Validity and Stability of a Computerized Metabolic System with Mixing Chamber

Ø. Foss1 , J. Hallén1
  • 1Institute of Sports and Biological Studies, Norwegian University of Sport and Physical Education, Oslo, Norway
Further Information

Publication History

Accepted after revision: July 25, 2004

Publication Date:
05 October 2004 (online)

Abstract

The aim of this study was to validate a computerized metabolic system with mixing chamber (Oxycon Pro, Erich Jaeger GmbH, Hoechberg, Germany) against the Douglas bag method (1) over a large range of ventilations and (2) for drift during shorter (25-min time trial) and longer (∼ three months) test periods. Eighteen well-trained/elite cyclists performed graded exercise tests, maximal oxygen uptake tests and time trial tests on an electromagnetic braked cycle ergometer. Respiratory variables were measured simultaneously, once or several times in every test by the Oxycon Pro and the Douglas bag method. (1) Overall, oxygen uptake was 0.8 % (0.03 l · min-1) lower with the Oxycon Pro than with the Douglas bag method with a coefficient of variation of 1.2 % (n = 802) (p < 0.05). (2) During the time trials, oxygen uptake measured with the Oxycon Pro gradually decreased from 0.5 % (0.02 l) lower than the Douglas bag method at 5 min to 1.0 % (0.05 l) lower at 25 min (p < 0.05). Over the period of three months of testing, oxygen uptake measured with the Oxycon Pro gradually increased from 1.1 % (0.04 l) lower than the Douglas bag method at the start to 0.5 % (0.02 l) lower at the end (p < 0.05). This study demonstrates that a computerized metabolic system with mixing chamber is an accurate system for measuring oxygen uptake. This applies to testing over a large range of ventilations and for stability both during shorter and longer test periods.

  • 1 Bassett Jr D R, Howley E T, Thompson D L, King G A, Strath S J, Mclaughlin J E, Parr B B. Validity of inspiratory and expiratory methods of measuring gas exchange with a computerized system.  J Appl Physiol. 2001;  91 218-224
  • 2 Beaver W L, Wasserman K, Whipp B J. On-line computer analysis and breath-by-breath graphical display of exercise function tests.  J Appl Physiol. 1973;  34 128-132
  • 3 Bland J M, Altman D G. Statistical methods for assessing agreement between two methods of clinical measurement.  Lancet. 1986;  1 307-310
  • 4 Carter J, Jeukendrup A E. Validity and reliability of three commercially available breath-by-breath respiratory systems.  Eur J Appl Physiol. 2002;  86 435-441
  • 11 Foss Ø, Hallén J. The most economical cadence increases with increasing workload.  Eur J Appl Physiol. 2004;  92 443-451
  • 5 Gore C J, Clark R J, Shipp N J, van der Ploeg G E, Withers R T. CPX/D underestimates V·O2 in athletes compared with an automated Douglas bag system.  Med Sci Sports Exerc. 2003;  35 1341-1347
  • 6 Jentjens R L, Moseley L, Waring R H, Harding L K, Jeukendrup A E. Oxidation of combined ingestion of glucose and fructose during exercise.  J Appl Physiol. 2004;  96 1277-1284
  • 7 Proctor D N, Beck K C. Delay time adjustments to minimize errors in breath-by-breath measurement of V·O2 during exercise.  J Appl Physiol. 1996;  81 2495-2499
  • 8 Rietjens G J, Kuipers H, Kester A D, Keizer H A. Validation of a computerized metabolic measurement system (Oxycon-Pro) during low and high intensity exercise.  Int J Sports Med. 2001;  22 291-294
  • 9 Salminen R, Aunola S, Malkia E, Vuori I. Computerized breath-by-breath analysis of respiratory variables during exercise.  Med Prog Technol. 1982;  9 27-32

Jostein Hallén

Norwegian University of Sport and Physical Education

PO Box 4014, Ullevål Stadion

0806 Oslo

Norway

Phone: + 4723262314

Fax: + 47 23 26 24 51

Email: jostein.hallen@nih.no

    >