Keywords meta-analysis - DOAC Dipstick - direct oral anticoagulants - point-of-care test
Introduction
The number of patients with non-valvular atrial fibrillation and thromboembolic events
is increasing, mainly because the population is aging.[1 ]
[2 ] Direct oral anticoagulants (DOACs) are preferred over vitamin-K antagonists for
preventing stroke in patients with non-valvular atrial fibrillation and for preventing
and treating venous thromboembolism[3 ]
[4 ] because DOACs cause fewer intracranial bleeds, have fewer interactions with food
and other drugs, and have a faster onset and offset of action.[5 ] Furthermore, regular drug monitoring is not required with DOACs because of their
more predictable pharmacodynamic and pharmacokinetic properties.[6 ]
DOACs can be detected in the laboratory by measuring the activated thromboplastin
time and prothrombin time, although this depends on drug levels and reagent's sensitivity.
Further tests include thrombin-specific clotting assays, such as the diluted thrombin
time test and the ecarin clotting time. The viscoelastic hemostatic methods using
thromboelastography is another test. DOACs can also be measured using chromogenic
substrates and liquid chromatography tandem mass spectrometry (LC-MS/MS).[6 ] However, in emergency situations (such as bleeding, urgent surgery or thrombolysis),
a fast, accessible and accurate point-of-care (POC) test is needed for detecting DOACs.[7 ]
A qualitative POC DOAC Dipstick test was developed to detect DOACs in the urine. This
was possible because 35–80% of dabigatran, apixaban, edoxaban and rivaroxaban are
excreted unchanged into the urine.[8 ] Differences between the two types of test strips were that prototype test strips
determine DXI and DTI on separate strips for analysis in separate urine samples while
the commercial DOAC Dipstick (DOASENSE GmbH, Heidelberg, Germany) has separate pads
for DXI and DTI on one test strip. Material and the technique for immobilization of
reagents on pads differed between the types of test strips. The medium for color identification
by the observer was urine sample and surface of the pad of DOAC Dipstick for prototype
and commercial versions of test strips, respectively. The correctness of investigator's
interpretation of the color was performed by trained laboratory personal[9 ] and liquid-chromatography mass spectrometry (LC-MS/MS)[10 ]
[11 ] for protype and commercial test strips, respectively. These differences could lead
to different performance characteristics such as sensitivity, specificity, accuracy,
negative and positive predictive values of the two types of test strips.
The aim of the investigation was to summarize available data in the literature on
test strips performance to see whether the test can confidently be used in clinical
practice. We systematically searched literature databases for studies investigating
the detection of DXI and DTI in patient urine samples using test strips and conducted
a meta-analysis to compare the performance of prototype and commercial type of test
strips. In addition, based on the results of our meta-analysis we performed simulations
to investigate the predictive values in populations with a lower proportion of DOAC
intake.
Methods
Search Strategy
A systematic literature search was performed between 1993 up to October 2020 to identify
relevant studies in PubMed (MEDLINE) and Cochrane Library databases. One additional
abstract[12 ] was found in the Wiley Online Library. The literature search was performed in collaboration
with librarians at the University of Heidelberg. The reference lists of all included
papers were hand-searched to identify other relevant articles. The search string is
listed in [Table 1 ].
