Thromb Haemost 1978; 39(02): 466-473
DOI: 10.1055/s-0038-1646706
Original Article
Schattauer GmbH Stuttgart

Quantitation of Radio-Labelled Vitamin K1 and Vitamin K1 Epoxide in Plasma by High Pressure Liquid Chromatography

Authors

  • Thorir D Bjornsson

    The Divisions of Clinical Pharmacology and Cardiology, Department of Medicine, Stanford University Medical Center, Stanford, California, USA
  • Sarah E Swezey

    The Divisions of Clinical Pharmacology and Cardiology, Department of Medicine, Stanford University Medical Center, Stanford, California, USA
  • Peter J Meffin

    The Divisions of Clinical Pharmacology and Cardiology, Department of Medicine, Stanford University Medical Center, Stanford, California, USA
  • Terrence F Blaschke

    The Divisions of Clinical Pharmacology and Cardiology, Department of Medicine, Stanford University Medical Center, Stanford, California, USA
Further Information

Publication History

Received 21 July 1977

Accepted 20 October 1977

Publication Date:
12 July 2018 (online)

Preview

Summary

A convenient, accurate and reproducible high pressure liquid chromatographic method for the quantitation of radio-labelled vitamin K1 and vitamin K1 epoxide in plasma is described. The method involves the determination of total ether extractable radioactivity, and a chromatographic separation to determine the relative quantities of radio-labelled vitamin K1 and vitamin K1 epoxide. The method is useful over a wide range of ratios of the two compounds, and has a coefficient of variation of approximately 5%.