J Neurol Surg A Cent Eur Neurosurg 2019; 80(03): 223-227
DOI: 10.1055/s-0038-1677518
Case Report
Georg Thieme Verlag KG Stuttgart · New York

Management of Pulse Generators in a Breast Cancer Patient with in Situ Subthalamic Nucleus Deep Brain Stimulation

Byung-chul Son
1   Department of Neurosurgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, the Republic of Korea
,
Joong-Seok Kim
2   Department of Neurology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, the Republic of Korea
,
Woo-Chan Park
3   Department of Surgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, the Republic of Korea
,
Hak-cheol Ko
1   Department of Neurosurgery, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, the Republic of Korea
› Author Affiliations
Further Information

Publication History

25 December 2017

25 May 2018

Publication Date:
01 February 2019 (online)

Abstract

Although deep brain stimulation (DBS) has been used for > 25 years in the treatment of movement disorders, no report has been published on the management of DBS pulse generators implanted in the anterior chest in patients with breast cancer who require mastectomy, radiotherapy, and future imaging studies.

We describe a 62-year-old female patient with advanced Parkinson's disease (PD) who was dependent on bilateral subthalamic nucleus (STN) DBS. She was diagnosed with cancer in her left breast. To avoid difficulties in imaging studies, surgery, and radiotherapy related to the breast cancer, bilateral pulse generators for STN DBS previously implanted in the anterior chest wall were repositioned to the anterior abdominal wall with replacement of long extension cables. During mastectomy and the relocation of the pulse generators, we were not aware of the risks of an open circuit and neuroleptic malignant-like syndrome due to our limited knowledge about how to manage DBS hardware.

Coincident breast cancer and the need for STN DBS is underreported. Considering the uncertainties in the management of pulse generators and the incidence of breast cancer, guidelines for handling DBS hardware in the setting of cancer are needed. More careful attention should be paid to performing magnetic resonance imaging in DBS-dependent patients with chronic PD.

 
  • References

  • 1 Benabid AL, Chabardes S, Mitrofanis J, Pollak P. Deep brain stimulation of the subthalamic nucleus for the treatment of Parkinson's disease. Lancet Neurol 2009; 8 (01) 67-81
  • 2 Zrinzo L, Yoshida F, Hariz MI. , et al. Clinical safety of brain magnetic resonance imaging with implanted deep brain stimulation hardware: large case series and review of the literature. World Neurosurg 2011; 76 (1-2): 164-172 ; discussion 69–73
  • 3 Wadasadawala T, Pandey A, Agarwal JP. , et al. Radiation therapy with implanted cardiac pacemaker devices: a clinical and dosimetric analysis of patients and proposed precautions. Clin Oncol (R Coll Radiol) 2011; 23 (02) 79-85
  • 4 Kamangar F, Dores GM, Anderson WF. Patterns of cancer incidence, mortality, and prevalence across five continents: defining priorities to reduce cancer disparities in different geographic regions of the world. J Clin Oncol 2006; 24 (14) 2137-2150
  • 5 Solan AN, Solan MJ, Bednarz G, Goodkin MB. Treatment of patients with cardiac pacemakers and implantable cardioverter-defibrillators during radiotherapy. Int J Radiat Oncol Biol Phys 2004; 59 (03) 897-904
  • 6 Last A. Radiotherapy in patients with cardiac pacemakers. Br J Radiol 1998; 71 (841) 4-10
  • 7 Lewin AA, Serago CF, Schwade JG, Abitbol AA, Margolis SC. Radiation induced failures of complementary metal oxide semiconductor containing pacemakers: a potentially lethal complication. Int J Radiat Oncol Biol Phys 1984; 10 (10) 1967-1969
  • 8 Levy RM. MRI-compatible neuromodulation devices: critical necessity or desirable adjunct?. Neuromodulation 2014; 17 (07) 619-626
  • 9 Son BC, Han SH, Choi YS. , et al. Transaxillary subpectoral implantation of implantable pulse generator for deep brain stimulation. Neuromodulation 2012; 15 (03) 260-266 ; discussion 266
  • 10 Reyes CV. Clear cell hidradenocarcinoma developing in pacemaker pocket. Pacing Clin Electrophysiol 2008; 31 (11) 1513-1515
  • 11 De Mattia L, Brieda M, Dametto E. A carcinoma of the breast mimicking a pacemaker pocket infection. Europace 2011; 13 (02) 220
  • 12 Mazzetti HA, Tentori MC. Breast cancer and pacemakers. Europace 2014; 16 (02) 303-305
  • 13 ASTM F2503–05. Standard practice for marketing medical devices and other items for safety in the magnetic resonance environment. Available at: https://www.astm.org/DATABASE.CART/HISTORICAL/F2503-05.htm
  • 14 Neurostimulation system: Libra deep brain stimulation system, neurostimulation system, Advanced Neuromodulation Systems, ANS, St. Jude Medical, Plano, TX. Available at: http://www.mrisafety.com/SafetyInfov.asp?SafetyInfoID=301
  • 15 Medtronic. MRI guidelines for Medtronic deep brain stimulation systems: physician's manual. Available at: https://www.zaloker-zaloker.si/assets/Uploads/MR-smernice.pdf
  • 16 Messina G, Rizzi M, Dones I, Franzini A. Cosmetic posterior implant of internal pulse generators in deep brain stimulation procedures: technical report. Neuromodulation 2014; 17 (08) 729-730
  • 17 Neuneier J, Barbe MT, Dohmen C. , et al. Malignant deep brain stimulation-withdrawal syndrome in a patient with Parkinson's disease. Mov Disord 2013; 28 (12) 1640-1641
  • 18 Mizuno Y, Takubo H, Mizuta E, Kuno S. Malignant syndrome in Parkinson's disease: concept and review of the literature. Parkinsonism Relat Disord 2003; 9 (Suppl. 01) S3-S9
  • 19 Munshi A, Wadasadawala T, Sharma PK. , et al. Radiation therapy planning of a breast cancer patient with in situ pacemaker—challenges and lessons. Acta Oncol 2008; 47 (02) 255-260
  • 20 Marbach JR, Sontag MR, Van Dyk J, Wolbarst AB. ; American Association of Physicists in Medicine. Management of radiation oncology patients with implanted cardiac pacemakers: report of AAPM Task Group No. 34. Med Phys 1994; 21 (01) 85-90
  • 21 Sundar S, Symonds RP, Deehan C. Radiotherapy to patients with artificial cardiac pacemakers. Cancer Treat Rev 2005; 31 (06) 474-486