Ultraschall Med 2001; 22(2): 91-95
DOI: 10.1055/s-2001-12853
ORIGINAL ARTICLE/ORIGINALARBEIT
© Georg Thieme Verlag Stuttgart · New York

Clinical Trial on the Accuracy of a Freehand and Sensor-Independent Three-Dimensional Power Doppler Ultrasound System Measuring Diameters, Volumes and Vascularity of Malignant Primaries of the Neck

Klinische Studie zur Messgenauigkeit eines drei-dimensionalen power Doppler Ultraschallsystems ohne externen Positionsgeber bei Durchmessern, Volumina und Vaskularisation maligner Primärtumoren des HalsesM. Keberle, M. Jenett, D. Hahn
  • Institut für Röntgendiagnostik, Universität Würzburg
Further Information

Publication History

Publication Date:
31 December 2001 (online)

Abstract

Aim: Diameters, volumes and vascularity of malignant primaries of the neck are of substantial importance for staging and prognosis. The purpose of this study was to determine the accuracy of a newly developed freehand and sensor-independent three-dimensional power Doppler ultrasound system (3D PDUS) in the assessment of these parameters. Method: 24 patients with squamous cell carcinomas of the neck underwent conventional ultrasound in B- and power Doppler-mode (US), 3D PDUS, and computed tomography (CT). Diameters (sagittal, longitudinal, and transverse) and volumes of the tumours were correlated with hand-segmentated CT images, whereas tumour vascularity was compared with US. Results: 3D PDUS and CT correlated highly regarding diameters and volumes (correlation coefficients r = 0.98/P < 0.001 and r = 0.98/P < 0.001, respectively). 3D PDUS and US correlated highly regarding vascularity (r = 0.92/P < 0.001). Conclusion: Although 3D PDUS is based on a freehand and sensorless data acquisition, it appears to be as accurate as CT or US for measurements of diameters, volumes and vascularity of neck malignancies. In spite of the excellent results it has to be mentioned that this technique needs a lot of training and its general use has to be employed with caution.

Klinische Studie zur Messgenauigkeit eines drei-dimensionalen power Doppler Ultraschallsystems ohne externen Positionsgeber bei Durchmessern, Volumina und Vaskularisation maligner Primärtumoren des Halses

Studienziel: Durchmesser, Volumen und Vaskularisation maligner Primärtumoren aus dem HNO-Bereich haben bezüglich des Stagings und der Prognose einen besonderen Stellenwert. Das Studienziel bestand darin, die Genauigkeit eines neuartigen, freihändigen, dreidimensionalen power Doppler Ultraschalls ohne externen Positionsgeber (3D PDUS) bei der Bestimmung dieser Parameter zu ermitteln. Methode: Bei 24 Patienten mit Plattenepithelkarzinomen aus dem HNO-Bereich wurden ein konventioneller B- und power Doppler-mode Ultraschall (US), 3D PDUS und eine Computertomographie (CT) durchgeführt. Durchmesser (sagittal, longitudinal und transversal) und das Volumen der Tumoren wurden mit manuell segmentierten CT-Bildern und die Tumorvaskularisation mit US korreliert. Ergebnisse: 3D PDUS und CT korrelierten sehr gut bezüglich der Durchmesser und des Volumens (Korrelationskoeffizienten jeweils r = 0,98/P < 0,001). 3D PDUS und US korrelierten ebenfalls sehr gut bezüglich der Vaskularisation (r = 0,92/P < 0,001). Schlussfolgerung: Trotz fehlender externer Positionsgeber scheint 3D PDUS ebenso genaue Messungen von Durchmessern, Volumina und der Vaskularisation von HNO-Tumoren zu ermöglichen wie die CT bzw. US. Trotz der exzellenten Ergebnisse muss darauf hingewiesen werden, dass diese Technik einen hohen Lernbedarf hat und nicht ohne Skepsis allgemein angewendet werden sollte.

