Introduction
The clinical conditions of edentulous jaws are nowadays considered a common problem
for patients and clinicians. The choice of the correct treatment plan, and its significance
for the patients’ quality of life, still remains a challenge due to the different
ridge atrophic shapes and to the several prosthodontic solutions. The partially or
totally edentulous patient experiences a condition of strong physical and social discomfort.
The loss of teeth has unfavorable impact on the oral esthetics and function, involving
also the quality of life of the patients.[1]
[2]
[3] In the past 20 years, the advent of the dental implant-based therapies quickly developed
as safe and predictable remedy for fully edentulous patients and as an alternative
to standard removable dentures. Several treatments can include prosthodontic rehabilitation
involving high or low numbers of dental implants, cemented or screw-retained dental
prosthesis, and dental implant-supported overdentures. Usually, the treatment choice
is related to the patient’s desires but at the same time it is chosen considering
the patients anatomical conditions. Many patients, especially the oldest ones affected
by severely atrophic ridges, feel uncomfortable with the removal of prosthesis, especially
of the lower jaw.[4] To overcome these clinical adverse conditions, two or four dental implants for full-arch
prosthesis rehabilitation could be recommended as first treatment option. In this
way, noninvasive surgery could be guaranteed with excellent functional and esthetic
long-term clinical results, enabling recovery of patients’ quality of life. The old
overdenture system can be defined as a removable dental prosthesis positioned over
one or more remaining natural teeth, the residual roots, and/or dental implants. The
dental implant–abutment connections have to be solid and secure, considering unfavorable
inclination due to the typical anatomical condition of the atrophic maxilla.[4] The present study reported a new way of stabilization of full-arch fixed prosthesis,
developed by moving a step further over the old concept of overdenture.
Today, the dental prosthesis could be classified as fixed with cemented or screw-retained
abutments. However, when it comes to determining the different holding systems available
in terms of favorable clinical outcomes, ease of maintenance, patient satisfaction
and preferences, cost to the patient, or ease of cleaning and removal, available literature
still does not provide reliable data as to which is the ideal retention system for
this kind of dental rehabilitation. When placed on the abutments, the prostheses,
fixed or not, have to guarantee esthetics and function. This last parameter is most
difficult to be achieved because the forces involved in the masticatory cycle are
not linear and not homogenous.[5]
[6] Recently, several published papers investigated the stress over the bone tissue
and over dental implants and prosthodontic elements during the chewing alternations,
and aimed to study the point of fracture as well as the wear of the material used.[7] The uniform dispersion of the stresses that occurred on prosthodontic components
during the chewing cycles is documented as being conditioned by the number and the
position of the dental implants, as well as by the basic material and by the singular
prosthodontic aspect and elements. One of the most common clinical problem associated
with implant-supported restoration is screw loosening and then loss of retention over
time.[8] The purpose of the presented research is to evaluate the prosthetic elements of
retention for full-arch fixed prosthesis placed over four dental implants by using
OT Equator (Rhein 83 S.R.L.; Bologna, Italy) attachments and OT Bridge (Rhein 83 S.R.L.)
components and the seeger to highlight potential failures related to any fracture
of the structural components or any overload on bone tissue, as well as to evaluate
the effectiveness of the seeger as an extra element of retention.
Therefore, the principal aim is to consider the grade of retention by using the passant
screws, screwed to all or to OT Equator attachments, that are tightened into the dental
implant (Osstem TSIII; Osstem Global Co., Ltd., Seoul, South Korea) according to the
all-on-four technique by using finite element analysis (FEA). A comparative analysis
has then been performed to finalize the systems with a type of implant (Osstem TSIII)
and “abutments” connected by metric threading.[9]
[10]
[11] Furthermore, in consequence of the absence of many parameters, pictures have been
utilized to recreate the files. Potential approximations have been considered due
to this operation.[12] The results of the tests were recorded as graphic and data, which were matched to
identify which was better among the ones studied. Finally, the von Mises stress test
has been set and evaluated.
