Application of Dual-Source-Computed Tomography in Pediatric Cardiology in Children Within the First Year of Life

Einsatz der Dual-Source-Computertomografie in der Kinderkardiologie bei Kindern im ersten Lebensjahr
P. Hausmann
1   Pediatric Cardiology, Friedrich-Alexander-University Erlangen-Nürnberg (FAU), Germany
,
A. Stenger
1   Pediatric Cardiology, Friedrich-Alexander-University Erlangen-Nürnberg (FAU), Germany
,
S. Dittrich
1   Pediatric Cardiology, Friedrich-Alexander-University Erlangen-Nürnberg (FAU), Germany
,
R. Cesnjevar
2   Congenital Heart Surgery, Friedrich-Alexander-University Erlangen-Nürnberg (FAU), Germany
,
A. Rüffer
2   Congenital Heart Surgery, Friedrich-Alexander-University Erlangen-Nürnberg (FAU), Germany
,
M. Hammon
3   Radiology, Friedrich-Alexander-University Erlangen-Nürnberg (FAU), Germany
,
M. Uder
3   Radiology, Friedrich-Alexander-University Erlangen-Nürnberg (FAU), Germany
,
O. Rompel
3   Radiology, Friedrich-Alexander-University Erlangen-Nürnberg (FAU), Germany
,
M. Glöckler
1   Pediatric Cardiology, Friedrich-Alexander-University Erlangen-Nürnberg (FAU), Germany
› Author Affiliations
Further Information

Publication History

16 May 2015

13 October 2015

Publication Date:
27 January 2016 (online)

Abstract

Purpose: To assess fields of application and value of dual source computed tomography (DSCT) for diagnostics and therapy in patients with congenital heart disease during their first year of life. Evaluation of image quality, surgical use and radiation exposure of 2nd and 3 rd generation DSCT.

Materials and methods: DSCT was applied in 118 cases between January 2012 and October 2014 for diagnostics of congenital heart defects. 2nd generation was used in 91 cases until April 2014 and 3 rd generation in 27 cases during the period thereafter. 3 D reconstructions of the image data were created for clinical diagnostics and planning of interventions. Image quality was assessed using a 4-point-scale. The visibility of the mammary arteries was analyzed, and signal-to-noise-ratio (SNR) and contrast-to-noise-ratio (CNR) were calculated. The usefulness of 3D-reconstructions for surgical planning was rated using a 5-point-scale. Radiation exposure and contrast dye consumption were determined. All cases were analyzed retrospectively.

Results: DSCT was successfully used in 118 cases. All image data obtained were interpretable. More than 60 percent of cases did not show any artifacts. The mammary arteries were visible down to the diaphragmatic arch in more than 80 percent of cases. Diagnostic value and surgical benefit were evaluated as “useful” or as “essential” in all cases. Median radiation dose was 0.4 mSv and 0.27 mSv for the 2nd and 3 rd generation DSCT, respectively. Consumption of contrast dye was 2 ml/kg in all cases.

Conclusion: DSCT is a modern and extremely helpful technique for diagnostics and planning of interventions in patients with complex congenital heart defects. Extracardiac vascular structures in particular can be depicted three-dimensionally at high resolution. The use of iterative reconstruction with 3 rd generation DSCT yielded image quality similar to that of 2nd generation DSCT at considerably reduced radiation exposure level compared to 2nd generation DSCT. 3 rd generation DSCT is a low risk, accurate and extremely fast technique for diagnosing unstable patients with CHD.

Key points:

• Expanded scope of indications for DSCT in diagnosing critically ill infants

• Effective radiation dose is considerably lower than 0.5 mSv

• Extremely rapid image acquisitions with high image quality

• Possibility of optimized 3D-based surgical planning

Citation Format:

• Hausmann P, Stenger A, Dittrich S et al. Application of Dual-Source-Computed Tomography in Pediatric Cardiology in Children Within the First Year of Life. Fortschr Röntgenstr 2016; 188: 179 – 187

Zusammenfassung

Ziel: Evaluation von Einsatzmöglichkeiten und Nutzen der Dual-Source-Computertomografie (DSCT) bei Säuglingen mit angeborenem Herzfehler (AHF). Bewertung von Bildqualität, chirurgischen Stellenwert und Strahlenbelastung der 2. und 3. Generation DSCT.

