RSS-Feed abonnieren
DOI: 10.1055/a-2631-7218
Economics, Ergonomics, Effectiveness: A Monocentric Comparison of Single-Use vs. Reusable Flexible Ureteroscopes in Challenging Kidney Stone Cases
Ökonomie, Ergonomie, Effektivität – ein monozentrischer Vergleich von Einweg- und wiederverwendbaren flexiblen Ureterorenoskopen bei anspruchsvollen NierensteinfällenAuthors

Abstract
Introduction
Hybrid strategies involving the use of both single-use flexible ureteroscopes (sufURS) and reusable flexible ureteroscopes (rfURS) are being implemented worldwide to extend the longevity of rfURS. We present a retrospective comparison of the clinical performance, ergonomics, and financial aspects between sufURS and rfURS.
Material & Methods
Challenging cases were preoperatively selected based on the presence of nephrolithiasis requiring laser lithotripsy and a steep preoperative infundibulopelvic angle (IPA) < 50°. All procedures were performed at the UKSH Campus Lübeck between January 2020 and March 2021 using sufURS (Pusen PU3022A) and rfURS (Olympus URF-V3). Postoperative surveys to evaluate the performance and ergonomics were completed by the surgeon and the assisting nurse.
Results
A total of 29 sufURS procedures were followed by 25 rfURS procedures. Stone characteristics in both groups were comparable (p > 0.05). No statistically significant differences in clinical outcomes were observed, including stone-free rate, median operation time, or the rate of Clavien-Dindo ≥ 2 complications (p > 0.05). Surgeons found sufURS to be significantly better suited for accessing the lower calyx. Assisting nurses rated sufURS as significantly easier to assemble and dismantle. The overall cost per case was comparable between sufURS (€749) and rfURS (€791).
Conclusions
Both sufURS and rfURS perform comparably for challenging nephrolithiasis cases. sufURS does not offer a clear financial advantage over traditional rfURS.
Zusammenfassung
Einleitung
Hybridstrategien, die die Verwendung sowohl von flexiblen Einweg-Ureterorenoskopen (sufURS) als auch von wiederverwendbaren flexiblen Ureterorenoskopen (rfURS) beinhalten, werden weltweit implementiert, um die Lebensdauer von rfURS zu verlängern. Wir präsentieren einen retrospektiven Vergleich der klinischen Leistung, Ergonomie und finanziellen Aspekte zwischen sufURS und rfURS.
Material & Methoden
Herausfordernde Fälle wurden präoperativ selektiert, als Nephrolithiasis die eine voraussichtliche Laserlithotripsie mit einem steilen präoperativen infundibulopelvinen Winkel (IPA) < 50° erforderten. Alle Eingriffe wurden am UKSH Campus Lübeck zwischen Januar 2020 und März 2021 unter Verwendung von sufURS (Pusen PU3022A) und rfURS (Olympus URF-V3) durchgeführt. Postoperative Umfragen zur Leistung und Ergonomie wurden vom Operateur und der assistierenden Krankenschwester durchgeführt.
Ergebnisse
Insgesamt wurden 29 sufURS-Eingriffe von 25 rfURS-Eingriffen gefolgt. Die Steincharakteristika in beiden Gruppen waren vergleichbar (p > 0.05). Es wurden keine statistisch signifikanten Unterschiede in klinischen Ergebnissen beobachtet, einschließlich Steinfreiheitsrate, mediane Operationszeit oder Rate von Clavien-Dindo ≥2 Komplikationen (p > 0.05). Operateure fanden sufURS signifikant besser geeignet, um den unteren Kelch zu erreichen. Assistierendes Pflegepersonal bewertete sufURS als signifikant einfacher zusammenzubauen und zu demontieren. Die Gesamtkosten pro Fall waren zwischen sufURS (€749) und rfURS (€791) vergleichbar.
Schlussfolgerungen
Sowohl sufURS als auch rfURS sind bei herausfordernden Nephrolithiasis-Fällen vergleichbar. sufURS bietet keinen klaren finanziellen Vorteil gegenüber traditionellen rfURS.
Publikationsverlauf
Eingereicht: 13. April 2024
Angenommen nach Revision: 04. Juni 2025
Artikel online veröffentlicht:
06. August 2025
© 2025. Thieme. All rights reserved.
Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
-
References
- 1 Butticè S, Sener TE, Netsch C. et al. LithoVue: A new single-use digital flexible ureteroscope. Cent European J Urol 2016; 69: 302-305
- 2 Usawachintachit M, Isaacson DS, Taguchi K. et al. A Prospective Case-Control Study Comparing LithoVue, a Single-Use, Flexible Disposable Ureteroscope, with Flexible, Reusable Fiber-Optic Ureteroscopes. J Endourol 2017; 31: 468-475
- 3 Van Compernolle D, Veys R, Elshout PJ. et al. Reusable, Single-Use, or Both: A Cost Efficiency Analysis of Flexible Ureterorenoscopes After 983 Cases. J Endourol 2021; 35: 1454-1459
- 4 Hennessey DB, Fojecki GL, Papa NP. et al. Single-use disposable digital flexible ureteroscopes: an ex vivo assessment and cost analysis. BJU Int 2018; 121: 55-61
- 5 Ozimek T, Cordes J, Gilbert N. et al. Laser fibre, rather than the stone, may harm the scope: retrospective monocentric analysis of 26 pre- and intraoperative factors of flexible ureteroscope (fURS) damage. World J Urol 2020; 38: 2035-2040
- 6 Ventimiglia E, Smyth N, Doizi S. et al. Can the introduction of single-use flexible ureteroscopes increase the longevity of reusable flexible ureteroscopes at a high volume centre?. World J Urol 2022; 40: 251-256
- 7 Ozimek T, Kramer MW, Hupe MC. et al. The Impact of Endourological Experience on Flexible Ureteroscopy Outcomes and Performance at Different Levels of Expertise: Retrospective Multifactorial Analysis. Urol Int 2020; 104: 452-458
- 8 Karim SS, Hanna L, Geraghty R. et al. Role of pelvicalyceal anatomy in the outcomes of retrograde intrarenal surgery (RIRS) for lower pole stones: outcomes with a systematic review of literature. Urolithiasis 2020; 48: 263-270
- 9 Dindo D, Demartines N, Clavien PA. Classification of surgical complications: a new proposal with evaluation in a cohort of 6336 patients and results of a survey. Ann Surg 2004; 240: 205-213
- 10 Bone RC, Balk RA, Cerra FB. et al. Definitions for sepsis and organ failure and guidelines for the use of innovative therapies in sepsis. The ACCP/SCCM Consensus Conference Committee. American College of Chest Physicians/Society of Critical Care Medicine. Chest 1992; 101: 1644-1655
- 11 Ozimek T, Schneider MH, Hupe MC. et al. Retrospective Cost Analysis of a Single-Center Reusable Flexible Ureterorenoscopy Program: A Comparative Cost Simulation of Disposable fURS as an Alternative. J Endourol 2017; 31: 1226-1230
- 12 Kam J, Yuminaga Y, Beattie K. et al. Single use versus reusable digital flexible ureteroscopes: A prospective comparative study. Int J Urol 2019; 26: 999-1005
- 13 Bragaru M, Multescu R, Geavlete P. et al. Comparison of Flexible Ureteroscope Performance between Reusable and Single-Use Models. J Clin Med 2023; 12: 1093
- 14 Dragos LB, Somani BK, Sener ET. et al. Which Flexible Ureteroscopes (Digital vs. Fiber-Optic) Can Easily Reach the Difficult Lower Pole Calices and Have Better End-Tip Deflection: In Vitro Study on K-Box. A PETRA Evaluation. J Endourol 2017; 31: 630-637
- 15 Carr T, Teucher U, Mann J. et al. Waiting for surgery from the patient perspective. Psychol Res Behav Manag 2009; 2: 107-119
- 16 Ozimek T, Kramer MW, Hupe MC. et al. The Impact of Endourological Experience on Flexible Ureteroscopy Outcomes and Performance at Different Levels of Expertise: Retrospective Multifactorial Analysis. Urol Int 2020; 104: 452-458
- 17 Legemate JD, Kamphuis GM, Freund JE. et al. Durability of Flexible Ureteroscopes: A Prospective Evaluation of Longevity, the Factors that Affect it, and Damage Mechanisms. Eur Urol Focus 2019; 5: 1105-1111