Keywords
BioFoam - hemostatic agent - surgical sealant - cardiovascular surgery
Introduction
Cardiac and cardiovascular surgical procedures typically involve closure of the repaired
or grafted tissues with sutures or staples which, despite careful application, can
often leave small gaps through which blood may ooze slowly or profusely. Failure to
rapidly and effectively control anastomotic bleeding can result in deleterious clinical
sequelae that may include anemia, hemodynamic instability, hypothermia, hypovolemia,
reduced oxygen delivery to tissues, impaired visualization of the surgical field,
and an increase in the duration of surgery.[1] Moreover, uncontrolled bleeding can result in the need for a blood transfusion or
reoperation and has been shown to be associated with an increase in the risk of mortality
and a higher cost of treatment.[1]
The common occurrence of anastomotic bleeding following cardiovascular surgery and
the potentially grave consequences for the patient have driven the development and
availability of an array of hemostatic methods. The methods range from simple application
of manual pressure at the anastomotic site to the topical use of mechanical sealants
and pharmacological agents. In addition to the application of pressure with a finger,
sponge, clips, or sutures, other mechanical approaches include electrosurgery, laser,
radio frequency energy, argon beam coagulation, ultrasonic scalpel, and ultrasonic
surgical aspirator.[1] Sealants that create a physical barrier to staunch the flow of blood from the anastomotic
site include gelatins, collagens, oxidized celluloses, synthetic glues, and glutaraldehyde-based
glues. Thrombin and fibrin are the main pharmacological agents applied topically to
promote coagulation of blood at the anastomotic site.[1] The choice of the hemostatic approach varies greatly between institutions and is
generally based on the specific nature of the anastomotic site, the extent and speed
of blood loss from the site, and the surgeon's experience and knowledge of available
hemostatic methods.[2]
Hemostatic agents have become established as a valuable adjunct to standard methods
of controlling anastomotic bleeding in liver and spleen surgical procedures and are
now increasingly used in cardiac and cardiovascular surgical procedures. The effectiveness
of hemostatic agents in cardiovascular surgery has been reported in several clinical
studies. In a prospective, multicenter, randomized study with 333 subjects, a fibrin
sealant was found to control bleeding within 5 minutes of application in 92.6% of
cases compared with only 12.4% of cases with the conventional agents.[3] Waragai et al reported a study with 112 subjects in which microporous polysaccharide
hemostatic (MPH) bandage was compared with standard manual compression. Significantly
fewer subjects treated with the MPH bandage required compression for 15 or more minutes
(p = 0.006) and significantly more subjects treated with MPH had a shorter time to hemostasis
compared with compression alone (p = 0.048).[4] In a study with 20 subjects undergoing carotid endarterectomy, the use of Quixil
surgical sealant was found to greatly reduce the time to hemostasis (p < 0.001) and the amount of blood loss (p < 0.001) when compared with the use of a standard topical hemostatic agent.[5] These published studies and others have established a precedent for the use of hemostatic
agents following a range of cardiovascular surgical procedures.
BioFoam surgical matrix further referred to here as BioFoam comprises two solutions
contained separately in a double-barreled syringe: one barrel contained the protein
bovine serum albumin (BSA) and sodium bicarbonate dissolved in water; the second barrel
contained the cross-linking agent glutaraldehyde and acetic acid dissolved in water.
BioFoam is generated when the plunger is pushed forcing the two solutions into a double-helix
design tip which ensures their thorough mixing. The rapid reaction of sodium bicarbonate
and acetic acid generates carbon dioxide which immediately expands the cross-linked
BSA–glutaraldehyde complex into a flexible, mixed cell foam approximately three times
the volume of the original solutions. Tests have shown that BioFoam has a burst strength
of over 200 mm Hg which enables it to withstand postoperative hypertension.[6] The mixed-cell foam presents both a mechanical barrier to stop blood from oozing
from the anastomotic site and a mixed-cell porous structure which allows entry of
blood into the matrix where it aggregates. The presentation of BioFoam as preprepared
solutions in a syringe that can be stored at room temperature enables its rapid deployment
during surgery, while mixture of the solutions in the application tip ensures minimal
wastage.