Material und Methoden: Die DSCT wurde zwischen 2012 und 2014 in 118 Fällen zur Diagnostik AHF angewendet. Bis April 2014 wurde die 2. Generation DSCT in 91 Fällen und die 3. Generation im darauffolgenden Zeitraum in 27 Fällen durchgeführt. Die gewonnenen Bilddaten wurden zur Diagnostik und Eingriffsplanung dreidimensional (3 D) aufbereitet. Zur Evaluation der Bildqualität wurde eine 4-Punkte-Skala verwendet. Die Visualisierung der A. mammaria, das Signal-Rausch-Verhältnis (SNR) sowie das Kontrast-Rausch-Verhältnis (CNR) wurden ausgewertet. Der chirurgische Nutzen der DSCT wurde anhand einer 5-Punkte-Skala beurteilt. Die Strahlendosis und der Kontrastmittelverbrauch wurden bestimmt.

Ergebnisse: Alle gewonnenen Bilddaten waren interpretierbar. In über 60 % der Fälle waren die Bilddaten artefaktfrei. Die A. mammaria war in über 80 % der Fälle mindestens bis zur Zwerchfellkuppel darstellbar. Die diagnostische Wertigkeit und der chirurgische Nutzen wurden immer als „nützlich“ oder „essenziell“ bezeichnet. Die Strahlenbelastung betrug bei der 2. Generation im Median 0,4 mSv und bei der 3. Generation 0,27 mSv. Der Kontrastmittelverbrauch war stets 2 ml/kg.

Schlussfolgerung: Die DSCT ist ein modernes, überaus hilfreiches Verfahren zur Diagnostik und Eingriffsplanung bei Patienten mit komplexen AHF. Insbesondere extrakardiale, vaskuläre Strukturen lassen sich in hoher Auflösung 3 D darstellen. Durch die iterative Rekonstruktion der Daten in der 3. Generation DSCT kann bei erheblich geringerer Strahlenbelastung im Vergleich zur 2. Generation DSCT eine gleich gute Bildqualität erzielt werden. Die 3. Generation DSCT ist damit ein schonendes, exaktes und extrem schnelles Verfahren zur Diagnostik instabiler kardiologischer Patienten.

