Qualitätskontrolle von Röntgenthoraxaufnahmen bei Kindern in diagnostischen Zentren mit und ohne kinderradiologische Kompetenz

Quality Control of Pediatric Chest X-Rays in Diagnostic Centers with and without Pediatric CompetenceC. D. Alt1 , D. Engelmann1 , J. P. Schenk1 , J. Troeger1
  • 1Abteilung Pädiatrische Radiologie, Universitätsklinik Heidelberg
Further Information

Publication History

eingereicht: 6.4.2005

angenommen: 8.12.2005

Publication Date:
25 January 2006 (online)

Zusammenfassung

Ziel: Der Strahlenschutz im pädiatrischen Röntgen hat eine hohe Bedeutung wegen besonderer Empfindlichkeit strahlensensibler Organe im frühen Lebensalter. Eine optimale Bildqualität dient dem Strahlenschutz und sollte daher angestrebt werden. In unserer Studie untersuchten wir die Bildqualität von Röntgenthoraxaufnahmen bei Kindern in diagnostischen Zentren (Universitätskliniken und Großkliniken). Anschließend beurteilten wir Qualitätsunterschiede zwischen Abteilungen ohne kinderradiologische Kompetenz (R) und Abteilungen mit kinderradiologischer Kompetenz (KR). Material und Methoden: Unser Kollektiv bestand aus 313 konventionellen/analogen Aufnahmen von 207 Patienten (192 p. a./a. p. und 121 seitlich, darunter 43 aus R, 258 aus KR und 12 weder aus R noch aus KR) und 38 digitalen Aufnahmen von 26 Patienten (25 p. a./a. p. und 13 seitlich, 1 aus R, 37 aus KR). Alle Patienten (0 - 18 Jahre) stammen aus der Nephroblastomstudie SIOP-93/01-GPOH. Begutachtet wurden alle initialen Röntgenthoraxaufnahmen, die uns nach Anforderung in der Zeit vom 3.4.2002 - 14.6.2002 zugesandt wurden. Untersuchte Parameter waren: Belichtung, Zentrierung/Positionierung, Einblendung und Bildschärfe. Die Bewertung erfolgte mittels eines Scores von 1 (bestes Ergebnis) bis 5 (schlechtestes Ergebnis). Daraus ergab sich der Gesamtscore (A = optimal, B = kleine Mängel, C = erhebliche Mängel, D = unbrauchbar). Zusätzliche Messungen galten der optischen Dichte, dem Bildzentrum und der relativen Feldgröße. Zusätzlich führten wir bei den konventionellen Aufnahmen statistische Tests durch mit dem Mann-Whitney-U-Test und der logistischen Regression. Die Studie wurde retrospektiv durchgeführt. Die Belichtung, die Bildschärfe und die optische Dichte wurden bei den digitalen Aufnahmen nicht analysiert. Ergebnisse: Bei allen konventionellen Aufnahmen zeigten die Parameter Zentrierung/Positionierung und Einblendung nur eine mäßige Qualität (Mittelwerte: 2,4 bzw. 2,8), die Belichtung und die Bildschärfe eine gute bis sehr gute Qualität (Mittelwerte: 1,9 bzw. 1,5). Die Qualität der Thoraxaufnahmen aus KR war vor allem bei jüngeren Patienten (bis 5 Jahre) besser als die Qualität der Thoraxaufnahmen aus R (Mittelwert Altersgruppe 0 - 1 Monat: KR = 1,7; Mittelwerte Altersgruppe 2 Monate - 2 Jahre: R = 2,4 und KR = 1,8; Mittelwerte Altersgruppe 3 - 5 Jahre: R = 2,5 und KR = 1,8). Schlussfolgerung: Trotz insgesamt guter Bildqualität zeigten sich vor allem im Bereich der Zentrierung/Positionierung und der Einblendung bei beiden Untersuchergruppen (R und KR) noch Qualitätsmängel. Dadurch konnten die geforderten strahlenhygienischen Anforderungen teilweise nicht erfüllt werden. Die Aufnahmen aus KR zeigten im Vergleich zu den Aufnahmen aus R eine höhere Qualität in diesem speziellen Studienkollektiv. Tägliche Qualitätskontrollen sind bei Röntgenthoraxaufnahmen bei Kindern zum Erreichen eines hohen Qualitätsstandards notwendig.

