Electrical stimulation of the cochlea to treat tinnitus has been explored for decades.
The effect of cochlear implantation on tinnitus varies significantly, ranging from
54% to 86% of adults reporting partial to complete tinnitus suppression (e.g., Souliere
et al, 1992[28]; Kou et al, 1994[29]; Tyler, 1994[29]; Ito, 1997[14]; Mo et al, 2002[19]). Despite these positive results, tinnitus will remain bothersome for some patients
(2-9%; Tyler and Kelsay, 1990[30]; Hazell et al, 1995[11]; Mo et al, 2002[19]) receiving a cochlear implant (CI). The goals of electrical stimulation for tinnitus
suppression are different than for improving speech perception, i.e., to provide ‘silence’
versus maximize audibility and speech understanding. Therefore, it is likely that
the optimal stimulation parameters for tinnitus suppression will differ from those
for speech perception (Zeng et al, 2011[35]). There are many reports on the effect of programming parameters on speech perception
of CI patients (e.g., Plant et al, 2002[21]; Koch et al, 2004[26]; Balkany et al, 2007[3]; Firszt et al, 2009[6]). These CI programming parameters include signal processing/coding strategy (e.g.,
ACE, FSP, and HiRes), stimulation mode of electrodes (e.g., bipolar versus monopolar),
stimulation rate (e.g., 125–5100 Hz), input frequency to electrode allocation table,
stimulation channels (e.g., 12 to 24), threshold levels (minimum level producing audible
sound, abbreviated T), and most comfortable levels (MCLs) (maximum output level, abbreviated
MCL, M, or C).
However, few large-scale studies have investigated the most salient programming parameters
for tinnitus suppression in CI patients who remain bothered by their tinnitus after
implantation. Dauman and Tyler (1993)[4] were the first to vary the current level of the electrical stimulation and document
the effects on tinnitus loudness for two CI patients. Rubinstein et al (2003)[26] developed a high-rate conditioner stimulus of 5000 Hz in an attempt to restore spontaneous
neural activity without producing audible sound. Tinnitus was successfully suppressed
in 30-50% of the participants using transtympanic electrodes (11 adults) and CIs (three
adults), and there was minimal detection of the conditioner stimulus. In three participants
who experienced tinnitus suppression, lowering the stimulation rate and current level
of the conditioner stimulus provided less tinnitus reduction than a high-rate conditioner
stimulus. By comparison, other researchers have reported tinnitus suppression with
low stimulation rates (e.g., <100 Hz; Zeng et al, 2011).[35]
More recently, Liu et al (2016)[18] assessed the effect of programming adjustments for CI patients with tinnitus. They
hypothesized that electrode impedance, an indicator of changes in the CI over time,
would decrease following programming adjustments and may alter tinnitus symptoms.
The patients were grouped by tinnitus onset: (a) 108 with tinnitus preimplantation,
(b) 88 with tinnitus after surgery, but before initial activation, and (c) 44 with
tinnitus one year postimplantation. Patients in each group were randomly assigned
to receive regular programming at 4, 6, 8, and 12 weeks postimplantation or programming
at four weeks postimplantation (i.e., at initial activation) with no subsequent programming.
Results revealed that electrode impedance was significantly lower in the programming
subgroups with pre- and postoperative tinnitus. In addition, a greater, more immediate
reduction was reported in tinnitus handicap for the groups who received regular programming
(Liu et al, 2016[18]). They concluded that regular CI programming lowered electrode impedance, which
effectively suppressed tinnitus in the patients who experienced it preoperatively,
postoperatively before initial activation, and postoperatively by at least one year
after initial activation. Programming adjustments were limited to T and C levels with
other parameters set consistently across participants: ACE strategy at 900 Hz stimulation
rate with 22 channels and eight maxima. Finally, previous studies found no significant
correlation between CI device type to tinnitus awareness, tinnitus distress, or tinnitus
handicap scores (Quaranta et al, 2008;[24] Andersson et al, 2009;[1] Gomersall et al, 2019[10]). In summary, studies indicate that regular programming adjustments, different programming
parameters, and various CI devices can result in tinnitus suppression, although an
exploration of the most salient programming parameters for tinnitus suppression is
needed.
