Semin intervent Radiol 2018; 35(04): 356-358
DOI: 10.1055/s-0038-1669469
Morbidity and Mortality
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

A Complication of Ultrasound-Guided Inferior Vena Cava Filter Placement

Benjamin McDowell
1   Department of Radiology, University of Illinois Health, Chicago, Illinois
,
William Bremer
1   Department of Radiology, University of Illinois Health, Chicago, Illinois
,
Charles E. Ray Jr.
1   Department of Radiology, University of Illinois Health, Chicago, Illinois
› Author Affiliations
Further Information

Publication History

Publication Date:
05 November 2018 (online)

A 45-year-old woman with acquired immunodeficiency syndrome (AIDS) presented to the emergency department with altered mental status and tachycardia. Initial computed tomography (CT) of the head revealed hydrocephalus; an external ventricular drain was placed after the patient's condition continued to decline. She subsequently developed tachycardia, after which a pulmonary embolism protocol chest CT was positive for multiple subsegmental pulmonary emboli. Considering the patient's poor cardiopulmonary reserve and history of recent surgery, interventional radiology (IR) fellow was consulted for emergent after-hours inferior vena cava filter (IVCF) placement.

Due to the patient's hemodynamic instability, a bedside IVCF placement was performed under transabdominal ultrasound (US) guidance in lieu of standard venographic IVCF placement. An IR fellow performed the procedure at bedside under the direct supervision of an IR attending. The fellow performing the procedure had no experience placing a bedside IVCF, and the supervising attending had not performed the procedure in several years.

Preprocedural US revealed what appeared to be a retroaortic left renal vein ([Fig. 1a]). The right common femoral vein was accessed with a 21-gauge needle. Using standard technique, the filter delivery sheath was advanced to the level thought to represent the retroaortic left renal vein. A Gunther Tulip IVCF (Cook Medical Inc., Bloomington, IL) was deployed under direct sonographic guidance ([Fig. 1b]), and appeared to be in normal position on postprocedural US.

Zoom Image
Fig. 1 (a) Transabdominal ultrasound in the axial plane demonstrating a presumed retroaortic left renal vein (arrows). (b) Ultrasound-guided delivery of the inferior vena cava (IVC) filter in the expected location of the infrarenal IVC. (c) Inferior vena cavagram demonstrating the filter in the proximal right common iliac vein. (d) Placement of a new IVC filter in the infrarenal IVC under direct fluoroscopy.

A follow-up abdominal radiograph revealed the filter to be in abnormal position in the projection of the right common iliac vein. The patient was subsequently transported to the IR suite the following morning for filter repositioning. Initial venography confirmed the filter to be in the proximal right common iliac vein ([Fig. 1c]). The malpositioned filter was removed under fluoroscopic guidance following a jugular vein approach with a loop snare device. A venogram following filter retrieval revealed no evidence of vascular injury. A new Gunther Tulip IVCF was placed in standard infrarenal IVC position under direct fluoroscopy ([Fig. 1d]).

No further complications of filter placement were noted during the patient's hospital stay. Magnetic resonance imaging demonstrated multiple ring-enhancing brain lesions, and further workup revealed a severely decreased CD4 count with cerebral spinal fluid studies positive for cytomegalovirus and Epstein–Barr virus. Findings were concerning for infection versus central nervous system lymphoma; however, the patient's tenuous clinical status precluded biopsy and given her severely immunocompromised state, chemotherapy could not be safely offered. After palliative care consultation and discussion with her family, the patient was transferred to a hospice facility.

 
  • References

  • 1 Caplin DM, Nikolic B, Kalva SP, Ganguli S, Saad WE, Zuckerman DA. ; Society of Interventional Radiology Standards of Practice Committee. Quality improvement guidelines for the performance of inferior vena cava filter placement for the prevention of pulmonary embolism. J Vasc Interv Radiol 2011; 22 (11) 1499-1506
  • 2 Gunn AJ, Iqbal SI, Kalva SP. , et al. Intravascular ultrasound-guided inferior vena cava filter placement using a single-puncture technique in 99 patients. Vasc Endovascular Surg 2013; 47 (02) 97-101
  • 3 Conners III MS, Becker S, Guzman RJ. , et al. Duplex scan-directed placement of inferior vena cava filters: a five-year institutional experience. J Vasc Surg 2002; 35 (02) 286-291
  • 4 Amankwah KS, Seymour K, Costanza M, Berger J, Gahtan V. Transabdominal duplex ultrasonography for bedside inferior vena cava filter placement: examples, technique, and review. Vasc Endovascular Surg 2009; 43 (04) 379-384
  • 5 Liu Y, Zhou H, Chen C. , et al. Assessment of the safety and efficacy of bedside ultrasound guidance for inferior vena cava filter placement in critically ill intensive care unit patients. Ultrasound Med Biol 2015; 41 (04) 929-935
  • 6 Van Ha TG. Complications of inferior vena caval filters. Semin Intervent Radiol 2006; 23 (02) 150-155
  • 7 Poletti PA, Becker CD, Prina L. , et al. Long-term results of the Simon nitinol inferior vena cava filter. Eur Radiol 1998; 8 (02) 289-294
  • 8 Jia Z, Wu A, Tam M, Spain J, McKinney JM, Wang W. Caval penetration by inferior vena cava filters: a systematic literature review of clinical significance and management. Circulation 2015; 132 (10) 944-952
  • 9 Milovanovic L, Kennedy SA, Midia M. Procedural and indwelling complications with inferior vena cava filters: frequency, etiology, and management. Semin Intervent Radiol 2015; 32 (01) 34-41
  • 10 Ebaugh JL, Chiou AC, Morasch MD, Matsumura JS, Pearce WH. Bedside vena cava filter placement guided with intravascular ultrasound. J Vasc Surg 2001; 34 (01) 21-26
  • 11 Glocker RJ, Awonuga O, Novak Z. , et al. Bedside inferior vena cava filter placement by intravascular ultrasound in critically ill patients is safe and effective for an extended time. J Vasc Surg Venous Lymphat Disord 2014; 2 (04) 377-382
  • 12 Passman MA, Dattilo JB, Guzman RJ, Naslund TC. Bedside placement of inferior vena cava filters by using transabdominal duplex ultrasonography and intravascular ultrasound imaging. J Vasc Surg 2005; 42 (05) 1027-1032
  • 13 Hodgkiss-Harlow K, Back MR, Brumberg R. , et al. Technical factors affecting the accuracy of bedside IVC filter placement using intravascular ultrasound. Vasc Endovascular Surg 2012; 46 (04) 293-299
  • 14 Trinh VT, Davies JM, Berger MS. Surgery for primary supratentorial brain tumors in the United States, 2000-2009: effect of provider and hospital caseload on complication rates. J Neurosurg 2015; 122 (02) 280-296
  • 15 Stavrakis AI, Ituarte PH, Ko CY, Yeh MW. Surgeon volume as a predictor of outcomes in inpatient and outpatient endocrine surgery. Surgery 2007; 142 (06) 887-899 , discussion 887–899
  • 16 Hu JC, Gold KF, Pashos CL, Mehta SS, Litwin MS. Role of surgeon volume in radical prostatectomy outcomes. J Clin Oncol 2003; 21 (03) 401-405