], [6].
Accurate preoperative ultrasound-guided wire marking is particularly tricky in smaller
focal findings with diameters of less than 1 cm. Neoadjuvant therapy concepts represent
new challenges for breast surgeons, radiologists and pathologists, as there is no
target for preoperative, ultrasound-guided wire marking in patients in complete pathological
remission (pCR).
This difficulty can be remedied using a marker clip to locate the primary breast tumor
during ultrasound-guided core needle biopsy prior to surgical therapy or neoadjuvant
chemotherapy. Clip marking of the primary tumor should be done together with the ultrasound-guided
core needle biopsy to reduce re-excision rates [7], [8], [9], [10].
Based on the findings of our previous studies and using a clip marking system we helped
to develop and establish [11], [12], we investigated the accuracy of an innovative clip marking method in patients with
breast cancer (cT2) scheduled to undergo neoadjuvant chemotherapy [13]. Our results confirmed that the innovation we had developed had precisely adapted
the stylet length of the marking system to the single-use breast biopsy system (HistoCore™).
Previously, when using the O-Twist Marker™ clip system, a pre-fabricated sliding spacer
was placed directly around the stylet, and the notch length of the spacer used during
marker clip placement (pushed forward through the coaxial biopsy needle in situ) had
to be varied for every intervention, depending on the respective length of the biopsy
needle. This approach was imprecise and semi-subjective because it was not possible
to adjust the fitted spacer precisely using only the marking ring engraved at intervals
of one centimeter. Our innovation offered the possibility of placing a precise marker
clip for every biopsy needle length without requiring a spacer. This creates a target
point for preoperative ultrasound-guided wire marking, which is particularly useful
in patients in pathological complete remission (pCR) after neoadjuvant chemotherapy
[7], [8], [9], [10].
If standard control mammography (two orthogonal planes, cranio-caudal and medio-lateral
oblique views) is done after stereotactic vacuum biopsy with clip marking [9], the question arises whether it is possible to detect dislocation of an intramammary
marker clip after ultrasound-guided core needle biopsy with clip placement. Based
on our previous experience [11], [12], [13], [14] with intramammary clip marking systems, the aim of our current study was to investigate
the accuracy of sonography and digital breast tomosynthesis to determine the location
of intramammary marker clips placed in small breast cancers with diameters of less
than 2 cm (cT1).
Material and Methods
Patient population
Between June 2013 and November 2013 50 patients with a suspicion of invasive breast
cancer and focal findings with diameters of less than 2 cm (cT1, BI-RADS™ 4/5) were
investigated at the University Breast Center Franconia, using complementary breast
diagnostics consisting of clinical examination, mammography (Selenia Dimensions3D™
[Hologic™]) and sonography (2-D, Acuson Antares, 13 MHz [Siemens™]). Lesion diameters
ranged from 0.4 to 1.6 cm (median 0.9 cm) and patient age range was 31–77 years (median
54 years).
Ultrasound-guided core needle biopsy and clip marking
Invasive breast cancer was confirmed histologically in all 50 patients by ultrasound-guided
core needle biopsy using a single-use breast biopsy system (HistoCore™, BIP™ Biomedizinische
Instrumente & Produkte GmbH, Germany) [11]. Patients then underwent breast-conserving surgery without prior neoadjuvant chemotherapy.
Intramammary clip marking was additionally done using a directly adapted clip system
based on the established O-Twist Marker™ system (BIP™ Biomedizinische Instrumente
& Produkte GmbH, Germany) [12], [13]. All ultrasound-guided core needle biopsies and clip markings were performed by
the same two experienced investigators (RSW, BB) to exclude potential inter-investigator
variability.
A single-use breast biopsy system (HistoCore™) was used with a 12-gauge, 10-cm outer
cannula and a needle advance of 18 or 25 mm. After careful disinfection of the skin
and administration of a local anesthetic, the single-use breast biopsy system (combination
of a coaxial cannula [11 gauge] and the core biopsy needle [12 gauge]) was placed
over the focal tumor. Core needle biopsy was carried out under ultrasound control
tangentially to the linear 13.0-MHz transducer. The length of the needle was documented
before and after the intervention on pictures and on video. Four or more core needle
biopsy specimens were obtained to secure sufficient material for histological diagnosis
and molecular-genetic testing. Using the coaxial needle (11 gauge) and the adapted
clip system (O-Twist Marker™), a clip was placed directly in the puncture site, i.e.
in the middle of the tumorous lesion, under “real-time” ultrasound guidance for subsequent
control investigations of clip localization.