Table 1
Search string for the meta-analysis
(“direct oral anticoagulant*”[tiab] OR doac*[tiab] OR”new oral anticoagulant*”[tiab]
OR Noac*[tiab] OR”Dabigatran”[mh] OR “Rivaroxaban”[mh] OR “apixaban”[nm] OR “Rivaroxaban”[nm]
OR “Dabigatran”[nm] OR “edoxaban” [nm] OR “apixaban”[tiab] OR “Rivaroxaban”[tiab]
OR “Dabigatran”[tiab] OR “edoxaban” [tiab]) AND (“Point-of-Care Testing”[Mesh] OR
Plasma[Mesh] OR Serum[Mesh] OR Urine[Mesh] OR “Point of care”[tiab] OR Plasma[tiab]
OR Serum[tiab] OR Urine[tiab] OR Dipstick*[tiab] OR “Blood Coagulation Tests”[Mesh]
OR “Coagulation”[tiab] OR “Mass Spectrometry”[Mesh] OR “Mass Spectrometry”[tiab] OR
“International Normalized Ratio*”[tiab] OR INR[tiab] OR “Partial Thromboplastin Time”[tiab]
OR aptt[tiab] OR Ptt[tiab] OR “Prothrombin Time*”[tiab] OR Pt[tiab] OR Thromboelastography[tiab]
OR Thromboelastometry[tiab] OR “Thrombin Time*”[tiab] OR “whole blood clotting”[tiab]
OR “chromogenic”[tiab] OR Hemoclot[tiab]) AND (“sensitivity and specificity”[Mesh]
OR sensitiv*[tiab] OR “predictive value*”[tiab] OR accurac*[tiab] OR diagnosis[Subheading:noexp]
OR diagnos*[tiab] OR specificity[tiab])
Inclusion criteria were the determination of DOACs in urine samples of patients treated
with rivaroxaban, apixaban, edoxaban and dabigatran. Duplicate publications, narrative
reviews, case reports, and studies that measured DOACs using coagulation and chromogenic
tests, chromatography methods or blood based POC tests were excluded. Studies were
screened and eligible studies were identified by one researcher (AM) and confirmed
by another (SH). Any discrepancies were resolved by discussion. Each included study
was evaluated for risk of bias using the checklist of methodological quality assessment
using the QUADAS-2 method.[13 ] Risk of bias was assessed for patient selection, index test, reference standard,
flow and timing. Applicability was granted for patient selection, index test, and
reference standard. The risk of bias was considered low if all two or three categories
were fulfilled, as high if one of the categories was unfulfilled, and unclear if more
than one category were not fulfilled, respectively. All studies independent of their
risk of bias were included in the analysis.[14 ]
Statistical Analysis
Statistical analyses were performed using SAS software, release 9.4 (Cary, USA) and
MetaDiSc software, release 1.4 (Madrid, Spain).[15 ] The qualitative data of the prototype and commercial test strips (true positive,
true negative, false positive, and false negative values) were presented for rivaroxaban,
apixaban, edoxaban, and dabigatran and stratified by study ([Table 2 ]).
Table 2
Data used for pooled analysis and meta-analysis
DOAC
Type of test strip
True positive
False positive
True negative
False negative
Study 1[20 ]
Rivaroxaban
Prototype
449
8
395
16
Dabigatran
480
4
476
0
Study 2[21 ]
Rivaroxaban
Prototype
77
2
50
1
Apixaban
64
3
62
1
Dabigatran
76
0
52
1
Study 3[22 ]
Rivaroxaban
Prototype
24
0
29
0
Apixaban
26
0
29
0
Dabigatran
29
0
29
0
Study 4[11 ]
Rivaroxaban
Commercial
147
8
421
3
Apixaban
160
10
Edoxaban
127
4
Dabigatran
427
3
448
2
DXI total
Prototype, commercial
1074
21
986
35
DTI total
Prototype, commercial
1012
7
1005
3
The sensitivity is defined as the proportion of true positive results in relation
to the population treated with a DOAC (factor Xa or thrombin inhibitor) and the specificity
as the proportion of true negative results in relation to the population of untreated
controls (not treated with a factor Xa or thrombin inhibitor). Sensitivity and specificity
of individual and pooled studies were analyzed using MetaDiSc. The sensitivities and
specificities of prototype and commercial test strips were compared using Chi-squared
test. If the presumptions of the Chi-squared test were not fulfilled, Fisher's exact
test was used alternatively. Test results were considered as statistically significant
at p-values below 0.05. Furthermore, odds ratios (OR) and 95% CI of sensitivity and
specificity derived from the two types of test strips were considered as relative
frequencies (not normally distributed data) and compared by Chi-squared test or Fisher's
exact test as appropriate. Heterogeneity between studies was calculated using chi-squared
heterogeneity test and the I2 index at a p-value of < 0.05. An I2 index value gauges heterogeneity - between 0 to 25% indicates insignificant heterogeneity; > 25%
to 50% low heterogeneity; > 50% to 75% moderate heterogeneity; and > 75% high heterogeneity.[14 ] The random effects model according to DerSimonian and Laird was used to analyze
pooled data[16 ]–this technique takes any heterogeneity between the studies into account. Forest
plots were created for sensitivity and specificity of studies showing weight by size
of points and in percent, values with 95% CI, I2 index and p-values for differences.[17 ] The area under the curve (AUC) of the summary receiver operating characteristic
(SROC) curve was calculated to assess the diagnostic accuracy of the meta-analysis.[18 ] Based on the sensitivity and specificity analyses, the accuracy, negative predictive
value (NPV) and positive predictive value (PPV) of test strips results were calculated
for the simulated prevalences of 1%, 10%, 30% and 60% based on Bayes' rule. The simulated
prevalence represents the simulated proportion of a population who take DOACs in a
given period of time.
