Reliability of Different Protocols for Elbow Flexion and Extension Strength Measurement via Hand-held Dynamometer
Zuverlässigkeit verschiedener Protokolle zur Messung der Ellenbogenflexion und -streckkraft mittels HanddynamometerAuthors

Abstract
Objectives
To investigate the reliability of elbow flexion and extension strength measurements via a hand-held dynamometer (HHD) in different positions and to determine the minimal detectable change (MDC).
Design
Reliability study.
Methods
Twenty healthy university students were evaluated at an exercise laboratory. Elbow flexion and extension strength were each measured by two physiotherapists in three different positions. Intra-class correlation coefficients (ICCs) were calculated to assess intra-rater reliability (ICC 3,3) and inter-rater reliability (ICC 3,3) and were used to calculate MDC.
Results
All measurement procedures have excellent intra- and inter-rater reliability scores in females (0.88–0.98) and males (0.81–0.95), except for moderate intra-rater reliability (0.72) for seated measurement of elbow flexion strength with the shoulder at 90-degree flexion in males. There is a systematic difference in inter-rater results for all flexion positions and intra-rater results for one flexion position in males (p<0.05). MDC95% ranged from 3.44 to 11.69 for intra-rater scores and from 6.19 to10.79 for inter-rater scores.
Conclusions
All procedures showed excellent reliability, except for moderate intra-rater reliability in seated measurement of elbow flexion strength with the shoulder at 90-degree flexion in males. However, elbow flexion measurement was prone to systematic differences in males. Muscle strength of the tested individual and testers should be considered.
Zusammenfassung
Ziele
Untersuchung der Zuverlässigkeit von Messungen der Ellenbogenbeugungs- und -streckkraft mithilfe eines Handdynamometers (HHD) in verschiedenen Positionen und Bestimmungen der minimalen nachweisbaren Änderungen (MDC).
Design
Zuverlässigkeitsstudie.
Powered by Editorial Manager and ProduXion Manager from Aries Systems Corporation
Methode
Zwanzig gesunde Universitätsstudenten wurden im Übungslabor untersucht. Messungen der Ellenbogenbeugung und -streckung wurden von jeweils 2 Physiotherapeuten in 3 verschiedenen Positionen durchgeführt. Für die Intrarater-Reliabilität (ICC 3,3) und die Interrater-Reliabilität (ICC 3,3) wurden klasseninterne Korrelationskoeffizienten (ICCs) berechnet und zur Berechnung des MDC verwendet.
Ergebnisse
Alle Messverfahren weisen bei Frauen (0,88–0,98) und Männern (0,81–0,95) ausgezeichnete Intra- und Inter-Bewerter-Zuverlässigkeitswerte auf, mit der Ausnahme, dass die Messung der Ellenbogenbeugungskraft im Sitzen bei 90°-Flexion der Schulter eine mäßige Intra-Bewerter-Zuverlässigkeit bei Männern aufweist (0,72). Es gibt einen systematischen Unterschied zwischen den Ergebnissen der Inter-Bewerter für alle Flexionspositionen und den Ergebnissen der Intra-Bewerter für eine Flexionsposition bei Männern (p<0,05). Der 95%-MDC für Intra-Bewerter-Scores schwankte zwischen 3,44 und 11,69 und für Inter-Bewerter-Scores zwischen 6,19 und 10,79.
Schlussfolgerung
Alle Verfahren zeigten eine ausgezeichnete Zuverlässigkeit, außer dass die Messung der Ellenbogenbeugungskraft im Sitzen bei 90-Flexion der Schulter bei Männern eine mäßige Intra-Reviewer-Zuverlässigkeit aufwies. Allerdings war die Messung der Ellenbogenbeugung bei Männern anfällig für systematische Unterschiede. Die Muskelkraft der getesteten Person und der Tester sollte berücksichtigt werden.
Publikationsverlauf
Eingereicht: 22. November 2023
Angenommen nach Revision: 26. Juni 2025
Artikel online veröffentlicht:
24. September 2025
© 2025. Thieme. All rights reserved.
Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
-
References
- 1 Follmer B, Dellagrana RA, de Lima LAP. et al. Analysis of elbow muscle strength parameters in Brazilian jiu-jitsu practitioners. Journal of Sports Sciences 2016; 35 (23) 2373-2379
- 2 Folland JP, Williams AG. The adaptations to strength training: Morphological and neurological contributions to increased strength. In Sports Medicine 2007; 145-168
- 3 Wilk KE, Macrina LC, Cain EL. et al. Rehabilitation of the Overhead Athlete’s Elbow. Sports Health 2012; 405: 404-414
- 4 Pocecco E, Ruedl G, Stankovic N. et al. Injuries in judo: A systematic literature review including suggestions for prevention. In British Journal of Sports Medicine 2013; 47 (18) 1139-1143
- 5 Morse KW, Eno JJ, Altchek DW. et al. Injuries of the Biceps and Superior Labral Complex in Overhead Athletes. In Current Reviews in Musculoskeletal Medicine 2019; 12 (02) 72-79
- 6 Schrama PPM, Stenneberg MS, Lucas C. et al. Intraexaminer reliability of hand-held dynamometry in the upper extremity: A systematic review. In Archives of Physical Medicine and Rehabilitation 2014; 95 (12) 2444-2469
- 7 Hayes K, Walton JR, Szomor ZL. et al. Reliability of 3 methods for assessing shoulder strength. Journal of Shoulder and Elbow Surgery 2002; 11 (01) 33-39
- 8 Stark T, Walker B, Phillips JK. et al. Hand-held dynamometry correlation with the gold standard isokinetic dynamometry: A systematic review. In PM and R 2011; 3 (05) 472-479
- 9 Coombes BK, Bisset L, Vicenzino B. Elbow flexor and extensor muscle weakness in lateral epicondylalgia. British Journal of Sports Medicine 2012; 46 (06) 449-453
- 10 Nesterenko S, Domire ZJ, Morrey BF. et al. Elbow strength and endurance in patients with a ruptured distal biceps tendon. Journal of Shoulder and Elbow Surgery 2010; 19 (02) 184-189
- 11 Kirsch RF, Acosta AM, Perreault EJ. et al. Measurement of isometric elbow and shoulder moments: Position-dependent strength of posterior deltoid-to-triceps muscle tendon transfer in tetraplegia. IEEE Transactions on Rehabilitation Engineering 1996; 4 (04) 403-409
- 12 Abdelzaher IE, Ababneh AF, Alzyoud JM. Isometric elbow extensors strength in supine- and prone-lying positions. Physiotherapy Theory and Practice 2013; 29 (01) 61-66
- 13 Andrews AW, Thomas MW, Bohannon RW. Normative values for isometric muscle force measurements obtained with hand-held dynamometers. Physical Therapy 1996; 76 (03) 248-59
- 14 Bohannon RW. Reference values for extremity muscle strength obtained by hand-held dynamometry from adults aged 20 to 79 years. Archives of Physical Medicine and Rehabilitation 1997; 78 (01) 26-32
- 15 Bohannon RW. Shoulder position influences elbow extension force in healthy individuals. Journal of Orthopaedic and Sports Physical Therapy 1990; 12 (03) 111-114
- 16 Fieseler G, Molitor T, Irlenbusch L. et al. Intrarater reliability of goniometry and hand-held dynamometry for shoulder and elbow examinations in female team handball athletes and asymptomatic volunteers. Archives of Orthopaedic and Trauma Surgery 2015; 135 (12) 1719-1726
- 17 Vermeulen HM, de Bock GH, van Houwelingen HC. et al. A comparison of two portable dynamometers in the assessment of shoulder and elbow strength. Physiotherapy 2005; 91 (02) 101-112
- 18 Richardson J, Stratford P, Cripps D. Assessment of reliability of the hand-held dynamometer for measuring strength in healthy older adults. Physiotherapy Theory and Practice 1998; 14 (01) 49-54
- 19 Romero-Franco N, Fernández-Domínguez JC, Montaño-Munuera JA. et al. Validity and reliability of a low-cost dynamometer to assess maximal isometric strength of upper limb: Low cost dynamometry and isometric strength of upper limb. Journal of Sports Sciences 2019; 37 (15) 1787-1793
