Introduction:
The purpose of this study was to evaluate the performance of the EUROArray HPV genotyping
assay (Euroimmun) against the HPV 3.5 LCD-Array Kit (Chipron) from liquid based cervical
cytology samples clinician collected from women undergoing cervical inspection at
the dysplasia consultation.
Methods:
DNA extracted from clinician-collected cervical brush specimens from 163 German women
were evaluated by EUROarray HPV and HPV 3.5 LCD-Array Kit for detection of HR (high-risk)-
and LR (low-risk)-HPV genotypes.
Genotype-specific agreements were assessed by Cohen's kappa statistic and Fisher's
z-test of significance between proportions.
Results:
Positivity for 17 HR-HPV types was 42.3% for EUROarray and 38.0% for HPV 3.5 LCD-Array.
Agreement between the EUROarray and the HPV 3.5 LCD-Array for detection of HR-HPV
was very good (κ= 0.83). For detection of HR-HPV, agreement with EUROarray was 92.0%.
Identical genotypes of HR-HPV were detected in 50 of 72 samples with any detected
HR-HPV (81.9%, κ= 0.47). EUROarray genotyping for HPV16 and/or 18 was highly concordant
with Chipron (relative agreement 96.3%, κ= 0.88).
Conclusion:
Detection of HR-HPV was not significantly different between EUROarray and HPV 3.5
LCD-Array. EUROarray was highly concordant with the HPV 3.5 LCD-Array evaluated for
detection of high-risk HPV.