The study was conducted according to the Preferred Reporting Items for Systematic
Reviews and Meta-Analyses (PRISMA) statement.[19 ]
Results
Identified Studies
1,081 potentially eligible studies were found in the database search. After removing
duplicate publications, narrative reviews, case reports, studies that did not detect
DOACs, and studies that did not detect DOACs in urine samples of patients treated
with DOACs, four studies were eligible for inclusion in the meta-analysis. The PRISMA
flow chart shows the exclusion and inclusion of studies ([Fig. 1 ]).
Fig. 1 PRISMA flow chart for selection of included studies.
Three studies used the prototype test strips. Study 1 was a single-center study (n = 465 rivaroxaban, n = 480 dabigatran).[20 ] Study 2 was an international collaborative study performed with urine samples of
patients treated with rivaroxaban (n = 78), apixaban (n = 65), and dabigatran (n = 77) including the results of day one of two days of testing by participating centers.[21 ] Study 3 evaluated urine samples of patients treated with rivaroxaban (n = 24), apixaban (n = 26), and dabigatran (n = 29) and of controls not treated with anticoagulants (n = 29)[22 ] of which preliminary results were reported.[12 ] Results of a positive or negative adjudication of colors of factor Xa and thrombin
inhibitor pads of the test strips by observers were compared with those of trained
laboratory personal. Study 4 used the commercial DOAC Dipstick in a multicentre trial
using urine samples of patients treated with rivaroxaban (n = 150), apixaban (n = 170), edoxaban (n = 131), and dabigatran (n = 429) ([Table 2 ]). In this study, the thrombin inhibitor pad of the DOAC Dipstick served as a negative
control if patients were treated with a DXI and vice versa. Therefore, a control group
not taking an anticoagulant was not required. Results of the visual adjudication of
colors of factor Xa and thrombin inhibitor pads of DOAC Dipstick by observers were
analyzed following dichotomization of the quantitative results of DOACs in urine at
a cut-off value of <30 ng/ml DOAC determined liquid chromatography mass spectrometry.[11 ] As reported, participants of all four studies were on stable treatment with DOACs
and were admitted to outpatient care units. All patients had a diagnosis of non-valvular
atrial fibrillation or venous thromboembolism. DOACs were given orally at doses of
10 mg od, 15 mg od, or 20 mg od (rivaroxaban), 2.5 mg and 5 mg bid (apixaban), 30 mg
and 60 mg od (edoxaban), and 110 mg and 150 mg bid (dabigatran). Urine samples were
collected at random time between 1.5 hour and 24 hour after administration of the
immediate prior dose. Patients had to have normal renal function for treatment with
a DOAC and had stable health conditions when investigated. Creatinine clearance was
not determined in patients.[11 ]
[20 ]
[21 ]
[22 ]
Risk of Bias
Study quality was evaluated using QUADAS-2 (Quality Assessment of Diagnostic Accuracy
Studies 2), a standardized tool for quality assessment of studies of diagnostic accuracy.
