Z Orthop Ihre Grenzgeb 2001; 139(2): 109-116
DOI: 10.1055/s-2001-15040
KNIEENDOPROTHETIK

Georg Thieme Verlag Stuttgart · New York

Navigation in der Knieendoprothetik - vorläufige klinische Erfahrungen und prospektiv vergleichende Studie gegenüber konventioneller lmplantationstechnik

Navigation in Knee Arthroplasty - Preliminary Clinical Experience and Prospective Comparative Study in Comparison with Conventional TechniqueR.  K.  Mielke, U.  Clemens, J.-H.  Jens, S.  Kershally
  • Abt. für Rheumaorthopädie, Nordwestdeutsches Rheumazentrum, St. Josef-Stift Sendenhorst
Further Information

Publication History

Publication Date:
31 December 2001 (online)

Zusammenfassung.

Fragestellung: Zeigt die Anwendung eines computerintegrierten Instrumentationssystems in der Knieendoprothetik (Navigationssystem) grundsätzlich günstige Resultate und wie fällt der Vergleich mit der konventionellen lnstrumentierung aus? Grundlage: Das OrthoPilot-Navigationssystem definiert die individuelle mechanische Beinachse über eine intraoperative kinematische Analyse. Infrarotdioden an festen Trägern, die am Beckenkamm, am distalen Femur, der proximalen Tibia und sprunggelenksnah fixiert sind, werden über eine 3D-Infrarotkamera, die mit einer UNIX work station verbunden ist, räumlich lokalisiert. Über ein integriertes Kalkulationsprogramm werden die Zentren von Hüfte, Knie und Sprunggelenk definiert. Anschließend werden LED-bestückte konventionelle Ausrichtinstrumente zur Festlegung der Hauptresektionsebenen an Femur und Tibia herangezogen. Methode: Die ersten sechzig mit dem OrthoPilot-System operierten Fälle wurden nachuntersucht. Zuzüglich wurde die Gruppe der ersten dreißig Fälle mit einer vergleichbaren konventionell operierten Kontrollgruppe verglichen. Die navigierte Gruppe schließt daher die sogenannte „learning curve” ein. Die Beinachsen und femoralen sowie tibialen Winkel wurden radiologisch bei der postoperativen 3-Monats-Kontrolle erfasst. Ergebnisse: Die radiologisch ermittelten Messwerte der OrthoPilot-Gruppe waren denen der manuellen Gruppe deutlich überlegen. Die Unterschiede waren jedoch statistisch bei den Parametern „mechanische Achse”, „femorale Achse, Sagittalebene” und „tibiale Achse, Frontalebene” nicht signifikant. Beim Parameter „tibiale Achse, Sagittalebene” war eine hoch signifikante Differenz zugunsten des Navitagionssystems zu beobachten. Der Parameter „femorale Achse, Frontalebene” zeigte ein nicht signifikant besseres Ergebnis in der manuellen Gruppe. Gesamthaft war eine leichte Tendenz zur valgischen Positionierung der femoralen Komponenten bei Benutzung des Navigationssystems feststellbar. Komplikationen, die einen Einfluss auf das klinische Nachuntersuchungsergebnis hatten, traten nicht auf Der zusätzliche Zeitaufwand für die Navigation wird zwischen 10 und 15 Minuten angesetzt. Schlussfolgerungen: Das OrthoPilot-Navigationssystem erleichtert eindeutig die akurate Ausrichtung knieendoprothetischer Komponenten in Femur und Tibia. Generell sind die radiologisch ermittelten Ergebnisse mit Einsatz der Navigation günstiger als mit Verwendung eines konventionellen Instrumentariums. Erhebliche Abweichungen von einer optimalen Ausrichtung können bei Anwendung der Navigation weitgehend vermieden werden. Mit zunehmender Erfahrung erwies sich das OrthoPilot-System als zunehmend zuverlässig und in der Anwendung sicher.

Navigation in Knee Arthroplasty - Preliminary Clinical Experience and Prospective Comparative Study in Comparison with Conventional Technique.

Objectives: The efficiency of a computer-integrated instrumentation system in knee arthroplasty was evaluated and compared with a conventional instrumentation system. Background: The OrthoPilot System defines the individual axis of the leg by means of an intraoperative kinematic analysis. LED's mounted on rigid bodies and screwed to the pelvis, femur and tibia are localized by a 3D infrared camera which is linked to a UNIX work station. The integrated calculation program leads to definition of the centres of hip, knee and ankle. Thereafter, LED-equipped alignment instruments allow definition of the femoral and tibial main resection planes. Method: The first sixty cases were included in the study. In addition, thirty cases each were entered into an OrthoPilot group and in a similar conventional control group. The navigated cohort consists of cases one to thirty, thus enclosing the “learning curve”. Results: Leg axes and femoral and tibial angles were assessed radiographically at the three-months postoperative control. Radiological measurements of the OrthoPilot group were clearly superior to those of the manual group. The differences, however, were not statistically different in the parameters “mechanical axis”, “femoral axis lat.” and “tibial axis ap.”. With regard to the parameter “tibial axis lat.” a significant difference in favour of the navigation system was observed. The measurements of “femoral axis ap.” were insignificantly better in the manual group. In general, a slight tendency towards valgus positioning of the femoral components when using the navigation system has to be discussed. Complications influencing the clinical outcome did not occur. Additional time for navigation is calculated in a range of ten to fifteen minutes. Conclusions: The OrthoPilot system clearly facilitates proper alignment of endoprosthetic components in femur and tibia. Generally, the obtained values representing endoprosthetic alignment are superior to conventional technique. Marked deviations from ideal alignment can almost be avoided by means of the navigation system. During the learning curve the OrthoPilot system gained in reliability and reproducability.