The purpose of this article was to review the available CI programming parameters
for tinnitus suppression and to consider possible clinical research designs for selecting
the optimal programming parameters for clinical applications. This information may
be helpful to audiologists who provide tinnitus management to their CI patients and
has direct implications in the development of therapeutic interventions for CI patients
with tinnitus.
Programming Parameters for Tinnitus Suppression
Signal Coding Strategy
Few studies have directly evaluated the impact of signal coding strategies on tinnitus.
Quaranta et al (2008)[24] found that more sophisticated and higher rate strategies (ACE, HiRes, and CIS) were
more effective than slower strategies (SAS and SPEAK) in suppressing tinnitus for
41 CI patients. Significant differences emerged between the signal coding strategies
when the CI was on but not when the CI was off. However, the effect on individual
patients was not considered because only group data were presented. A review of 32
studies (Quaranta et al, 2004[23]) found that multichannel CIs were more effective for tinnitus suppression than single-channel
CIs. As suggested by Quaranta et al (2008)[24], more complex stimulation may provide additional temporal information to stimulate
the auditory nerve fibers and the central auditory pathway and result in tinnitus
suppression. Greater auditory input from the CI has the potential to mask tinnitus,
effectively changing the neural representation within the brain and allowing patients
to refocus their attention away from tinnitus. Therefore, although few studies have
evaluated signal-coding strategies for tinnitus suppression, the literature suggests
that more sophisticated and higher rate strategies appear to be more effective.
Rate of Stimulation
Low and high rates of electrical stimulation are effective in suppressing tinnitus
in CI patients, although the specific results vary across studies. Dauman and Tyler
(1993)[4] reported that a stimulation rate of 125 Hz at 20% of the stimulus loudness was effective
in suppressing tinnitus for two CI patients. Furthermore, they found that less electrical
current was needed at 125 Hz to suppress the tinnitus compared with rates of 80, 250,
and 500 Hz. Hazell et al (1989)[12] demonstrated that tinnitus could be suppressed with a low, 100-Hz sinusoid in six
CI patients. Zeng et al (2011)[35] reported that low-rate stimulation (<100 Hz) was effective in suppressing tinnitus
for one CI patient with unilateral hearing loss and severe tinnitus. In that study,
they investigated various high-rate stimulation rates and patterns, but complete tinnitus
suppression resulted when using a low, 100-Hz stimulation rate. Early reports also
suggested that the optimal stimulation rate for tinnitus suppression might be different
for different patients (Kuk et al, 1989[17]). Electrical stimulation using rates from 62 to 8000 Hz was effective in reducing
tinnitus for 5 of 10 participants using a ball electrode on the eardrum.
Rubinstein et al (2003)[26] developed a novel high-rate conditioner stimulus (i.e., 5000 Hz) presented along
with the CI signal to suppress tinnitus in patients using both intracochlear and transtympanic
electrodes (Rubinstein and Tyler, 2004[26]). In the development of the conditioner stimulus, multiple rates of stimulation
were compared, but a high-rate stimulus was found to be themost effective in suppressing
tinnitus when presented in conjunctionwith the signal fromthe CI (Rubinstein et al,
2003[26]). Although these studies were conducted with a small number of participants, the
results fromindividual participants indicate that audiologists should try programs
with low- and high-stimulation rates for their CI patients with bothersome tinnitus.
Several findings resulted from these studies on rate of stimulation for tinnitus suppression.
First, the length of electrical stimulation for tinnitus suppression ranged from 3
(Zeng et al, 2011[35]) to 15 minutes (Rubinstein et al, 2003[26]), suggesting that a slow, central adaptation process may be involved for some patients.
Once electrical stimulation was turned off, the effects on tinnitus varied significantly.
In some patients, a ‘rebound’ of tinnitus occurred that was worse than baseline (Zeng
et al, 2011[35]). For other patients, residual inhibition of the tinnitus lasted from 45 minutes
to three days (Rubinstein et al, 2003[26]). Across studies, it was observed that tinnitus suppression diminished steadily
over time with continuous stimulation. When this occurred, changes to the stimulus
parameters (i.e., increasing current level) were required to suppress tinnitus successfully.
Finally, some studies (Rubinstein et al, 2003;[26] Zeng et al, 2011[35]) used custom research platforms to deliver the electrical stimulation due to the
limitations in the stimulus parameters within the manufacturer's sound processors
and software. It is anticipated that more options for programming CIs for tinnitus
suppression will be available from CI manufacturers as more research is conducted
with a greater number of CI patients with tinnitus.
