Endoscopy 2007; 39(11): 948-951
DOI: 10.1055/s-2007-966858
Original article

© Georg Thieme Verlag KG Stuttgart · New York

Accuracy of measurements of the length of esophageal lesions: An experimental study

D.  Polymeros1 , K.  Triantafyllou1 , C.  Tzathas1 , T.  Emmanuel1 , E.  Tsironi1 , S.  D.  Ladas1
  • 1Hepato-Gastroenterology Unit, Second Propaedeutic Department of Medicine, Medical School, Athens University, Attikon University General Hospital, Athens, Greece
Further Information

Publication History

submitted 9 December 2006

accepted after revision 12 June 2007

Publication Date:
16 November 2007 (online)

Background and study aims:Hiatal hernia, Barrett’s esophagus and erosive esophagitis are defined and classified by measuring their length during endoscopy. The primary aim of our study was to evaluate the accuracy of length measurement of esophageal lesions with the conventional gastroscope marked at 5-cm intervals, and the secondary aim was to test the performance of a modified gastroscope marked at 1-cm intervals.

Methods: 24 senior endoscopists carried out endoscopy on a plastic mannequin. The esophagus was randomly replaced by one of a set of seven plastic tubes, each tube having two colored rings which were 1, 1.5, 2, 2.5, 3, 3.5 or 4 cm apart. The endoscopists were asked to measure the distance between the mouth-guard and each of the two ”esophageal” rings during endoscope withdrawal, with a precision of 0.5 cm. All participants endoscoped all seven tubes blindly, first using the conventional scope and then the modified scope.

Results: Using the conventional gastroscope, measurements were overall incorrect in 67.9 % (95 % CI 61 - 75.1) of cases and incorrect by at least 1 cm in 21 % (95 % CI 14.8 - 27.2) of cases. These percentages were significantly reduced by using the modified gastroscope (47.6 %, 95 % CI 40.1 - 55.2 and 7.1 %, 95 % CI 3.2 - 11.0, respectively) (P < 0.001). Overall accuracies in measuring lengths of ≥ 2 cm and ≥ 3 cm, corresponding to hiatal hernia and Barrett’s esophagus definitions, were 83 % and 94 % - 95 % with the conventional and modified gastroscopes, respectively (P < 0.05).

Conclusions: Our data suggest that estimation of the length of esophageal lesions using conventional endoscopes is inaccurate. The accuracy of measurements is significantly improved with the use of an endoscope marked at intervals of 1 cm.

References

  • 1 Hill L D, Kozarek R A, Kraemer S J. et al . The gastroesophageal flap valve: in vitro and in vivo observations.  Gastrointest Endosc. 1996;  44 541-547
  • 2 Armstrong D, Bennett J R, Blum AL. et al . The endoscopic assessment of esophagitis: a progress report on observer agreement.  Gastroenterology. 1996;  111 85-92
  • 3 Sharma P, McQaid K, Dent J. et al . AGA Chicago Workshop. A critical review of the diagnosis and management of Barrett’s esophagus: the AGA Chicago Workshop.  Gastroenterology. 2004;  127 310-330
  • 4 Sharma P, Dent J, Armstrong D. et al . The development and validation of an endoscopic grading system for Barrett’s esophagus: The Prague C & M criteria.  Gastroenterology. 2006;  131 1392-1399
  • 5 Avidan B, Sonnenberg A, Schnell T. et al . Hiatal hernia size, Barrett’s length, and severity of acid reflux are all risk factors for esophageal adenocarcinoma.  Am J Gastroenterol. 2002;  97 1930-1936
  • 6 Kim S L, Waring J P, Spechler S J. et al . Diagnostic inconsistencies in Barrett’s esophagus.  Gastroenterology. 1994;  107 945-949
  • 7 Dekel R, Wakelin D E, Wendel C. et al . Progression or regression of Barrett’s esophagus - is it all in the eye of the beholder?.  Am J Gastroenterol. 2003;  98 2612-2615
  • 8 Eisen G M, Montgomery E A, Azumi N. et al . Qualitative mapping of Barrett’s metaplasia: a prerequisite for intervention trials.  Gastrointest Endosc. 1999;  50 814-818
  • 9 Wakabayashi T, Nakazawa S, Yoshino J. et al . A new method of real-time endoscopic measurement with an electric catheter.  . 1994;  26 466-469

S. D. Ladas, MD

Hepato-Gastroenterology Unit

Attikon University General Hospital

1 Rimini Street

12462 AthensGreece

Fax: + 30-210-5326422

Email: sdladas@hol.gr

    >