We have developed a new endotriptor that can be used through the instrument channel
of the duodenoscope. The coiled metal sheaths are 53 cm long and 2.4 mm in diameter
(Figure [1 ]
a), and fit over the remaining wire of the basket catheter. “Remaining wire” here means
the length of wire between the entrance of the instrument channel port on the duodenoscope
and the cut end of the wire. The sheaths are designed to be slightly shorter than
the “remaining wire”. This allows easy insertion of a sheath over the wire and into
the endoscope. An Olympus JF-200 duodenoscope and an FG 22-Q-1 or 403Q basket catheter
(Olympus, Tokyo, Japan), the length and diameter of which are suitable for the new
endotriptor, were used.
When basket impaction occurs with a bile stone in the distal common bile duct, the
basket catheter is cut at the handle. Only the tube sheath should then be pulled out,
leaving the wire inside the endoscope. Three coiled metal sheaths from the endotriptor
are inserted over the basket wire and into the duodenoscope. The end of the sheath
is then gripped with a vise (Figure [1 ]
b). The endotriptor can be pulled out with the basket catheter after the stone has
been crushed. In 20 patients who had basket impaction after endoscopic sphincterotomy
and underwent lithotripsy with the new endotriptor, the stones were crushed successfully
in all cases. There were no complications such as biliary perforation or bleeding.
Figure 1 a The newly developed endotriptor consists of three coiled metal sheaths 53 cm long
and 2.4 mm in diameter. b The end of the endotriptor is gripped in a vise.
The endotriptors currently available cannot be used without withdrawing the duodenoscope
from the patient [1]. The duodenoscope then has to be reinserted to extract the fragments of a stone
immediately after traditional lithotripsy. With the new endotriptor, the required
maneuvers can be carried out without removing the duodenoscope - a considerable advantage.
There is no chance of injuring the mucosa of the upper gastrointestinal tract with
the new endotriptor, as it passes through the endoscope. In addition, the distal common
bile duct is never at risk of being perforated by the tip of the new endotriptor,
as the impacted basket can be returned to the upper bile duct by pushing the endotriptor.
Endoscopy_UCTN_Code_TTT_1AR_2AH