The present study was conducted to investigate the safety and effectiveness of BioFoam
Surgical Matrix as a surgical adjunct to achieving anastomotic hemostasis following
cardiovascular surgery ([Figs. 1] and [2]).
Fig. 1 BioFoam on needle hole bleeding.
Fig. 2 BioFoam on anastomotic suture line.
Materials and Methods
A prospective, multicenter, single-arm study was conducted to assess the safety and
effectiveness of BioFoam used as a surgical adjunct to achieve anastomotic hemostatis
following cardiovascular surgery. A total of 75 subjects were treated with BioFoam
at two centers: 15 at the German Heart Centre, Munich, Germany; and 58 at the Ospedale
San Raffaele, Milan, Italy.
The study was conducted in compliance with the protocol, the International Conference
on Harmonization—Good Clinical Practice , and all applicable regulatory requirements.
Participating investigating sites were responsible for complying with applicable regional
or national regulations governing the conduct of postmarketing surveillance studies.
The study was conducted in accordance with the World Medical Association Declaration
of Helsinki adopted by the 18th WMA General Assembly in June 1964 and all amendments
thereafter. The respective ethics committee of each clinical center approved the study
protocol and all subjects provided their written informed consent before study enrolment.
The study was listed on ClinicalTrials.gov (NCT02164201).
Patients included in the study were ≥ 18 years of age and scheduled to undergo an
elective cardiovascular surgical procedure which could include but was not limited
to thoracic aortic aneurysm, aortic valve replacement, and Type A aortic dissection
(where the use of BioFoam was limited to the anastomotic site). Intraoperatively,
the inclusion criterion was generalized oozing from the anastomotic repair site following
use of standard repair procedures (such as sutures and staples) and for which the
surgeon considered it necessary to use a hemostatic agent.
Exclusion criteria were: known hypersensitivity to albumin, bovine products, or glutaraldehyde;
an active infection (either systemic or in the repair region); a pathology or underlying
disease which made them unsuitable for clinical investigation as judged by the Investigator;
a coagulation disorder; abnormal calcium metabolism (e.g., chronic renal failure or
hyperparathyroidism); life expectancy less than that required for the follow-up duration;
were pregnant, planning to become pregnant during the follow-up period, or actively
breast feeding; or immunocompromised. No subjects were entered into the study on an
“emergency use” basis. Subjects were also excluded if they experienced a major intraoperative
bleeding incidence as defined by the American College of Surgeons Advanced Trauma
Life Support Class II, III, or IV Hemorrhage.
The intent to treat population included all subjects enrolled and the per protocol
cohort included subjects for whom there were no major protocol deviations including
failure to meet any of the preoperative or intraoperative inclusion criteria, or any
informed consent violation.
The hemostatic agent under investigation in this study was BioFoam Surgical Matrix,
a combination of BSA and glutaraldehyde with foaming agents, as previously described.
Investigators who were experienced and skilled in performing cardiovascular surgery
were trained on the use of BioFoam per the Instructions for Use before participation
in the study.
After completion of surgery, if a hemostatic agent was required to stop oozing from
the anastomotic repair site, a temporary dry field was created with a swab or suction.
BioFoam was then immediately applied to the anastomotic site and the region was closely
observed at 1, 3, 5, 7, and 10 minutes. Time to hemostasis was recorded. All subjects
were followed up for 30 days following surgery and assessed either during a patient
visit to the hospital or by the surgeon directly speaking to the patient by phone.
The primary effectiveness end point was achievement of hemostasis (yes/no) of the
repaired anastomotic site at 3 minutes after a single application of BioFoam. The
secondary effectiveness end point was time to hemostasis assessed at 1, 5, 7, or 10
minutes (yes/no) after a single application of BioFoam ([Table 1]). When hemostasis was achieved, observation was continued for a further 1 minute
to confirm cessation of anastomotic bleeding. Immediate hemostasis was considered
to have taken place when bleeding stopped within 1 minute after a single application
of BioFoam and there was no oozing during a further 2 minutes of observation.
