Endoscopy 2007; 39(7): 637-642
DOI: 10.1055/s-2007-966571
Original article

© Georg Thieme Verlag KG Stuttgart · New York

Prospective nonrandomized comparison of two modes of argon beamer (APC) tumor desobstruction: effectiveness of the new pulsed APC versus forced APC

A.  Eickhoff1 , R.  Jakobs1 , D.  Schilling1 , D.  Hartmann1 , U.  Weickert1 , M.  D.  Enderle2 , J.  C.  Eickhoff3 , J.  F.  Riemann1
  • 1Medical Department C, Klinikum Ludwigshafen GmbH, Ludwigshafen, Germany
  • 2ERBE Elektromedizin GmbH, Tübingen, Germany
  • 3Department of Biostatistics and Medical Informatics, University of Wisconsin, Madison, Wisconsin, USA
Further Information

Publication History

submitted 27 July 2006

accepted after revision 9 March 2007

Publication Date:
05 July 2007 (online)

Background and study aim: Argon plasma coagulation (APC) has become an established noncontact method of tumor palliation in a variety of locations. The present prospective study evaluated a new APC system (APC-2) using amplified power settings and different application modes, such as intermittent energy delivery (pulsed APC) in comparison with the conventional technique (forced APC).

Patients and methods: A total of 100 patients with esophageal, gastric, or rectal tumors were alternately (but not randomized) enrolled and treated with either pulsed APC (n = 46) or forced APC (n = 54). Parameters to assess the palliative effect were: amount of lumen restoration (⅓, ⅔, complete), objective planimetry, stenosis length, treatment time, and number of APC sessions.

Results: Overall response rate was similar in both groups (pulsed 83 %, forced 87 %), the same was found in the subgroups with different amounts of lumen restoration and for the other objective parameters. However, the tumor debulking effect was achieved in a significantly shorter median treatment time with forced compared with pulsed APC (13.6 vs. 18.2 minutes, P = 0.03), with a similar number of treatment sessions in both groups. Complications also occurred with similar frequency in both groups.

Conclusions: There was no significant difference in overall local tumor response between the two modes of APC application. However, data from this nonrandomized study suggest a faster achievement of response with forced APC. A combination of both modes may be superior.

References

  • 1 Brand E, Pearlman N. Electrosurgical debulking of ovarian cancer: a new technique using the argon beam coagulator.  Gynecol Oncol. 1990;  39 115-118
  • 2 Grund K E, Storek D, Farin G. Endoscopic argon plasma coagulation (APC) first clinical experiences in flexible endoscopy.  Endosc Surg Allied Technol. 1994;  2 42-46
  • 3 Grund K E, Zindel C, Farin G. Argon plasma coagulation through a flexible endoscope. Evaluation of a new therapeutic method after 1606 uses.  Dtsch Med Wochenschr. 1997;  122 432-438
  • 4 Johanns W, Luis W, Janssen J. et al . Argon plasma coagulation (APC) in gastroenterology: experimental and clinical experiences.  Eur J Gastroenterol Hepatol. 1997;  9 581-587
  • 5 Vargo J J. Clinical applications of the argon plasma coagulator.  Gastrointest Endosc. 2004;  59 81-88
  • 6 Canard J M, Vedrenne B. Clinical application of argon plasma coagulation in gastrointestinal endoscopy: has the time come to replace the laser?.  Endoscopy. 2001;  33 353-357
  • 7 Manner H, May A, Faerber M. et al . The effect of argon plasma coagulation on porcine liver tissue using the new VIO system in comparison to APC 300 and Nd: YAG laser.  Gastrointest Endosc. 2005;  61 T1366
  • 8 Frank J, Queck J, Koukaki S. et al . New argon plasma coagulation modes and first unicentric clinical experiences in gastrointestinal endoscopy.  Endo Heute. 2006;  19 15-22
  • 9 Sumiyama K, Kaise M, Kato M. et al . New generation argon plasma coagulation in flexible endoscopy: ex vivo study and clinical experience.  J Gastroenterol Hepatol. 2006;  21 1122-1128
  • 10 Robertson G S, Thomas M, Jamieson J. et al . Palliation of oesophageal carcinoma using the argon beam coagulator.  Br J Surg. 1996;  83 1769-1771
  • 11 Akhtar K, Byrne J P, Bancewicz J, Attwood S E. Argon beam plasma coagulation in the management of cancers of the esophagus and stomach.  Surg Endosc. 2000;  14 1127-1130
  • 12 Heindorff H, Wojdemann M, Bisgaard T, Svendsen L B. Endoscopic palliation of inoperable cancer of the oesophagus or cardia by argon electrocoagulation.  Scand J Gastroenterol. 1998;  33 21-23
  • 13 Sherbakov A M. Endoscopic argon-plasma coagulation application in palliative treatment of recurrent stomach cancer at the esophageal anastomosis.  Vopr Onkol. 2005;  51 230-231
  • 14 Wahab P J, Mulder C J, den Hartog G, Thies J E. Argon plasma coagulation in flexible gastrointestinal endoscopy: pilot experiences.  Endoscopy. 1997;  29 176-181
  • 15 Manner H, May A, Faerber M. et al . Safety and efficacy of a new high power argon plasma coagulation system (hp-APC) in lesions of the upper gastrointestinal tract.  Dig Liver Dis. 2006;  38 471-478
  • 16 Jakobs R, Miola J, Eickhoff A. et al . Endoscopic laser palliation for rectal cancer - therapeutic outcome and complications in eighty-three consecutive patients.  Z Gastroenterol. 2002;  40 551-556

A. Eickhoff, MD 

Medical Department C

Klinikum Ludwigshafen

Bremserstr. 79

67063 Ludwigshafen

Germany

Fax: +49-621-5034114

Email: eickhofa@klilu.de

    >