Electrode Location and Number
Studies have found that the electrode location affects the current that is needed
to suppress tinnitus. In addition, the optimal location for tinnitus suppression appears
to vary from patient to patient, although these results are limited to reports from
single participants. Dauman and Tyler (1993)[4] reported that the middle electrodes required less current to suppress tinnitus compared
with more apical or basal electrodes for two patients using intracochlear devices.
Specifically, data from one tinnitus patient with only 11 active electrodes (only
electrodes 21 to 11 were available because of a severe labyrinthine fracture) revealed
that the electrodes positioned in the middle of the array (e.g., 13 and 15) required
considerably less current to suppress tinnitus than those at the end (e.g., 21). Other
researchers suggested that stimulation using the entire electrode array was more effective
in reducing tinnitus compared with stimulation of only the basal electrodes (Punte
et al, 2013[22]). Furthermore, Zeng et al (2011)[35] found that the four most apical electrodes with a low-rate of stimulation (20-100 Hz)
were more effective in suppressing high-pitched tinnitus than the basal electrodes.
It has been suggested that tinnitus suppression using the low-rate stimulation in
the apex of the cochlea rather than the base indicates that a certain place mechanism
might be related to restored inhibition within the auditory system (Zeng et al, 2011[35]).
In an attempt to determine the electrodes that are most effective for tinnitus suppression,
some researchers (Rubinstein et al, 2003;[26] Rubinstein and Tyler, 2004[25]) have matched electrodes to the patient's tinnitus pitch. Using this approach, Rubinstein
et al (2003)[26] found that tinnitus suppression occurred with the pitch-matched electrodes of 7
and 14 located in the middle of the array. In addition, Rubinstein and Tyler (2004)[25] reported on a case study of a patient with unilateral hearing loss and severe tinnitus
due to Meniere's disease who received a CI. One month postactivation, the patient
matched the pitch of her tinnitus to an apical electrode, and the high-rate conditioner
stimulus was effective in suppressing her low-frequency tinnitus. Results across studies
are variable with regard to the optimal electrode location for tinnitus suppression
likely because of the small number of participants included in each study. With this
limitation in mind, preliminary results indicate that measuring the pitch of the patient's
tinnitus might be a necessary step for tinnitus suppression via a CI.
In addition, decreasing or increasing the number of electrodes in the patient's program
may alter the tinnitus percept. First, many studies have reported that speech perception
(e.g., vowel and consonant perception) improves as the number of electrodes is increased
up to about eight electrodes, and adding additional electrodes above 8 to 12 produces
less improvement (Fishman et al, 1997;[7] Dorman et al, 2000;[5] Friesen et al, 2001[8]). More recently, Schvartz-Leyzac et al (2017)[27] found that performance improved for more challenging speech tests when 20 electrodes
were activated compared with only eight electrodes. Consistent with these findings,
one study found that some bilateral CI patients showed a gradual degradation in performance
on more complex tasks such as sound localization and speech-in-noise perceptionwhen
the number of electrodeswere gradually reduced from 20 to 1 (Perreau et al, 2010[20]). Multiple studies have observed large individual differences in performance when
the number of CI electrodes is reduced (Fishman et al, 1997;[7] Frijns et al, 2003;[9] Perreau et al, 2010;[20] Schvartz-Leyzac et al, 2017[27]). In addition, many factors may influence performance with reduced electrodes such
as the signal coding strategy (ACE versus CIS), channel interaction, and stimulation
mode (tripolar versus monopolar). Thus, a more conservative approach is advocated
when deactivating electrodes tomaintain optimal CI performance.
From our informal discussions with audiologists, many audiologists will eliminate
CI electrodes if the patient reports that high- or low-pitched sounds interfere with
or worsen their tinnitus. For example, if stimulation from a basal electrode exacerbates
a patient's tinnitus, then this electrode might be eliminated from the patient's program.
As a result, the frequency allocation table is changed because those frequencies are
assigned to the remaining electrodes, which might stimulate different neural fibers.