Materials and Methods
The principal factors that regulate the precision of the FEA method have been detected.
In association with these, the detailed shape of the system and the peri-implant bone
anatomy to be shaped, the perimeter conditions and constraints, the material mechanical
features, the load conditions “repeated on time related to masticatory cycle,” the
bone–implant contact, the test of convergence, and the validation of the model have
been involved in this analysis.
Solid representation of the jaws, dental implants, and prosthodontic elements were
recreated from Roster images, which are processed through a three-dimensional computer-aided
design (3D CAD) in finite element method (FEM). The search process was then separated
into the following two moments: the preprocessing—finite element model’s building
step—and the postprocessing—converting and modeling of solutions.[9]
[13] The following tests incorporated the tests about the dental prosthesis marked “Rhein
83 S.R.L.” with the following configuration:
1. Four OT Equator–OT Bridge assembled with four passant retaining screws: Test 1
2. Four OT Equator–OT Bridge assembled with three out of four passant retaining screws:
Test 2
3. Four OT Equator–OT Bridge assembled with two out of four passant retaining screws:
Test 3
4. Four OT Equator–OT Bridge assembled without retaining screws: Test 4
This type of retention system (OT Equator) was born as an evolution of the spherical
system. The component to be highlighted is also represented by the presence of the
seeger. The subequatorial component allows the housing inside the cylindrical abutment
of an interchangeable acetal ring, the seeger, and represents a systematic alternative
to screwed and cemented solutions.
The threaded hole with a closed bottom on the attachment does not communicate with
the implant, avoiding coronal–apical bacterial infiltrations. Thanks to this, there
are also great advantages in overcoming implant disparallelisms, even in extreme cases
of over 80°, without resorting to components such as multiunit abutment or drilling
techniques. The real revolution offered by the OT Equator is versatility with the
use of the same identical abutment, both for the removable and fixed prosthesis solution
([Fig. 1]).
Fig. 1 OT Equator and OT bridge systems details: elastic seeger with mounter (A); elastic seeger positioning (B); exocad screenshot about screw connection (C); final prosthesis structure (D).
Reverse Engineering
The model sizes were created from the prosthodontic elements and the pictures were
made real by the small details of their physico-chemical features published in the
literature and by the brand drafts. The creating moment has been developed by SolidWork,
by which the information is screened from the physical system to a mathematical clone,
deduced from the same number of variables and “filtering out” the delaying ones ([Fig. 2]). To create a homogeneous and adequate mesh of the whole model, the discretization
tests revealed the difficulty in finding optimal solutions with a small number of
elements. For this reason, the mesh has been refined in areas of particular interest.
The areas where there is a coupling have also been thickened, to obtain a better response
along the contact interface between the components. For all the tests, the load condition
considered is shown in [Table 1]. To ensure that a real situation was considered, the frictions between the various
components were also considered ([Table 2]).
Table 1
Summary of the type and location of the loads applied to the system
Test
|
Loading conditions
|
Type of load
|
Compression load
|
Location
|
Molars
|
Condition
|
Osseointegration
|
Table 2
Summary of the type of contact and the numerical value used
Connection
|
Type of coupling
|
Coefficient clutch
|
Cancellous bone/cortical bone
|
Bonded
|
–
|
Cortical bone/fixture
|
Bonded
|
–
|
Cancellous bone/fixture
|
Bonded
|
–
|
Fixture/OT Equator
|
Frictional
|
0.2
|
OT Equator/Seeger
|
Frictional
|
0.35
|
OT Equator/abutment
|
Frictional
|
0.2
|
Seeger/abutment
|
Frictional
|
0.35
|
Abutment/bridge
|
Bonded
|
–
|
Fig. 2 Reverse engineering on a dental implant.