References

  • 1 Pameijer F A, Balm A JM, Hilgers F JM, et al. Variability of tumor volumes in T3-staged head and neck tumors.  Head & Neck. 1997;  19 6
  • 2 Martinez-Gimeno C, Rodriguez E M, Vila C N, et al. Squamous cell carcinoma of the oral cavity: a clinicopathologic scoring system for evaluating risk of cervical lymph node metastasis.  Laryngoscope. 1995;  105 728
  • 3 Dietz A, Delorme S, Rudat V, Zuna I, Conradt C, Vanselow B, Weidauer H. Prognostic assessment of sonography and tumor volumetry in advanced cancer of the head and neck by use of Doppler ultrasonography.  Otolaryngol Head Neck Surg. 2000;  122 596
  • 4 Macchiarini P, Fontanini G, Hardin M J, et al. Relation of neovascularisation to metastasis of non-small-cell lung cancer.  Lancet. 1992;  340 145
  • 5 Williams J K, Carlson G W, Cohen C, et al. Tumor angiogenesis as a prognostic factor in oral cavity tumors.  Am J Surg. 1994;  168 373
  • 6 Penfold C N, Partridge M, Rojas R, et al. The role of angiogenesis in the spread of oral squamous cell carcinoma.  Br J Oral Maxillofac Surg. 1996;  34 37
  • 7 Alcalde R E, Shintant S, Yoshihama Y, et al. Cell proliferation and tumor angiogenesis in oral squamous cell carcinoma.  Anticancer Res. 1995;  15 1417
  • 8 Shields L E, Lowery C, Deforge C, et al. Technology and early clinical experience with real time 3D ultrasound.  Electromedica. 1998;  66 84
  • 9 Keberle M, Jenett M, Betz T, et al. Neue Möglichkeiten in Diagnostik und Bilddokumentation mit 3D Power Doppler Ultraschall am Beispiel von Mundbodenkarzinomen.  Ultraschall in Med. 2000;  21 26
  • 10 Klimek L, Schreiber J, Amadee R G, et al. Three-dimensional ultrasound evaluation in the head and neck.  Otolaryngol Head Neck Surg. 1998;  118 267
  • 11 Sader R, Zeilhofer H F, Deppe H, et al. Geräte- und transducerunabhängige 3D-Sonographie im Mund-Kiefer-Gesichtsbereich.  Ultraschall in Med. 1995;  16 269
  • 12 Fleischer A C, Wojcicki W E, Donnelly E F, et al. Quantified color Doppler sonography of tumor vascularity in an animal model.  J Ultrasound Med. 1999;  18 547
  • 13 Merz E. Aktuelle technische Möglichkeiten der 3D-Sonographie in der Gynäkologie und Geburtshilfe.  Ultraschall in Med. 1997;  18 190
  • 14 Kirbach D, Whittingham T A. 3D ultrasound - the Kretz-technik Voluson® approach.  Europ J Ultrasound. 1994;  1 85
  • 15 Fenster A. Dreidimensionale Rekonstruktionstechniken. In: Bogdahn U, Becker G, Schlachetzki F (Ed) Echosignalverstärker und transkranielle Farbduplex-Sonographie. Berlin, Wien; Blackwell Wissenschafts-Verlag 1998: 121-140
  • 16 Ohishi H, Hirai T, Yamada R, et al. Three-dimensional power Doppler sonography of tumor vascularity.  J Ultrasound Med. 1998;  17 619
  • 17 Friemel B H, Bohs L N, Nightingale K R, et al. Speckle decorrelation due to two-dimensional flow gradients.  IEEE UFFC. 1998;  45 317
  • 18 Fornage B D, Atkinson E N, Nock L F, Jones P H. US with extended field of view: Phantom-tested accuracy of distance measurements.  Radiology. 2000;  214 579
  • 19 Delorme S, Peschke P, Zuna I, et al. Sensitivity of color Doppler sonography: an experimental approach.  Ultrasound Med Biol. 1999;  25 541
  • 20 Rubin J M, Bude R O, Carson P L, et al. Power Doppler US : A potentially useful alternative to mean frequency-based color Doppler US.  Radiology. 1994;  190 853
  • 21 Bude R O, Rubin J M, Adler R S. Power versus conventional color Doppler sonography: comparison in the depiction of normal intrarenal vasculature.  Radiology. 1994;  192 777
  • 22 Delorme S. Beurteilung der Tumorvaskularisation mit der Dopplersonographie.  Radiologe. 1998;  38 335
  • 23 Mende U, Zöller J, Dietz A, Wannenmacher M, Born I A, Maier H. Die Sonographie im Primärstaging von Kopf-Hals-Tumoren.  Radiologe. 1996;  36 207

Dr. Marc  Keberle

Institut für Röntgendiagnostik
der Universität Würzburg

Josef-Schneider-Straße 2
97080 Würzburg


Email: E-mail:marc.keberle@mail.uni-wuerzburg.de

    >