Interobserver Variability of 64-Slice Computed Tomography for the Quantification of Non-Calcified Coronary Atherosclerotic Plaque

Interobserver-Variabilität der nichtinvasiven 64-Zeilen-Computertomografie bei der Quantifizierung nichtkalzifizierter KoronarplaquesT. Pflederer1 , M. Schmid1 , D. Ropers1 , U. Ropers1 , S. Komatsu2 , W. G. Daniel1 , S. Achenbach1
  • 1Department of Internal Medicine 2, University of Erlangen
  • 2Cardiovascular Division, Osaka Police Hospital
Further Information

Publication History

received: 11.12.2006

accepted: 1.3.2007

Publication Date:
09 May 2007 (online)

Zusammenfassung

Ziel: Die Mehrzeilen-Computertomografie (MDCT) erlaubt eine verlässliche Detektion und Quantifizierung von kalzifizierten Plaques in den Koronargefäßen. Bezüglich der Zuverlässigkeit der MDCT zur Quantifizierung nichtkalzifizierter Plaques existieren bislang nur wenige Angaben. Wir ermittelten die Interobserver-Variabilität (IOV) der 64-Zeilen-CT bei der Bestimmung des Volumens nichtkalzifizierter Plaques in den drei Koronararterien. Material und Methoden: Datensätze von 41 Patienten wurden ausgewertet, bei denen sich mittels MDCT (Sensation 64, Siemens Forchheim, Deutschland, 330 ms Rotation, 0,6 mm Kollimation, 60 ml Kontrastmittel i. v., Flussrate 5 ml/s) nichtkalzifizierte Plaques im proximalen Segment mindestens eines Koronargefäßes nachweisen ließen. Die Bildqualität der Datensätze wurde nach einer 4-stufigen Skala beurteilt. Die Plaquevolumina wurden unabhängig durch zwei geblindete Untersucher ermittelt. Dies erfolgte mittels manueller Flächenbestimmung in aufeinander folgenden Querschnitten durch die jeweilige nichtkalzifizierte Plaque. Dabei wurde eine Schichtdicke von 1 mm sowie ein Inkrement von 0,5 mm verwendet. Anschließend wurde die IOV bestimmt und der Einfluss von Bildqualität und Plaquegröße untersucht. Ergebnisse: Das mittlere Volumen der nichtkalzifizierten Plaques betrug im RIVA 157 ± 85 mm3, im RCX 76 ± 43 mm3 und in der RCA 133 ± 80 mm3 (RIVA vs. RCX: p < 0,01; RIVA vs. RCA: p = 0,33; RCX vs. RCA: p < 0,01). Für RIVA, RCX und RCA zeigte sich eine mittlere, absolute Interobserverdifferenz bei der Bestimmung der Plaquevolumina von 23 ± 15 mm3, 20 ± 9 mm3 und 38 ± 21 mm3. Dies entsprach einer mittleren Interobserver-Variabilität von 17 ± 10 % (RIVA), 29 ± 13 % (RCX) und 32 ± 13 % (RCA) (RIVA vs. RCX: p < 0,01; RIVA vs. RCA: p < 0,01; RCX vs. RCA: p = 0,87). Es zeigte sich eine signifikante inverse Korrelation zwischen IOV und der Größe der Plaquevolumina selbst (r = - 0,48; p = 0,01). Die IOV war abhängig von der Bildqualität: Die höchste Bildqualität stellte sich im Bereich des RIVA dar (2,4 ± 0,5), während RCX (2,1 ± 0,7) und RCA (2,0 ± 0,6) eine schlechtere Bildqualität aufwiesen. Schlussfolgerung: Die 64-Zeilen-CT weist bei der Quantifizierung nichtkalzifizierter Plaquevolumina eine substanzielle Interobserver-Variabilität (IOV) auf. Die IOV im Bereich des RIVA ist signifikant niedriger als in RCX und RCA. Dies scheint im Vergleich zu den anderen Koronararterien vor allem in einem größeren mittleren Plaquevolumen sowie in einer besseren Bildqualität im Bereich des RIVA begründet zu sein.

Abstract

Purpose: Multidetector CT (MDCT) is a reliable tool to assess and quantify calcified plaque in coronary arteries. Only very limited information is available concerning the accuracy of MDCT for evaluating non-calcified plaque. We determined the interobserver variability for measuring non-calcified plaque volumes in the three main coronary arteries using 64-slice computed tomography. Materials and Methods: We retrospectively evaluated data sets of 41 patients who received a 64-slice CT scan (Sensation 64, Siemens Forchheim, Germany, 330 msec rotation, 0.6 mm collimation, 60 ml contrast agent i. v. at 5 ml/sec) due to suspected stable coronary artery disease. The patients showed presence of non-calcified plaque in the proximal part of at least one main coronary artery. The image quality was defined on the basis of a 4-point rating scale. Two independent and blinded investigators measured the plaque volume of the non-calcified plaque by manually tracing plaque areas in contiguous cross-sectional reconstructions rendered along the vessel centerline using a slice thickness of 1 mm and an increment of 0.5 mm. The interobserver variability was evaluated and the influence of plaque volume and image quality on interobserver variability was determined. Results: The mean volume of non-calcified plaque was 157 ± 85 mm3, 76 ± 43 mm3 and 133 ± 80 mm3 for the LAD, LCX and RCA, respectively (LAD vs. LCX: p < 0.01; LAD vs. RCA: p = 0.33; LCX vs. RCA: p < 0.01). There was a mean absolute difference in plaque volume between the two observers of 23 ± 15 mm3, of 20 ± 9 mm3 and of 38 ± 21 mm3, which corresponds to a mean interobserver variability of 17 ± 10 %, 29 ± 13 % and 32 ± 13 % for the LAD, LCX and RCA, respectively (LAD vs. LCX: p < 0.01; LAD vs. RCA: p < 0.01; LCX vs. RCA: p = 0.87). A significant inverse correlation between interobserver variability and the extent of the plaque volume (r = - 0.48; p = 0.01) was found. Interobserver variability was dependent on image quality: The highest image quality was observed in the LAD (2.4 ± 0.5), while the image quality in the LCX (2.1 ± 0.7) and the RCA (2.0 ± 0.6) was lower. Conclusion: Interobserver variability for the quantification of non-calcified plaque volumes in 64-slice MDCT is substantial. Interobserver variability in the LAD was significantly lower than in the LCX and the RCA. This might be due to a larger mean plaque volume and better image quality in the LAD than in other coronary arteries.