Table 1
Time to hemostasis
Time to hemostasis
(min)
|
Number of subjects
(%)
n = 74[a]
|
1
|
40 (54.1)
|
3
|
22 (29.7)
|
5
|
7 (9.4)
|
7
|
0 (0)
|
10
|
0 (0)
|
> 10
|
1 (1.4)
|
N/A[b]
|
4 (5.4)
|
a One subject was excluded as the time to hemostasis had not been recorded, although
the achievement of hemostasis was reported for this patient.
b These subjects required reoperation for bleeding within 24 hours after the surgery.
None of these events was considered by the Investigator to be associated with BioFoam.
The following assessments were made to evaluate the safety of BioFoam during this
study: intraoperative blood product administration; use of alternative means to achieve
hemostasis after application of BioFoam; the incidence of reoperation for bleeding;
the total time of the operative procedure; the total hospitalization time; any additional
hospitalization or surgical procedure through to the final follow-up period; the incidence
of procedure complications; or adverse events (AEs) through to the final follow-up
period. All AEs and serious adverse events (SAEs) were recorded and causal relationship
with BioFoam was judged by the Investigator ([Tables 2] and [3]).
Table 2
Adverse events
Type of event
|
Number of subjects (%)
n = 75
|
Atrial fibrillation
|
11 (14.6)
|
Fever[a]
|
1 (1.3)
|
Bradycardia
|
1 (1.3)
|
a Patient with fever also required reoperation for bleeding.
Table 3
Serious adverse events
Type of event
|
Number of subjects (%)
n = 75
|
Air in chest drain
|
1(1.3)
|
Left ventricular apex fissure[a]
|
1 (1.3)
|
Left arm ischemia[a]
|
1(1.3)
|
Reoperation for bleeding[b]
|
4 (5.3)
|
Pseudoaneurysm[a]
|
1(1.3)
|
Pericardial Effusion
|
1(1.3)
|
Wound Infection
|
1(1.3)
|
Asthenia and palpitations
|
1(1.3)
|
Sepsis
|
1(1.3)
|
a These events all occurred in the same patient.
b One patient returned to theater for bleeding on two occasions.
In addition to the standard measures of intraoperative status and the above-stated
measures, the following measures specific to the use of BioFoam were recorded: estimated
total amount of BioFoam applied; confirmation of proper coverage of the resected area
with BioFoam; central venous pressure at the time of BioFoam application; and intraoperative
blood loss.
Results
A total of 95 subjects were screened and consented to participate in this prospective,
multicenter, single-arm study. Of these 95 subjects, 20 (21%) did not meet intraoperative
inclusion criteria. The remaining 75 subjects were treated with BioFoam. Two of the
subjects at the German Heart Centre were lost to follow-up. As such, 73 subjects completed
the study per protocol.
The baseline characteristics and medical history of the 75 subjects enrolled in the
study are presented in [Table 4]. All subjects were Caucasian.
Table 4
Subject demographics and medical history
Characteristics
|
Value (%)
|
Age
|
Mean ± SD
|
67.91 ± 9.2
|
Range (min, max)
|
35–83
|
Gender
|
Female
|
50 (66.6)
|
Male
|
25 (33.4)
|
Hypertension
|
53 (70.6)
|
Aortic aneurysm
|
20 (26.6)
|
Coronary artery disease
|
19 (25.3)
11 (14.7) Unknown
|
Smoking
|
41 (54.6)
|
Diabetes mellitus
|
10 (13.3)
|
Diabetes type
|
Insulin dependent
|
4/10 (40.0)
|
Noninsulin dependent
|
6/10 (60.0)
|
Chronic obstructive pulmonary disease
|
12 (16)
1 (1.3) Unknown
|
Visceral occlusive disease
|
2 (2.7)
1 (1.3) Unknown
|
Chronic renal failure
|
3 (4)
|
Congestive heart failure
|
6 (8)
|
Cerebrovascular disease
|
3 (4)
|
Peripheral vascular disease
|
12 (16)
1 (1.4) Unknown
|
History of stroke
|
1 (1.4)
2 (2.6) Unknown
|
Malperfusion syndrome (thoracoabdominal)
|
3 (4)
|
NYHA functional class
|
1
|
29(38.6)
|
2
|
25 (33.4)
|
3
|
13 (17.3)
|
4
|
6 (8)
|
Unknown
|
2 (2.7)
|
Aortic insufficiency
|
None
|
25 (33.3)
|
Mild
|
15 (20)
|
Moderate
|
24 (32)
|
Severe
|
9 (12)
|
Unknown
|
2 (2.7)
|
Other relevant cardiac or vessel diseases
|
48 (64)
|
Abbreviations: NYHA, New York Heart Association; SD, standard deviation.
