Comparison of Ultrasound Attenuation Imaging Using a Linear versus a Conventional Convex Probe: A Volunteer Study

Authors

  • O Hänni

    1   University Hospital Zurich, Institute for Anesthesiology and Perioperative Medicine, Zurich, Schweiz
  • L Ruby

    2   University Hospital Zurich, Diagnostic and Interventional Radiology, Zurich, Schweiz
  • C Paverd

    2   University Hospital Zurich, Diagnostic and Interventional Radiology, Zurich, Schweiz
  • F A Huber

    2   University Hospital Zurich, Diagnostic and Interventional Radiology, Zurich, Schweiz
  • T Frauenfelder

    2   University Hospital Zurich, Diagnostic and Interventional Radiology, Zurich, Schweiz
  • A Martin

    2   University Hospital Zurich, Diagnostic and Interventional Radiology, Zurich, Schweiz
 

Background Ultrasound attenuation imaging (ATI) is a valuable tool for assessing liver tissue and diagnosing hepatic steatosis, a condition affecting 20-30% of the Western population. This study compares ATI measurements using a linear probe versus a conventional convex probe in healthy volunteers to evaluate the feasibility and reliability of the linear probe for liver assessment.

Methods Thirty-three healthy volunteers (19 females, 14 males, aged 24.5-63.3 years) were recruited. Ultrasound measurements were performed using a TUS-AI800 scanner with both convex and linear probes. Measurements were taken at 1 cm below the liver capsule, with repeated sessions 4-5 weeks apart to assess repeatability. Statistical analysis included unpaired t-tests and calculations of intraobserver variability.

Results The ATI values for the linear probe ranged from 0.49 to 0.85 dB/cm/MHz, which is comparable to the convex probe's range of 0.50 to 0.89 dB/cm/MHz. The depth had a significant impact on the measurements; for the convex probe, attenuation values decreased with greater depths, while for the linear probe, they increased. The size of the region of interest (ROI) affected the measurements obtained with the convex probe but did not have an effect on those from the linear probe. Intraobserver variability was found to be 6.1% for the convex probe and 5.0% for the linear probe.

Conclusion ATI measurements using a linear probe are feasible and yield values comparable to those obtained with a convex probe, although depth and ROI size significantly influence the results. The linear probe shows promise for superficial liver assessments and other superficial tissues. Further research is needed to establish protocols and validate findings in patients with liver disease.