All studies were open label. All patients had diagnoses of non-valvular atrial fibrillation
and venous thromboembolism. They were included consecutively for studies 1, 3, and
4 (low risk of bias). Study 2 used samples of selected patients (unclear risk). The
index test was compared in studies 1, 2 and 3 to visual analysis of test pads by laboratory
trained personal (low risk) and in study 4 to LC-MS/MS (low risk). The flow and timing
of the index test and reference standard was performed the same day (studies 1 and
3, low risk) and was variable for studies 2 and 4. However, DOACs stored at -24°C
in urine samples are stable over 24 months (https://doasense.de/files/ENGLISH_IFU_DOASENSE-Control-Urines_DOASENSE-WI7-5-8-EN-Rev02.pdf ) resulting in an adjudication of low risk of bias. Applicability assessment was the
same as for study quality ([Table 3 ]).
Table 3
Risk of bias of the trials assessed by Quandas-2 as a standardized tool for quality
assessment of studies of diagnostic accuracy
Study, Reference
Risk of bias
Applicability concerns
Patient selection
Index test
Reference standard
Flow and timing
Patient selection
Index test
Reference standard
Study 1[20 ]
low
low
low
low
low
low
low
Study 2[21 ]
unclear
low
low
unclear
unclear
low
low
Study 3[22 ]
low
low
low
low
low
low
low
Study 4[11 ]
low
low
low
low
low
low
low
Comparison of Prototype and Commercial Test Strips
The results of the individual studies were summarized as true positive, false positive,
true negative and false negative detection of factor Xa and thrombin inhibitors in
urine of patients treated with the DXIs rivaroxaban, apixaban and edoxaban and the
DTI dabigatran ([Table 2 ]).
The sensitivity, specificity and accuracy ranged between 0.941 and 0.998 at all simulated
prevalences (1%, 10%, 30% and 60%) for all DOACs and both prototype and commercial
test strips. The PPV decreased with decreasing prevalence as expected for DOACs. In
contrast, the NPV increased up to 0.999 for all DOACs and both types of test strips
with decreasing prevalence. No clinically relevant differences were found between
prototype and commercial test strips ([Table 4 ]).
Table 4
Simulated comparison of sensitivity, specificity, PPV, NPV and accuracy for the prototype
and commercial test strips at a prevalence between 1% and 60%
Prototype test strip
Commercial test strip
Prevalence
1%
10%
30%
60%
1%
10%
30%
60%
DXI
Sensitivity
0.973
0.962
Specificity
0.978
0.981
PPV
0.304
0.828
0.949
0.985
0.343
0.851
0.957
0.987
NPV
0.999
0.997
0.988
0.960
0.999
0.996
0.984
0.946
Accuracy
0.978
0.977
0.976
0.975
0.981
0.979
0.976
0.970
DTI
Sensitivity
0.998
0.995
Specificity
0.993
0.993
PPV
0.586
0.940
0.984
0.995
0.602
0.943
0.985
0.996
NPV
0.999
0.999
0.999
0.997
0.999
0.999
0.998
0.993
Accuracy
0.993
0.993
0.994
0.996
0.993
0.993
0.994
0.996
Rivaroxaban
Sensitivity
0.970
0.980
Specificity
0.979
0.981
PPV
0.322
0.839
0.953
0.986
0.347
0.854
0.957
0.987
NPV
0.999
0.996
0.987
0.956
0.999
0.998
0.991
0.970
Accuracy
0.979
0.978
0.977
0.974
0.981
0.981
0.981
0.981
Apixaban
Sensitivity
0.989
0.941
Specificity
0.968
0.981
PPV
0.238
0.775
0.930
0.979
0.338
0.849
0.956
0.987
NPV
0.999
0.999
0.995
0.983
0.999
0.993
0.975
0.918
Accuracy
0.968
0.970
0.974
0.981
0.981
0.977
0.969
0.957
The frequencies of correct positive, correct negative, false positive and false negative
detection of factor Xa and thrombin inhibitors using prototype and commercial test
strips are shown in [Table 5 ]. The specificity of pooled data on DXI and DTI detection by prototype and commercial
test strips was not significantly different. The sensitivity of prototype and commercial
version of test strips was also not significantly different for DXI and DTI ([Table 5 ]).