Location control of marker clips using sonography and digital breast tomosynthesis
Location of the intramammary marker clip was controlled by sonography (2-D, Acuson
Antares, 13 MHz [Siemens™]) and digital breast tomosynthesis (Selenia Dimensions 3D™
[Hologic™]) approximately 30 minutes after the intervention. A compression bandage
was used in all patients to minimize hematoma formation. As part of our study, digital
breast tomosynthesis with the same radiation dose was performed in each patient (after
informed consent) instead of control mammography (two orthogonal planes, cranio-caudal
and medio-lateral oblique view), as recommended in the German interdisciplinary S3
Guideline for the Diagnosis, Therapy and Follow-up Care of Breast Cancer [6], [15]. Measurements were also carried out by the same two investigators (RSW, BB) to reduce
inter-investigator variability.
Statistical analysis
SPSS 18.0 was used for statistical analysis. Data were initially analyzed descriptively
and subsequently assessed for statistically relevant differences between investigated
groups. The level of significance was p < 0.05.
Results
Comparison of sonography versus digital breast tomosynthesis to locate intramammary
marker clips
After placing the marker clip in the center of each tumor, the location of the marker
clip was controlled using both ultrasound and digital breast tomosynthesis ([Figs. 1] and [2]). No dislocation of the marker clip after removal of the clip applicator was observed
during ultrasound monitoring.
Fig. 1 a and b a Ultrasound-guided core needle biopsy of a breast cancer lesion (cT1). The biopsy
needle is precisely in the center of the focal tumor (red arrow) (2D, Acuson Antares,
13 MHz (Siemens™), magnification: 3×). b Ultrasound localization of the intramammary marker clip. The biopsy needle tip is
sited directly above the focal tumor (red arrows) (2D, Acuson Antares, 13 MHz [Siemens™],
magnification: 3×) after placement of the marker clip (red circle). Virtual extension
of the direction of dislocation of the marker clip (light blue arrow); position of
the marker clip (purple circle); dislocation of the marker clip (3D reconstruction)
by 2 mm along the x-axis (white arrow), by 1 mm along the y-axis (yellow arrow), by
3 mm along the z-axis (green arrows) (dislocation).
Fig. 2 Localization of the intramammary marker clip using digital breast tomosynthesis in
the same patient as in [Fig. 1]. The sharply delineated marker clip (circle) is located directly in the center of
the lesion; no dislocation has occurred (Selenia Dimensions 3D™ [Hologic™] with CAD™
[R2™]). Secondary findings include three areas with micro-calcifications (dotted lines).
In 42 of 50 patients (84 %), ultrasound (measurement of the tumor without the echo-rich
margin) detected no dislocation of the marker clip. In eight patients (16 %) ultrasound
control indicated a maximum dislocation of 7 mm along the x-, y- or z-axis ([Table 1]).
Table 1 Measured dislocation of the intramammary marker clip in mm (along the x-, y- and
z-axis); dislocation was observed in 8 patients using ultrasound and in 2 patients
using digital breast tomosynthesis (DBT). The intratumorous location of the marker
clip was confirmed by specimen x-ray (SX) and histopathological examination (Histo.)
(out of a total of 50 examined patients).
Patient no.
|
Ultrasound
|
DBT
|
SX
|
Histo.
|
|
X
|
Y
|
Z
|
X
|
y
|
z
|
|
|
1
|
1
|
2
|
3
|
0
|
0
|
0
|
✓
|
✓
|
2
|
5
|
3
|
3
|
0
|
0
|
0
|
✓
|
✓
|
3
|
4
|
6
|
5
|
0
|
0
|
0
|
✓
|
✓
|
4
|
5
|
7
|
6
|
0
|
0
|
0
|
✓
|
✓
|
5
|
1
|
2
|
2
|
0
|
0
|
0
|
✓
|
✓
|
6
|
1
|
3
|
5
|
0
|
0
|
0
|
✓
|
✓
|
7
|
7
|
1
|
2
|
0
|
0
|
0
|
✓
|
✓
|
8
|
1
|
6
|
4
|
0
|
1
|
3
|
✓
|
✓
|
9
|
0
|
0
|
0
|
1
|
1
|
2
|
✓
|
✓
|
Digital breast tomosynthesis showed precise placement without dislocation of the marker
clip in 48 patients (96 %); a maximum dislocation of 3 mm along the x-, y- or z-axis
was found in two patients (4 %) ([Table 1]). In six patients (12 %) ultrasound wrongly indicated dislocation of the marker
clip. The reason for this was probably that the tumor size in these patients was between
0.4 and 0.7 cm (median 0.6 cm).
