Semin Thromb Hemost 1995; 21(S 02): 20-24
DOI: 10.1055/s-0032-1313598
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Hematological Parameters Influencing the Thrombostat 4000

Dietmar Söhngen
,
Eckhard Hatistein
,
Axel Heyll
,
Gerald Meckenstock
,
Simone Wienen
,
Wolfgang Schneider
Weitere Informationen

Publikationsverlauf

Publikationsdatum:
09. Mai 2012 (online)

Preview

Abstract

The effects of platelet counts, hematocrit, and leukocyte counts were studied on the closure times of the Thrombostat 4000 (in-vitro bleeding time, IVBT). Closure times became longer with platelet counts <50 × 109/L; an inverse linear correlation could be established. Hematocrit was also inversely correlated with the closure time. At constant platelet counts a hematocrit of 55% yielded an immediate closure of the filter, while with a hematocrit <15% no closure times could be measured. At constant platelet counts and hematocrits, nomonuclear and polymorphonuclear leukocytes also influenced closure times; increased counts resulted in shorter closure times. Leukocytes from a patient with chronic myelocytic leukemia had the same effects.