Although it may be detrimental for speech perception when spectral resolution is decreased
(Schvartz-Leyzac et al, 2017[27]), eliminating certain electrodes may be helpful for some patients to suppress their
tinnitus. Conversely, for some CI patients with bothersome tinnitus, adding electrodes
back into the CI program might be indicated if they were deactivated because of performance
and sound quality issues. Reactivating certain electrodes may provide a low-level
background sound to mask tinnitus and change the neural excitation patterns that produce
tinnitus (Tyler et al, 2008a[31]). Regardless, speech recognition performance should be monitored when changing the
number of electrodes in a CI patient's program or when using low-level background
sound to mask tinnitus in CI patients (Rubinstein et al, 2003;[26] Tyler et al, 2015[33]). If the audiologist takes a more conservative approach to programming, adjusting
T and C levels as a first attempt to maintain speech perception would be advised before
eliminating CI electrodes.
Threshold (T-Levels)
Based on our discussions with audiologists and CI manufacturers, decreasing T-levels
have been recommended in CI fittings for patients with bothersome tinnitus because
a lower threshold setting may provide low-level background sound to mask the patient's
tinnitus. Van de Heyning et al (2008)[34] implemented this clinical approach by setting T-levels to 0 current units. The 22
patients in their study had unilateral hearing loss with severe tinnitus and received
a standard-length CI for tinnitus relief. If the tinnitus percept was still present
when T-levels were decreased to 0 current unit, then the levels were globally raised
to 10% of the MCLs. The study revealed that fitting the unilateral CI in this manner
was effective in reducing the patients' tinnitus loudness and handicap scores (Van
de Heyning et al, 2008[34]). Moreover, Liu et al (2016)[18] adjusted global T-levels using standard threshold measures for the 234 CI patients
included in their study. Liu et al (2016)[18] reported that adjusting T-levels on electrodes that were pitch matched to the patient's
tinnitus was not successful compared with a standard fitting procedure. No further
details were provided. Finally, other groups have modified T-levels on individual
electrodes on the array when stimulating the cochlea for tinnitus suppression (e.g.,
Rubinstein et al, 2003;[26] Zeng et al, 2011[35]). There appears to be many methods for adjusting T-levels for tinnitus suppression,
and more research is needed to determine which approach is beneficial.
We have known for years that low-level background sounds and amplification from a
hearing aid can be effective for many patients in reducing the prominence of tinnitus
(e.g., Hazell et al, 1985[13]). Low-level background noise and amplified sound not only interferes with the tinnitus-related
neural activity coded by the brain but also allows the listener to concentrate on
external sounds instead of tinnitus.
MCLs
Several studies have shown that electrode current level influences tinnitus loudness.
Rubinstein et al (2003)[26] found that higher current levels (equating to MCLs) were associated with rapid changes
in tinnitus loudness for one CI patient whose tinnitus was suppressed with the conditioner
stimulus. Higher current levels also elicited complete tinnitus suppression in many
participants. Zeng et al (2011)[35] found a similar result where tinnitus was suppressed at higher current levels near
the MCL for one CI patient. In that study, loudness scaling of the patient's tinnitus
was completed by first obtaining a baseline measurement of the patient's tinnitus
loudness, and then setting the MCL on selected electrodes for tinnitus suppression.
Finally, Dauman and Tyler (1993)[4] showed that the current level needed to suppress tinnitus was also dependent on
several factors, including the electrode position in the cochlea and the interelectrode
distance. Specifically, less current was needed for tinnitus suppression when the
electrode distance was greater (bipolar 13 versus bipolar 11; Dauman and Tyler, 1993[4]).
For some tinnitus patients, decreasing the MCL on certain electrodes (i.e., the basal
electrodes that stimulate the cochlea at the same frequency of tinnitus) could reduce
tinnitus loudness. This might be effective for patients who experience changes to
the pitch or quality of their tinnitus with CI use (Souliere et al, 1992[28]). These results on MCLs and tinnitus loudness are limited to a few participants,
and more research from a greater number of CI patients with tinnitus is needed. In
sum, the results currently suggest that the CI current level should be sufficient
to suppress the patient's tinnitus, but not too high to cause loudness discomfort,
pain, facial sensations, or other complications (Zeng et al, 2015[36]).
Possible Clinical Research Designs
Given the lack of large-scale studies that systematically investigate CI programming
for tinnitus suppression, we have considered possible research designs for clinical
applications. First, in a daily alternating method, the audiologist fits two programs
with different programming parameters and patients alternate daily between these programs
to determine the optimal setting for tinnitus suppression. This strategy has been
applied to CI and tinnitus patients and is a feasible method for selecting optimal
programming parameters (Tyler et al, 2008b;[32] 2015[33]). Some patients required at least three months to determine their preference using
the daily alternating method, and preferences changed over time from those at initial
activation (Tyler et al, 2008b[32]).