The difference in the accuracy of the two imaging techniques to locate marker clips
was statistically significant (p < 0.05). The intratumorous location of the marker
clip was confirmed both by specimen x-ray done intraoperatively and by the subsequent
histopathological examination of the tumor. All patients underwent breast-conserving
therapy, and specimen x-rays were correlated with findings of digital breast tomosynthesis
and histopathological examination of the specimens.
Discussion
The aim of this study was to investigate the accuracy of different imaging techniques
when determining the precise position of marker clips placed directly in the center
of intramammary lesions with diameters of less than 2 cm (cT1). Our recent development,
which combines a single-use core biopsy needle with a precisely adapted clip marker
system for the placement of marker clips through a coaxial needle in situ, allows
clips to be placed accurately even in small lesions (cT1). Control imaging using ultrasound
and digital breast tomosynthesis to verify the position of the marker clip additionally
confirmed the high diagnostic accuracy of digital breast tomosynthesis.
Ultrasound-guided core needle biopsy is currently the standard approach for the diagnostic
workup of unclear lesions in the breast. Several studies have shown that an identical
or higher degree of diagnostic certainty can be achieved with ultrasound-guided core
needle biopsy compared to open biopsy of palpable and non-palpable findings, as the
latter are associated with false-negative rates of between 0.3 and 8.2 % [16], [17]. Other studies have shown that local rates of recurrence and the intervals to tumor
recurrence did not differ between patients diagnosed using ultrasound-guided core
needle biopsy and those diagnosed by primary surgical intervention [18], [19], [20], [21].
The advantages of ultrasound-guided core needle biopsy, which can be carried out with
little expenditure of time, are its limited invasiveness and lower costs [22], [23]. Knowledge of the tumorʼs histological characteristics allows better planning of
surgical operations, if surgery is required, and a more targeted intervention, particularly
with regard to axillary sentinel lymph node biopsy. This is reflected in the lower
rates of follow-up surgeries for incomplete tumor resection [24]. As around three quarters of masses detected on imaging are benign, this can prevent
unnecessary surgical interventions, provided that the assessment of the images of
the lesion concurs with the histological findings.
Although ultrasound is a well-established diagnostic method, the range of indications
and the value of digital breast tomosynthesis has not yet been precisely determined
[15], [25], [26], [27].
Digital image acquisition, image processing and image reproduction allows many sequences
of images to be acquired within a short time. The compilation of sequentially acquired
tomograms is used to create three-dimensional images of the breast, so-called digital
breast tomosynthesis [25]. The radiation exposure parameters for every plane are selected to ensure that total
radiation exposure corresponds to the radiation dose of two-plane mammography. Various
reconstruction algorithms are used to display the breast as a series of slices at
different depths or as a freely rotatable 3-dimensional image. Digital breast tomosynthesis
thus redresses one of the limitations of mammography, namely, the reduction of the
three-dimensional breast to a two-dimensional image [25]. Digital breast tomosynthesis should therefore be able to avoid false-positive or
false-negative findings which are the result of superimpositions created with mammography.
This should improve detection rates in mammographically denser breasts [15]. Digital breast tomosynthesis also appears to be superior to both sonography and
mammography for the determination of the size of breast tumors [15]. Other authors have shown that the recall rate for mammography screening was reduced
when mammography was combined with tomosynthesis [28], [29], [30], [31].
Our findings show that determination of the location of marker clips placed after
ultrasound-guided core needle biopsy is significantly more precise using digital breast
tomosynthesis compared to ultrasound. One reason for this could be because digital
breast tomosynthesis is standardized compared to elastography or ultrasound, where
the application is semi-subjective, irrespective of whether it is 2-D, 3-D, or 4-D
imaging [32], [33]. It remains to be seen whether modern three-dimensional ultrasound systems such
as the Automated Breast Ultrasound Scanner (ABVS) will improve the results of conventional
two-dimensional ultrasound even more [34], [35].
Conclusion
Digital breast tomosynthesis is a good example of recent developments in breast imaging
techniques. These developments continue to improve the quality of diagnostics and
the therapy of breast cancer while reducing the rate of surgical re-excisions, particularly
for small breast cancer tumors or after neoadjuvant chemotherapy with subsequent pathological
complete remission (pCR). The findings of our study on locating intramammary marker
clips has expanded the range of indications for digital breast tomosynthesis.