Table 5
Sensitivity and specificity data: Frequencies of correct positive, correct negative,
false positive and false negative results by pooled data using prototype and commercial
test strips to detect DXI and DTI
Prototype test strip
Commercial test strip
Total
p Value
Sensitivity
DXI
True positive
640
434
1074
0.3334*
False negative
18
17
35
DTI
True positive
585
427
1012
0.5768°
False negative
1
2
3
Specificity
DXI
True negative
565
421
986
0.6730*
False positive
13
8
21
DTI
True negative
557
448
1005
1.0000°
False positive
4
3
7
*= Chi-squared test, °= Fisher's Exact test
The sensitivity and specificity of the pooled data from prototype and commercial test
strips were also not significantly different for DXIs and DTI, with values between
0.962 and 0.998 (p-values between 0.3334 and 1.0000). The ORs of sensitivity were
higher for DXI and DTI tests using prototype test strips and the ORs of specificity
were higher for DXI and DTI of the commercial test trips (all not significant), respectively
([Table 6 ]). The sub-analysis for rivaroxaban and apixaban also revealed high sensitivity and
specificity values between 0.941 and 0.989 for both types of test strips and ORs were
all not significantly different ([Table 6 ]). Edoxaban could not be evaluated because data were only available from study 4.[11 ]
Table 6
Sensitivity, specificity and OR with 95%CIs for detection of DXIs, DTI, rivaroxaban
and apixaban by prototype and commercial test strips
Prototype test strip
value (95% CI)
Commercial test strip value (95% CI)
OR (95% CI)
p Value
DXI
Sensitivity
0.973 (0.957; 0.984)
0.962 (0.940; 0.978)
0.718 (0.366; 1.409)
0.3334*
Specificity
0.978 (0.962; 0.988)
0.981 (0.972; 0.988)
1.211 (0.497; 2.948)
0.6730*
DTI
Sensitivity
0.998 (0.991; 1.000)
0.995 (0.983; 0.999)
0.365 (0.033; 4.038)
0.5768°
Specificity
0.993 (0.982; 0.998)
0.993 (0.981; 0.999)
1.072 (0.239; 4.816)
1.0000°
Rivaroxaban
Sensitivity
0.970 (0.952; 0.982)
0.980 (0.943; 0.996)
1.515 (0.438; 5.238)
0.7800°
Specificity
0.979 (0.962; 0.990)
0.981 (0.964; 0.992)
1.110 (0.434; 2.839)
0.8271*
Apixaban
Sensitivity
0.989 (0.940; 1.000)
0.941 (0.895; 0.971)
0.178 (0.022; 1.411)
0.1033°
Specificity
0.968 (0.910; 0.993)
0.981 (0.964; 0.992)
1.735 (0.452; 6.666)
0.4254°
*= Chi-squared test, °= Fisher's Exact test.
An OR > 1 indicates that DOAC Dipstick has a higher sensitivity or specificity compared
with prototype test strip.
Meta-analysis
The forest plots in [Fig. 2 ] show the sensitivity and specificity of DXI and DTI (dabigatran, rivaroxaban, and
apixaban) detection in four studies. The sensitivity and specificity were 0.968 and
0.979 for DXI detection, respectively ([Fig. 2A ]) and both 0.993 for DTI ([Fig. 2B ]). The sub-analysis of rivaroxaban ([Fig. 2C ]) and apixaban ([Fig. 2D ]) data showed sensitivity and specificity values between 0.958 and 0.980. Inconsistency
values ranged from 0% to 58.3%.
Fig. 2 Forest plot showing sensitivity (upper panel) and specificity (lower panel) analysis
results for studies with DXI ([Figure 2A ]), DTI ([Figure 2B ]), rivaroxaban ([Figure 2C ]), and apixaban ([Figure 2D ]) using the prototype (green) and commercial (blue) test strips. Pooled data are
shown in red. Values are presented with 95% CI. Size of the circles represents the
weight of the studies. P-values were determined using the Chi-squared test. Inconsistency/Heterogeneity
of studies is shown in %. (A) Forest plots DXI, sensitivity (upper panel), specificity
(lower panel). (B) Forest plots DTI, sensitivity (upper panel), specificity (lower
panel). (C) Forest plots rivaroxaban, sensitivity (upper panel), specificity (lower panel). (D)
Forest plots apixaban, sensitivity (upper panel), specificity (lower panel).