References

  • 1 Arad Y, Spadaro L A, Goodman K. et al . Prediction of coronary events with electron beam computed tomography.  J Am Coll Cardiol. 2000;  36 1253-1260
  • 2 Achenbach S, Ropers D, Pohle K. et al . Influence of lipid-lowering therapy on the progression of coronary artery calcification: a prospective evaluation.  Circulation. 2002;  106 1077-1082
  • 3 Guerci A D, Spadaro L A, Goodman K J. et al . Comparison of electron computed tomography scanning and conventional risk factor assessment for the prediction of angiographic coronary artery disease.  J Am Coll Cardiol. 1998;  32 673-679
  • 4 Hoffmann U, Moselewski F, Nieman K. et al . Noninvasive assessment of plaque morphology and composition in culprit and stable lesions in acute coronary syndrome and stable lesions in stable angina by multidetector computed tomography.  J Am Coll Cardiol. 2006;  47 1655-1662
  • 5 Leber A W, Knez A, White C W. et al . Composition of coronary atherosclerotic plaques in patients with acute myocardial infarction and stable angina pectoris determined by contrast-enhanced multislice computed tomography.  Am J Cardiol. 2003;  91 714-718
  • 6 Hausleiter J, Meyer T, Hadamitzky M. et al . Prevalence of noncalcified coronary plaques by 64-slice computed tomography in patients with an intermediate risk for significant coronary artery disease.  J Am Coll Cardiol. 2006;  48 312-318
  • 7 Leber A W, Knez A, Becker A. et al . Accuracy of multidetector spiral computed tomography in identifying and differentiating the composition of coronary atherosclerotic plaque: a comparative study with intracoronary ultrasound.  J Am Coll Cardiol. 2004;  43 1241-1247
  • 8 Achenbach S, Moselewski F, Ropers D. et al . Detection of calcified and non-calcified coronary atherosclerotic plaque by contrast-enhanced, submillimeter multidetector spiral computed tomography: a segment-based comparison with intravascular ultrasound.  Circulation. 2004;  109 14-17
  • 9 Moselewski F, Ropers D, Pohle K. et al . Comparison of measurement of cross-sectional coronary atherosclerotic plaque and vessel areas by 16-slice multidetector computed tomography versus intravascular ultrasound.  Am J Cardiol. 2004;  94 1294-1297
  • 10 Leber A W, Becker A, Knez A. et al . Accuracy of 64-slice computed tomography to classify and quantify plaque volumes in the proximal coronary system.  J Am Coll Cardiol. 2006;  47 672-677
  • 11 Weber C, Begemann P, Wedegärtner U. et al . Koronarkalkquantifizierung und Koronarangiographie mittels Mehrzeilendetektorspiral-CT - Klinische Erfahrungen.  Fortschr Röntgenstr. 2005;  177 50-59
  • 12 Kovacs A, Probst C, Sommer T. et al . CT-Koronarangiographie bei Patienten mit Vorhofflimmern.  Fortschr Röntgenstr. 2005;  177 1655-1662
  • 13 Dewey M, Hoffmann H, Hamm B. Multislice CT coronary angiography: Effect of sublingual nitroglycerine on the diameter of coronary arteries.  Fortschr Röntgenstr. 2006;  178 600-604
  • 14 Ferencik M, Nieman K, Achenbach S. Non-calcified and calcified coronary plaque detection by contrast-enhanced multi-detector computed tomography: a study of interobserver agreement.  J Am Coll Cardiol. 2006;  47 207-209
  • 15 Jensen L O, Thayssen P, Pedersen K E. et al . Low variation and high reproducibility in plaque volume with intravascular ultrasound.  Int J Cardiol. 2004;  97 463-469
  • 16 Hong M K, Mintz G S, Lee S W. et al . The site of plaque rupture in native coronary arteries.  J Am Coll Cardiol. 2005;  46 261-265
  • 17 Leschka S, Alkadhi H, Plass A. et al . Accuracy of MSCT coronary angiography with 64-slice technology: first experience.  Eur Heart J. 2005;  26 1482-1487
  • 18 Ropers D, Rixe J, Anders K. et al . Usefulness of multidetector row spiral computed tomography with 64- × 0.6-mm and 330-msec rotation for the noninvasive detection of significant coronary artey stenoses.  Am J Cardiol. 2006;  97 343-348

Dr. Tobias Pflederer

Department of Internal Medicine 2, University of Erlangen

Ulmenweg 18

91054 Erlangen

Phone: ++49/91 31/8 53 55 31

Fax: ++49/91 31/8 53 55 32

Email: tobiaspflederer@web.de

    >