In total, 105 procedures were performed in the 75 patients. The types of surgery performed
are shown in [Table 5]. All subjects required the use of an adjunct hemostat to control generalized anastomotic
oozing following standard repair procedures (such as sutures and staples).
Table 5
Type of surgery performed
Type of procedure
|
Number of subjects (%)
n = 75
|
Mitral valve replacement
|
3 (4.0)
|
Redo mitral valve replacement
|
1 (1.3)
|
Redo aortic valve replacement
|
1(1.3)
|
Aortic valve replacement
|
27 (36)
|
Aortic valve replacement with CABG
|
7 (9.4)
|
Aortic valve replacement with aneurysm repair
|
10 (13.5)
|
Aortic valve replacement with mitral valve replacement
|
5 (6.8)
|
Aneurysm repair
|
10 (13.4)
|
Aortic stenosis
|
1 (1.3)
|
Aortic valve replacement + aneurysm repair + CABG
|
1 (1.3)
|
Aortic valve replacement + atrial fibrillation ablation
|
1 (1.3)
|
Mitral valve replacement + tricuspid valve
|
1 (1.3)
|
Mitral valve replacement + tricuspid valve replacement + aortic repair
|
1 (1.3)
|
Mitral valve replacement + tricuspid valve replacement + ACVB
|
1 (1.3)
|
ACVB + left ventricular aneurysm
|
1 (1.3)
|
ACVB
|
1 (1.3)
|
Mitral valve replacement + ACVB + atrial septal aneurysm
|
1 (1.3)
|
Mitral valve annuloplasty + LAA occlusion
|
1 (1.3)
|
Aortic valve replacement + marrows myomectomy
|
1 (1.3)
|
Abbreviations: ACVB, aortocoronary venus bypass; CABG, coronary artery bypass graft;
LAA, left atrial appendage.
The time to hemostasis was measured in 74 of the 75 subjects treated with BioFoam.
Of these 74 subjects, 62 (84%) met the primary effectiveness end point of hemostasis
achievement within 3 minutes. The secondary effectiveness end point of hemostasis
within 10 minutes was achieved in 69 (93%) of the 74 subjects. The results for primary
and secondary effectiveness end points are shown in [Table 1].
Hemostasis was achieved after 10 minutes in one of the five subjects who did not achieve
the secondary effectiveness end point. Of the remaining four subjects, one required
the use of additional sutures; two required the use of alternative adhesives—Tachosil
was applied in one case and BioGlue in the other; and no additional procedure or reintervention
was recorded for the fourth subject.
Less than 5 mL of BioFoam was used in 72 of the 75 subjects (96%) treated with BioFoam.
In 74 (98.6%) of the 75 subjects, the volume of BioFoam applied was considered by
the Investigator to have adequately covered the oozing area. BioFoam did not adequately
cover the surgical site in one (1.3%) of the 75 subjects. In this subject, although
10 mL of BioFoam was applied to the left atrial appendage during mitral valve repair
over a tissue to tissue anastomosis, the surgeon noted that the BioFoam did not adhere
to the tissue. Tachosil was applied to the tissue but hemostasis could not be adequately
maintained and the subject underwent reoperation for bleeding.
