Semin Thromb Hemost 1995; 21(S 02): 20-24
DOI: 10.1055/s-0032-1313598
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Hematological Parameters Influencing the Thrombostat 4000

Dietmar Söhngen
,
Eckhard Hatistein
,
Axel Heyll
,
Gerald Meckenstock
,
Simone Wienen
,
Wolfgang Schneider
Further Information

Publication History

Publication Date:
09 May 2012 (online)

Abstract

The effects of platelet counts, hematocrit, and leukocyte counts were studied on the closure times of the Thrombostat 4000 (in-vitro bleeding time, IVBT). Closure times became longer with platelet counts <50 × 109/L; an inverse linear correlation could be established. Hematocrit was also inversely correlated with the closure time. At constant platelet counts a hematocrit of 55% yielded an immediate closure of the filter, while with a hematocrit <15% no closure times could be measured. At constant platelet counts and hematocrits, nomonuclear and polymorphonuclear leukocytes also influenced closure times; increased counts resulted in shorter closure times. Leukocytes from a patient with chronic myelocytic leukemia had the same effects.

 
  • References

  • 1 Ivy AC, Nelson D, Bucher GR. The standardization of certain factors in the cutaneous “venostasis” bleeding time technique. J Lab Clin Med 26: 1812-1822 1941;
  • 2 Mielke CH, Kaneshiro MM, Maher IA, Weiner JM, Rapaport SI. The standardized normal Ivy bleeding time and its prolongation by aspirin. Blood 34: 204-215 1969;
  • 3 Niewenhuis HK, Sixma JJ. Bleeding time measurements. In: Harker LA, Zimmerman TS, (Eds): Measurements of Platelet Function . Churchill Livingstone; New York: 1983: 26-45
  • 4 Rodgers RPC, Levin J. A clinical reappraisal of the bleeding time. Semin Thromb Hemost 16: 1-144 1990;
  • 5 Didisheim P, Claypool DA, Stropp JQ, Baker RW. The Teflon tubing bleeding time in man. Thromb Haemost 46 (131 (abst)) 1981;
  • 6 Uchiyama S, Stropp JQ, Claypool DA, Didisheim P, Dewanjee MK. Filter bleeding time: A new in vitro test of hemostasis. 1. Evaluation in normal and thrombocytopenic subjects. Thromb Res 31: 99-116 1983;
  • 7 Kratzer MAA, Born GVR. Simulation of primary haemostasis in vitro. Haemostasis 15: 357-362 1985;
  • 8 Kratzer MAA, Belucci S, Caen JP. Detection of abnormal platelet functions with an in vitro model of primary haemostasis. Haemostasis 15: 363-370 1985;
  • 9 Kretschmer V, Schikor B, Söhngen D, Dietrich G. In vitro bleeding test—A simple method for the detection of aspirin effects on platelet function. Thromb Res 56: 593-602 1989;
  • 10 Dietrich G, Kretschmer V, Schikor B, Söhngen D. Der In-vitro-Blutungstest (IVBT)—Eine empfindliche Methode zum Nachweis von Plättchenfunktionsstörungen bei Thrombozytenspendem.. In: Kretschmer V, Stangel W, (Eds) Contributions to Infusion Therapy and Clinical Nutrition . vol 26. Karger; Basel: 1990: 142-146