The area under the curve (AUC) of summary receiver operating characteristic (SROC)
curve analysis showed AUC values of 0.9957 for DXI, 0.9990 for DTI dabigatran, 0.9964
for rivaroxaban, and 0.9947 for apixaban. Edoxaban data were only available for the
commercial DOAC Dipstick test so are not reported.
Discussion
The present analysis demonstrates that the small-scale laboratory prototype and the
large-scale produced commercial test strips have comparable performances despite the
multiple differences of productions techniques and methods to prove the correctness
of the visual color assessment of the observers. Moreover, the meta-analysis results
quantified the evidence of diagnostic accuracy for the DXI and DTI with its sensitivity
and specificity from commercial DOAC Dipstick results in existing studies. Accordingly,
the results of the meta-analysis should increase the confidence in the validity of
DOAC Dipstick to qualitatively detect DOACs in urine samples of patients treated with
apixaban, edoxaban, rivaroxaban, and dabigatran.[11 ]
Other POCT tests able to determine the presence of DOACs from blood samples have also
been reported in the literature. Thromboelastographic methods,[23 ] global coagulation assays [24 ], specific coagulation POCT testing methods using ecarin reagent for determination
of dabigatran [25 ], a dielectric microsensor after recalcification of a small amount whole blood sample
[26 ] and others.[27 ] The performance of the urine based DOAC Dipstick to identify DOACs on the studied
populations was at least as high as those obtained obtained with specific testing
from patients' plasma and whole blood samples using chromogenic assays for DXI or
tests based on ecarin reagent for dabigatran measurements.
The results of this meta-analysis confirm these high sensitivity and specificity of
the prototype and commercial test strips and thereby increases the confidence in the
results of the DOAC Dipstick test- supporting their use in emergency care medicine
and other medical conditions where rapid medical decision-making processes are required.[11 ]
[27 ] Importantly, test results need to be interpreted only in connection with the patient's
clinical situation. Examples for clinical indications are patients with acute ischemic
stroke to help in thrombolysis or mechanical decision-making, acute or haemorrhagic
for deciding to use an antidote, before an acute major urgent surgical intervention,
before epidural anesthesia, to confirm stopping of a DOAC before a required temporary
interruption or to check adherence to therapy.[11 ]
The comparative simulated prevalence analysis confirmed that the prototype and commercial
test strips have comparable performances with no clinically relevant differences.
Measures of sensitivity and specificity indicate the quality of a diagnostic test,
but the question still remains whether DOACs are really present in a patient's system.
In these situations, predictive values are important since they indicate the magnitude
to which the test result can be relied on to rule out clinically important concentrations
of drug. The PPV and NPV represent the proportions of positive or negative test results
that were identified correctly. However, these values depend on the fraction of people
evaluated who are taking a DOAC at the time of analysis. In a clinical context, the
PPV should be interpreted with caution because if only a few patients are taking a
DOAC, the PPV will be lower than was seen in this study. The PPV also depends on the
prevalence in a specific clinical context. In the studies included in our meta-analysis,
outpatients with stable DOAC therapy were included – in “real world use” the PPV may
vary by disease (such as infections, malignancy, emergency care, major operations,
thromboembolism and major bleeding, among others).
In contrast, the NPV increased with decreasing prevalence, indicating a high accuracy
of the test strips and suggesting a higher probability of no DOACs being present in
a patient's system with increasing numbers of patients being analyzed. This is likely
unless DOAC excretion into the urine is reduced by nephropathy, or if food, drugs
and drug metabolites change the urine's color.[28 ]
[29 ]
[30 ]
[31 ] In these circumstances, the creatinine pad and urine color pad are important controls
for accurate DOAC Dipstick results.[11 ]
[32 ]
In this analysis, we combined the results of four studies in one random-effect meta-analysis.
