Laser-Sintered versus Conventional Cobalt-Chromium Bars in Implant-Supported Complete Mandibular Overdentures: An In Vitro Strain Analysis Comparative Study

Abstract
Objective The aim of this study was to compare the effect of laser-sintered and conventional cobalt-chromium bars on strain distribution around implants in mandibular implant-supported overdentures.
Materials and Methods A three-dimensional (3D) epoxy resin model for completely edentulous mandible was used for this study. Two implants were placed in the canine region, and the 3D model was scanned by a laboratory scanner. Twelve cobalt-chromium bars were divided into two groups. Group 1 includes six bars that were designed by software and created using the laser sintering method. Group 2 includes six bars that were made using computer-aided design and computer-aided manufacturing to make a wax model, fabricated by a conventional lost wax-casting technique. Each implant was surrounded by four channels in the epoxy resin model, and four strain gauges were cemented. A universal testing machine was used to apply the load, and a strain measuring analysis was performed.
Results Null hypothesis was rejected as there was significant difference in strain distribution around implants between both groups for the microstrains around the implants in bar-retained implant-supported prostheses, in both bilateral and unilateral results.
Conclusion The laser sintering method showed better results in comparison with conventional methods in fabrication of bar-retained implant-supported prostheses due to less trauma to the implant and supporting structures.
Keywords
cobalt-chromium bars - implant supported overdenture - direct 3D printing - strain gauge analysisPublication History
Article published online:
27 January 2025
© 2025. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/)
Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India
-
References
- 1 Al-Rafee MA. The epidemiology of edentulism and the associated factors: a literature Review. J Family Med Prim Care 2020; 9 (04) 1841-1843
- 2 Martins AMC, Guimarães LS, Campos CH. et al. The effect of complete dentures on edentulous patients' oral health-related quality of life in long-term: a systematic review and meta-analysis. Dent Res J (Isfahan) 2021; 18 (01) 65
- 3 Gupta A, Felton DA, Jemt T, Koka S. Rehabilitation of edentulism and mortality: a systematic review. J Prosthodont 2019; 28 (05) 526-535
- 4 Kutkut A, Bertoli E, Frazer R, Pinto-Sinai G, Fuentealba Hidalgo R, Studts J. A systematic review of studies comparing conventional complete denture and implant retained overdenture. J Prosthodont Res 2018; 62 (01) 1-9
- 5 Leão RS, Moraes SLD, Vasconcelos BCE, Lemos CAA, Pellizzer EP. Splinted and unsplinted overdenture attachment systems: a systematic review and meta-analysis. J Oral Rehabil 2018; 45 (08) 647-656
- 6 Saceleanu A, Florido-Suarez N, Jimenez-Marco C, Mirza-Rosca J. Relation between composition, structure and properties of different dental alloys. Microsc Microanal 2022; 28 (01) 684-688
- 7 Presotto AGC, Cordeiro JM, Presotto JGC. et al. Feasibility of 3D printed Co–Cr alloy for dental prostheses applications. J Alloys Compd 2021; 862 (01) 171-181
- 8 Okano H, Tasaka A, Matsunaga S. et al. Effects of hollow structures added by selective laser sintering on the mechanical properties of Co-Cr alloy. J Prosthodont Res 2023; 67 (03) 460-467
- 9 Goguta L, Lungeanu D, Negru R, Birdeanu M, Jivanescu A, Sinescu C. Selective laser sintering versus selective laser melting and computer aided design–computer aided manufacturing in double crowns retention. J Prosthodont Res 2021; 65 (03) 371-378
- 10 Matta RE, Riegger K, Trägler H. et al. Establishment of a new biomechanical measurement method for surface deformation of bone by force application via dental implants—a pilot study. Appl Sci (Basel) 2021; 11 (16) 1-13
- 11 Tribst JPM, Dal Piva AM, Borges AL. Biomechanical tools to study dental implants: a literature review. Braz Den J 2016; 19 (04) 5-11
- 12 ELsyad MA, Dayekh MA, Khalifa AK. Locator versus bar attachment effect on the retention and stability of implant-retained maxillary overdenture: an in vitro study. J Prosthodont 2019; 28 (02) 627-636
- 13 Gordon JC. Misch's Contemporary Implant Dentistry. Implant Dent 2019; 28 (06) 526-527
- 14 Ghodsi S, Pirmoazen S, Beyabanaki E, Rostami M, Alikhasi M. The effect of milling metal versus milling wax on implant framework retention and adaptation. J Prosthodont 2019; 28 (02) e739-e743
- 15 Aalaei S, Abedi P, Niknami S, Taghavi F, Damghani FT. The effect of attachment types and implant level on the stress distribution in a mandibular overdenture: a 3D Finite Element Analysis. Braz Dent Sci 2022; 25 (03) 1-8
