A patient suffering from chronic alcoholic Child B cirrhosis underwent oesophagogastroduodenoscopy
because of anemia by means of a Pentax EG-2730. Esophageal varices with cherry red
spots were found. Multiband ligation (Saeed Multi-Band Ligator MBL-6; Wilson-Cook
Medical, Winston-Salem, North Carolina, USA) was carried out. The Opti-Vu barrel with
preloaded bands was attached to the tip of the endoscope, with the barrel advanced
onto the tip as far as possible. The banding was performed successfully without difficulty.
However, the Opti-Vu barrel was then found to be missing from the tip of the endoscope.
The endoscope was inserted again, and the barrel was found in the upper end of the
esophagus (Figure [1]). The barrel was finally retrieved by means of a foreign-body forceps (Figure [2]). There was no bleeding or trauma related to this detached barrel or its removal
and the patient recovered uneventfully. We found later that the model we used (MBL-6)
was suitable for an endoscope with an outer diameter of 9.5 - 13 mm. The endoscope
we used, however, had an outer diameter of only 9 mm.
Figure 1 The detached Opti-Vu barrel was found in the upper end of the esophagus on re-introduction
of the endoscope.
Figure 2 The detached barrel was successfully removed using a foreign-body forceps.
There are very few reports on complications with this new ligator. When this multiple-band
ligator is used, the trigger cords that are attached to the transparent tubing are
also detached when all six bands are fired. With a conventional band ligator, the
string is still firmly attached to the barrel, especially after the band is fired,
by further tightening of the trigger cord.
This case reminded us that multiband ligators should be properly checked for compatibility
with the endoscope. It is also advisable to check for a firm fit between the barrel
and the endoscope. When using multiband ligators, we should pay the same attention
to checking whether the barrel is still on the tip of the endoscope after the procedure,
as we do to checking the number of gauzes recovered after a laparotomy.