Rofo 2000; 172(12): 1006-1010
DOI: 10.1055/s-2000-9218
ABDOMEN
ORIGINALARBEIT
© Georg Thieme Verlag Stuttgart · New York

Tissue Harmonic Imaging Sonographie Wird die Bildqualität bei Routine-Ultraschalluntersuchungen im Abdomen verbessert?

M. Stiskal1 , R. Steinbach1 , G. Obholzer1 , W. Frank2 , H. Fischer3 , H. Czembirek1
  • 1Krankenhaus der Stadt Wien-Lainz, Zentralröntgeninstitut und Schnittbildzentrum
  • 2Institut für Statistik der Universität Wien
  • 3Boltzmann Institut f. klin. Onkologie u. photodynamische Therapie im Krankenhaus der Stadt Wien-Lainz
Further Information

Publication History

Publication Date:
31 December 2000 (online)

Zusammenfassung.

Ziel: Beurteilung des Einflusses des Tissue harmonic imaging (THI) mit Puls-Inversions-Technik auf die Bildqualität bei Routine-Ultraschalluntersuchungen des Abdomens im Vergleich mit der konventionellen Sonographie. Material und Methoden: Bei 25 Patienten wurden 398 Ultraschallbilder des Abdomens in einer prospektiven Doppelblindstudie ausgewertet. 49 % der Bilder wurden im THI-, 51 % im Basis-Modus angefertigt. Bildparameter, anatomische Region und Schnittebenen waren aufgrund einer umschaltbaren Ultraschallsonde konstant. Bildkontrast, Organabgrenzbarkeit, Gefäßdarstellung und Zystenartefakte wurden von vier Radiologen nach einer dreiteiligen Skala bewertet. Die Interobservervariabilität wurde bestimmt. Ergebnisse: Sämtliche Gruppenergebnisse im THI-Modus waren statistisch signifikant besser als die im Basis-Modus. Eine sehr hohe Übereinstimmung der Interrater bestand bei den Bildparametern „Organgrenzen” und „Bildkontrast”. Eine deutlich geringere Übereinstimmung ergab sich bei den Bildparametern „Gefäßdarstellung” und „Zystenartefakte”. Schlussfolgerungen: Bildkontrast und Abgrenzbarkeit von Organen gegenüber benachbarten Strukturen sowie Reduktion der Artefakte in zystischen Strukturen und Gefäßen werden mit dem THI-Verfahren gegenüber der konventionellen Sonographie unter Nutzung der Puls-Inversions-Technik signifikant verbessert.

Tissue Harmonic Imaging Sonography. Is the image quality in abdominal ultrasound improved?

Purpose: Evaluation and statistical analysis of the influence of THI with pulse-inversion technology on image quality in abdominal ultrasound in comparison to conventional sonography. Material and methods: In 25 patients 398 ultrasound images of the abdomen were evaluated in a prospective, double blind study. 49 % of the images were obtained with THI sonography, 51 % with conventional sonography, respectively. Image parameter, anatomical region and image plane were kept constant by using a switchable ultrasound probe. Image contrast, delineation of organs, detection of vessels and artifacts in cystic structures were evaluated by four radiologists using a 3-grade scale. The interobserver variability was determined. Results: The results of all groups showed statistically significant improvement compared to conventional sonography. High agreement of the evaluaters was achieved for image contrast and delineation of organs, less agreement in detection of vessels and artifacts in cystic structures, respectively. Conclusion: Image contrast and delineation of organs to neighboring structures as well as reduction of artifacts in cystic structures and vessels are statistically significantly improved with THI sonography using pulse-inversion technology in comparison to conventional sonography.

Literatur

  • 1 Schoelgens C. Native™ Tissue Harmonic Imaging.  Radiologe. 1998;  38 420-423
  • 2 Transquart F, Grenier N, Eder V, Pourcelot L. Clinical use of ultrasound tissue harmonic imaging.  Ultrasound Med Biol. 1999;  25 889-894
  • 3 Desser T S, Jeffrey R B, Lane M J, Ralls P W. Tissue Harmonic Imaging: Utility in Abdominal and Pelvic Sonography.  J Clin Ultrasound. 1999;  27 135-141
  • 4 Shapiro R S, Wagreich J, Parsons R B, Stancato-Pasik A, Yeh H C, Lao R. Tissue Harmonic Imaging Sonography: Evaluation of Image Quality Compared with Conventional Sonography.  AJR Am J Roentgenol. 1998;  171 1203-1206
  • 5 Hann L E, Bach A M, Cramer L D, Siegel D, Yoo H H, Garcia R. Hepatic sonography: comparison of tissue harmonic and standard sonography techniques.  AJR Am J Roentgenol. 1999;  173 201-206
  • 6 Tanaka S, Oshikawa O, Sasaki T, Ioka T, Tsukuma H. Evaluation of tissue harmonic imaging for the diagnosis of focal liver lesions.  Ultrasound Med Biol. 2000;  26 183-187
  • 7 Czembirek H. Ultraschall wird lesbar: Standardisierte Dokumentation in der Sonographie des Abdomens. Wien New York; Springer-Verlag 1988
  • 8 Cohen J. A Coefficient of Agreement for Normal Scales.  Educational and Psychological Measurement. 1960;  20 37-46
  • 9 Powers J, Averkion M, Hope-Simpson D. Pulse Inversion Harmonics and other Contrast Methods. ATL Ultrasound, Bothell, Washington 1998: 1-4
  • 10 Haberkorn U, Rudat V, Layer G, Zuna I, Lorenz A, Van Kaick G. Der Einfluß von Bauchwandzusammensetzung und Bauchwanddicke auf das Ultraschall-B-Bild.  Fortschr Röntgenstr. 1991;  155 327-331
  • 11 Haberkorn U, Zuna I, Lorenz A. et al . Echographic tissue characterization in diffuse parenchymal liver disease.  Ultrasound Imaging. 1990;  12 155

Dr. M. Stiskal

Zentralröntgeninstitut und Schnittbildzentrum Krankenhaus der Stadt Wien-Lainz

Wolkersbergenstraße 1 1130 Wien Österreich

Phone: 0043 180110-2610

Fax: 0043 180110-2614

Email: stm@roe.khl.magwien.gv.at

    >