Abstract

Purpose: Radiation protection in pediatric radiology is very important because of the particular sensitivity of radiosensitive organs in younger patients. Optimized image quality supports radiation protection and should be targeted. In our study we examined the quality of pediatric chest X-rays at diagnostic centers (university hospitals and other large clinics). We then evaluated differences in image quality in departments without pediatric competence (R) and departments with pediatric competence (PR). Materials and Methods: Our study was based on 313 conventional chest X-rays from 207 patients (192 p. a./a. p. and 121 lateral, 43 from R, 258 from PR and 12 neither from R nor KR) and 38 digital chest X-rays from 26 patients (25 p. a./a. p. and 13 lateral, 1 from R and 37 from PR). All patients (age 0 - 18 years) are from Nephroblastoma-Study SIOP-93/01-GPOH. We examined all initial chest X-rays, which were sent to us for evaluation upon request between 4/3/2002 and 6/14/2002. The examined parameters were: exposure, centering of the X-rays/patient positioning, collimation and sharpness. The X-rays were evaluated on a scale from 1 (best result) to 5 (worst result), resulting in an overall score of A = optimum, B = minor problems, C = major problems, or D = unusable. The optical density, the center of the image and the relative field size were also measured. Statistical tests (Mann-Whitney-U and log regression) were carried out on the conventional images. The study was performed retrospectively. The exposure, sharpness and optical density of the digital X-rays were not analyzed. Results: In the case of all conventional X-rays, the quality of the centering of the X-rays/patient positioning and collimation was moderate (average scale value: 2.4 and 2.8), and the quality of the exposure and sharpness was good and very good (average scale value: 1.9 and 1.5). The quality of the chest X-rays in departments with additional pediatric radiological expertise was better mainly in the case of younger patients (younger than 5 years) than departments without additional pediatric radiological expertise (average scale value in age group 0 - 1 month: PR = 1.7; average scale value in age group 2 months - 2 years: R = 2.4 and PR = 1.8; average scale value in age group 3 - 5 years: R = 2.5 and PR = 1.8). Conclusion: Despite the good overall image quality, the quality of the centering of the X-rays/patient positioning and collimation was insufficient in both examiner groups (R and PR). For this reason, some radiation protection requirements could not be fulfilled. X-rays from PR were higher quality than X-rays from R in this special study group. Day-to-day quality checks are necessary for pediatric chest X-rays in order to achieve a high quality standard.

Literatur

  • 1 Bundesärztekammer . Leitlinien der Bundesärztekammer zur Qualitätssicherung in der Röntgendiagnostik.  21.4.1995;  , www.bundesaerztekammer.de
  • 2 Röntgenverordnung 2003.  , www.uni-wuerzburg.de/strahlenschutzstelle/strl/roev.html
  • 3 Cook J V, Kyriou J C, Pettet A. et al . Key factors in the optimization of paediatric X-ray practice.  BJR. 2001;  178 1032-1040
  • 4 Fendel H, Stieve F E. Strahlenschutz in der Kinderradiologie, NCRP-Bericht Nr. 68. Berlin; H Hoffmann Verlag 1983
  • 5 Mönig M, Wenz W. Auswirkungen der neuen Röntgenverordnung auf die Röntgendiagnostik.  Radiologe. 1988;  178 12-19
  • 6 Tasch C, Dütting T, Zieger B. et al . Qualitätskontrolle in der ambulanten röntgendiagnostischen Versorgung von Kindern und Jugendlichen am Beispiel der konventionellen Röntgenaufnahme - Pilotstudie.  Fortschr Röntgenstr. 1999;  178 121-125
  • 7 Engelmann D, Dütting T, Wunsch R. et al . Qualität der ambulanten Röntgenthoraxuntersuchung des Kindes - eine Pilotstudie.  Radiologe. 2001;  178 442-446
  • 8 Study Group on Quality Criteria Development in Paediatrics of the European Commission .European Guidelines on Quality Criteria for Diagnostic Radiographic Images in Peadiatrics,. EUR 16 261 1996
  • 9 Fendel H, Stieve F E. Vernünftige diagnostische Bildgebung bei Kindern, Bericht einer Studiengruppe der WHO. Berlin; H Hoffmann Verlag 1990
  • 10 Kaul A. Kleine Strahlendosen: Risiko und Nutzen.  Radiologe. 1991;  178 225-226
  • 11 Kereiakes J G. Handbook of radiation doses. Boca Raton Florida; CRC Press 1980
  • 12 Krug B, Böttge M, Coburger S. et al . Qualitätskontrolle der ambulanten bildgebenden Diagnostik in Nordrhein-Westfalen, Teil 1.  Fortschr Röntgenstr. 2003;  178 46-57
  • 13 Fendel H, Schneider K, Schöfer H. et al .Wirksamkeit von Strahlenanwendungen in der Kinderheilkunde. Abschlussbericht BMU St.Sch. 4024/887 LMU München; 1986
  • 14 Meier-Duis H, Mödder U. Qualitätssicherung in der Radiologie. Berlin; Blackwell Wissenschaft 1992
  • 15 Lewentat G, Bohndorf K. Analyse verworfenen Filmmaterials als Element der Qualitätssicherung in der Diagnostischen Radiologie.  Fortschr Röntgenstr. 1997;  178 376-381
  • 16 Busch H P. Das Krankenhaus im Wandel - Die radiologische Abteilung auf dem Weg zum Dienstleistungszentrum Radiologie.  Fortschr Röntgenstr. 2004;  178 1546-1548
  • 17 Lehmkuhl L, Mulzer J, Teichgraeber U. et al . Evaluation der Abbildungsqualität unterschiedlicher Befundmodalitäten in der digitalen Radiologie.  Fortschr Röntgenstr. 2004;  178 1031-1038

Céline Désirée Alt

Abteilung Pädiatrische Radiologie, Universitätsklinik Heidelberg

69120 Heidelberg

Phone: ++ 49/62 21/ 56 23 30

Fax: ++ 49/62 21/ 56 29 98

Email: celinealt@gmx.de

    >