Overall 12 (16%) of the 75 subjects experienced one AE ([Table 2]) and 13 SAEs occurred in 10 (13.3%) of 75 subjects ([Table 3]). The most common AE reported was atrial fibrillation which occurred in 11 (14.6%)
of the 75 subjects. None of the AEs or SAEs reported in this study were considered
by the Investigators to be related to the use of BioFoam.
Four (5.3%) of the 75 subjects required reoperation for bleeding within 24 hours of
surgery. None of these events was considered by the Investigator to be associated
with an area of BioFoam application. There were no observations of bleeding in any
subject at discharge and no reoperation for bleeding following discharge.
Blood products (red blood cells, platelets, and fresh frozen plasma) were administered
to 14 (18.6%) of the 75 subjects. A further 5 (6.6%) subjects were given banked autologous
blood and the majority, 57 (75.7%) of the 75 subjects, required only a cell saver.
A mean operation time of 218.2 (±72.2) minutes was observed in this study.
Discussion
Patients who undergo cardiovascular surgical procedures which have been repaired with
sutures or staples often have anastomotic bleeding that must be rapidly and effectively
controlled by the surgeon to avoid deleterious clinical consequences. The clinical
sequelae may include anemia, hemodynamic instability, hypothermia, hypovolemia, reduced
oxygen delivery to tissues, impaired visualization of the surgical field, an increase
in the duration of surgery, blood transfusion, and reoperation.[1]
BioFoam has been developed to offer surgeons a new option for achieving rapid and
effective anastomotic hemostasis—with a convenient prefilled syringe that can be stored
at room temperature. Initially registered in 2009 for use in liver and spleen surgery,
BioFoam received approval in 2012 for an extended indication in cardiovascular surgery.
The convenient packaging that obviates the need for preparation prior to use and the
stability of BioFoam at room temperature are intended to enable rapid deployment.[7] Speed of application of the hemostat can be an important consideration when bleeding
is profuse and the patient's prognosis is under threat.
This prospective, multicenter, single arm study demonstrated the safety and effectiveness
of BioFoam as a surgical adjunct to achieve anastomotic hemostatis following cardiovascular
surgery in 75 subjects at two centers (15 subjects at the German Heart Centre, Munich,
Germany, and 58 subjects at the Ospedale San Raffaele, Milan, Italy). The primary
effectiveness end point of hemostasis achieved within 3 minutes was observed in 62
(84%) of the 75 subjects ([Table 1]). Hemostasis was achieved in a further five subjects within 10 minutes after application
of BioFoam. The effectiveness of BioFoam in a broad range of cardiovascular surgical
settings in this study supports its general use for addressing anastomotic bleeding
following cardiovascular surgery.
BioFoam was well tolerated in this study. None of the AEs in 12 (16%) subjects or
the 13 SAEs in 10 (13.3%) of the 75 subjects were considered by the Investigator to
be related to the use of BioFoam. Four (5.3%) of the seventy-five subjects required
reoperation for bleeding within 24 hours of surgery. The safety outcomes for this
study were comparable to those observed in studies with established surgical sealants
used in cardiovascular surgery.[8]
Blood products (red blood cells, platelets, and fresh frozen plasma) were given to
14 (18.6%) of the 75 subjects and a further 5 (6.6%) required banked autologous blood.
The majority of subjects—57 (75.7%) of the 75 subjects—required only a cell saver.
A low requirement for blood products, such as that observed in this study, has been
shown to be associated with better outcomes for patients.[9]
The mean operation time of 218.2 (±72.2) minutes observed in this study is similar
to that observed with BioGlue, a nonfoaming formulation sealant produced by CryoLife.[10]
It is acknowledged that this study does have limitations. As this was not a randomized
study, there is no comparison to any “other agent” or a comparison to “no agent,”
so although the results are promising, there is no direct proof of superiority versus
other methods of controlling bleeding. While protamine was administered to most of
the patients during surgery, use of protamine before or after BioFoam application
was not recorded. As such, it is not possible to ascertain the possible contribution
of protamine to cessation of bleeding in this study.
Overall, the results of this study support the safety and effectiveness of BioFoam
Surgical Matrix when used as an adjunct for anastomotic hemostasis following cardiovascular
surgery.