Finite Element Methods
Therefore, after having these 3D CAD pictures, the FEA jaw-implant-prosthesis has
been created by using Ansys Workbench. A 3D linear static parametric simulation has
been developed underlining the ratio (stress and strain) between bone tissue and prosthodontic
pieces, and fixture and OT Equator retainers.
Characteristic of the Materials Involved in the Study
The identical force has been directed to the different implants and the consequent
strength dispersion has been recorded. The features of the materials have been classified
in terms of Young’s modulus, Poisson’s ratio, and density. The numerous physical characteristics
of the components have been recorded regarding the occlusal and lateral strengths.
The titanium alloy (Ti6Al4V) under evaluation can be classified as homogeneous, linear,
and isotropic, while the bone tissues (cortical and cancellous) that should be anisotropic
have been classified as orthotropic. Therefore, the numerous alteration parameters
in the three space vectors in response to the stress have been recorded.[13]
[14]
[15]
[16]
[17]
[18]
Results
Analyzed with the other manuscripts published to date in the scientific literature,
this study is presenting a reproduction as detailed as possible: interaction between
the surfaces of nonpenetration with friction and preloading of the passant screw.
In the past, it was common to apply a “joint” relation between the sections and not
to consider preload stress, nevertheless, to the detriment of the truthfulness of
the data. In this specific study, the authors obtained a balance for achieving outcomes
that are as similar to reality.[9]
[18]
[19]
[20]
[21]
[22]
[23]
A CAD drawing of each element has been developed and then assembled in a unique model
with relative adaption. At the same time, the purpose of the investigation is to research
the global tension on the four groups. A compression vertical load of 800 N has been
applied to the model.
Test 1 Results
From the analysis of the results from “Test 1” ([Figs. 3]
[4]
[5]
[6]), it can be seen that the maximum value reached on the prosthesis is equal to 291
MPa (tension lower than that of yield) while for the bone the value found is equal
to 83 MPa. By preparing a detail of the four prostheses, removing the bone from the
analysis, it can be seen that the results remain unchanged. From a preliminary analysis,
it can be seen how the two posterior prostheses are more stressed than the anterior
ones. The bridge is loaded, with greater intensity, on the left side (the prostheses
are not mounted symmetrically). It can be noted that the most stressed element is
the connecting screw. In particular, the most stressed screw is the left front, since
the load is unloaded more on it.
Fig. 3 Von Mises stress related to the connection screws.
Fig. 4 Von Mises stress related to the implant.
Fig. 5 Total movement of the system on Test 1(A), 2(B), 3(C), 4(D).
Fig. 6 Von Mises stress related to test 1(A), 2(B), 3(C), 4(D).
For the comparison of the four system configurations, it is useful to evaluate the
displacements of the system to characterize its lability. It is easy to read/note
that the displacement calculated as the result of the three movements along the three
axes x, y, and z is greater in correspondence with the implant, on the rear left. Once the overall
displacements of the system have been calculated, they are assessed along each individual
axis. From the evaluation of the displacements along the x-axis, it can be noted that the greatest displacement along the increasing x is located on the left posterior stump. Moving on to the evaluation of the displacements
along the y-axis, a displacement can always be noticed along the positive y in the same area. This points out that from the set compression load, twisting moments
arise along the x and y axes. Ultimately, the displacements along the z-axis are checked, to note if there is also a bending of the implant beyond the torsion.
This analysis shows that there is a displacement along the positive z precisely on the anteriorly mounted prostheses.
The results obtained from “Test 1” show that the most stressed elements are the connection
screws. It can also be noted that the displacement modules along the three axes are
contained around 0.1 mm. It can be concluded that the implant is fixed, and resists
under a static load of 400 N per molar.