Deutscher Artikel/German Article

 
  • References

  • 1 Lindinger A, Schwedler G, Hense HW. Prevalence of Congenital Heart Defects in Newborns in Germany: Results of the First Registration Year of the PAN Study (July 2006 to June 2007). Klin Padiatr 2010; 222: 321-326
  • 2 Silva KP, Rocha LA, Leslie AT et al. Newborns with congenital heart diseases: epidemiological data from a single reference center in Brazil. J Prenat Med 2014; 8: 11-16
  • 3 Schultz AH, Localio AR, Clark BJ et al. Epidemiologic features of the presentation of critical congenital heart disease: implications for screening. Pediatrics 2008; 121: 751-757
  • 4 Tumanyan MR, Filaretova OV, Chechneva VV et al. Repair of Complete Atrioventricular Septal Defect in Infants with Down Syndrome: Outcomes and Long-Term Results. Pediatr Cardiol 2014; DOI: 10.1007/s00246-014-0966-7.
  • 5 Padalino MA, Cavalli G, De Franceschi M et al. Surgical outcomes of total anomalous pulmonary venous connection repair: a 22-year experience. J Card Surg 2014; 29: 678-685
  • 6 Nakayama Y, Hiramatsu T, Iwata Y et al. Surgical results for functional univentricular heart with total anomalous pulmonary venous connection over a 25-year experience. Ann Thorac Surg 2012; 93: 606-613
  • 7 Glöckler M, Halbfaß J, Koch A et al. Preoperative assessment of the aortic arch in children younger than 1 year with congenital heart disease: utility of low-dose high-pitch dual-source computed tomography. A single-centre, retrospective analysis of 62 cases. Eur J Cardiothorac Surg 2014; 45: 1060-1065
  • 8 Bharucha T, Mertens L. Recent advances in pediatric echocardiography. Expert review of cardiovascular therapy 2013; 11: 31-47
  • 9 Kuettner A, Gehann B, Spolnik J et al. Strategies for dose-optimized imaging in pediatric cardiac dual source CT. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin 2009; 181: 339-348
  • 10 Lee T, Tsai IC, Fu YC et al. Using multidetector-row CT in neonates with complex congenital heart disease to replace diagnostic cardiac catheterization for anatomical investigation: initial experiences in technical and clinical feasibility. Pediatr Radiol 2006; 36: 1273-1282
  • 11 Achenbach S, Barkhausen J, Beer M et al. Consensus recommendations of the German Radiology Society (DRG), the German Cardiac Society (DGK) and the German Society for Pediatric Cardiology (DGPK) on the use of cardiac imaging with computed tomography and magnetic resonance imaging. Fortschr Röntgenstr 2012; 184: 345-368
  • 12 Achenbach S, Barkhausen J, Beer M et al. Consensus recommendations of the German Radiology Society (DRG), the German Cardiac Society (DGK) and the German Society for Pediatric Cardiology (DGPK) on the use of cardiac imaging with computed tomography and magnetic resonance imaging. Fortschr Röntgenstr 2012; 184: 345-368
  • 13 Khatri S, Varma SK, Khatri P et al. 64-slice multidetector-row computed tomographic angiography for evaluating congenital heart disease. Pediatr Cardiol 2008; 29: 755-762
  • 14 Ou P, Celermajer DS, Calcagni G et al. Three-dimensional CT scanning: a new diagnostic modality in congenital heart disease. Heart 2007; 93: 908-913
  • 15 Taylor AM, Dymarkowski S, Hamaekers P et al. MR coronary angiography and late-enhancement myocardial MR in children who underwent arterial switch surgery for transposition of great arteries. Radiology 2005; 234: 542-547
  • 16 Glockler M, Halbfass J, Koch A et al. Preoperative assessment of the aortic arch in children younger than 1 year with congenital heart disease: utility of low-dose high-pitch dual-source computed tomography. A single-centre, retrospective analysis of 62 cases. European journal of cardio-thoracic surgery 2014; 45: 1060-1065
  • 17 Ben Saad M, Rohnean A, Sigal-Cinqualbre A et al. Evaluation of image quality and radiation dose of thoracic and coronary dual-source CT in 110 infants with congenital heart disease. Pediatr Radiol 2009; 39: 668-676
  • 18 Bongartz GGS, Jurik AG, Leonardi M et al. European Guidelines on Quality Criteria for Computed Tomography. 2015; Accessed february 18, 2015
  • 19 Ellis AR, Mulvihill D, Bradley SM et al. Utility of computed tomographic angiography in the pre-operative planning for initial and repeat congenital cardiovascular surgery. Cardiology in the young 2010; 20: 262-268
  • 20 Deak PD, Smal Y, Kalender WA. Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 2010; 257: 158-166
  • 21 Brady SL, Kaufman RA. Investigation of American Association of Physicists in Medicine Report 204 size-specific dose estimates for pediatric CT implementation. Radiology 2012; 265: 832-840
  • 22 Shinkawa T, Anagnostopoulos PV, Johnson NC et al. Early results of the “clamp and sew” Fontan procedure without the use of circulatory support. Ann Thorac Surg 2011; 91: 1453-1459
  • 23 Navabi MA, Rastegar SM, Kiani A et al. Avoiding cardiopulmonary bypass in extracardiac cavopulmonary connection: does it really matter?. J Thorac Cardiovasc Surg 2010; 139: 1183-1188
  • 24 Chen X, Qu Y, Peng ZY et al. Clinical value of multi-slice spiral computed tomography angiography and three-dimensional reconstruction in the diagnosis of double aortic arch. Experimental and therapeutic medicine 2014; 8: 623-627
  • 25 Mertens L, Seri I, Marek J et al. Targeted neonatal echocardiography in the neonatal intensive care unit: practice guidelines and recommendations for training. Eur J Echocardiogr 2011; 12: 715-736
  • 26 Flohr TG, McCollough CH, Bruder H et al. First performance evaluation of a dual-source CT (DSCT) system. European radiology 2006; 16: 256-268
  • 27 Brenner DJ, Hall EJ. Computed tomography – an increasing source of radiation exposure. N Engl J Med 2007; 357: 2277-2284
  • 28 Yoon H, Kim MJ, Yoon CS et al. Radiation dose and image quality in pediatric chest CT: effects of iterative reconstruction in normal weight and overweight children. Pediatr Radiol 2014; DOI: 10.1007/s00247-014-3176-9.
  • 29 Singh S, Kalra MK, Shenoy-Bhangle AS et al. Radiation dose reduction with hybrid iterative reconstruction for pediatric CT. Radiology 2012; 263: 537-546
  • 30 Hou Y, Xu S, Guo W et al. The optimal dose reduction level using iterative reconstruction with prospective ECG-triggered coronary CTA using 256-slice MDCT. European journal of radiology 2012; 81: 3905-3911