- 16 Dikova T. Properties of Co-Cr dental alloys fabricated using additive technologies. In: Leszek A. ed. Biomaterials in Regenerative Medicine. London, UK: IntechOpen; 2018: 141-159
- 17 Lowe LG, Shcherbukhin VM. An implant-supported, cobalt-chromium milled bar and nonflanged attachment-retained overdenture to rehabilitate the edentulous mandible. J Prosthet Dent 2009; 102 (01) 46-51
- 18 Abdelhafiz SH, Abdelkader SH, Azer A. Marginal and internal fit evaluation of metal copings fabricated by selective laser sintering and CAD/CAM milling techniques: in-vitro study. Alexandria Dent J 2022; 47 (02) 134-139
- 19 Ahmed N, Abbasi MS, Haider S. et al. Fit accuracy of removable partial denture frameworks fabricated with cad/cam, rapid prototyping, and conventional techniques: a systematic review. BioMed Res Int 2021; 2021 (01) 3194433
- 20 Almufleh B, Emami E, Alageel O. et al. Patient satisfaction with laser-sintered removable partial dentures: a crossover pilot clinical trial. J Prosthet Dent 2018; 119 (04) 560-567.e1
- 21 Abd El-Rahman N, El-Sharkawy A, Mohy El-Din M, Saad M. Strain evaluation of peek cantilever bars around implants assisted mandibular overdentures (in vitro study). Alex Dent J 2022; 47 (03) 138-145
- 22 Assunção WG, Barão VA, Tabata LF, Gomes ÉA, Delben JA, dos Santos PH. Biomechanics studies in dentistry: bioengineering applied in oral implantology. J Craniofac Surg 2009; 20 (04) 1173-1177
- 23 El-Abd MF, El-Sheikh MM, El-Gendy MN. Effect of position of single implant with two different attachments on stress distribution of mandibular complete overdenture (in-vitro study). Tanta Dental Journal 2018; 15: 63-69
- 24 Rady N, Zaki M, Abd Elhamid S. Influence of different bar attachment materials on stress distribution around two inter foraminal dental implants: an in vitro study. Egypt Dent J 2023; 69 (01) 457-464
- 25 Othman D, Kashef N, El Gendy MM. In vitro stress analysis of class II Kennedy conventional design and side plate with OT attachment. Tanta Dent J 2023; 20 (03) 161-168
- 26 Khurana N, Rodrigues S, Shenoy S. et al. a comparative evaluation of stress distribution with two attachment systems of varying heights in a mandibular implant-supported overdenture: a three-dimensional finite element analysis. J Prosthodont 2019; 28 (02) e795-e805
- 27 Kortam S. Peri-implant strains with palate less locator retained maxillary implant overdentures reinforced with either metal or poly ether-ether ketone (peek) materials. an invitro study. Egypt Dent J 2020; 66 (02) 1225-1234
- 28 Atia GA, Shalaby HK, Roomi AB. et al. Macro, micro, and nano-inspired bioactive polymeric biomaterials in therapeutic, and regenerative orofacial applications. Drug Des Devel Ther 2023; 17: 2985-3021
- 29 Zhou Y, Li N, Yan J, Zeng Q. Comparative analysis of the microstructures and mechanical properties of Co-Cr dental alloys fabricated by different methods. J Prosthet Dent 2018; 120 (04) 617-623
- 30 Udomsawat C, Rungsiyakull P, Rungsiyakull C, Khongkhunthian P. Comparative study of stress characteristics in surrounding bone during insertion of dental implants of three different thread designs: a three-dimensional dynamic finite element study. Clin Exp Dent Res 2018; 5 (01) 26-37
- 31 Yoo JS, Kwon K-R, Noh K, Lee H, Paek J. Stress analysis of mandibular implant overdenture with locator and bar/clip attachment: comparative study with differences in the denture base length. J Adv Prosthodont 2017; 9 (03) 143-151
- 32 Khorshid H. A comparison between two different types of attachments in implant retained mandibular overdentures (a stress analysis study). Egypt Dent J 2018; 64 (02) 1893-1902
- 33 Helaly OA, Mohamed AQ, El Quoriaty HA. Retention and stress distribution of implant retained mandibular complete overdenture with locator versus ball and socket attachments (an in vitro study). Al-Azhar J Dent Sci 2023; 26 (01) 1-7
- 34 Elsyad MA, Elhddad AA, Khirallah AS. The effect of implant diameter on strain around implants retaining a mandibular overdenture with Locator attachments: an in vitro study. Dent Mater J 2016; 35 (06) 938-945
- 35 Revilla-León M, Sadeghpour M, Özcan M. A review of the applications of additive manufacturing technologies used to fabricate metals in implant dentistry. J Prosthodont 2020; 29 (07) 579-593
- 36 Kassapidou M, Stenport VF, Johansson CB. et al. Cobalt chromium alloys in fixed prosthodontics: investigations of mechanical properties and microstructure. J Prosthet Dent 2023; 130 (02) 255.e1-255.e10
- 37 Alageel O, Abdallah MN, Alsheghri A, Song J, Caron E, Tamimi F. Removable partial denture alloys processed by laser-sintering technique. J Biomed Mater Res B Appl Biomater 2018; 106 (03) 1174-1185
- 38 Hong M-H, Lee D-H, Hanawa T, Kwon T-Y. Comparison of microstructures and mechanical properties of 3 cobalt-chromium alloys fabricated with soft metal milling technology. J Prosthet Dent 2022; 127 (03) 489-496