Literatur

  • 01 Bargren  J H, Blaha  J D, Freeman  M AR. Alignment in total knee arthroplasty.  Clin Orthop. 1983;;  173 178-183
  • 02 Delp  S L, Stulberg  S D, Davies  B L, Picard  F, Leitner  F. Computer assisted knee replacement.  Clin Orthop. 1998;;  354 49-56
  • 03 DiGioia  A M, Jaramaz  B, Colgan  B D. Computer assisted orthopaedic surgery, image guided and robotic assistive technologies.  Clin Orthop. 1998;;  354 31-39
  • 04 Hood  R W, Vanni  M, Insall  J N. The correction of knee alignment in 225 consecutive total condylar knee replacements.  Clin Orthop. 1981;;  160 94-105
  • 05 Hvid  I, Nielsen  S. Total condylar knee arthroplasty.  Acta Orthop Scand. 1984;;  55 160-165
  • 06 Jeffery  R S, Morris  R W, Denham  R A. Coronal alignment after total knee replacement.  J Bone Joint Surg. 1991;;  73B 709-714
  • 07 Jiang  C, Insall  J N. Effect of rotation on the axial alignment of the femur.  Clin Orthop. 1989;;  248 50-56
  • 08 Jonsson  B, Astrom  J. Alignment and long-term clinical results of a semi-constrained knee prosthesis.  Clin Orthop. 1988;;  226 124-128
  • 09 Knutson  K, Lindstrand  A, Lidgren  L. Survival of knee arthroplasties, a nation-wide multicenter investigation of 8000 cases.  J Bone Joint Surg. 1986;;  68B 795-803
  • 10 Krackow  K A, Serpe  L, Phillips  M J, Bayers-Thering  M, Mihalko  R M. A new technique for determining proper mechanical axis alignment during total knee arthroplasty: Progress toward computer assisted TKA.  Orthopedics. 1999;;  22 698-702
  • 11 Leitner  F, Picard  F, Minfelde  R, Schulz  H J, Clinquin  P, Saragaglia  D. Computer assisted knee surgical total replacement.  In: CVRMed-MRCAS, J. Troccaz, E. Grimson, R. Mösges (Hrsg) Springer 1997;: 630-638
  • 12 Lotke  P A, Ecker  M L. Influence of positioning of prosthesis in total knee replacement.  J Bone Joint Surg. 1977;;  59A 77-79
  • 13 Mielke  R K, Clemens  U, Kershally  S. Computer integrated instrumentation in knee arthroplasty, a comparative study of conventional and computerized technique.  In: Proceedings. 4th Annual North American Program on Computer Assisted Orthopaedic Surgery 93-96, Pittsburgh; 2000
  • 14 Nafei  A, Kristensen  O, Knudson  H M, Hvid  I, Jensen  J. Survivorship analysis of cemented total condylar knee arthoplasty.  J Arthroplasty. 1996;;  11 7-10
  • 15 Oswald  M H, Jakob  R P, Schneider  E, Hoogewoud  H-M. Radiological analysis of normal axial alignment of femur and tibia in view of total knee arthroplasty.  J Arthoplasty. 1993;;  8 419-426
  • 16 Picard  F, Saragaglia  D, Montbarbon  E, Chaussard  C, Leitner  F, Raoult  O. Computer assisted knee arthroplasty - preliminary clinical results with the OrthoPilot System.  In: Proceedings. 4th International CAOS Symposium, Davos; 1999
  • 17 Picard  F, Moody  J, Jaramaz  B, DiGioia  A M, Nikou  C, LaBarca  R S. A classification proposal for computer-assisted knee systems.  In: Proceedings. 4th Annual North American Program on Computer Assisted Orthopaedic Surgery 89 - 90, Pittsburgh; 2000
  • 18 Picard  F, DiGioia  A M, Sell  D, Moody  J, Jaramaz  B, Nikou  C, LaBarca  R S. KneeNav-TKR: Concept and clinical application.  In: Proceedings. 4th Annual North American Program on Computer Assisted Orthopaedic Surgery 99 - 100, Pittsburgh; 2000
  • 19 Ranawat  C S, Flynn  W F, Saddler  S, Hansraj  K H, Maynhard  M J. Long-term results of total condylar knee artbroplasty. A 15-years survivorship study.  Clin Orthop. 1993;;  286 94-102
  • 20 Rand  J A, Coventry  M B. Ten-year evaluation of geometric total knee arthroplasty.  Clin Orthop. 1988;;  232 168-173
  • 21 Ritter  M A, Faris  P M, Keating  E M, Meding  J B. Postoperative alignment of total knee replacement its effect on survival.  Clin Orthop. 1994;;  299 153-156
  • 22 Swanson  K E, Stocks  G W, Warren  P D, Hazel  M R, Janssen  H F. Does axial limb rotation affect the alignment measurements in deformed limbs?.  Clin Orthop. 2000;;  371 246-252
  • 23 Scuderi  G R, Insall  J N, Windsor  R E, Moran  M C. Survivorship of cemented knee replacement.  J Bone Joint Surg. 1989;;  71B 798-803
  • 24 Tew  M, Waugh  W. Tibial-femoral alignment and the results of knee replacement.  J Bone Joint Surg. 1985;;  67B 551-556
  • 25 Wright  J G, Treble  N, Feinstein  A R. Measurement of lower limb alignment using long radiographs.  J Bone Joint Surg. 1991;;  73B 721-723

Prof. Dr. med. Rolf K. Mielke

Abt. für Rheumaorthopädie
Nordwestdeutsches Rheumazentrum St.
Josef-Stift

Westtor 7

48324 Sendenhorst

Email: E-mail: mielke@st-josef-stift.de

    >