- 20 Riemann BL, Lininger MR. Statistical Primer for Athletic Trainers: The Essentials of Understanding Measures of Reliability and Minimal Important Change. Journal of Athletic Training 2018; 53 (01) 98-103
- 21 Daloia LMT, Leonardi-Figueiredo MM, Martinez EZ. et al. Isometric muscle strength in children and adolescents using Hand held dynamometry: reliability and normative data for the Brazilian population. Brazilian Journal of Physical Therapy 2018; 22 (06) 474-483
- 22 Walter S, Eliasziw M, Donner A. Sample size and optimal designs for reliability studies. Statistics in Medicine 1998; 17 (01) 101-110
- 23 Kendall FP, Kendall E. Geise Provance. Muscles, testing and function. 3rd ed. The Netherlands: Houten; 2000
- 24 West AM, Scarborough DM, McInnis KC. et al. Strength and Motion in the Shoulder, Elbow, and Hip in Softball Windmill Pitchers. PM & R: the journal of injury, function, and rehabilitation 2019; 11 (12) 1302-1311
- 25 Van Aswegen H, Eales C, Richards G. Establishing the reliability of test procedures in order to investigate the effect of penetrating trunk trauma and prolonged mechanical ventilation on the recovery of adult survivors. JSAJoP 2007; 63 (01) 22-27
- 26 Fleiss JL. The Design and Analysis of clinical experiments. Hoboken; NJ: Wiley & Sons;: 2003
- 27 Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. The Journal of Strength & Conditioning Research 2005; 19 (01) 231-240
- 28 Lima-Dos-Santos AL, Domingos-Gomes JR, Dantas Andrade OS. et al. Health-related physical fitness of military police officers in Paraiba, Brazil. Rev Bras Med Trab 2018; 16: 429-435
- 29 Kilmer DD, McCrory MA, Wright NC. et al. Hand-held dynamometry reliability in persons with neuropathic weakness. Arch Phys Med Rehabil 1997; 78: 1364-1368
- 30 Dowman L, McDonald CF, Hill CJ. et al. Reliability of the hand held dynamometer in measuring muscle strength in people with interstitial lung disease. Physiotherapy 2016; 102: 249-255
- 31 Kotte SHP, Viveen J, Koenraadt KLM. et al. Normative values of isometric elbow strength in healthy adults: a systematic review. Shoulder Elbow 2018; 10: 207-215
- 32 Wikholm JB, Bohannon RW. Hand-held Dynamometer Measurements: Tester Strength Makes a Difference. J Orthop Sports Phys Ther 1991; 13: 191-198
- 33 Stone CA, Nolan B, Lawlor PG. et al. Hand-held dynamometry: tester strength is paramount, even in frail populations. J Rehabil Med 2011; 43: 808-811
- 34 Scott DA, Bond EQ, Sisto SA. et al. The intra- and interrater reliability of hip muscle strength assessments using a handheld versus a portable dynamometer anchoring station. Arch Phys Med Rehabil 2004; 85: 598-603
- 35 Wagner H, Buchecker M, von Duvillard SP. et al. Kinematic comparison of team handball throwing with two different arm positions. Int J Sports Physiol Perform 2010; 5: 469-483
- 36 Günzkofer F, Engstler F, Bubb H. et al. Isometric elbow flexion and extension joint torque measurements considering biomechanical aspects. In: First International Symposium on Digital Human Modeling, 14–15. 2011
- 37 Bohannon RW. Manual muscle testing: does it meet the standards of an adequate screening test?. Clinical Rehabilitation 2005; 19 (06) 662-667
- 38 Dodds RM. et al. Grip strength across the life course: normative data from twelve British studies. PLOS ONE 2014; 9 (12) e113637
- 39 Taylor RL, Casey JJ. Quadriceps Torque Production on the Cybex II Dynamometer as Related to Changes in Lever Arm Length. Journal of Orthopaedic & Sports Physical Therapy 1986; 8 (03) 147-152