Rofo 2012; 184(7): 643-649
DOI: 10.1055/s-0032-1312727
Technik und Medizinphysik
© Georg Thieme Verlag KG Stuttgart · New York

Optimizing Image Quality and Dose for Digital Radiography of Distal Pediatric Extremities Using the Contrast-to-Noise Ratio

Optimierung von Bildqualität und Dosis für die digitale Radiografie distaler kindlicher Extremitäten auf Grundlage des Kontrast-Rausch-Verhältnisses
R. Hess
1   University of Applied Science, Hamburg
,
U. Neitzel
2   Philips Healthcare, Clinical Science Diagnostic X-ray, Hamburg
› Author Affiliations
Further Information

Publication History

05 November 2011

31 March 2012

Publication Date:
22 May 2012 (online)

Abstract

Purpose: To investigate the influence of X-ray tube voltage and filtration on image quality in terms of contrast-to-noise ratio (CNR) and dose for digital radiography of distal pediatric extremities and to determine conditions that give the best balance of CNR and patient dose.

Materials and Methods: In a phantom study simulating the absorption properties of distal extremities, the CNR and the related patient dose were determined as a function of tube voltage in the range 40 – 66 kV, both with and without additional filtration of 0.1 mm Cu/1 mm Al. The measured CNR was used as an indicator of image quality, while the mean absorbed dose (MAD) – determined by a combination of measurement and simulation – was used as an indicator of the patient dose.

Results: The most favorable relation of CNR and dose was found for the lowest tube voltage investigated (40 kV) without additional filtration. Compared to a situation with 50 kV or 60 kV, the mean absorbed dose could be lowered by 24 % and 50 %, respectively, while keeping the image quality (CNR) at the same level.

Conclusion: For digital radiography of distal pediatric extremities, further CNR and dose optimization appears to be possible using lower tube voltages. Further clinical investigation of the suggested parameters is necessary.

Zusammenfassung

Ziel: Ziel der Untersuchung war es, den Einfluss von Röntgenröhrenspannung und Vorfilterung auf Bildqualität, ausgedrückt durch das Kontrast-Rausch-Verhältnis (CNR), und die Dosis bei digitalen Röntgenaufnahmen distaler kindlicher Extremitäten zu untersuchen und die Bedingungen zu bestimmen, bei denen das beste Verhältnis von CNR und Patientendosis erreicht wird.

Material und Methoden: Mittels eines Phantoms, welches die Absorptionseigenschaften kindlicher distaler Extremitäten simuliert, wurden das CNR und die damit verknüpfte Patientendosis für Röhrenspannungen von 40 bis 66 kV mit und ohne Zusatzfilter von 0,1 mm Cu/1 mm Al bestimmt. Das gemessene CNR diente als Maß für die Bildqualität, während die mittlere absorbierte Dosis (MAD) aus einer Kombination von Messung und Simulation als Maß für die Patientendosis ermittelt wurde.

Ergebnisse: Das vorteilhafteste Verhältnis von CNR und Dosis ergab sich für die niedrigste untersuchte Röhrenspannung (40 kV) und ohne Zusatzfilterung. Im Vergleich zu Aufnahmen bei 50 kV und 60 kV konnte die mittlere absorbierte Dosis bei Beibehaltung der Bildqualität (des CNR) um 24 % bzw. 50 % gesenkt werden.

Schlussfolgerung: Durch Verwendung niedrigerer Röhrenspannungen erscheint für die digitale Radiographie distaler kindlicher Extremitäten eine weitergehende Optimierung von CNR und Dosis möglich. Die Relevanz der Ergebnisse für klinische Anwendungen muss weiter überprüft werden.

