Endoscopy 2013; 45(01): 42-47
DOI: 10.1055/s-0032-1325960
Original article
© Georg Thieme Verlag KG Stuttgart · New York

Esophageal self-expandable stent material and mesh grid density are the major determining factors of external beam radiation dose perturbation: results from a phantom model

B. K. Abu Dayyeh
1   Division of Gastroenterology & Hepatology, Mayo Clinic, Rochester, Minnesota, USA
,
J. J. VanDamme
2   Division of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA
,
R. C. Miller
2   Division of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA
,
T. H. Baron
1   Division of Gastroenterology & Hepatology, Mayo Clinic, Rochester, Minnesota, USA
› Author Affiliations
Further Information

Publication History

submitted 29 May 2012

accepted after revision 24 September 2012

Publication Date:
19 December 2012 (online)

Background: Self-expandable esophageal stents are increasingly used for palliation or as an adjunct to chemoradiation for esophageal neoplasia. The optimal esophageal stent design and material to minimize dose perturbation with external beam radiation are unknown. We sought to quantify the deviation from intended radiation dose as a function of stent material and mesh density design.

Methods: A laboratory dosimetric film model was used to quantify perturbation of intended radiation dose among 16 different esophageal stents with varying material and stent mesh density design.

Results: Radiation dose enhancement due to stent backscatter ranged from 0 % to 7.3 %, collectively representing a standard difference from the intended mean radiation dose of 1.9 (95 % confidence interval [CI] 1.5 – 2.2). This enhancement was negligible for polymer-based stents and approached 0 % for the biodegradable stents. In contrast, all metal alloy stents had significant radiation backscatter; this was largely determined by the density of mesh design and not by the type of alloy used.

Conclusions: Stent characteristics should be considered when selecting the optimal stent for treatment and palliation of malignant esophageal strictures, especially when adjuvant or neo-adjuvant radiotherapy is planned.

 
  • References

  • 1 Simard EP, Ward EM, Siegel R et al. Cancers with increasing incidence trends in the United States: 1999 through 2008. CA Cancer J Clin 2012; 62: 118-128 Epub 2012 Jan 4
  • 2 Enzinger PC, Mayer RJ. Esophageal cancer. N Engl J Med 2003; 349: 2241-2252
  • 3 Blot WJ, McLaughlin JK. The changing epidemiology of esophageal cancer. Semin Oncol 1999; 26: 2-8
  • 4 Urschel JD, Vasan H. A meta-analysis of randomized controlled trials that compared neoadjuvant chemoradiation and surgery to surgery alone for resectable esophageal cancer. Am J Surg 2003; 185: 538-543
  • 5 Cunningham D, Allum WH, Stenning SP et al. Perioperative chemotherapy versus surgery alone for resectable gastroesophageal cancer. N Engl J Med 2006; 355: 11-20
  • 6 Homs MY, Steyerberg EW, Eijkenboom WM et al. Single-dose brachytherapy versus metal stent placement for the palliation of dysphagia from oesophageal cancer: multicentre randomised trial. Lancet 2004; 364: 1497-1504
  • 7 Bower M, Jones W, Vessels B et al. Nutritional support with endoluminal stenting during neoadjuvant therapy for esophageal malignancy. Ann Surg Oncol 2009; 16: 3161-3168
  • 8 Griffiths EA, Gregory CJ, Pursnani KG et al. The use of biodegradable (SX-ELLA) oesophageal stents to treat dysphagia due to benign and malignant oesophageal disease. Surg Endosc 2012; 26: 2367-2375 Epub 2012 Mar 7
  • 9 Homann N, Noftz MR, Klingenberg-Noftz RD et al. Delayed complications after placement of self-expanding stents in malignant esophageal obstruction: treatment strategies and survival rate. Dig Dis Sci 2008; 53: 334-340
  • 10 Sharma P, Kozarek R. Role of esophageal stents in benign and malignant diseases. Am J Gastroenterol 2010; 105: 258-273 quiz 274
  • 11 Sgourakis G, Gockel I, Radtke A et al. The use of self-expanding stents in esophageal and gastroesophageal junction cancer palliation: a meta-analysis and meta-regression analysis of outcomes. Dig Dis Sci 2010; 55: 3018-3030
  • 12 Chen YK, Schefter TE, Newman F. Esophageal cancer patients undergoing external beam radiation after placement of self-expandable metal stents: is there a risk of radiation dose enhancement?. Gastrointest Endosc 2011; 73: 1109-1114
  • 13 Atwood TF, Hsu A, Ogara MM et al. Radiotherapy dose perturbation of esophageal stents examined in an experimental model. Int J Radiat Oncol Biol Phys 2012; 82: 1659-1664
  • 14 Li XA, Chibani O, Greenwald B et al. Radiotherapy dose perturbation of metallic esophageal stents. Int J Radiat Oncol Biol Phys 2002; 54: 1276-1285
  • 15 Park JY, Shin JH, Song HY et al. Airway complications after covered stent placement for malignant esophageal stricture: special reference to radiation therapy. AJR Am J Roentgenol 2012; 198: 453-459