Another method to evaluate the effectiveness of CI parameters for tinnitus suppression
in a clinical setting is by changing only one parameter at a time during an at-home
trial (see [Table 1]). Audiologists start in step 1 by administering a pretrial tinnitus measure, such
as a questionnaire or psychoacoustic measurement of tinnitus, to document the effectiveness
of the current CI parameters in suppressing tinnitus. In step 2, the audiologist adjusts
a single parameter in their CI program and provides the patient with an at-home trial
for 2-3 months. We recommend changing only one single parameter at a time to isolate
the effects of adjusting electrical stimulation for tinnitus suppression. The patient
returns to the clinic after 2-3 months, and at step 3, the same tinnitus measure would
be re-administered after the trial to document the effectiveness of the new program
in suppressing tinnitus. A speech recognition test is administered in steps 1 and
3 to ensure that speech perception is not significantly affected by the programming
adjustments. In our studies, word recognition tests are presented at a normal conversational
level (i.e., 60 dB HL) in the sound field with and without sound therapy. There are
several advantages to these research designs alternating daily among programming parameters
or providing an at-home trial with a new tinnitus suppression program. Specifically,
these are within-subject clinical research designs such that each patient serves as
their own control and individual differences within the heterogeneous tinnitus population
are considered.
Table 1
A Within-Subject Clinical Design for Evaluating the Effectiveness of CI Programming
Parameters for Tinnitus
Step
|
Action
|
Description
|
1
|
Administer pretrial tinnitus measure, speech recognition test
|
Document current effectiveness of CI to suppress tinnitus
|
2
|
Adjust CI programming
|
At-home trial for two months
|
3
|
Administer posttrial tinnitus measure, speech recognition test
|
Document effectiveness of new CI programming to suppress tinnitus
|
Finally, the timing of the implementation of this protocol likely depends on each
patient. For patients with severe tinnitus, this protocol may be appropriate to begin
at initial activation, whereas if the tinnitus is not bothersome, a waiting period
(e.g., after three months of CI use) may be warranted. Research that includes more
participants with varying patient characteristics and tinnitus symptoms is needed
to better understand which patients will benefit, when intervention should begin,
and which parameters yield the best results for tinnitus suppression.
Conclusions
It is clear from the research literature that adjusting a CI to reduce the prominence
of tinnitus is complex. The optimal stimulus parameters are likely to be different
for different patients (Tyler et al, 2008a[31]), and currently, the results are limited to a few participants across studies. There
is no straightforward advice that can be given at this time for how to program a CI
to suppress tinnitus. For patients with preoperative tinnitus that remains bothersome
after implantation, the research suggests programming the CI using a sophisticated
signal coding strategy at high and low rates of stimulation and setting the current
at a sufficient level to suppress tinnitus, but not causing discomfort. Then, the
audiologist could consider adjusting the electrode number (e.g., eliminate or reactivate
particular electrodes), T-levels (e.g., set at 0 versus measured T-levels), andMCL
adjustments at a later time using the proposed clinical research designs. Audiologists
should explore some of these parameters for tinnitus suppression systematically when
working with their CI patients. Single-participant experiments can be very helpful.
There are some CI patients who report tinnitus as a complication from CI surgery.
In that group, tinnitus is largely experienced in the implanted ear and remains after
surgery. A small portion of these patients experience improvements in their tinnitus
once the CI is activated (Arts et al, 2015[2]). Therefore, if the patient presents with tinnitus after the surgery and it improves
with CI use, there would be no need to further evaluate programming parameters for
tinnitus suppression. However, an observation period may be beneficial to ensure that
the tinnitus improves for patients who have tinnitus as a complication from CI surgery.
We suggest the following as next steps in the clinical management of CI patients with
tinnitus:
-
Document what audiologists are currently doing to manage their CI patients with tinnitus
such as implementation of tinnitus management protocols, administration of questionnaires,
testing with tinnitus psychoacoustic measurements, and counseling.
-
Determine the successful approaches for managing tinnitus in CI patients, including
CI programming parameters and sound therapy options.
-
Develop a clinical protocol for programming CIs for tinnitus for busy clinics.