In contrast to the fixed effects model, the random effects model assesses both intra-
and inter-study variance and provides wider confidence intervals and a better estimate
of the effect size.[33 ] The meta-analysis showed a pooled sensitivity and specificity of >96% for DXIs (rivaroxaban,
apixaban and edoxaban) and DTI (dabigatran). Regarding the accuracy of both types
of test strips, the area under the curve values of the SROC were 0.9957 for DXIs and
0.9990 for DTIs. In addition, the overall heterogeneity of sensitivity and specificity
values for all DOACs were categorized as low (between 0% to 58.3%) according Higgins
et al.[14 ] By combining the four eligible studies in one meta-analysis, this analysis provides
evidence for the generalizability of study results using the commercial version of
test strips.
Limitations of the present analysis need to be considered. All studies were performed
under the guidance of the same investigator. However, the tests were evaluated by
different participants in the studies which corresponds to the clinical application.
The four included studies used not completely identical study designs and investigated
different DOACs: apixaban was investigated in three of the four studies[11 ]
[21 ]
[22 ] and edoxaban in one study.[11 ] This means the exploratory power of edoxaban was low in the meta-analysis. However,
the one study investigating edoxaban was a large-scale study[11 ] that provided enough data to evaluate the accuracy of edoxaban detection by the
DOAC Dipstick. A further potential limitation was the heterogeneity between studies;
however, we found heterogeneity to be low to moderate with values between 0% and 58%
when compared with the literature.[34 ] Future studies should further investigate this heterogeneity using subgroup analysis,
sensitivity analysis or meta-regression.[35 ]
The DOAC Dipstick was indicated as a useful tool for detecting DOACs in emergency
care by the NICE Guidance document.[36 ] Furthermore, the DOAC Dipstick was mentioned in a Guideline document for the treatment
of the femoral fracture as a suitable on-site test in patients with acute hip-fracture
on handling further anticoagulant medication.[37 ] Several investigator-initiated studies are ongoing in acute major orthopaedic surgery,
in patients with ischemic stroke to support rapid medical decision processes and adherence
to their therapy for validation of DOAC Dipstick test. They include investigations
of plasma levels of rivaroxaban, edoxaban, apixaban and dabigatran compared with qualitative
DOAC Dipstick results in more heterogenous patient populations. The DOAC Dipstick
may also be used to manage patients who need to be switched from DOAC therapy immediately
to low-molecular weight heparin upon admission to hospital such as upon hospitalization
for COVID-19.[38 ]
In conclusion, this study shows the robustness of the DOAC Dipstick in detecting DOACs
in patient urine samples, thereby increase the confidence that this test is suitable
for use in clinical practice. The analysis highlighted the high accuracy of the DOAC
Dipstick in detecting rivaroxaban and apixaban. Simulation of prevalence analysis
showed the NPV is very high, which is important when intake of DOACs is unknown. Further
studies are ongoing to validate the DOAC Dipstick in various clinical emergency situations
and to compare DOAC levels quantified in blood with DOAC levels detected qualitatively
in urine samples.
What is Known about the Topic?
Rapid and accurate determination of oral direct factor Xa (DXI) and thrombin inhibitors
(DTI) remains challenging in emergency medical situations.
Several studies have been published on the detection of DXI and DTI in patient urine
samples by prototype and commercial test strips.
Differences between the two types of test strips include that DOAC Dipstick test determines
DXI and DTI on one strip only, materials and techniques for immobilization of reagents
and medium for color identification.
What Does this Paper Add?
No statistically significant differences were found for sensitivity and specificity
between both versions of test strips increasing the confidence in the high performance
characteristics of the commercial version of DOAC Dipstick.
In this systematic review and meta-analysis, both included types of test strips demonstrated
an over 96% sensitivity and specificity for detecting DXI and DTI in a single test
only from patients' urine samples.
The simulation of the prevalence showed the very high negative predictive value that
is important in a general population and in specific patients when intake of DOACs
in unknown.