Test 2 Results
From the analysis of the results from “Test 2” ([Fig. 3]
[4]
[5]
[6]), it can be seen that the maximum value reached on the prosthesis is equal to 314
MPa (tension lower than that of yield) while for the bone the value found is equal
to 118 MPa. A different distribution of the tensions can be seen, given the fact that
the implants are constrained on three surfaces and no longer on four. Furthermore,
the rear left screw is more loaded than in the previous case, even if the peaks are
found in the front right one.
Moving on to the evaluation of the displacements along the y-axis, a displacement can always be noticed along the negative y in the same area. This points out that from the set compression load, twisting moments
arise along the x and y axes.
Ultimately, the displacements along the z-axis are checked, to note if there is also a bending of the implant beyond the torsion.
From this analysis, it appears that there is a displacement along the positive z precisely on the prostheses mounted anteriorly, especially in the nonanchored area.
The results obtained from “Test 2” show that the most stressed elements are the connection
screws. It can also be noted that the modules of the movements along the three axes
are contained around 0.2 mm. It can be concluded that the implant is once again fixed,
and resists under a static load of 400 N for molars.
Test 3 Results
From the analysis of the results from “Test 3,” as shown in [Figs. 3]
[4]
[5]
[6], it can be seen that the maximum value reached on the prosthesis is equal to 346
MPa (lower tension than the yield strength) while for the bone the value found is
equal to 153 MPa.
By preparing a detail of the four prostheses, removing the bone from the analysis,
it can be seen that the results remain unchanged. From a preliminary analysis, it
can be seen how the two posterior prostheses are more stressed than the anterior ones.
The bridge is loaded, with greater intensity, on the left side (the prostheses are
not mounted symmetrically).
A graph was drawn relating to the trend of the voltages generated at the equator interface/connection
screw, to examine how the voltage state in this area varies. It can be noted that
the most stressed element is the connecting screw. In particular, the most stressed
screw is the one on the rear left, since the load is more discharged on it.
A different distribution of the tensions can be seen, given the fact that the implants
are constrained on two surfaces and no longer on four. Furthermore, the rear left
screw is more loaded than in the previous case. Moving on to the evaluation of the
displacements along the y-axis, a displacement can always be noticed along the negative y in the same area. This points out that from the set compression load, twisting moments
arise along the x and y axes.
Ultimately, the displacements along the z-axis are checked, to note if there is also a bending of the implant beyond the torsion.
This analysis shows that there is a displacement along the positive z precisely on the anteriorly mounted prostheses.
The results obtained from “Test 3” show that loading the system with 87% of the total
load leads to a condition in which the resultant of the movements reaches a value
around 0.3 mm, a value for which it could have system instability.
Test 4 Results
From the analysis of the results from “Test 4,” as shown in [Figs. 3]
[4]
[5]
[6], it can be seen that the maximum value reached on the prosthesis is equal to 208
MPa (lower than the yield stress) while for the bone the value found is equal to 83
MPa.
By preparing a detail of the four prostheses, removing the bone from the analysis,
it can be seen that the results remain unchanged. From a preliminary analysis, it
can be seen how the two posterior implants are more stressed than the anterior ones.
The bridge is loaded, with greater intensity, on the left side (the prostheses are
not mounted symmetrically).
It can be noted that the most stressed element is the bridge. A different distribution
of the strain can be seen. Furthermore, using this setup, it can be noted that the
OT Equator attachment is more stressed than in previous cases, since as it is no longer
anchored with the connection screws; the stump by rotating applies pressure on the
OT Equator attachments.
For the comparison of the four system configurations, it is useful to evaluate the
displacements of the system to characterize its lability. It is easy to read/note
that the displacement calculated as the result of the three movements along the three
axes x, y, and z is greater in correspondence with the two front implants.
Once the overall displacements of the system have been calculated, they are assessed
along each individual axis. From the evaluation of the displacements along the x-axis, it can be noted that the greatest displacement along the increasing x is located on the left posterior stump.
Ultimately, the displacements along the z-axis are checked, to note if there is also a bending of the implant beyond the torsion.
