Subscribe to RSS
DOI: 10.1055/s-2000-8164
A Fibreoptic Scoring System to Assess the Position of Laryngeal Mask Airway Devices. Interobserver Variability and a Comparison Between the Standard, Flexible and Intubating Laryngeal Mask Airways
Publication History
Publication Date:
31 December 2000 (online)
Summary.
We determined the interobserver reliability of a fibreoptic scoring system for assessing the position of the laryngeal mask airway (LMA), the flexible laryngeal mask airway (FLMA) and the intubating laryngeal mask airway (ILM). We also compare fibreoptic position between the devices. Thirty anaesthetised adult patients were studied in random order in a triple crossover manner. Two observers blinded to each others findings scored the fibreoptic position as follows: 4, only vocal cords visible; 3, vocal cords plus posterior epiglottis visible; 2, vocal cords plus anterior epiglottis visible; 1, vocal cords not seen. Interobserver reliability was examined using intraclass correlation coefficient (ICC). Interobserver reliability was graded as excellent for the LMA (ICC = 0.89), FLMA (ICC = 0.87) and ILM (ICC = 0.79). Fibreoptic scores were higher for the LMA and FLMA compared with the ILM (both p<0.001). We conclude that interobserver reliability for the fibreoptic scoring system is excellent for the LMA, FLMA and ILM. Fibreoptic position is better for the LMA and FLMA compared with the ILM. Fibreoptic scoring has potential utility for research and clinical practice with laryngeal mask devices.
Ein fiberoptisches Score-System zur Beurteilung der Position von Larynxmasken. Interindividuelle Untersuchervariabilität und ein Vergleich zwischen Standard-, Flexibler und Intubationslarynxmaske.
Wir bestimmten die interindividuelle Verlässlichkeit eines fiberoptischen Score Systems für die Beurteilung von Standard-Larynxmasken (LMA), flexiblen Larynxmasken (FLMA) und Intubationslarynxmasken (ILM). Wir vergleichen auch die fiberoptische Positionierung zwischen den verschiedenen Masken. Dreißig anästhesierte erwachsene Patienten wurden nach dem Zufallsprinzip in einem triple cross-over untersucht. Zwei Untersucher, die in ihren Ergebnissen gegeneinander blind waren, bewerteten die fiberoptische Position wie folgt: 4, nur Stimmbänder sichtbar; 3, Stimmbänder plus hintere Epiglottis sichtbar; 2, Stimmbänder plus vordere Epiglottis sichtbar; 1, Stimmbänder nicht sichtbar. Die interindividuelle Verlässlichkeit wurde nach dem Intraclass Korrelationskoeffizient (ICC) beurteilt. Die interindividuelle Verlässlichkeit wurde bei der LMA als hervorragend eingestuft (ICC = 0,89). FLMA (ICC = 0,87) und ILM (ICC = 0,79). Die fiberoptischen Scores waren bei der LMA und FLMA höher als bei der ILM (beide p < 0,001). Wir schließen aus diesen Ergebnissen, dass die interindividuelle Verlässlichkeit für das fiberoptische Score-System für LMA, FLMA und ILM als hervorragend eingestuft werden kann. Die fiberoptische Position ist bei LMA und FLMA besser als bei der ILM. Das fiberoptische Score-System ist von potenziellem Nutzen für Untersuchungen und den klinischen Alltag mit Larynxmasken.
Key words:
Laryngeal mask airway devices - Interobserver variability
Schlüsselwörter:
Larynxmasken - Interindividuelle Variabilität
References
- 1 Brimacombe J, Brain A IJ, Berry A. The laryngeal mask airway: review and practical guide. London; WB Saunders Company Limited 1997
- 2 Berry A M, Brimacombe J, McManus K F, Goldblatt M. An evaluation of the factors influencing selection of the optimal size of laryngeal mask airway in normal adults. Anaesthesia. 1998; 53 565-570
- 3 Joshi S, Sciacca R R, Solanki D R, Young W L, Mathru M M. A prospective evaluation of clinical tests for placement of laryngeal mask airways. Anesthesiology. 1998; 89 1141-1146
- 4 Brimacombe J, Keller C. Performance of the Size 5 Laryngeal Mask Airway in Males and Females. Anaesthesiologie Intensivmedizin Notfallmedizin Schmerztherapie (In Press)
- 5 Yanagimoto M. [Effect of position of the epiglottis on volume-pressure curve during use of the LM]. Journal of Japanese Society for Clinical Anesthesia (Nihon rinsho-masui Gakkai Shi). 1992; 12 738-741
- 6 Kapila A, Addy E V, Verghese C, Brain A IJ. The intubating laryngeal mask airway: an initial assessment of performance. Br J Anaesth. 1997; 79 710-713
- 7 Brimacombe J, Berry A. The laryngeal mask airway - anatomical and physiological implications. Acta Anaesthesiol Scand. 1996; 4 201-209
- 8 Brimacombe J, Berry A. A proposed fiber-optic scoring system to standardize the assessment of laryngeal mask airway position. Anesth Analg. 1993; 76 457
- 9 Brimacombe J. Laryngoscopy through the LMA - a useful skill to acquire. Anaesth Intens Care. 1992; 2 535
- 10 Brimacombe J R, Brain A IJ, Berry A M. The laryngeal mask instruction manual for anaesthesia. Henley-on-Thamas; Intavent Research Ltd. 1999
- 11 Brain A IJ. LMA - FastrachTM Instruction Manual. The Laryngeal Mask Company Ltd 1997
-
12 Rosner B.
Hypothesis testing: categorical data. In Fundamentals of Biostatistics. Belmont; Wadsworth Publishing Company 1995: 423-426 -
13 Fleiss J L.
Reliability of measurements. In The Design and Analysis of Clinical Experiments. New York; John Wiley and Sons 1986: 2-32 - 14 Keller C, Brimacombe J. The influence of head and neck position on oropharyngeal leak pressure and cuff position with the flexible and the standard laryngeal mask airway. Anesth Analg. 1999; 88 913-916
- 15 Keller C, Brimacombe J. Pharyngeal mucosal pressures, airway sealing pressures and fiberoptic position with the intubating versus the standard laryngeal mask airway. Anesthesiology. 1999; 9 1001-1006
- 16 Keller C, Brimacombe J, Keller K, Morris R. A comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patients. Br J Anaesth. 1999; 82 286-287
- 17 Brimacombe J R, Berry A, White P F. The laryngeal mask airway: limitations and controversies. International Anesthesiology Clinics. 1998; 36 155-182
- 18 Brimacombe J. Analysis of 1500 laryngeal mask uses by one anaesthetist in adults undergoing routine anaesthesia. Anaesthesia. 1996; 51 76-80
- 19 Brimacombe J, Keller C. Comparison of the flexible and standard laryngeal mask airways. Can J Anaesth. 1999; 46 558-563
Dr. J. Brimacombe
Dept. of Anaesthesia and Intensive Care, Cairns Base Hospital
The Esplanade, Cairns 4870, Australia
Email: jbrimacombe@north.net.au