 
  • References

  • 1 Stephan G, Schneider K, Panzer W et al. Enhanced yield of chromosome aberrations after CT examinations in paediatric patients. Int J Rad Biol 2007; 83: 281-287
  • 2 ICRP Publication 73: Radiological protection and safety in medicine. Ann ICRP 1996; 26
  • 3 ICRP Publication 103: The 2007 Recommendations of the International Commission on Radiological Protection. Ann ICRP 2007; 37
  • 4 Nicholson RA, Thornton A, Akpan M. Radiation dose reduction in paediatric fluoroscopy using added filtration. Br J Radiol 1995; 68: 296-300
  • 5 Mooney R, Thomas PS. Dose reduction in a paediatric x-ray department following optimization of radiographic technique. Br J Radiol 1998; 71: 852-860
  • 6 Hansson B, Finnbogason T, Schuwert P et al. Added copper filtration in digital paediatric double-contrast colon examinations: effects on radiation dose and image quality. Eur Radiol 1997; 7: 1117-1122
  • 7 Seifert H, Jesberger HJ, Schneider G et al. Dose reduction in thorax radiography in simulated neonates with additional filtration and digital luminescence radiography. Acta Radiol 1998; 39: 514-519
  • 8 Fenner JW, Morrison GD, Kerry J et al. A practical demonstration of improved technique factors in paediatric fluoroscopy. Br J Radiol 2002; 75: 596-602
  • 9 Nickoloff EL, Berdon WE, Lu ZF et al. Pediatric high KV/filtered airway radiographs: comparison of CR and film-screen systems. Pediatr Radiol 2002; 32: 476-484
  • 10 Onnasch DGW, Schemm A, Kramer HH. Optimization of radiographic parameters for paediatric cardiac angiography. Br J Radiol 2004; 77: 479-487
  • 11 Wiltz HJ, Petersen U, Axelsson B. Reduction of absorbed dose in storage phosphor urography by significant lowering of tube voltage and adjustment of image display parameters. Acta Radiol 2005; 46: 391-395
  • 12 Gislason AJ, Davies AG, Cowen AR. Dose optimization in pediatric cardiac x-ray imaging. Med Phys 2010; 37: 5258-5269
  • 13 Brosi P, Stuessi A, Verdun FR et al. Copper filtration in pediatric digital X-ray imaging: its impact on image quality and dose. Radiol Phys Technol 2011; 4: 148-155
  • 14 Tapiovaara MJ, Sandborg M, Dance DR. A search for improved technique factors in paediatric fluoroscopy. Phys Med Biol 1999; 44: 537-559
  • 15 Tingberg A, Sjöström D. Search for optimal tube voltage for image plate radiography. Proc SPIE 2003; 5034: 187-196
  • 16 Samei E, Dobbins 3rd JT, Lo JY et al. A framework for optimising the radiographic technique in digital x-ray imaging. Radiat Prot Dosimetry 2005; 114: 220-229
  • 17 Doyle P, Martin CJ, Gentle D. Application of contrast-to-noise ratio in optimizing beam quality for digital chest radiography: comparison of experimental measurements and theoretical simulations. Phys Med Biol 2006; 51: 2953-2970
  • 18 Geijer H, Norrman E, Persliden J. Optimizing the tube potential for lumbar spine radiography with a flat-panel detector. Br J Radiol 2009; 82: 62-68
  • 19 Birch R, Marshall M. Computation of Bremsstrahlung x-ray spectra and comparison with spectra measured with a Ge(Li) detector. Phys Med Biol 1979; 24: 505-517
  • 20 Leitlinie der Bundesärztekammer zur Qualitätssicherung in der Röntgendiagnostik – Qualitätskriterien röntgendiagnostischer Untersuchungen. Gemäß Beschluss des Vorstandes der Bundesärztekammer vom 23. November 2007. http://www.bundesaerztekammer.de/downloads/LeitRoentgen2008Korr2.pdf (abgerufen am 15.9.2011)
  • 21 European Guidelines on quality criteria for diagnostic radiographic images in pediatrics – EUR 16261, Office for official publications of the European Communities. Luxembourg: 1996
  • 22 Granfors PR, Aufrichtig R. Performance of a 41×41 cm² amorphous silicon flat panel x-ray detector designed for angiographic and R&F imaging applications. Med Phys 2003; 30: 2715-2726
  • 23 Huda W. Assessment of the problem: Pediatric doses in screen-film and digital radiography. Pediatr Radiol 2004; 34: 173-S182
  • 24 Tapiovaara M, Lakkisto M, Servomaa A. PCXMC: A PC-based Monte Carlo program for calculating patient doses in medical x-ray examinations. Report STUK-A139. Helsinki: Finnish Centre for Radiation and Nuclear Safety; 1997 http://www.stuk.fi/sateilyn_kaytto/ohjelmat/PCXMC/en_GB/pcxmc/