This analysis shows that there is a displacement along the positive z precisely on the anteriorly mounted prostheses.
The results obtained from “Test 4” show that loading the system with l2% of the total
load leads to a condition in which the resultant of the movements reaches a value
around 0.4 mm. It can be concluded that this system is unstable. It is therefore not
advisable to use such a conformation.
Even all the developed prosthodontic elements showed a typical system associated to
the chewing cycles; the most stressed component of the fixture is represented by the
connecting screw.[1]
[24]
[25]
[26]
[27] Bar overdenture is a common and well-adaptable treatment option because of its stress
distribution, but its cost is high and patients sometimes choose to have higher results
with lower expenses ([Figs. 4]
[5]
[6]).[15]
[16]
[17]
[28]
Discussion
Recently, several published papers highlighted how in rehabilitative dentistry the
parameters of knowledge relating to screw retention, prosthodontic properties, and
the dental implant osteointegration phenomena are fundamental for having a long-term
clinical success. A further analysis about the biomechanical parameters of the oral
cavity anatomy and physiology is important for the knowledge of the bone mechanical
properties as well as for a precise data on the jawbone shape and size.[29]
[30] In the past 20 years, biomaterial internal geometry and structure have largely assisted
in the integration of FEM in the product realization process.[5]
[31]
The advantages of FEM in the biomedical area are several. The most actual is connected
to the chance of FEM that can enable early device efficiency testing prior to costly
prototyping and bench testing.[32]
[33] For the four tests, it was chosen to apply a load of different entity. This opportunity
is dictated by the fact that by eliminating one anchor at a time, it could be seen
that the displacements and tensions have an increasingly accentuated variation. Paying
attention to the displacements, related to the system, it can be seen that in the
last test, where it was chosen to stress the system with a load that is 12% of the
total, the displacement of the system is four times greater than the first test in
which all the prostheses were anchored with connection screws.
It can be noted that although in the last test it was chosen to stress the system
with a load that is 12% of the total, the resulting stress is 60% compared with the
maximum stress calculated in the four tests.[34]
[35] In the field of dentistry, the shape of prosthodontic instruments for retained overdenture
base and for fixing full-arch denture has been amply studied in the recent literature,
for researching the integration and the wear influenced by the masticatory cycles.
The OT Equator attachments have been recently studied because it is on trade just
by 2007. This retainer can be applied both for overdenture with direct connection
and for overdenture placed over a secondary structure.[36] A review and meta-analysis performed by Keshk et al[37] published on 2017 about full-arch dental fixation systems and about their conditioning
over the peri-implant bone loss showed how there are no statistically significant
anomalies between the kinds of overdenture attachments with regard to marginal bone
loss, bleeding index, gingival index, and plaque index.[37]
[38]
[39]
[40]
[41]
[42]
[43]
[44] The real innovation in this method of attachment lies precisely in the presence
of an acetal ring, called seeger ([Fig. 1]), which guarantees stability and retention to the prosthesis even in the absence
of screwing systems, thus promoting a distribution of forces.
Conclusion
It could therefore be concluded that the first two configurations, respectively those
in which there are four connection screws and three connection screws, are safe, since
the stresses generated are lower than the yield points of the material. Test 3, in
which only two connecting screws are used, highlights the possible instability of
the system. In this case, although the load is 87% compared with the loads applied
in the previous tests, the tensions that arise are 1.5 times higher than the first
two tests. Finally, the last test highlights, as previously discussed, the possible
instability of the system due to the failure to anchor with implant connection screws.
The screw methods are a guarantee of fixation; however, the presence of the seeger
significantly improves the stability and the retention of the whole prosthesis, as
demonstrated by the tests. Higher, long-term stability could be expected using four
and three screws. The new shape and the presence of seeger could be useful for increasing
the stability of a full-arch prosthesis system delivered on four implants, particularly
when one out of four screws is excluded. Clinical studies are recommended to confirm
those results.